Quantitative T1 and Effective Proton Density (PD*) mapping in children and adults at 7T from an MP2RAGE sequence optimised for uniform T1-weighted (UNI) and FLuid And White matter Suppression (FLAWS) contrasts | medRxiv /* */ /* */ <!-- <!-- /*! * yepnope1.5.4 * (c) WTFPL, GPLv2 */ (function(a,b,c){function d(a){return"[object Function]"==o.call(a)}function e(a){return"string"==typeof a}function f(){}function g(a){return!a||"loaded"==a||"complete"==a||"uninitialized"==a}function h(){var a=p.shift();q=1,a?a.t?m(function(){("c"==a.t?B.injectCss:B.injectJs)(a.s,0,a.a,a.x,a.e,1)},0):(a(),h()):q=0}function i(a,c,d,e,f,i,j){function k(b){if(!o&&g(l.readyState)&&(u.r=o=1,!q&&h(),l.onload=l.onreadystatechange=null,b)){"img"!=a&&m(function(){t.removeChild(l)},50);for(var d in y[c])y[c].hasOwnProperty(d)&&y[c][d].onload()}}var j=j||B.errorTimeout,l=b.createElement(a),o=0,r=0,u={t:d,s:c,e:f,a:i,x:j};1===y[c]&&(r=1,y[c]=[]),"object"==a?l.data=c:(l.src=c,l.type=a),l.width=l.height="0",l.onerror=l.onload=l.onreadystatechange=function(){k.call(this,r)},p.splice(e,0,u),"img"!=a&&(r||2===y[c]?(t.insertBefore(l,s?null:n),m(k,j)):y[c].push(l))}function j(a,b,c,d,f){return q=0,b=b||"j",e(a)?i("c"==b?v:u,a,b,this.i++,c,d,f):(p.splice(this.i++,0,a),1==p.length&&h()),this}function k(){var a=B;return a.loader={load:j,i:0},a}var l=b.documentElement,m=a.setTimeout,n=b.getElementsByTagName("script")[0],o={}.toString,p=[],q=0,r="MozAppearance"in l.style,s=r&&!!b.createRange().compareNode,t=s?l:n.parentNode,l=a.opera&&"[object Opera]"==o.call(a.opera),l=!!b.attachEvent&&!l,u=r?"object":l?"script":"img",v=l?"script":u,w=Array.isArray||function(a){return"[object Array]"==o.call(a)},x=[],y={},z={timeout:function(a,b){return b.length&&(a.timeout=b[0]),a}},A,B;B=function(a){function b(a){var a=a.split("!"),b=x.length,c=a.pop(),d=a.length,c={url:c,origUrl:c,prefixes:a},e,f,g;for(f=0;f<d;f++)g=a[f].split("="),(e=z[g.shift()])&&(c=e(c,g));for(f=0;f<b;f++)c=x[f](c);return c}function g(a,e,f,g,h){var i=b(a),j=i.autoCallback;i.url.split(".").pop().split("?").shift(),i.bypass||(e&&(e=d(e)?e:e[a]||e[g]||e[a.split("/").pop().split("?")[0]]),i.instead?i.instead(a,e,f,g,h):(y[i.url]?i.noexec=!0:y[i.url]=1,f.load(i.url,i.forceCSS||!i.forceJS&&"css"==i.url.split(".").pop().split("?").shift()?"c":c,i.noexec,i.attrs,i.timeout),(d(e)||d(j))&&f.load(function(){k(),e&&e(i.origUrl,h,g),j&&j(i.origUrl,h,g),y[i.url]=2})))}function h(a,b){function c(a,c){if(a){if(e(a))c||(j=function(){var a=[].slice.call(arguments);k.apply(this,a),l()}),g(a,j,b,0,h);else if(Object(a)===a)for(n in m=function(){var b=0,c;for(c in a)a.hasOwnProperty(c)&&b++;return b}(),a)a.hasOwnProperty(n)&&(!c&&!--m&&(d(j)?j=function(){var a=[].slice.call(arguments);k.apply(this,a),l()}:j[n]=function(a){return function(){var b=[].slice.call(arguments);a&&a.apply(this,b),l()}}(k[n])),g(a[n],j,b,n,h))}else!c&&l()}var h=!!a.test,i=a.load||a.both,j=a.callback||f,k=j,l=a.complete||f,m,n;c(h?a.yep:a.nope,!!i),i&&c(i)}var i,j,l=this.yepnope.loader;if(e(a))g(a,0,l,0);else if(w(a))for(i=0;i (function(w,d,s,l,i){w[l]=w[l]||[];w[l].push({'gtm.start':new Date().getTime(),event:'gtm.js'});var f=d.getElementsByTagName(s)[0];var j=d.createElement(s);var dl=l!='dataLayer'?'&l='+l:'';j.src='//www.googletagmanager.com/gtm.js?id='+i+dl;j.type='text/javascript';j.async=true;f.parentNode.insertBefore(j,f);})(window,document,'script','dataLayer','GTM-P4HH5NV'); Skip to main content Home About Submit ALERTS / RSS Search for this keyword Advanced Search Quantitative T1 and Effective Proton Density (PD*) mapping in children and adults at 7T from an MP2RAGE sequence optimised for uniform T1-weighted (UNI) and FLuid And White matter Suppression (FLAWS) contrasts View ORCID Profile Ayşe Sıla Dokumacı , View ORCID Profile Katy Vecchiato , Raphael Tomi-Tricot , View ORCID Profile Michael Eyre , View ORCID Profile Philippa Bridgen , Pierluigi Di Cio , View ORCID Profile Chiara Casella , View ORCID Profile Tobias C. Wood , View ORCID Profile Jan Sedlacik , Tom Wilkinson , Sharon L. Giles , View ORCID Profile Joseph V. Hajnal , View ORCID Profile Jonathan O’Muircheartaigh , View ORCID Profile Shaihan J. Malik , View ORCID Profile David W. Carmichael doi: https://doi.org/10.1101/2024.06.28.24307535 Ayşe Sıla Dokumacı 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Ayşe Sıla Dokumacı For correspondence: ayse.dokumaci{at}kcl.ac.uk Katy Vecchiato 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 3 Department of Forensic and Neurodevelopmental Sciences, Institute of Psychiatry, Psychology and Neuroscience, King’s College London , London, United Kingdom 4 Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Katy Vecchiato Raphael Tomi-Tricot 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 5 MR Research Collaborations, Siemens Healthcare Limited , Frimley, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site Michael Eyre 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Michael Eyre Philippa Bridgen 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Philippa Bridgen Pierluigi Di Cio 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site Chiara Casella 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 4 Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Chiara Casella Tobias C. Wood 7 Department of Neuroimaging, Institute of Psychiatry, Psychology and Neuroscience, King’s College London , London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Tobias C. Wood Jan Sedlacik 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 8 Robert Steiner MR Unit, Medical Research Council Laboratory of Medical Sciences, Hammersmith Hospital Campus , Du Cane Road, London, UK 9 Mansfield Centre for Innovation, Imaging Sciences, Institute of Clinical Sciences, Imperial College London, Hammersmith Hospital Campus , Du Cane Road, London, UK Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Jan Sedlacik Tom Wilkinson 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site Sharon L. Giles 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 10 Guy’s and St Thomas’ NHS Foundation Trust , London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site Joseph V. Hajnal 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 4 Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Joseph V. Hajnal Jonathan O’Muircheartaigh 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom 3 Department of Forensic and Neurodevelopmental Sciences, Institute of Psychiatry, Psychology and Neuroscience, King’s College London , London, United Kingdom 4 Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 11 MRC Centre for Neurodevelopmental Disorders, King’s College London , London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Jonathan O’Muircheartaigh Shaihan J. Malik 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for Shaihan J. Malik David W. Carmichael 1 Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London , London, United Kingdom 2 London Collaborative Ultra high field System (LoCUS), London, United Kingdom Find this author on Google Scholar Find this author on PubMed Search for this author on this site ORCID record for David W. Carmichael Abstract Full Text Info/History Metrics Supplementary material Data/Code Preview PDF Abstract Introduction Quantitative MRI is important for non-invasive tissue characterisation. In previous work we developed a clinically feasible multi-contrast protocol for T 1 -weighted imaging based on the MP2RAGE sequence that was optimised for both children and adults. It was demonstrated that a range of Fluid And White Matter Suppression (FLAWS) related contrasts could be produced while maintaining T 1 -weighted uniform image (UNI) quality, a challenge at higher field strengths. Here we introduce an approach to use these images to calculate effective proton density (PD*) and quantitative T 1 relaxation maps especially for shorter repetition times (TR MP2RAGE ) than those typically used previously. Methods T 1 and PD* were estimated from the analytical equations of the MP2RAGE signal derived for partial Fourier acquisitions. The sensitivity of the fitting results was evaluated with respect to the TR MP2RAGE and B 1 + effects on both excitation flip angles and inversion efficiency and compared to vendor T 1 maps which do not use B 1 + information. Data acquired for a range of individuals (aged 10-54 years) at the shortest TR MP2RAGE (4000ms) were compared across white matter (WM), cortical grey matter, and deep grey matter regions. Results The T 1 values were insensitive to the choice of different TR MP2RAGE . The results were similar to the vendor T 1 maps if the B 1 + effects