Raw SWCNT material, CoMoCAT SG65 and SG65i (Sigma-Aldrich) was used for the large-scale preparation of OCC-SWCNTs. The SWCNTs were dissolved in chlorosulfonic acid (Sigma-Aldrich, 99.9%) at a concentration of ~4 mg ml −1 with magnetic stirring, followed by the addition of an aniline derivative at different molar ratios relative to the SWCNT carbon, and equimolar amounts of sodium nitrite (≥97.0%, Sigma-Aldrich). The aniline derivatives tested for these experiments include 4-amino-2-fluorobenzoic acid (97%, Sigma-Aldrich), 3,4,5-trifluoroaniline (98%, Sigma-Aldrich) and N , N -diethyl-p-phenylenediamine (97%, Sigma-Aldrich). The SWCNT–superacid mixture was then added drop-by-drop into Nanopure water with vigorous stirring ( Safety note: the neutralization process is aggressive; a significant amount of heat and acidic smog can be generated. Personal protective equipment, including goggles/facial mask, lab coats, and acid-resistant gloves, are necessary. The neutralization must be performed in a fume hood). The resulting OCC-SWCNTs instantly precipitate out from the solution. The precipitates were then filtered on an anodic aluminum oxide filtration membrane with a pore size of 0.02 μm (Whatman Anodisc inorganic filter membrane), thoroughly rinsed with Nanopure water and then dried in a vacuum oven.
The OCC-SWCNTs were stabilized by 3.5 mg ml −1 ssDNA in phosphate buffered saline (PBS). The OCC-SWCNT were individually dispersed by ultrasonication at 6 W for 60 min using a probe-tip sonicator (Sonics & Materials) at 4 °C for 1 h. The DNA to SWCNT mass ratio is 5 to 1. Then the OCC-DNA solutions were centrifuged at 100,000 g and 4 °C for 30 min. The 80% supernatant was dialysed against PBS for 36 h to remove free DNA (Spectra-Por, Float-A-Lyzer, MWCO = 1MDa). The absorption spectra of the dialysed solutions were collected with a UV-Vis-NIR spectrophotometer (Jasco). After subtracting absorption background, the optical density at (6,5) E 11 (~1,000 nm) was used to estimate the relative OCC-DNA concentration 48 . The OCC-DNAs were kept at 4 °C until used (up to 6 months) as the OCC-DNAs remained colloidally stable.
For the training set data collection, we used the OCC-DNAs that were synthesized within 6 months prior to testing with patient serum samples (1 week to 6 months old). For the test set, we used freshly prepared OCC-DNAs (less than 2 weeks old). The OCC-DNA concentration was adjusted to 0.325 mg l −1 in PBS. We introduced 20 μl of a patient serum sample to 80 μl of OCC-DNAs in a 96-well plate (Corning) to make the OCC-DNA concentration of 0.26 mg l −1 in each well. OCC-DNAs in 100 μl PBS (0.26 mg l −1 ) was also prepared to compare the relative changes in sensor response in serum for feature vector construction (see Data preprocessing in Methods ). The OCC-DNA was incubated at room temperature for 2 h and in a cold room (4 °C) after the spectral acquisition at 2 h time point. Data were taken at three time points during incubation: 2 h, 24 h and 72 h.
To test sensor sensitivity to serum biomarkers, OCC-DNA complexes were added to a 96-well plate at a concentration of 0.26 mg l −1 in a 100 μl total volume of 20% FBS (Gibco). In triplicate, the following were added into wells at biologically relevant concentrations: 0–352,000 U ml −1 recombinant human CA125/MUC16 (R&D Systems), 0–100 nM recombinant human HE4 (RayBiotech), 0–100 nM recombinant human YKL40 (R&D Systems), 0–50 nM recombinant human mesothelin (BioLegend), 0–1,000 μM creatinine (≥98%, anhydrous, Fisher Scientific) or 0–200 μM bilirubin (≥97%, Fisher Scientific). Experiments were performed with the same time points as above. All experiments were performed in triplicate.
