Measurement of endometrial thickness with transvaginal ultrasonography: is it reproducible?
article
OA: closed
CC0
⤵ 2 in-corpus citations
AI-generated summary
Transvaginal ultrasonography demonstrated excellent intraobserver and good interobserver agreement for measuring endometrial thickness, though measurements near critical cut-offs warrant caution.
One-sentence paraphrase of the abstract; not a substitute for reading it. No clinical advice. How this works
Abstract
To assess the intraobserver and interobserver variability of transvaginal ultrasonographic measurement of endometrial thickness, two investigators (M.V., M.-F. D.) independently evaluated two successive measurements in 55 patients with a 5.0 MHz transvaginal transducer. Each observer was blinded to his or her own and to the other's results. Measurements were taken at the maximal thickness in the longitudinal uterine axis. The intraclass coefficients evaluating the agreement between two measures of a same observer were 94% and 97%; the coefficient was 94% between the two investigators. Intraobserver kappa values (agreement on endometrium 5 mm) were 0.70 and 0.81, respectively; the interobserver kappa value was 0.74. Transvaginal sonographic measurement of endometrial thickness has excellent intraobserver and good interobserver agreement. However, in clinical practice, an endometrial thickness measuring +/- 1 mm of a cut-off value should be interpreted with caution, especially when this may modify the clinical management.
My notes (saved in your browser only)
Citation neighborhood (sparse)
Too few in-corpus citations on either side for a chart; here are the lists.
Cites (2)
- Sonographic depiction of normal and abnormal endometrium with histopathologic correlation. 1986
- Sonographic depiction of the endometrium during normal cycles 1986
Cited by (2)
References (12)
- Sonographic depiction of normal and abnormal endometrium with histopathologic correlation. via openalex
- Sonographic depiction of the endometrium during normal cycles via openalex
- W2023983090 via openalex
- W2043242099 via openalex
- W2060164720 via openalex
- W2137320515 via openalex
- W2157500339 via openalex
- W2171820490 via openalex
- W2316834452 via openalex
- W1936017713 via openalex
- W2413558923 via openalex
- W1985797139 via openalex
Cited by (2)
Source provenance
- openalex
- last seen: 2026-06-10T17:14:06.276822+00:00
- unpaywall
- last seen: 2026-09-13T06:26:10.529621+00:00
License: CC0
· commercial use OK