on the excitation flip angle and inversion efficiency were not included in the fits. T 1 values varied over development into adulthood, especially for the deep grey matter regions whereas only a very small difference was observed for WM T 1 . Effective PD maps were produced which did not show a significant difference between children and adults for the age range included. Conclusion We produced PD* maps and improved the accuracy of T 1 maps from an MP2RAGE protocol that is optimised for UNI and FLAWS-related contrasts in a single scan at 7T by incorporating the excitation flip angle and inversion efficiency related effects of B 1 + in the fitting. This multi-parametric protocol made it possible to acquire high resolution images (0.65mm iso) in children and adults within a clinically feasible duration (7:18 min:s). The combination of analytical equations utilizing B 1 + maps led to T 1 fits that were consistent at different TR MP2RAGE values. Average WM T 1 values of adults and children were very similar (1092ms vs 1117ms) while expected reductions in T 1 with age were found for GM especially for deep GM. 1 Introduction Quantitative MRI has an important role in the measurement of tissue microstructure and its alteration as part of healthy development and ageing ( Deoni, 2010 ; Eminian et al., 2018 ; Lutti et al., 2014 ) and in a wide range of pathology ( Adler et al., 2017 ; McDowell et al., 2018 ). Many different approaches exist for longitudinal relaxation constant (T 1 ) mapping (A. G. Teixeira et al., 2020 ; Brix et al., 1990 ; Christensen et al., 1974 ; Crawley & Henkelman, 1988 ; Deichmann & Haase, 1992 ; Deoni, 2007 ; Deoni et al., 2003 ; Frahm et al., 1986 ; Fram et al., 1987 ; Graumann et al., 1986 ; Gupta, 1977 ; Hahn, 1950 ; Helms et al., 2008 ; Henderson et al., 1999 ; Homer & Beevers, 1985 ; Leipert & Marquardt, 1976 ; Look & Locker, 1968 , 1970 ; Lutti et al., 2014 ; Ma et al., 2023 ; Wright et al., 2008 ); however, a fast and widely used approach at 7 Tesla (T) for human brain is the Magnetization Prepared 2 Rapid Gradient Echoes (MP2RAGE) ( Marques et al., 2010 ) sequence. The MP2RAGE sequence ( Marques et al., 2010 ), which is an extension of the Magnetization Prepared RApid Gradient Echo (MPRAGE) sequence ( Marques et al., 2010 ; Mugler & Brookeman, 1990 ; Van de Moortele et al., 2009 ), has been used at 7T to produce 3D structural T 1 -weighted brain images and T 1 maps. This sequence acquires two Gradient Echo (GRE) images (INV1 and INV2) at two different inversion times (TI 1 and TI 2 ) following a non-selective adiabatic inversion pulse where the time between two consecutive inversion pulses is called TR MP2RAGE ( Marques et al., 2010 ). These two images are combined to form the uniform T 1 -weighted image (UNI) which is insensitive to receive field (B 1- ) variability, proton density (PD), and T 2 * effects ( Marques et al., 2010 ; Van de Moortele et al., 2009 ). T 1 maps are typically produced using the UNI signal from look up tables (LUTs) which assume uniform B 1 + ( Marques et al., 2010 ). Nonetheless, derived images and quantitative maps are still sensitive to transmit field (B1 + ) inhomogeneity effects ( Marques et al., 2010 ). Even though the MP2RAGE scan parameters can be optimised to reduce the B1 + effects, some residual B 1 + bias is likely to be present in the T 1 maps ( Marques & Gruetter, 2013 ; Van de Moortele et al., 2009 ). To overcome this problem, Marques et al. ( Marques & Gruetter, 2013 ) employed the MP2RAGE sequence( Marques et al., 2010 ) and SA2RAGE sequence ( Eggenschwiler et al., 2012 ) that is used to calculate B1 + values assuming a single T 1 value over the whole brain together ( Marques & Gruetter, 2013 ). Two 2D look up tables relating either the T 1 value to the MP2RAGE signal intensity and the B 1 + value or the B 1 + value to the MP2RAGE signal intensity and the T 1 value were utilized iteratively to estimate both T 1 and B 1 + ( Marques & Gruetter, 2013 ). The B 1 + effect was considered only for the excitation pulses and not for the adiabatic inversion pulse ( Marques & Gruetter, 2013 ). A study investigating the inter-site variability compared two datasets one of which was acquired with a low B 1 + -sensitive protocol like the one by Marques and Gruetter ( Marques & Gruetter, 2013 ) while using a parallel transmit head coil for improved B 1 + homogeneity ( Haast et al., 2021 ). This demonstrated that B 1 + inhomogeneity affects both the T 1 values ( Marques & Gruetter, 2013 ) and segmentation and cortical surface calculation ( Haast et al., 2018 ), with differing values with and without B 1 + inhomogeneity correction ( Haast et al., 2021 ). The MP2RAGE sequence can also be optimised to produce GM-dominant FLuid And White Matter Suppression (FLAWS) images ( Beaumont et al., 2019 , 2021 ; Martin et al., 2023 ; Massire et al., 2021 ; Müller et al., 2022 ; Tanner et al., 2012 ; Urushibata et al., 2019 ), which can be beneficial for deep GM visualisation and potentially improve lesion detection in multiple sclerosis and focal cortical dysplasia ( Chen et al., 2018 ; Martin et al., 2023 ; Massire et al., 2021 ; Müller et al., 2022 ). It would be advantageous to be able to obtain T 1 maps from these protocols, but T 1 estimation is challenging due to the low grey matter (GM) signal in the bias-reduced UNI image which is used to calculate T 1 unless additional bias-reduced FLAWS-related images such as the FLAWS hc is used for this purpose ( Beaumont et al., 2021 ). A further clinically relevant quantitative value is proton density (PD) which is defined as the number of MR visible hydrogen protons in tissue ( Mezer et al., 2016 ; Tofts, 2003 ). PD values have been shown to be altered in multiple sclerosis ( Engström et al., 2014 ; Gracien et al., 2017 ), brain oedema in hepatic encephalopathy ( Shah et al., 2008 ), and peritumoral oedema in malignant gliomas ( Blystad et al., 2017 ). Quantitative PD maps can be obtained from MP2RAGE protocols; however, this has not been performed to the best of our knowledge. A number of studies have proposed to obtain T 1 maps by adding an additional GRE block to the MP2RAGE sequence ( Marques et al., 2010 ) named as MP3RAGE: Hung et al. ( Hung et al., 2013 ) determined T 1 by fitting the inversion efficiency ( eff ) of the pulse as a free parameter if the pulse profile information was not available; Rioux et al. ( Rioux et al., 2014 ) obtained T 1 and flip angle (FA) information by using three different combinations of the images acquired at different inversion times using a LUT as in Marques et al. ( Marques et al., 2010 ); and Olsson et al. ( Olsson et al., 2022 ) fitted T 1 and B 1 + simultaneously by using closed-form approximations of the signal equations for small FAs while assuming an eff value of 0.96. In this study, we aimed to determine if high quality T 1 and effective PD (PD*) maps could be generated from a 7T protocol with shorter than conventional repetition times (TR MP2RAGE ), previously optimised to generate UNI and FLAWS-related contrasts in a single scan ( Dokumacı et al., 2023 ). The term effective PD has been used because the T 2 * effects have not been considered( Cercignani et al., 2018 ; Weiskopf et al., 2013 ). To achieve this aim, analytical equations that account for partial Fourier acquisitions were derived and fitting was performed using the analytical equations in combination with B 1 + maps while investigating the effects of B 1 + both on the excitation flip angles and the inversion efficiency ( eff) . Results using protocols with different TR MP2RAGE values were compared. Data from children and adults using the short TR MPRAGE (=4000ms) were acquired and changes in T 1 with age were investigated. 2 Methods 2.1 Theory The time between two consecutive inversion pulses in an MP2RAGE sequence, TR MP2RAGE , includes two GRE blocks, the delay (TA) following the first inversion pulse until the start of the first GRE block, the delay (TB) between the two GRE blocks, and the remaining time (TC) following the second GRE block until the application of the next inversion pulse ( Marques et al., 2010 ). Following the approach of Marques et al.