Fluorescence emission spectra of OCC-DNAs were acquired using a home-built near-infrared fluorescence spectroscopy apparatus consisting of a tunable white-light laser source, inverted microscope and InGaAs NIR detector. A SuperK EXTREME supercontinuum white-light laser source (NKT Photonics) was used with a VARIA variable bandpass filter accessory, capable of tuning the output to 500–825 nm, set to a bandwidth of 20 nm centred at 575 nm. The light path was shaped and fed into the back of an inverted IX-71 microscope (Olympus), where it passed through a ×20 NIR objective (Olympus) and illuminated the samples in a 96-well plate. Emission from the OCC-DNAs was collected through the ×20 objective and passed through a dichroic mirror (875 nm cutoff, Semrock). The light was f/# matched to the spectrometer using several lenses and injected into a Shamrock 303i spectrograph (Andor, Oxford Instruments) with a slit width of 100 μm, which dispersed the emission using an 86 g mm −1 grating with 1.35 μm blaze wavelength. The spectral range was 723–1,694 nm with a resolution of 1.89 nm. The light was collected by an iDus 1.7 μm InGaAs (Andor, Oxford Instruments) with an exposure time of 10 s. An HL-3-CAL-EXT halogen calibration light source (Ocean Optics) was used to correct for wavelength-dependent features in the emission intensity arising from the spectrometer, detector and other optics. An Hg/Ne pencil-style calibration lamp (Newport) was used to calibrate the spectrometer wavelength. Background subtraction was conducted using a well in a 96-well plate filled with PBS or 20% FBS, depending on the experiment. Following acquisition, the data were processed with custom code written in Matlab that applied the aforementioned spectral corrections and background subtraction and was used to fit the data with Lorentzian functions.
Waste samples (269) were collected from female patients diagnosed with ovarian and other cancers under a Memorial Sloan Kettering Cancer Center Institutional Review Board approved protocol. From this sample set, 56 specimens were collected from patients diagnosed with high-grade serous ovarian cancer, 71 from healthy donors, 56 from patients with other gynaecologic diseases, 61 from patients with non-gynaecologic diseases and 25 from patients in remission. There was no statistically significant difference in age distribution for each group. Diagnoses were identified from a chart review of each patient; all diagnoses included histology and were confirmed by a gynaecologic oncology attending physician. Patient demographics, diagnosis and biomarker levels are available in Supplementary Information .
Serum concentrations of CA125 and HE4 were determined on the Abbott Architect i2000 analyser (Abbott Diagnostics) using a chemiluminescent microparticle immunoassay. YKL40 was analysed using a singleplex immunoassay on the Protein Simple Ella system. The Abbott C8000 analyser was used to determine the concentrations of creatinine by quantitating the formation of creatinine picrate in alkaline conditions, and bilirubin was analysed by the formation of azobilirubin using the diazo reagent under specified conditions.
Quantities representing the sensor response to patient serum were acquired by the Lorentzian fitting of OCC-DNA fluorescence spectra: E 11 intensity, E 11 − intensity, E 11 wavelength and E 11 − wavelength. The average value of triplicates was used as feature data for machine learning processes. Feature values were defined as a difference in sensor response acquired from patient serum and PBS. Specifically, the E 11 peak position feature, dwl, was defined as the wavelength difference between the E 11 peak in the patient sample (wl) and PBS (wl 0 ): dwl = wl – wl 0 . The E 11 peak intensity feature, dint, was normalized as dint = int/int 0 , where int and int 0 are the E 11 peak intensity in serum and PBS, respectively. Similarly, we defined E 11 − peak related features, dwl* and dint*, indicating the relative E 11 − peak position and intensity. We additionally considered the relative change in E 11 − to E 11 intensity: Δint = (int*/int)(int 0 */int 0 ) −1 – 1, and the wavelength difference between two peaks: Δwl = dwl* – dwl, to check whether the addition of these features would create a larger variance in HGSOC prediction.
We normalized each feature vector to be in the range of −1 and 1 to balance the feature contribution to the model. The imbalance in the size of each group was corrected by upscaling minority species (SMOTE: Synthetic Minority Oversampling Technique) 49 so that the prediction models were not biased by groups with larger sample sizes. For the biomarker prediction models, we divided the data into normal versus high biomarker level groups on the basis of clinical references (CA125, 50 U ml −1 ; HE4, 150 pM; YKL40, 1,650 pM) and corrected the group size using SMOTE.