( Marques et al., 2010 ), the MP2RAGE steady-state signal for the longitudinal magnetization and the modified signal equations accounting for the partial Fourier acquisition for the GRE blocks were derived as( Dokumaci AS, 2022 ): In Eqn. (1) , EA = exp(-TA/T 1 ), EB = exp(-TB/T 1 ), EC = exp(-TC/T 1 ), and E1 = exp(-TR GRE /T 1 ). TR GRE is the time between the small flip angle excitations in the GRE blocks (n in total). The efficiency of the inversion pulse is denoted as eff and defined as ( Marques et al., 2010 ) 𝑀 𝑧,𝑖𝑛𝑣 = −𝑒𝑓𝑓𝑀 𝑧 (0) where M z,inv is the inverted longitudinal magnetisation and M z (0) is the initial longitudinal magnetisation. Equations ( 2 ) and ( 3 ) account for the partial Fourier factor of 6/8 in the first phase encoding (partition) direction, acquired in the innermost k-space loop. This results in a shift of the k-space centre from the middle of the n phase encoding blocks to (n/3 + 1) in the sequence implementation. To fit T 1 and M 0 (PD) a nonlinear least-squares solver ( lsqnonlin ) in MATLAB (R2022b, The Mathworks, Natick, Massachusetts) was used to minimise the cost function: for f 1 = s 1 - PD*k 1 and f 2 = s 2 - PD*k 2 where s 1 and s 2 are the measured GRE TI1 and GRE TI2 signals, respectively with the theoretical values of PD*k 1 and PD*k 2 from Eqns. (2) and (3). The term seq represents the imaging parameters TR GRE , TI 1 , TI 2 , TR MP2RAGE , n, α 1 , α 2 whose values are known. The terms k 1 and k 2 are not functions of PD but depend on T 1 and the sequence parameters. The fitting “options” for the lsqnonlin function were MaxIterations = 100000; MaxFunctionEvaluations = 400000; FunctionTolerance = 1e-20; FiniteDifferenceStepSize = 1e-2; and OptimalityTolerance = 1e-10. The T 1 search range was 200ms to 30000ms while for the unscaled PD 0 to 100000 was used. The initial values for T 1 and PD were 1500ms and 20000, respectively. 2.2 Different Fits for Calculating T 1 To explore the weighting of the two likely factors contributing to spatial effects (inversion efficiency ( eff ) and B 1 + variability) different fits were performed using the nonlinear least-squares algorithm: Fit 1 excluded B 1 + information in the excitation flip angles (α 1 and α 2 ) and assumed eff =1 throughout the whole 3D image. Fit 2 included B 1 + information in the excitation flip angles (α 1 and α 2 ) but still assumed eff =1. Lastly, Fit 3 included both the B 1 + information and variable eff . For all fits, the phase information was used to determine the polarity of the magnetization due to T 1 recovery at the first inversion time (TI 1 ). The effect of TR MP2RAGE on the fits was investigated. In addition, T 1 maps from the vendor which are determined from a LUT without using B 1 + information ( Marques et al., 2010 ) were included for comparison. The function colorbarpzn ( He, 2024 ) running in MATLAB was used to create the blue-white-red colormaps and colorbars. 2.3 Data Acquisition All data was obtained under institutional ethical approval (REMAS 8700 and 18/LO/1766) and following informed consent. For all scans, a MAGNETOM Terra 7T system (Siemens Healthcare, Erlangen, Germany) using a single transmit (1Tx/32Rx) coil (Nova Medical) was employed. The summary of scan protocols is given in Table 1 . Most of the data were acquired using the protocol optimized in our previous study ( Dokumacı et al., 2023 ) that confers the benefits of short scan time, high resolution, and high-quality FLAWS and UNI contrast options from a single scan. Firstly, data were obtained with different TR MP2RAGE values in a small group of 4 healthy adults (32±2 years, 1f) to assess the invariance of the T 1 estimate to the TR MP2RAGE ( Table 1a ). Secondly, data from 7 healthy children (12±2 years, 2f) and 2 more adults making it 6 adults in total (35±9 years, 1f) were obtained to determine if expected age-related T 1 changes could be observed; TR MP2RAGE =4000ms was used for all the scans in Table 1b except for Protocol 3 which had a low sensitivity to B 1 + with a longer TR MP2RAGE ( Marques et al., 2010 ) (TR MP2RAGE =8000ms). View this table: View inline View popup Download powerpoint Table 1. TI 1 /TI 2 is the time between the inversion pulse and the centre of k-space for the 1 st /2 nd GRE block ( Marques et al., 2010 ). Parallel imaging was employed using GRAPPA ( Griswold et al., 2002 ). a) Scans obtained in 4 healthy adults: a partial Fourier factor of 6/8 was used in both phase encoding directions. b) Scan parameters applied in healthy children and adults. Protocol 3 had low sensitivity to B 1 + but a longer TR MP2RAGE . A partial Fourier factor of 6/8 was used only in one direction (slice partial Fourier) for Protocols 2 and 3. To determine the effect of B 0 inclusion on the eff of the inversion pulse, a B 0 map (GRE sequence with 3 echoes TE 1 /TE 2 /TE 3 = 1.02ms/2.26ms/4.08ms and TR=10ms) was acquired in one subject. In addition to the B 0 map, MP2RAGE data using 2 different protocols (Protocols 2 and 3) and 2 different B 1 + maps (sat_tfl and AFI ( Yarnykh, 2007 )) were obtained where sat_tfl refers to the B 1 + method that uses a slice-selective preconditioning pulse in combination with a turbo FLASH readout ( Chung et al., 2010 ; Fautz et al., 2008 ). 2.4 T 1 and PD* estimation and analysis Data preprocessing consisted of the following steps performed in MATLAB (R2022b, The Mathworks, Natick, Massachusetts): 1) using the FLAWS hco images a brain mask was produced to limit the fit to brain and surrounding cerebrospinal fluid, 2) B 1 + map processing, 3) B 1 - bias field calculation for PD* map generation, and 4) determining the inversion efficiency ( eff ) maps which are explained in detail below. 2.4.1 Producing the brain mask FLAWS hco images were generated using the INV1 and INV2 images ( Beaumont et al., 2021 ; Dokumacı et al., 2023 ; Tanner et al., 2012 ) and converted to NIfTI using the dicm2nii function ( Li et al., 2016 ). A brain mask was calculated from a segmentation performed on the FLAWS hco image in SPM12 ( Penny et al., 2007 ). Regions were assigned to be within the brain if they had probability of p >0.99 of being either white matter (WM), grey matter (GM), or cerebrospinal fluid (CSF). The resulting binary mask was edited to produce a closed mask with the application of the imclose and imfill morphological operators in MATLAB. 2.4.2 B 1 + map processing B 1 + maps were smoothed using the hMRI toolbox ( Tabelow et al., 2019 ) and co-registered and resliced using the FLAWS hco image ( Beaumont et al., 2021 ; Dokumacı et al., 2023 ; Tanner et al., 2012 ) as the reference and the B 1 + magnitude images as the source in SPM12 ( Penny et al., 2007 ). The relative B 1 + values for each voxel were multiplied by the nominal flip angles to define the actual flip angles subsequently used in the fitting process. 2.4.3 B 1 - related bias field calculation Maps of the B 1 - related signal bias in the PD* fits were produced using SPM12( Penny et al., 2007 ) using the following parameters in the Segment module: extremely light regularisation (0.00001), Bias FWHM of 30 mm cutoff, and 3 different tissue types were chosen because the fits were masked. 2.4.4. Determining the inversion efficiency (eff) maps The adiabatic inversion pulse was a hyperbolic secant ( Baum et al., 1985 ; Silver et al., 1984 ) whose eff was determined using Bloch simulations for each spatial location for all subjects independently using the RF pulse information exported from the DICOM files in combination with the B 1 + maps and, where available, B 0 maps. 2.5 Data Analysis in FreeSurfer To determine the T 1 and PD* values for different brain regions, UNI images were segmented in FreeSurfer 7.3 ( Fischl, 2012 ). For this purpose, the UNI images were masked with the same closed masks that were used in the fitting process. Masked images were denoised using the DenoiseImage function ( Manjón et al., 2010 ) in Advanced Normalization Tools (ANTs)( Avants et al., 2009 ). A variant of the N3 Bias Correction algorithm( Sled et al., 1997 , 1998 ), N4 Bias Correction algorithm ( N4BiasFieldCorrection function) ( Tustison et al., 2010 ) implemented in the ANTs ( Avants et al., 2009 ) package was used with the default parameters (spline spacing b = 180; convergence c = [50x50x50x50, 0]). The masks were part of the inputs for both functions. Images were reoriented to standard orientation using the fslreorient2std function in FSL (FMRIB Software Library) ( Jenkinson et al., 2012 ; Smith et al., 2004 ). These images were segmented using the recon-all function ( Collins et al., 1994 ; Dale et al., 1999 ; Fischl, 2004 ; Fischl et al., 2001 , 2002 ; Fischl, Sereno, & Dale, 1999 ; Fischl, Sereno, Tootell, et al., 1999 ; Fischl & Dale, 2000 ; Non-Uniform Intensity Correction . http://Www.Bic.Mni.Mcgill.ca/Software/N3/Node6.Html , n.d.; Sled et al., 1997 , 1998 ) of FreeSurfer( Fischl, 2012 ) with -all (to perform all including subcortical segmentations) and -cm (to conform to the maximum resolution) options. All segmentations were checked via visual assessment by