Using algorithms implemented in ‘Scikit-Learn’ 50 , we created models on the basis of decision tree, logistic regression, artificial neural networks, random forest and SVM for binary classification. Hyperparameters for each model were optimized using Bayesian optimization, implemented in the HyperOpt library 35 . The loss function to minimize in the hyperparameter optimization was set to (1 – F -score). F -score (or F 1 -score) is a measure of accuracy in binary classification and is calculated from the harmonic mean of the positive predictive value (PPV) and sensitivity: 2/(sensitivity −1 + PPV −1 ). To rule out possible overfitting in the machine learning process, model performance was evaluated using 10-fold cross-validation. In the cross-validation process, stratified shuffle split validation was used to randomly partition the data set into 10 subsamples. In each partition, 9 of the 10 subsamples were used to train the model, while a single subsample was used to test the trained model. The average F -score of the 10-fold cross-validation was used to assess model performance. The trained models were then tested with an independent set of patient sera ( N = 54), sampled from different patients (test set), as external validation. SVR was used to construct the regression models of HGSOC serum biomarkers with 10-fold cross-validation. The loss function in the hyperparameter optimization was (1 – r 2 ). For SVM and SVR, a radial basis function kernel was used and the hyperparameter optimization was performed for the regularization parameter (cost) and the kernel coefficient (gamma), with the maximum iteration of 1,000. The hyperparameter space of each machine learning algorithm for model optimization is noted in Supplementary Information .
Results
We synthesized an array of OCC-DNA nanosensors by introducing the OCCs to the (6,5)-SWCNT via diazonium chemistry 33 and encapsulating them with a library of ssDNA to solubilize the nanosensors in biofluids. The ssDNA sequences were chosen on the basis of the recognition sequences of DNA that form specific wrapping patterns on the SWCNT surface 34 to result in diverse, highly defined surface morphologies to confer disparate sensitivities to the local environment 26 , 27 . Ten different OCC-DNA nanosensors were synthesized from the combinations of three OCCs and four DNA sequences ( Table 1 ). Each OCC-DNA nanosensor featured a pair of emission peaks that depend on the chemical nature of the OCC and DNA sequence. We used 575 nm excitation to selectively excite the (6,5)-SWCNT ( Fig. 2a , and Supplementary Figs. 1 and 2 ), resulting in emission at ~1,000 nm from the (6,5) nanotube species E 11 band and a peak falling between 1,110 to 1,170 nm, depending on the aryl functional group. The latter is denoted as the E 11 − band or ‘OCC peak’.
To determine a minimal set of OCC-DNA combinations that provide the most diverse responses from the patient samples, we measured the fluorescence spectral responses of the OCC-DNAs to serum samples from HGSOC patients and healthy individuals. Four serum samples of the two conditions were incubated with ten different OCC-DNAs for 2 h, and the fluorescence spectra of the OCC-DNA complexes were acquired. For each OCC-DNA nanosensor, we analysed four different spectral features of the OCC-DNA nanosensors that were modulated in response to interactions with analytes in serum: E 11 and E 11 − intensity (int and int*) and wavelength (wl and wl*). From these data, we identified the sensors that gave statistically significant differences in response to healthy versus cancer groups in parametric t -tests (quantified by P value; Fig. 2b and Extended Data Fig. 1 ). Our hypothesis was that OCC-DNAs that perform well independently would make good choices when used in combination. Six OCC-DNA nanosensors exhibiting E 11 or E 11 − peak wavelengths with statistically significant differences between HGSOC and healthy groups ( P < 0.10) were selected for the sensor array used in the subsequent parts of this study (highlighted in Table 1 and Extended Data Fig. 1 ). The selection/reduction of features improves the training speed and model performance by eliminating redundant features in the data set.
We initially exposed the OCC-DNA sensor array to 215 patient serum samples and constructed a data set comprising the spectral feature changes caused by the serum environment. Specifically, the size of the data matrix was N sa × ( N f × N OCC-DNA ), where N sa is the number of serum samples, N f is the number of features per OCC-DNA and N OCC-DNA is the number of different OCC-DNA complexes in the array. The set of serum samples was collected from 49 HGSOC, 51 other gynaecologic diseases (such as endometriosis and low-grade ovarian carcinoma), 29 non-gynaecologic cancer, 25 cancer patients in remission including 7 HGSOC, and 61 healthy donors ( Supplementary Table 1 ). The fluorescence spectra were collected at three time points during incubation: 2 h, 24 h and 72 h.
The average of triplicate sensor responses was used for the data analysis. We note that the variation of each measurement from the averaged triplicates was small for all the OCC-DNA peaks ( Supplementary Fig. 3 ). The variations in dwl and dwl* showed narrow Gaussian distributions, with standard deviations ranging from 3.72 to 5.37%. The maximum variation in the same sample was less than 15% (<0.3 nm). The analysis confirmed that our measurement can reliably identify the small spectral shifts. This is likely because OCC-DNAs exhibit relatively narrow bandwidths (35–80 meV), making small spectral shifts much easier to resolve compared with conventional fluorophores (>100 meV).
All four spectroscopic variables—int, int*, wl, wl*—measured from the OCC-DNA nanosensor array, exhibited statistically significant differentiation between HGSOC and healthy groups, but the data did not delineate a clear difference between HGSOC and other disease conditions ( Fig. 2c and Extended Data Fig. 2 ). Principal component analysis (PCA) was performed on the spectroscopic data ( N f = 4) upon a 2 h incubation from all combinations of OCC-DNA sensors ( N OCC-DNA = 6). The first two principal components accounted for 87.5% of the total variance (principal component loadings listed in Supplementary Table 2 ). Similar to Fig. 2c , healthy samples showed the differentiable signatures from the disease samples, denoted by their segregation into separate regions in the PCA plot ( Fig. 2d ), but HGSOC could not be separated from the other disease conditions.
To differentiate HGSOC from the other disease conditions, we next trained machine learning models using the sensor responses and clinical diagnostic results ( Fig. 3 and Extended Data Fig. 3 ). The algorithms were used for binary classification of sensor responses: HGSOC vs other diseases + healthy (the differentiation of HGSOC from all other samples). The set of features chosen for the classification task were the spectroscopic variables dint, dint*, dwl and dwl* collected from the OCC-DNA sensor array. For robustness, we investigated five standard machine learning algorithms with nested levels of optimization processes: model hyperparameters, model choice and multilevel validation. We tested supervised machine learning algorithms—logistic regression, decision tree, artificial neural networks (ANN), random forest (RF) and support vector machine (SVM)—while tuning models’ hyperparameters with Bayesian optimization 35 ( Supplementary Table 2 ). The averaged F-score in 10-fold cross-validation was used to assess the model performance (see Methods ).
We first examined the machine learning algorithm that most accurately classifies HGSOC ( Fig. 3a ). We compared the averaged F -scores of the machine learning algorithms using OCC-DNA combinations within the sensor array. We assessed combinations of OCC-DNA nanosensor responses, up to 6 at a time, out of the 6 originally selected OCC-DNAs (1 ≤ N OCC-DNA ≤ 6), for 63 total possible combinations for each incubation duration (see Supplementary Table 3 ). We found that SVM resulted in the best F -scores among the 5 machine learning algorithms that we tested ( Extended Data Fig. 3 ). Thus, we used SVM models for subsequent optimizations of the HGSOC classifier.
For our second optimization, we compared the differences in model performance using sensor responses measured under different durations of incubation with the serum samples ( N f = 3 × 63). In all the tested machine learning algorithms, there were no statistically significant differences between incubation times ( Extended Data Fig. 3 ). We found that combining datasets obtained over multiple incubation durations could improve the model performance, but the performance was only marginally better than using 2 h of serum incubation ( Fig. 3a and Supplementary Table 3 ). Thus, we used the 2 h data set for subsequent model development for simplicity.
Thirdly, we examined which spectroscopic variables in the set of feature vectors optimize F -scores. We compared three combinations of spectral variables, involving the E 11 − to E 11 intensity ratio (Δint), the wavelength difference between E 11 − and E 11 peaks (Δwl), dwl, dwl*, dint, dint*, and combinations thereof ( N f = (2, 4 or 6) × 63; Fig. 3b ). The SVM models trained with the data set of 2 variables, Δint and Δwl, resulted in lower F -scores. We found no statistically significant difference between 4 and 6 variables in the F -scores of the optimized SVM models potentially because Δint and Δwl are derivatives of the others. Thus, we used the 4 variables for further investigations.