using the Subcortical Automatic Segmentation (aseg) and Automatic Cortical Parcellation (aparc) files produced by FreeSurfer overlaid on original T 1 -weighted images. If the UNI image segmentation was not successful, the steps above were applied to the FLAWS hco images to be used as the FreeSurfer input. In one subject, parameters used for all other subjects did not result in a good segmentation and therefore were adjusted - a spline spacing of b = 100 and convergence c = [1000, 0] were used to get an improved result. The intersection of the FreeSurfer and SPM masks were used for WM while adding another constraint to include the values with T 1 ≤ 2000 ms to avoid contamination from CSF due to imperfect segmentation; for cortex the additional constraint was T 1 ≤ 2500 ms. For deep GM regions, only the FreeSurfer segmentations were used because of the high constraint in the SPM segmentation ( p >0.99) leading to region of interests including very few voxels (in the extreme cases without any voxels) but, like the cortex, voxels with T 1 ≤ 2500 ms were included in the masks. Regional T 1 mean values and standard deviations (SD) were calculated for the group of subjects who had the TR MP2RAGE =4000ms protocols to see how the values within different brain regions change by age and to establish values in children at 7T. 3 Results The pulse profile with the inversion efficiency simulated using the B 0 and B 1 + maps from the scan where Protocols 2 and 3 were compared are given in Figure 1 . As seen in Figure 1a-b , the inversion pulse is very efficient for a large range of B 1 + and off-resonance values. For this slice, the inversion efficiency mean value ± standard deviation was 0.995 ± 0.023 with a minimum value of 0.477. Without the inclusion of B 0 information in the inversion efficiency calculations, the inversion efficiency for this slice was simulated as 0.995 ± 0.023 with a minimum of 0.478. Inversion efficiency over the 3D volume was calculated as 0.990 ± 0.048 and 0.990 ± 0.047 with and without the inclusion of the B 0 map, respectively. Download figure Open in new tab Figure 1. a) Pulse profile using a range of B 0 and B 1 + values for this pulse which is insensitive to a large range of B 0 and B 1 + values b) Inversion Efficiency ( eff ) map using c) B 0 map from a slice acquired in this scan and d) the corresponding B 1 + map (sat_tfl) e) the corresponding UNI image which shows a big change at the back of the cerebellum exactly where the eff deviates from 1. The differences between the vendor T 1 map which is based on a LUT that does not use B 1 + information and the fit that uses B 1 + information demonstrated the expected spatial effect in T 1 due to the spatial variation in the B 1 + field ( Fig. 2 ). Different fits (Fits 1-3) were performed to investigate the contribution of the eff and B 1 + variability to these spatial effects. Download figure Open in new tab Figure 2. Vendor T 1 map (a), the fit (b), and their difference for WM and GM (c) demonstrating the spatial effect related to B 1 + non-uniformity. Figure 3 shows the histograms of WM and cortex from the data of one adult subject using Protocol 1 for these different fits and the vendor T 1 map. Supplementary Figure 1 shows the histograms from the deep GM regions. Table 2 lists the means and standard deviations for 4 adults including this subject (Adult 1) for WM and cortex. Download figure Open in new tab Figure 3. WM and cortex histograms for the vendor T 1 map and different fits in one adult (Adult 1). Fit 1 does not consider the B 1 + information like the vendor map. Fit 2 considers the B 1 + information but assumes the inversion efficiency to be uniform ( eff = 1) for each pixel. Fit 3 uses both the B 1 + and eff maps for each pixel. Fit 1 produced the closest results to those of the vendor map. The results from Fits 2 and 3 were similar because of the high inversion efficiency of the adiabatic inversion pulse ( eff = 0.994 ± 0.028 over all subjects). View this table: View inline View popup Download powerpoint Table 2. Means and SDs for WM and cortex T 1 values (ms) using different fits for 4 adults (Protocol 1). As in Fig.3 , the vendor map results are closest to those obtained with Fit 1 for all subjects. Likewise, Fits 2 and 3 provided very similar results due to the high inversion efficiency of the adiabatic pulse. In all subjects, the mean and median values of the vendor map and Fit 1 were the closest as both ignored the B 1 + information. For this highly efficient hyperbolic secant adiabatic inversion pulse with an eff of 0.994 ± 0.028 over 13 subjects, the results from Fit 2 and Fit 3 were very similar. Supplementary Table 1 shows the T 1 results using different fits in deep GM regions where a similar trend is observed. To investigate the effect of TR MP2RAGE on T 1 fits, 4 different TR MP2RAGE protocols ( Table 1a ) were compared using the Fit 3 estimation (accounting for B 1 + effects on both the excitation flip angles and the eff ). In Figure 4 , the T 1 means and standard deviations for different brain regions with different TR MP2RAGE values are shown. The standard deviation at each TR MP2RAGE corresponds to the physiological variation of the T 1 over the whole brain. The horizontal dotted lines correspond to the mean across TR MP2RAGE . The mean values and the uncertainty of the average using these different protocols are given above the lines. The uncertainty of the average is very small for each brain region in all adults which indicates the high precision of the T 1 fits across TR MP2RAGE . Download figure Open in new tab Figure 4. T 1 measurements using Fit 3 and protocols with different TR MP2RAGE values in 4 adults. The horizontal dotted lines correspond to the mean of the 4 measurements and the mean and uncertainty of the average is given above the line which indicates the small effect of these TR MP2RAGE on the fits. TR MP2RAGE 5000 and TR MP2RAGE 5000* correspond to the protocols 5000 (1) and 5000 (2) in Table 1a , respectively. After investigating the effect of TR MP2RAGE on the fits, data acquired at TR MP2RAGE =4000ms were compared in children and adults. Figure 5 shows the results from different protocols (see Table 1b for protocol summaries) for WM, cortex and deep GM structures in children and adults. The difference between children and adult T 1 s was significant for GM regions especially for caudate and putamen. WM values were very similar for both populations. Download figure Open in new tab Figure 5. T 1 fits (Fit 3) using different protocols (triangles correspond to Protocol 1 and disks correspond to Protocol 2 in Table 1b ) with the same TR MP2RAGE value (4000ms) in 7 children and 6 adults for different brain regions. The T 1 values of the children are generally higher compared to those of the adults for the GM regions especially for caudate and putamen. WM values of the children (12 ± 2 years) and adults (35 ± 9 years) are very similar. The plots for PD* values in different brain regions using different protocols in children and adults are shown in Figure 6 . There is not a prominent difference between children and adult PD* values. Table 3 summarizes the means and standard deviations for T 1 and PD* values of the children and adult populations. Download figure Open in new tab Figure 6. PD* values (Fit 3) using different protocols (triangles correspond to Protocol 1 and disks correspond to Protocol 2 in Table 1b ) with the same TR MP2RAGE value (4000ms) in 7 children and 6 adults for different brain regions. There is not a large difference between the children and adult PD* values although thalamus PD* values suggest a very slight decrease by age. The PD* values were scaled with respect to the WM mean which was assumed to be 0.69( Weiskopf et al., 2013 ). No T 2 * correction was made which would include a similar multiplication factor for WM and GM. View this table: View inline View popup Download powerpoint Table 3. Means and standard deviations (SD) of T 1 , PD*, and B 1 + in different brain regions of children and adults. The standard deviations reflect the physiological variation of tissue T 1 across the brain. The difference between the T 1 values of children and adults is significant for GM regions while the WM T 1 values are very close. The PD* results of children and adults are very similar as well. T 1 values were fitted (Fit 3) over the 3D image using 2 different B 1 + maps for Protocol 2 (TR MP2RAGE =4000ms) and it was also compared with Protocol 3 (TR MP2RAGE =8000ms) which is highly insensitive to B 1 + changes. The B 1 + sensitivities of all protocols used in this study are given in Supplementary Figure 2. The B 1 + maps (sat_tfl and AFI) are shown in Supplementary Figure 3 for a slice with the corresponding T 1 fits using Protocol 2. Figure 7 demonstrates representative images and parameters maps that were produced using a single acquisition in under 7.5 minutes from a healthy child. The vendor T 1 map and the T 1 fit are displayed in the same range (0 to 4095 ms). The PD* map was scaled such that the WM mean would be equal to 0.69. No T 2 * correction was made for the PD* images. The T 2 * effect is expected to be similar for WM and GM. Download figure Open in new tab Figure 7. Different contrasts and parameter maps at 0.65 mm isotropic resolution were generated using a single MP2RAGE acquisition in 7:18 (min:s) from a healthy child. The vendor T 1 map and the T 1 fit are displayed in the same T 1 range (0 to 4095 ms). The PD* image has not been corrected for T 2 * effects which would be a small factor due to the short TE (around 3 ms) and similar for WM and GM. 4 Discussion and Conclusion In this study, we improved the accuracy of T 1 maps and derived PD* images produced from our multi-contrast protocol that is optimised for UNI and FLAWS-related contrasts in a single scan ( Dokumacı et al., 2023 ) at 7T by including B 1 + effects for both the excitation flip angles and the inversion pulse in the fitting. This protocol made it possible to acquire high resolution images (0.65mm iso) in children and adults thanks to its short scan duration (7:18 min:s). T 1 fitting results were consistent at different TR MP2RAGE values (4000 ms, 4500 ms, and 5000 ms) for different brain regions (See Fig. 4 ). The fits resulted in a mean adult WM T 1 value of 1092 ± 63 ms which was slightly lower than the values found in literature with the mean range of 1100-1400 ms ( Marques et al., 2010 ; Marques & Norris, 2018 ; Wright et al., 2008 ). Mean T 1 WM value for children (12 ±2 years) was 1117 ± 74 ms. This result being very similar to the adult T 1 is not surprising as the myelination is relatively mature after 2-3 years of age with much smaller amplitude development into adulthood ( Barkovich, 2000 ). Cortex and deep GM regions generally had slightly lower values compared to the literature adult T 1 values ( Caan et al., 2019 ) although they were similar and demonstrates the same pattern of lower T 1 values in deep GM regions such as the thalamus. It was observed that the GM T 1 values of children were higher than those of adults especially in deep GM regions such as caudate and putamen. This was consistent with the expected reductions in T 1 with age from literature ( Cho et al., 1997 ; Gracien et al., 2017 ). Cho and colleagues ( Cho et al., 1997 ) demonstrated at 1.5T that T 1 vs age relationship follows a quadratic curve with minimum T 1 values observed for deep GM between ages of 38-48 while the inflection in the curve does not happen before 60 years for cortex. Figure 5 shows similar trends for deep GM and cortex with a higher T 1 in the younger subjects. No significant difference was observed between the mean PD* values of children and adults ( Figure 6 and Table 3 ). This is similar to the results from previous studies for this age range ( Hagiwara et al., 2021 ; Saito et al., 2012 ). The PD* values in the thalamus ( Figure 6 ) hints a very small decrease with age, but a larger study is required to confirm this result. Furthermore, it is possible that these small changes are related to small residual T 2 *-related changes in the PD* measure, despite the relatively short TE. A study by Callaghan and colleagues( Callaghan et al., 2014 ) found negative correlations between effective proton density and age in the putamen, pallidum, caudate nucleus, and the red nucleus with an echo time of 8.45ms at 3T. It was confirmed that the B 1 + has a large effect on the T 1 values as found by Haast and colleagues ( Haast et al., 2021 ) which necessitates future inclusion of the B 1 + in T 1 fits. This is particularly relevant where the protocol is sensitive to B 1 + variations at shorter TR MP2RAGE values which are a consequence of the requirements of short scan durations and high-resolution. The comparison of the B 1 + corrected high-resolution images (0.65 mm isotropic) to lower-resolution images (1 mm isotropic) acquired with a low-B 1 + -sensitive protocol by Marques and Gruetter ( Marques & Gruetter, 2013 ) revealed that by using any suitable B 1 + map, the residual effects of B 1 + variability on T 1 maps can be removed. The accuracy of the B 1 + map is important as it might affect the T 1 results significantly for protocols that are more sensitive to B 1 + variations. T 1 values fitted using two different B 1 + mapping methods in one subject were comparable ( Table 4 and Supplementary Figure 3). It could be beneficial to use protocols that are less sensitive to B 1 + variations while satisfying the short scan duration criterion for instance by combining CS methods ( Candes & Wakin, 2008 ; Donoho, 2006 ; Lustig et al., 2007 ; Mussard et al., 2020 ; Puy et al., 2012 ; Trotier et al., 2019 , 2022 ; Vasanawala et al., 2010 ) with longer TR MP2RAGE values. View this table: View inline View popup Download powerpoint Table 4. T 1 value comparison (Fit 3) in the same subject using different B 1 + maps (sat_tfl and AFI) and different protocols. Protocol 3 (TR MP2RAGE =8000ms) is less sensitive to the variations in B 1 + . Quantitative results are also affected by the accuracy of the segmentations. We used an additional precaution to prevent contamination by CSF voxels by limiting the masks for T 1 values ≤ 2000 ms for WM and ≤ 2500 ms for GM. T1 restriction was found to have a negligible effect on all the values except for hippocampus which is an area more prone to segmentation errors (Supplementary Table 2). One limitation of our study was that the B 0 map was available for only one subject. However, due to the very high inversion efficiency of this hyperbolic secant adiabatic pulse ( eff = 0.994 ± 0.028 over all subjects), the simulated B 0 effect in the inversion efficiency using the map acquired in this subject was negligible. As seen in Figure 1 , the efficiency deviates from 1 when B 1 + is very low and it is not so related to B 0 . No separate correction was made for the exponential term −𝑇𝐸/𝑇∗ but TE was very short (about 3 ms). Another limitation was that the signal equations assumed the conventional mono-exponential decay ( Labadie et al., 2014 ; Metere et al., 2017 ). In conclusion, it was possible to produce quantitative T 1 and PD* maps in children and adults at 7T in addition to the UNI and FLAWS-related contrasts from a single scan by using newly derived analytical equations required for partial Fourier acquisitions while incorporating the B 1 + inhomogeneity effect both on the excitation flip angles and the inversion efficiency. Data Availability The datasets that support the findings of this study are available upon request for any reasonable scientific purposes. Competing Interests Raphael Tomi-Tricot is an employee at Siemens Healthineers. Author Contributions A.S.D: Conceptualization, Methodology, Software, Validation, Formal Analysis, Investigation, Resources, Data Curation, Writing – Original Draft, Writing – Review & Editing, Visualization. K.V: Investigation, Resources, Data Curation, Writing – Review & Editing. R.T-T: Methodology, Software, Data Curation, Writing – Review & Editing. M.E: Software, Investigation, Resources, Data Curation, Writing – Review & Editing. P.B: Investigation, Resources, Data Curation, Project Administration. P.D.C: Investigation, Resources, Data Curation. C.C: Investigation, Resources, Data Curation, Writing – Review & Editing. T.C.W: Methodology, Software, Validation, Writing – Review & Editing. J.S: Software, Data Curation, Writing – Review & Editing. T.W: Software, Data Curation. S.L.G: Resources, Writing – Review & Editing, Project Administration. J.V.H: Methodology, Validation, Writing – Review & Editing, Visualization. J.O.M: Methodology, Writing – Review & Editing, Visualization, Supervision, Project Administration, Funding Acquisition. S.J.M: Conceptualization, Methodology, Software, Validation, Formal Analysis, Writing – Review & Editing, Visualization, Supervision, Funding Acquisition. D.W.C: Conceptualization, Methodology, Validation, Formal Analysis, Investigation, Resources, Writing – Original Draft, Writing – Review & Editing, Visualization, Supervision, Project Administration, Funding Acquisition. Acknowledgments The authors would like to acknowledge David Leitão, Ronald Mooiweer, and Oral Ersoy Dokumaci for