We then investigated the impact of the number of different OCC-DNA sensors in the array on the F -score (1 ≤ N OCC-DNA ≤ 6; Fig. 3c ). When more OCC-DNA elements were added to the sensor array, the F -scores tended to increase systematically. The trend was the same regardless of which machine learning algorithm was used ( Extended Data Fig. 3 ). The best SVM model was trained by the spectral response of 5 OCC-DNAs: NEt 2 *CTTC 3 TTC, NEt 2 *(TAT) 4 , 3F*(TAT) 4 , 3F*(AT) 15 and 3F*(GT) 15 . The averaged cross-validation score of the SVM model was 93.9% sensitivity, 95.2% specificity, and an F -score of 0.945 differentiating HGSOC from all other disease + healthy samples. Small variances in F -score and sensitivity (<0.1) in cross-validations suggest that the optimized models are generalizable within the sample set.
Lastly, we examined whether tuning the hyperparameters to maximize the F β score can improve sensitivity at a high specificity ( Fig. 3d ). The F β score is the weighted harmonic mean of PPV and sensitivity, and β is chosen such that sensitivity is considered β-times as important as PPV. At decreasing β from 3 to 0.2, sensitivity at 98% specificity systematically increased, although the improvement was statistically non-significant between μ values of 0.2, 0.5, 0.8 and 1 in this study ( Supplementary Table 4 ). The best-performing prediction model was the sensor array combination of 4-N(C 2 H 5 ) 2 *C T 2 C 3 T 2 C, 4-N(C 2 H 5 ) 2 *(TAT) 4 , 3,4,5-F 3 *(TAT)4, 3,4,5-F 3 *(AT) 15 and 3,4,5-F 3 *(GT) 15 , and yielded 87% sensitivity at 98% specificity when PPV and sensitivity were equally weighted ( β = 1).
To further assess the robustness of the sensor array and algorithm, we synthesized a new batch of OCC-DNAs under the same condition and collected the sensor array response data to an independent test set of 54 patient samples ( N sa = 54). To evaluate the model performance in various medical conditions, the test set was sampled from different patients, comprising 7 HGSOC, 5 other gynaecologic diseases, 32 non-gynaecologic diseases and 10 heathy patients. With this new sample set, the optimized SVM model resulted in 100% sensitivity at 98% specificity and an F -score of 0.978. These values are consistent with the cross-validation scores and gave a similar receiver operating characteristic (ROC) curve ( Fig. 3e ), indicating that the model did not overfit the data.
The risk of bias in the study was evaluated on the basis of Prediction model Risk Of Bias Assessment, PROBLAST 36 ( Appendix 1 in Supplementary Information ). The risk of bias scored low in terms of predictors, outcomes and analyses. In participants, the tool resulted in the finding of no systematic differences between training and cross-validation sets. However, the limited medical record of healthy donors and the enriched fraction of breast cancers in the non-HGSOC group of the test set may introduce systematic bias in participant selection and the validation of machine learning models, respectively. For clinical translation of the technology, these risks of bias must be taken into account.
We also endeavoured to account for chemical interferents and background chronic conditions that could confer a bias in the sensor response. From a patient chart review, we identified chronic diseases and most common medications administered to the patients ( Extended Data Fig. 4 ). We found that 75% of HGSOC and 68% of other disease patients suffered from at least one chronic condition, and the relative abundance between these disease groups was similar. Regarding the medications, we statistically assessed the contribution of each interferent to the sensor results using a multivariate regression model ( Supplementary Table 5 ). The regression model determined a linear correlation between the sensor response and medication, using estimated parameters and errors. The adjusted R 2 of the regression model ranged from −0.045 to 0.233, indicating weak linear correlations of each sensor response to medications. We confirmed that the sensor array technology accurately classified the disease status regardless of medication and chronic conditions, as evidenced by high F -scores of the HGSOC prediction models ( Supplementary Table 3 ). The analysis suggested no indications that such interferents reduced the specificity of HGSOC detection.