valuable discussions. This research was supported by GOSHCC Sparks Grant V4419, King’s Health Partners, in part by the Medical Research Council (UK) (grants MR/ K006355/1 and MR/LO11530/1) and Medical Research Council Center for Neurodevelopmental Disorders, King’s College London (MR/N026063/1), and by core funding from the Wellcome EPSRC Centre for Medical Engineering at King’s College London [WT203148/Z/16/Z]. J.O.M, K.V, and C.C were funded by a Sir Henry Dale Fellowship jointly by the Wellcome Trust and the Royal Society (206675/Z/17/Z). C.C was also funded by a grant from GOSHCC (VC1421). M.E was funded by Action Medical Research (GN2835) and the British Paediatric Neurology Association. This research was funded in whole, or in part, by the Wellcome Trust [WT203148/Z/16/Z and 206675/Z/17/Z] and by the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ NHS Foundation Trust and King’s College London and/or the NIHR Clinical Research Facility. The views expressed are those of the author(s) and not necessarily those of the NHS, the NIHR or the Department of Health and Social Care. For the purpose of open access, the author has applied a CC BY public copyright licence to any Author Accepted Manuscript version arising from this submission. References ↵ A. G. Teixeira , R. P. , Neji , R. , Wood , T. C. , Baburamani , A. A. , Malik , S. J. , & Hajnal , J. V. ( 2020 ). Controlled saturation magnetization transfer for reproducible multivendor variable flip angle T 1 and T 2 mapping . Magnetic Resonance in Medicine , 84 ( 1 ), 221 – 236 . doi: 10.1002/mrm.28109 OpenUrl CrossRef ↵ Adler , S. , Lorio , S. , Jacques , T. S. , Benova , B. , Gunny , R. , Cross , J. H. , Baldeweg , T. , & Carmichael , D. W . ( 2017 ). Towards in vivo focal cortical dysplasia phenotyping using quantitative MRI . NeuroImage: Clinical , 15 , 95 – 105 . doi: 10.1016/j.nicl.2017.04.017 OpenUrl CrossRef PubMed ↵ Avants , B. B. , Tustison , N. , & Song , G . ( 2009 ). Advanced normalization tools (ANTS) . Insight j , 2 ( 365 ), 1 – 35 . OpenUrl ↵ Barkovich , A. J . ( 2000 ). Concepts of myelin and myelination in neuroradiology . AJNR. American Journal of Neuroradiology , 21 ( 6 ), 1099 – 1109 . OpenUrl PubMed Web of Science ↵ Baum , J. , Tycko , R. , & Pines , A . ( 1985 ). Broadband and adiabatic inversion of a two-level system by phase-modulated pulses . Physical Review A , 32 ( 6 ), 3435 – 3447 . doi: 10.1103/PhysRevA.32.3435 OpenUrl CrossRef PubMed ↵ Beaumont , J. , Gambarota , G. , Saint-Jalmes , H ., Acosta , O. , Ferré , J. , Raniga , P. , & Fripp , J. ( 2021 ). High- resolution multi-T 1 –weighted contrast and T 1 mapping with low sensitivity using the fluid and white matter suppression (FLAWS) sequence at 7T . Magnetic Resonance in Medicine , 85 ( 3 ), 1364 – 1378 . doi: 10.1002/mrm.28517 OpenUrl CrossRef ↵ Beaumont , J. , Saint-Jalmes , H. , Acosta , O. , Kober , T. , Tanner , M. , Ferré , J. C. , Salvado , O. , Fripp , J. , & Gambarota , G . ( 2019 ). Multi T1-weighted contrast MRI with fluid and white matter suppression at 1.5 T . Magnetic Resonance Imaging , 63 , 217 – 225 . doi: 10.1016/j.mri.2019.08.010 OpenUrl CrossRef ↵ Blystad , I. , Warntjes , J. B. M. , Smedby , Ö. , Lundberg , P. , Larsson , E.-M. , & Tisell , A . ( 2017 ). Quantitative MRI for analysis of peritumoral edema in malignant gliomas . PLOS ONE , 12 ( 5 ), e0177135 . doi: 10.1371/journal.pone.0177135 OpenUrl CrossRef PubMed ↵ Brix , G. , Schad , L. R. , Deimling , M. , & Lorenz , W. J . ( 1990 ). Fast and precise T1 imaging using a TOMROP sequence . Magnetic Resonance Imaging , 8 ( 4 ), 351 – 356 . doi: 10.1016/0730-725X(90)90041-Y OpenUrl CrossRef PubMed Web of Science ↵ Caan , M. W. A. , Bazin , P. , Marques , J. P. , de Hollander , G. , Dumoulin , S. O. , & van der Zwaag , W. ( 2019 ). MP2RAGEME: T 1 , T 2 *, and QSM mapping in one sequence at 7 tesla . Human Brain Mapping , 40 ( 6 ), 1786 – 1798 . doi: 10.1002/hbm.24490 OpenUrl CrossRef ↵ Callaghan , M. F. , Freund , P. , Draganski , B. , Anderson , E. , Cappelletti , M. , Chowdhury , R. , Diedrichsen , J. , FitzGerald , T. H. B. , Smittenaar , P. , Helms , G. , Lutti , A. , & Weiskopf , N . ( 2014 ). Widespread age-related differences in the human brain microstructure revealed by quantitative magnetic resonance imaging . Neurobiology of Aging , 35 ( 8 ), 1862 – 1872 . doi: 10.1016/j.neurobiolaging.2014.02.008 OpenUrl CrossRef PubMed ↵ Candes , E. J. , & Wakin , M. B . ( 2008 ). An Introduction To Compressive Sampling . IEEE Signal Processing Magazine , 25 ( 2 ), 21 – 30 . doi: 10.1109/MSP.2007.914731 OpenUrl CrossRef Web of Science ↵ Cercignani , M. , Dowell , N. G. , & Tofts , P. S. ( 2018 ). Quantitative MRI of the brain: principles of physical measurement. ↵ Chen , X. , Qian , T. , Kober , T. , Zhang , G. , Ren , Z. , Yu , T. , Piao , Y. , Chen , N. , & Li , K . ( 2018 ). Gray-matter-specific MR imaging improves the detection of epileptogenic zones in focal cortical dysplasia: A new sequence called fluid and white matter suppression (FLAWS) . NeuroImage: Clinical , 20 , 388 – 397 . doi: 10.1016/j.nicl.2018.08.010 OpenUrl CrossRef PubMed ↵ Cho , S. , Jones , D. , Reddick , W. E. , Ogg , R. J. , & Steen , R. G . ( 1997 ). Establishing norms for age-related changes in proton T1 of human brain tissue in vivo . Magnetic Resonance Imaging , 15 ( 10 ), 1133 – 1143 . doi: 10.1016/S0730-725X(97)00202-6 OpenUrl CrossRef PubMed Web of Science ↵ Christensen , K. A. , Grant , D. M. , Schulman , E. M. , & Walling , C . ( 1974 ). Optimal determination of relaxation times of fourier transform nuclear magnetic resonance. Determination of spin-lattice relaxation times in chemically polarized species . The Journal of Physical Chemistry , 78 ( 19 ), 1971 – 1977 . doi: 10.1021/j100612a022 OpenUrl CrossRef ↵ Chung , S. , Kim , D. , Breton , E. , & Axel , L . ( 2010 ). Rapid B 1 + mapping using a preconditioning RF pulse with TurboFLASH readout . Magnetic Resonance in Medicine , 64 ( 2 ), 439 – 446 . doi: 10.1002/mrm.22423 OpenUrl CrossRef PubMed ↵ Collins , D. L. , Neelin , P. , Peters , T. M. , & Evans , A. C . ( 1994 ). Automatic 3D intersubject registration of MR volumetric data in standardized Talairach space . Journal of Computer Assisted Tomography , 18 ( 2 ), 192 – 205 . OpenUrl CrossRef PubMed Web of Science ↵ Crawley , A. P. , & Henkelman , R. M . ( 1988 ). A comparison of one-shot and recovery methods in T1 imaging . Magnetic Resonance in Medicine , 7 ( 1 ), 23 – 34 . doi: 10.1002/mrm.1910070104 OpenUrl CrossRef PubMed ↵ Dale , A. M. , Fischl , B. , & Sereno , M. I . ( 1999 ). Cortical Surface-Based Analysis . NeuroImage , 9 ( 2 ), 179 – 194 . doi: 10.1006/nimg.1998.0395 OpenUrl CrossRef PubMed Web of Science ↵ Deichmann , R. , & Haase , A . ( 1992 ). Quantification of T1 values by SNAPSHOT-FLASH NMR imaging . Journal of Magnetic Resonance ( 1969 ) , 96 (3), 608–612. doi: 10.1016/0022-2364(92)90347-A OpenUrl CrossRef Web of Science ↵ Deoni , S. C. L . ( 2007 ). High-resolution T1 mapping of the brain at 3T with driven equilibrium single pulse observation of T1 with high-speed incorporation of RF field inhomogeneities (DESPOT1-HIFI) . Journal of Magnetic Resonance Imaging , 26 ( 4 ), 1106 – 1111 . doi: 10.1002/jmri.21130 OpenUrl CrossRef PubMed Web of Science ↵ Deoni , S. C. L . ( 2010 ). Quantitative Relaxometry of the Brain . Topics in Magnetic Resonance Imaging , 21 ( 2 ), 101 – 113 . doi: 10.1097/RMR.0b013e31821e56d8 OpenUrl CrossRef PubMed ↵ Deoni , S. C. L. , Rutt , B. K. , & Peters , T. M . ( 2003 ). Rapid combined T 1 and T 2 mapping using gradient recalled acquisition in the steady state . Magnetic Resonance in Medicine , 49 ( 3 ), 515 – 526 . doi: 10.1002/mrm.10407 OpenUrl CrossRef PubMed ↵ Dokumaci AS , V. K. T.-T. R. B. P. E. M. W. T. C. C. S. J. W. T. G. S. H. J. M. S. O. J. C. DW . ( 2022 ). Evaluation of quantitative T1 and PD mapping at 7T from an MP2RAGE Sequence optimised to obtain UNI and FLAWS contrast images in a single scan . In Proc. Intl. Soc. Magn. Reson. Med . (Vol. 30 , p. 1244 ). OpenUrl ↵ Dokumacı , A. S. , Aitken , F. R. , Sedlacik , J. , Bridgen , P. , Tomi-Tricot , R ., Mooiweer , R. , Vecchiato , K. , Wilkinson , T. , Casella , C. , Giles , S. , Hajnal , J. V. , Malik , S. J. , O’Muircheartaigh , J. , & Carmichael , D. W. ( 2023 ). Simultaneous Optimization of MP2RAGE T 1 -weighted (UNI) and FLuid And White matter Suppression (FLAWS) brain images at 7T using Extended Phase Graph (EPG) Simulations . Magnetic Resonance in Medicine , 89 ( 3 ), 937 – 950 . doi: 10.1002/mrm.29479 OpenUrl CrossRef ↵ Donoho , D. L . ( 2006 ). Compressed sensing . IEEE Transactions on Information Theory , 52 ( 4 ), 1289 – 1306 . doi: 10.1109/TIT.2006.871582 OpenUrl CrossRef Web of Science ↵ Eggenschwiler , F. , Kober , T. , Magill , A. W. , Gruetter , R. , & Marques , J. P . ( 2012 ). SA2RAGE: A new sequence for fast B 1 + -mapping . Magnetic Resonance in Medicine , 67 ( 6 ), 1609 – 1619 . doi: 10.1002/mrm.23145 OpenUrl CrossRef PubMed ↵ Eminian , S. , Hajdu , S. D. , Meuli , R. A. , Maeder , P. , & Hagmann , P . ( 2018 ). Rapid high resolution T1 mapping as a marker of brain development: Normative ranges in key regions of interest . PLOS ONE , 13 ( 6 ), e0198250 . doi: 10.1371/journal.pone.0198250 OpenUrl CrossRef ↵ Engström , M. , Warntjes , J. B. M. , Tisell , A. , Landtblom , A.-M. , & Lundberg , P . ( 2014 ). Multi-Parametric Representation of Voxel-Based Quantitative Magnetic Resonance Imaging . PLoS ONE , 9 ( 11 ), e111688 . doi: 10.1371/journal.pone.0111688 OpenUrl CrossRef PubMed ↵ Fautz , H. P. , Vogel , M. , Gross , P. , Kerr , A. , & Zhu , Y . ( 2008 ). B1 mapping of coil arrays for parallel transmission . Proceedings of the 16th Annual Meeting of ISMRM, Toronto, Canada , 1247 . ↵ Fischl , B . ( 2004 ). Automatically Parcellating the Human Cerebral Cortex . Cerebral Cortex , 14 ( 1 ), 11 – 22 . doi: 10.1093/cercor/bhg087 OpenUrl CrossRef PubMed Web of Science ↵ Fischl , B . ( 2012 ). FreeSurfer . NeuroImage , 62 ( 2 ), 774 – 781 . doi: 10.1016/j.neuroimage.2012.01.021 OpenUrl CrossRef PubMed Web of Science ↵ Fischl , B. , & Dale , A. M . ( 2000 ). Measuring the thickness of the human cerebral cortex from magnetic resonance images . Proceedings of the National Academy of Sciences , 97 ( 20 ), 11050 – 11055 . doi: 10.1073/pnas.200033797 OpenUrl Abstract / FREE Full Text ↵ Fischl , B. , Liu , A. , & Dale , A. M . ( 2001 ). Automated manifold surgery: constructing geometrically accurate and topologically correct models of the human cerebral cortex . IEEE Transactions on Medical Imaging , 20 ( 1 ), 70 – 80 . doi: 10.1109/42.906426 OpenUrl CrossRef PubMed Web of Science ↵ Fischl , B. , Salat , D. H. , Busa , E. , Albert , M. , Dieterich , M. , Haselgrove , C. , van der Kouwe , A. , Killiany , R. , Kennedy , D. , Klaveness , S. , Montillo , A. , Makris , N. , Rosen , B. , & Dale , A. M. ( 2002 ). Whole Brain Segmentation . Neuron , 33 ( 3 ), 341 – 355 . doi: 10.1016/S0896-6273(02)00569-X OpenUrl CrossRef PubMed Web of Science ↵ Fischl , B. , Sereno , M. I. , & Dale , A. M . ( 1999 ). Cortical Surface-Based Analysis . NeuroImage , 9 ( 2 ), 195 – 207 . doi: 10.1006/nimg.1998.0396 OpenUrl CrossRef PubMed Web of Science ↵ Fischl , B. , Sereno , M. I. , Tootell , R. B. , & Dale , A. M . ( 1999 ). High-resolution intersubject averaging and a coordinate system for the cortical surface . Human Brain Mapping , 8 ( 4 ), 272 – 284 . doi: 10.1002/(sici)1097-0193(1999)8:43.0.co;2-4 OpenUrl CrossRef PubMed Web of Science ↵ Frahm , J. , Haase , A. , & Matthaei , D . ( 1986 ). Rapid Three-Dimensional MR Imaging Using the FLASH Technique . Journal of Computer Assisted Tomography , 10 ( 2 ), 363 – 368 . doi: 10.1097/00004728-198603000-00046 OpenUrl CrossRef PubMed Web of Science ↵ Fram , E. K. , Herfkens , R. J. , Johnson , G. A. , Glover , G. H. , Karis , J. P. , Shimakawa , A. , Perkins , T. G. , & Pelc , N. J . ( 1987 ). Rapid calculation of T1 using variable flip angle gradient refocused imaging . Magnetic Resonance Imaging , 5 ( 3 ), 201 – 208 . doi: 10.1016/0730-725X(87)90021-X OpenUrl CrossRef PubMed Web of Science ↵ Gracien , R.-M. , Nürnberger , L. , Hok , P. , Hof , S.-M. , Reitz , S. C. , Rüb , U. , Steinmetz , H. , Hilker-Roggendorf , R. , Klein , J. C. , Deichmann , R. , & Baudrexel , S . ( 2017 ). Evaluation of brain ageing: a quantitative longitudinal MRI study over 7 years . European Radiology , 27 ( 4 ), 1568 – 1576 . doi: 10.1007/s00330-016-4485-1 OpenUrl CrossRef PubMed ↵ Graumann , R. , Fischer , H. , & Oppelt , A . ( 1986 ). A new pulse sequence for determining T 1 and T 2 simultaneously . Medical Physics , 13 ( 5 ), 644 – 647 . doi: 10.1118/1.595867 OpenUrl CrossRef PubMed ↵ Griswold , M. A. , Jakob , P. M. , Heidemann , R. M. , Nittka , M. , Jellus , V. , Wang , J. , Kiefer , B. , & Haase , A . ( 2002 ). Generalized autocalibrating partially parallel acquisitions (GRAPPA) . Magnetic Resonance in Medicine , 47 ( 6 ), 1202 – 1210 . doi: 10.1002/mrm.10171 OpenUrl CrossRef PubMed Web of Science ↵ Gupta , R. K . ( 1977 ). A new look at the method of variable nutation angle for the measurement of spin-lattice relaxation times using fourier transform NMR . Journal of Magnetic Resonance (1969 ) , 25 ( 1 ), 231 – 235 . doi: 10.1016/0022-2364(77)90138-X OpenUrl CrossRef ↵ Haast , R. A. M. , Ivanov , D. , & Uludağ , K . ( 2018 ). The impact of correction on MP2RAGE cortical T 1 and apparent cortical thickness at 7 T . Human Brain Mapping , 39 ( 6 ), 2412 – 2425 . doi: 10.1002/hbm.24011 OpenUrl CrossRef ↵ Haast , R. A. M. , Lau , J. C. , Ivanov , D. , Menon , R. S. , Uludağ , K. , & Khan , A. R . ( 2021 ). Effects of MP2RAGE B1+ sensitivity on inter-site T1 reproducibility and hippocampal morphometry at 7T . NeuroImage , 224 , 117373 . doi: 10.1016/j.neuroimage.2020.117373 OpenUrl CrossRef ↵ Hagiwara , A. , Fujimoto , K. , Kamagata , K. , Murata , S. , Irie , R. , Kaga , H. , Someya , Y. , Andica , C. , Fujita , S. , Kato , S. , Fukunaga , I. , Wada , A. , Hori , M. , Tamura , Y. , Kawamori , R. , Watada , H. , & Aoki , S . ( 2021 ). Age-Related Changes in Relaxation Times, Proton Density, Myelin, and Tissue Volumes in Adult Brain Analyzed by 2-Dimensional Quantitative Synthetic Magnetic Resonance Imaging . Investigative Radiology , 56 ( 3 ), 163 – 172 . doi: 10.1097/RLI.0000000000000720 OpenUrl CrossRef ↵ Hahn , E. L . ( 1950 ). Spin Echoes . Physical Review , 80 ( 4 ), 580 – 594 . doi: 10.1103/PhysRev.80.580 OpenUrl CrossRef Web of Science ↵ Helms , G. , Dathe , H. , Kallenberg , K. , & Dechent , P . ( 2008 ). High-resolution maps of magnetization transfer with inherent correction for RF inhomogeneity and T 1 relaxation obtained from 3D FLASH MRI . Magnetic Resonance in Medicine , 60 ( 6 ), 1396 – 1407 . doi: 10.1002/mrm.21732 OpenUrl CrossRef PubMed Web of Science ↵ Henderson , E. , McKinnon , G. , Lee , T.-Y. , & Rutt , B. K . ( 1999 ). A fast 3D Look-Locker method for volumetric T1 mapping . Magnetic Resonance Imaging , 17 ( 8 ), 1163 – 1171 . doi: 10.1016/S0730-725X(99)00025-9 OpenUrl CrossRef PubMed ↵ Homer , J. , & Beevers , M. S . ( 1985 ). Driven-equilibrium single-pulse observation of T1 relaxation. A reevaluation of a rapid “new” method for determining NMR spin-lattice relaxation times . Journal of Magnetic Resonance ( 1969 ) , 63 (2), 287–297. doi: 10.1016/0022-2364(85)90318-X OpenUrl CrossRef ↵ Hung , W. , Chen , P. , Chuang , T. , Chang , H. , & Wu , M . ( 2013 ). High resolution volumetric T1 mapping using a novel MP3RAGE method . International Society for Magnetic Resonance in Medicine Annual Meeting , 2353 . ↵ Jenkinson , M. , Beckmann , C. F. , Behrens , T. E. J. , Woolrich , M. W. , & Smith , S. M . ( 2012 ). FSL . NeuroImage , 62 ( 2 ), 782 – 790 . doi: 10.1016/j.neuroimage.2011.09.015 OpenUrl CrossRef PubMed Web of Science ↵ Labadie , C. , Lee , J. , Rooney , W. D. , Jarchow , S. , Aubert-Frécon , M ., Springer , C. S. , & Möller , H. E. ( 2014 ). Myelin water mapping by spatially regularized longitudinal relaxographic imaging at high magnetic fields . Magnetic Resonance in Medicine , 71 ( 1 ), 375 – 387 . doi: 10.1002/mrm.24670 OpenUrl CrossRef ↵ Leipert , T. K. , & Marquardt , D. W . ( 1976 ). Statistical analysis of NMR spin-lattice relaxation times . Journal of Magnetic Resonance ( 1969 ) , 24 (2), 181–199. doi: 10.1016/0022-2364(76)90027-5 OpenUrl CrossRef ↵ Li , X. , Morgan , P. S. , Ashburner , J. , Smith , J. , & Rorden , C . ( 2016 ). The first step for neuroimaging data analysis: DICOM to NIfTI conversion . Journal of Neuroscience Methods , 264 , 47 – 56 . doi: 10.1016/j.jneumeth.2016.03.001 OpenUrl CrossRef PubMed ↵ Look , D. C. , & Locker , D. R . ( 1968 ). Nuclear Spin-Lattice Relaxation Measurements by Tone-Burst Modulation . Physical Review Letters , 20 ( 18 ), 987 – 989 . doi: 10.1103/PhysRevLett.20.987 OpenUrl CrossRef ↵ Look , D. C. , & Locker , D. R . ( 1970 ). Time saving in measurement of NMR and EPR relaxation times . Review of Scientific Instruments , 41 ( 2 ), 250 – 251 . OpenUrl CrossRef ↵ Lustig , M. , Donoho , D. , & Pauly , J. M . ( 2007 ). Sparse MRI: The application of compressed sensing for rapid MR imaging . Magnetic Resonance in Medicine , 58 ( 6 ), 1182 – 1195 . doi: 10.1002/mrm.21391 OpenUrl CrossRef PubMed Web of Science ↵ Lutti , A. , Dick , F. , Sereno , M. I. , & Weiskopf , N . ( 2014 ). Using high-resolution quantitative mapping of R1 as an index of cortical myelination . NeuroImage , 93 , 176 – 188 . doi: 10.1016/j.neuroimage.2013.06.005 OpenUrl CrossRef PubMed ↵ Ma , Y.