To test the utility of the SVM model relative to conventional diagnostic methods, we compared conventional biomarker-based HGSOC detection and histology results to the F -score predicted by the SVM model. We measured known biomarkers in the patient serum samples, including CA125, HE4 and YKL40, creatinine, and bilirubin by immunoassays (see Methods ). We assessed the diagnostic accuracy of serum HGSOC biomarkers in these patients ( Fig. 3f ). Although the differences in serum CA125, HE4, and YKL40 levels, with respect to the clinical references, were statistically significant between HGSOC, healthy, and other (non-HGSOC) diseases ( Extended Data Fig. 5 ), false-positive rates were high. For example, CA125-based screening with 50 U ml −1 cutoff resulted in 65.3% sensitivity, 88.3% specificity and an F -score of 0.621 in our patient sample set. The logistic regression of additional biomarkers marginally improved the HGSOC prediction ( Fig. 3f ). PCA plots of HGSOC biomarkers CA125, HE4 and YKL40 showed that these markers failed to differentiate HGSOC from other diseases, while the healthy individuals’ samples clustered together ( Fig. 3g ). Clinical trials using these biomarkers showed similar results 4 , 37 . These results confirmed that our perception/sensor-based technology substantially outperformed established serum biomarker-based classification, and the accuracy was much closer to diagnosis by a physician (using pathology, imaging and so on).
To better understand the molecular basis for the sensor-based HGSOC fingerprint, we investigated the sensor response to serum biomarkers ( Fig. 4 ). We measured the spectral response of the OCC-DNA nanosensors upon single analyte titration with bilirubin, creatinine and HGSOC serum biomarkers, including CA125, HE4, YKL40 and mesothelin in 20% foetal bovine serum (FBS) ( Fig. 4a - d and Extended Data Fig. 6 ). We found that several OCC-DNA spectral responses correlated with CA125, HE4, YKL40 and bilirubin concentrations, while mesothelin and creatinine showed no quantitative correlations with the sensor responses. We surmise that, because of this correlation, the inclusion of biomarker-dependent spectroscopic variables in the training data set improved F -scores for HGSOC identification. We then assessed the relative contribution of each spectral parameter to the model performance by an ablation study—individually dropping each spectroscopic variable from the analysis ( Extended Data Fig. 7 ). On analysis of feature importance, we identified that 3F*(GT) 15 and 3F*(TAT) 4 were the most important OCC-DNA nanosensors. We also found that the same feature in different sensor arrays can improve or reduce the prediction scores ( Extended Data Fig. 7 ). For instance, the E 11 − intensity (dint*) of NEt 2 *CTT has the highest positive feature importance (improved F -score by 0.067) in the sensor array of 4-N(C 2 H 5 ) 2 *CT 2 C 3 T 2 C, while the same feature reduced the F -score by 0.018 in the sensor array combination of 4-N(C 2 H 5 ) 2 *CT 2 C 3 T 2 C and 3,4,5-F 3 *(GT) 15 . Overall, the biomarker-dependent features scored highly, indicating that such features improved the SVM model performance ( Fig. 4e ). The observations confirmed that (1) OCC-DNA fluorescence transduces broad types of subtle differences in physicochemical properties of physisorbed molecules and (2) known serum biomarkers make up part of the disease fingerprint. However, the use of biomarker-dependent features exclusively did not result in optimal F -scores. The inclusion of certain features that showed no quantitative correlation with known biomarkers improved the model performance. These experiments suggest that the OCC-DNA nanosensor array results may be due, at least in part, to the transduction of heretofore unidentified biomarkers.
To further investigate the correlation between serum biomarker levels and the response of the nanosensor array, we assessed whether the sensor array responses could be used to train an SVM model to identify abnormal levels of known biomarkers in the patient samples. First, we trained an SVM classification model to detect elevated CA125 by dividing the patient sera into groups on the basis of threshold for suspicion of malignancy; normal (0–50 U ml −1 ) vs high (>50 U ml −1 ) CA125. The CA125 training resulted in high F -scores (>0.92) for all possible sensor array combinations ( Fig. 4f , g and Supplementary Table 6 ). We similarly assessed HE4 and YKL40 with respect to their clinical references of 150 pM for HE4 and 1,650 pM for YKL40, and we developed binary classification models to differentiate abnormal levels using the SVM algorithm. Both HE4 and YKL40 classifications resulted in high F -scores (0.89–0.98 and 0.81–0.93, respectively) for the detection of abnormal biomarker levels.
We additionally investigated whether support vector regression (SVR) models can quantitatively predict serum biomarker levels using the sensor array ( Fig. 4h ). The best CA125 regression model, using the three OCC-DNAs NEt 2 *(TAT) 4 , 3F*(TAT) 4 and 3F*(AT) 15 resulted in an average R -squared ( r 2 ) value of 0.719 ( Fig. 4i ). We note that the prediction error in the normal concentration range (<50 U ml −1 ) was larger than in the high concentration range. This can be attributed to the fact that the detection limit in the single titration experiment was close to the clinical reference of CA125. The SVR models of HE4 and YKL40 were also constructed, resulting in r 2 values of 0.55 and 0.56, respectively. The SVR models suggest that the known biomarkers influence the sensor responses, but the models were not sensitive enough to reliably predict the exact biomarker levels.
We assessed the contribution of each spectral parameter to the biomarker classification and regression models ( Extended Data Fig. 8 ). Most of the spectral parameters had positive relative importance on average, indicating that including such features improved the positive predictive value and sensitivity of the biomarker identification. A positive correlation of the feature importance to F -score (for binary classification) and r 2 (for regression) was stronger for the biomarker-dependent variables that were identified in the single-analyte experiments ( Fig. 4a - d ). Regarding bilirubin, however, although OCC-DNA fluorescence responses quantitatively correlated with its concentration over biologically relevant ranges ( Extended Data Fig. 6 ), we failed to optimize a good SVR model for detection due to the small variance of the biomarker levels within the patient samples. The SVR model performance for serum creatinine was poor due to a lack of quantitative correlation between sensor response and creatinine concentrations in the single-titrant experiment ( Extended Data Fig. 6 ).
Discussion
We constructed a nanosensor-array technology, composed of OCC-DNA elements and coupled with machine-learning algorithms, to investigate the potential to identify HGSOC in patient sera. The array was composed of multiple OCC moieties and DNA sequences, which together offer a rich design space for modulating the morphology and chemistry of the exposed nanotube surface. Our DNA sequence selection was based on the recognition sequences that form specific wrapping patterns on the nanotube surface. These sequences were originally selected to isolate individual (n,m) species/chiralities of nanotubes 34 . We reasoned that the recognition sequences of DNAs would confer the greatest diversity of interactions with the serum milieu, which is important to establish an OCC-DNA library for screening disease-specific sensor responses. We based this rationale on the findings that ssDNA encapsulates SWCNTs via π-π stacking interactions, and certain DNA sequences can behave like a ‘molecular mask’ that defines the shape and size of the exposed surface 26 . Their characteristic surface structures are responsible for diverse physicochemical properties of the OCC-DNAs 27 , leading to different protein corona compositions 38 , 39 . Different morphologies determined by OCCs and DNA thereby contribute to the selectivity of the nanotube surfaces to the serum milieu. The fluorescence modulation of SWCNTs is caused by several mechanisms, including Fermi level shifting through modulation of the immediate redox environment and exciton disruption in response to binding events, which change SWCNT intensity, and solvatochromic (wavelength) shifting due to perturbation of the local dielectric environment, including shifts due to modulation of the local electrostatic environment 24 , 40 . OCC fluorescence, on the other hand, is molecularly specific and extremely sensitive to the local chemical environment of the atomic defect sites 32 , 41 . Interactions between HGSOC serum biomarkers and OCC-DNA hybrids elicited diverse spectral responses of the sensor array that enabled sufficient differentiation of signals from other sera.
The sensor technology was used to identify HGSOC with high positive and negative predictive values. Model performance of the sensor technology exceeds the results of the current best clinical screening test using longitudinal CA125 and second-line transvaginal ultrasonography 9 (87% vs 84% clinical sensitivities at 98% specificity). However, considering the fact that specimens obtained from symptomatic individuals at diagnosis were used for the development and assessment of the technology, prediction outcomes will be lower in the clinical screening setting in which specimens are obtained in asymptomatic individuals before clinical diagnosis. Further studies, in the context of screening the general population, are warranted to evaluate the ability of the technology to identify pre-invasive and early-invasive disease.
This sensor technology exhibits several unique potential advantages for clinical applications. First, this method could be rapidly adapted to the detection of many diseases/conditions. The array could be used to train an algorithm to recognize nearly any disease when given enough data from the sensor responses to the appropriate patient serum samples. Second, this technology could supplement or replace the use of known biomarkers when there are issues with selectivity in conventional multi-analyte tests. Due to the potential to iteratively modify the sensor array and machine learning algorithms and to additively augment training set size, the selectivity may be increasingly optimized. Third, this sensor technology can be used in a high-throughput fashion to facilitate the screening of large populations. Fourth, because the technology does not rely on antibody-based molecular recognition elements, the sensors could be more robust than existing methods ( Supplementary Fig. 4 ), enabling use in resource-limited settings and in technologies such as point-of-care and wearable/implantable devices 42 . Lastly, the sensor technology has the potential to be developed into an inexpensive and rapid screening tool that produces a single, easy-to-interpret test result in primary-care settings. The materials needed for the sensor cost approximately $5 per sample because of the small amount of OCC-DNAs needed for screening (<5 ng). The cost of the sensor measurement would also diminish if measured via high-throughput instruments, and the potential for the use of very low sample volumes is substantial.
This work employed machine perception to detect disease fingerprints using an array of optical nanosensors. The study carefully investigated the attributes and molecular mechanism that resulted in the excellent accuracy of the machine-learning-aided nanosensor array. It is important to note that the best-performing HGSOC prediction model ( Fig. 3d ) included the spectroscopic variables that were not sensitive to the known biomarkers, and their relative importance was much more significant than the biomarker-related variables ( Extended Data Fig. 7 ). This suggests that there exist potential biomarkers or combinations thereof that are either unknown or not part of conventional screening approaches but were captured by the OCC-DNA sensor array. Information detailing which biomarkers and molecular interactions primarily result in the disease fingerprint is unknown and largely cannot be determined by current machine-learning methods 43 . We believe that it may be possible, with extensive investigations, to use quantitative proteomics aided by the nanosensor array as a discovery tool 44 - 46 . Such investigation could be used to facilitate biomarker-discovery efforts 47 and to uncover new information related to disease pathophysiology.
Extended Data
Four spectral parameters –intensity and wavelength changes of the E 11 and E 11 – peaks– were extracted from fluorescence spectra of four serum samples in each group. Each sample was measured in triplicate. Horizontal lines denote the median. Six OCC-DNA nanosensors, with p-values of the spectroscopic features lower than 0.10, were selected for the sensor array.
Four spectral parameters, a , dint , b , dint* , c , dwl , and d , dwl* , were extracted from fluorescence spectra of the sensor array after 2-hour serum incubation. Each sample was measured in triplicate.
The classification was divided as HGSOC versus other gynecologic diseases and benign groups. The blue line is the logarithmic regression of the median F-score.
a , Fraction of medication dose for HGSOC and other disease patients. b , Chronic conditions, and prevalence thereof, in patients measured in this study. Comorbidity was identified based on the patients’ medication information. c , Anti-cancer drugs or prescription drugs whose occurrence differed by 0.1 or higher between HGSOC and other disease groups.
a, CA125, b, HE4, and c, YKL40. The serum protein levels were quantified by automated immunoassay. Dotted lines indicate the clinical reference of each biomarker for HGSOC diagnosis. The error bars denote median ± 95% CI.
The fluorescence spectra were obtained 2 hours after the incubation. Vertical dashed lines indicate the clinical reference of each serum biomarker for HGSOC screening.
a, Feature importance of each spectral parameter, used to train the SVM models, of all OCC-DNA sensors in the arrays tested in this work. Solid lines indicate the median feature importance. b, Correlation of averaged F-score with the averaged feature importance of each spectroscopic variable. Vertical dashed lines indicate F-score when all four spectroscopic variables ( dint, dint*, dwl , and dwl* ) of the OCC-DNA were included as feature vectors in the model development.
For the binary classification models (top rows), samples were divided into two groups–abnormal vs. normal levels of serum biomarkers–based on the clinical references (CA125: 50 U/mL, HE4: 150 pM, YKL40: 1650 pM) and assessed the prediction accuracy of abnormal levels of each biomarker. Feature importance of the prediction models shows which spectral parameters most impacted the model performance using an ablation study. Biomarker dependent variables that were identified in Extended Data Fig. 4 are highlighted in bold. Vertical dashed lines indicate F-score when all four spectroscopic variables ( dint, dint*, dwl , and dwl* ) of the OCC-DNA were included as feature vectors in the model development.