-J. , Moazamian , D. , Port , J. D. , Edjlali , M. , Pruvo , J.-P. , Hacein-Bey , L. , Hoggard , N. , Paley , M. N. J. , Menon , D. K. , Bonekamp , D. , Pravatà , E. , Garwood , M. , Danesh-Meyer , H. , Condron , P. , Cornfeld , D. M. , Holdsworth , S. J. , Du , J. , & Bydder , G. M . ( 2023 ). Targeted magnetic resonance imaging (tMRI) of small changes in the T1 and spatial properties of normal or near normal appearing white and gray matter in disease of the brain using divided subtracted inversion recovery (dSIR) and divided reverse subtracted inversion recovery (drSIR) sequences . Quantitative Imaging in Medicine and Surgery , 13 ( 10 ), 7304 – 7337 . doi: 10.21037/qims-23-232 OpenUrl CrossRef ↵ Manjón , J. V. , Coupé , P. , Martí-Bonmatí , L ., Collins , D. L. , & Robles , M. ( 2010 ). Adaptive non-local means denoising of MR images with spatially varying noise levels . Journal of Magnetic Resonance Imaging , 31 ( 1 ), 192 – 203 . doi: 10.1002/jmri.22003 OpenUrl CrossRef PubMed ↵ Marques , J. P. , & Gruetter , R . ( 2013 ). New Developments and Applications of the MP2RAGE Sequence -Focusing the Contrast and High Spatial Resolution R1 Mapping . PLoS ONE , 8 ( 7 ), e69294 . doi: 10.1371/journal.pone.0069294 OpenUrl CrossRef PubMed ↵ Marques , J. P. , Kober , T. , Krueger , G. , van der Zwaag , W. , Van de Moortele , P.-F. , & Gruetter , R. ( 2010 ). MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field . NeuroImage , 49 ( 2 ), 1271 – 1281 . doi: 10.1016/j.neuroimage.2009.10.002 OpenUrl CrossRef PubMed Web of Science ↵ Marques , J. P. , & Norris , D. G . ( 2018 ). How to choose the right MR sequence for your research question at 7 T and above? NeuroImage , 168 , 119 – 140 . doi: 10.1016/j.neuroimage.2017.04.044 OpenUrl CrossRef ↵ Martin , A. , Emorine , T. , Megdiche , I. , Créange , A. , Kober , T. , Massire , A. , & Bapst , B . ( 2023 ). Accurate Diagnosis of Cortical and Infratentorial Lesions in Multiple Sclerosis Using Accelerated Fluid and White Matter Suppression Imaging . Investigative Radiology , 58 ( 5 ), 337 – 345 . doi: 10.1097/RLI.0000000000000939 OpenUrl CrossRef ↵ Massire , A. , Seiler , C. , Troalen , T. , Girard , O. M. , Lehmann , P. , Brun , G. , Bartoli , A. , Audoin , B. , Bartolomei , F. , Pelletier , J. , Callot , V. , Kober , T. , Ranjeva , J.-P. , & Guye , M . ( 2021 ). T1-Based Synthetic Magnetic Resonance Contrasts Improve Multiple Sclerosis and Focal Epilepsy Imaging at 7 T . Investigative Radiology , 56 ( 2 ), 127 – 133 . doi: 10.1097/RLI.0000000000000718 OpenUrl CrossRef ↵ McDowell , A. R. , Shelmerdine , S. C. , Carmichael , D. W. , & Arthurs , O. J . ( 2018 ). High resolution isotropic diffusion imaging in post-mortem neonates: a feasibility study . The British Journal of Radiology , 91 ( 1092 ), 20180319 . doi: 10.1259/bjr.20180319 OpenUrl CrossRef ↵ Metere , R. , Kober , T. , Möller , H. E. , & Schäfer , A . ( 2017 ). Simultaneous Quantitative MRI Mapping of T1, T2* and Magnetic Susceptibility with Multi-Echo MP2RAGE . PLOS ONE , 12 ( 1 ), e0169265 . doi: 10.1371/journal.pone.0169265 OpenUrl CrossRef ↵ Mezer , A. , Rokem , A. , Berman , S. , Hastie , T. , & Wandell , B. A . ( 2016 ). Evaluating quantitative proton- density-mapping methods . Human Brain Mapping , 37 ( 10 ), 3623 – 3635 . doi: 10.1002/hbm.23264 OpenUrl CrossRef PubMed ↵ Mugler , J. P. , & Brookeman , J. R . ( 1990 ). Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE) . Magnetic Resonance in Medicine , 15 ( 1 ), 152 – 157 . doi: 10.1002/mrm.1910150117 OpenUrl CrossRef PubMed Web of Science ↵ Müller , J. , La Rosa , F. , Beaumont , J. , Tsagkas , C. , Rahmanzadeh , R. , Weigel , M. , Bach Cuadra , M. , Gambarota , G. , & Granziera , C. ( 2022 ). Fluid and White Matter Suppression . Investigative Radiology , 57 ( 9 ), 592 – 600 . doi: 10.1097/RLI.0000000000000877 OpenUrl CrossRef ↵ Mussard , E. , Hilbert , T. , Forman , C. , Meuli , R. , Thiran , J. , & Kober , T . ( 2020 ). Accelerated MP2RAGE imaging using Cartesian phyllotaxis readout and compressed sensing reconstruction . Magnetic Resonance in Medicine , 84 ( 4 ), 1881 – 1894 . doi: 10.1002/mrm.28244 OpenUrl CrossRef PubMed Non-Uniform Intensity Correction . http://www.bic.mni.mcgill.ca/software/N3/node6.html . (n.d.). ↵ Olsson , H. , Andersen , M. , Kadhim , M. , & Helms , G . ( 2022 ). MP3RAGE: Simultaneous mapping of T 1 and B1+ in human brain at 7T . Magnetic Resonance in Medicine , 87 ( 6 ), 2637 – 2649 . doi: 10.1002/mrm.29151 OpenUrl CrossRef ↵ Penny , W. , Friston , K. , Ashburner , J. , Kiebel , S. , & Nichols , T . ( 2007 ). Statistical Parametric Mapping: The Analysis of Functional Brain Images . doi: 10.1016/B978-0-12-372560-8.X5000-1 OpenUrl CrossRef ↵ Puy , G. , Vandergheynst , P. , Gribonval , R. , & Wiaux , Y . ( 2012 ). Universal and efficient compressed sensing by spread spectrum and application to realistic Fourier imaging techniques . EURASIP Journal on Advances in Signal Processing , 2012 ( 1 ), 6 . doi: 10.1186/1687-6180-2012-6 OpenUrl CrossRef ↵ Rioux , J. A. , Saranathan , M. , & Rutt , B. K . ( 2014 ). Simultaneous whole-brain t1 and flip angle mapping with MP3RAGE . Proceedings of the 22nd Annual Meeting of ISMRM, Milan, Italy , 3216 . ↵ Saito , N. , Watanabe , M. , Sakai , O. , & Jara , H . ( 2012 ). Human lifespan age-related changes of the brain proton density by quantitative MRI . Proceedings of the 20th Annual Meeting of ISMRM , 780 . ↵ Shah , N. J. , Neeb , H. , Kircheis , G. , Engels , P. , Häussinger , D. , & Zilles , K . ( 2008 ). Quantitative cerebral water content mapping in hepatic encephalopathy . NeuroImage , 41 ( 3 ), 706 – 717 . doi: 10.1016/j.neuroimage.2008.02.057 OpenUrl CrossRef PubMed ↵ Silver , M. S. , Joseph , R. I. , & Hoult , D. I . ( 1984 ). Highly selective and π pulse generation . Journal of Magnetic Resonance (1969) , 59 ( 2 ), 347 – 351 . doi: 10.1016/0022-2364(84)90181-1 OpenUrl CrossRef ↵ Sled , J. G. , Zijdenbos , A. P. , & Evans , A. C. ( 1997 ). A Comparison of Retrospective Intensity Non-uniformity Correction Methods for MRI . Information Processing in Medical Imaging. https://api.semanticscholar.org/CorpusID:27760780 ↵ Sled , J. G. , Zijdenbos , A. P. , & Evans , A. C . ( 1998 ). A nonparametric method for automatic correction of intensity nonuniformity in MRI data . IEEE Transactions on Medical Imaging , 17 ( 1 ), 87 – 97 . doi: 10.1109/42.668698 OpenUrl CrossRef PubMed Web of Science ↵ Smith , S. M. , Jenkinson , M. , Woolrich , M. W. , Beckmann , C. F. , Behrens , T. E. J. , Johansen-Berg , H. , Bannister , P. R. , De Luca , M. , Drobnjak , I. , Flitney , D. E. , Niazy , R. K. , Saunders , J. , Vickers , J. , Zhang , Y. , De Stefano , N. , Brady , J. M. , & Matthews , P. M. ( 2004 ). Advances in functional and structural MR image analysis and implementation as FSL . NeuroImage , 23 , S208 – S219 . doi: 10.1016/j.neuroimage.2004.07.051 OpenUrl CrossRef PubMed Web of Science ↵ Tabelow , K. , Balteau , E. , Ashburner , J. , Callaghan , M. F. , Draganski , B. , Helms , G. , Kherif , F. , Leutritz , T. , Lutti , A. , Phillips , C. , Reimer , E. , Ruthotto , L. , Seif , M. , Weiskopf , N. , Ziegler , G. , & Mohammadi , S . ( 2019 ). hMRI – A toolbox for quantitative MRI in neuroscience and clinical research . NeuroImage , 194 , 191 – 210 . doi: 10.1016/j.neuroimage.2019.01.029 OpenUrl CrossRef ↵ Tanner , M. , Gambarota , G. , Kober , T. , Krueger , G. , Erritzoe , D. , Marques , J. P. , & Newbould , R . ( 2012 ). Fluid and white matter suppression with the MP2RAGE sequence . Journal of Magnetic Resonance Imaging , 35 ( 5 ), 1063 – 1070 . doi: 10.1002/jmri.23532 OpenUrl CrossRef PubMed ↵ Tofts , P . ( 2003 ). Quantitative MRI of the Brain (P. Tofts , Ed.). Wiley . doi: 10.1002/0470869526 OpenUrl CrossRef ↵ Trotier , A. J. , Dilharreguy , B. , Anandra , S. , Corbin , N. , Lefrançois , W. , Ozenne , V. , Miraux , S. , & Ribot , E. J . ( 2022 ). The Compressed Sensing MP2RAGE as a Surrogate to the MPRAGE for Neuroimaging at 3 T . Investigative Radiology , 57 ( 6 ), 366 – 378 . doi: 10.1097/RLI.0000000000000849 OpenUrl CrossRef PubMed ↵ Trotier , A. J. , Rapacchi , S. , Faller , T. L. , Miraux , S. , & Ribot , E. J . ( 2019 ). Compressed-Sensing MP2RAGE sequence: Application to the detection of brain metastases in mice at 7T . Magnetic Resonance in Medicine , 81 ( 1 ), 551 – 559 . doi: 10.1002/mrm.27438 OpenUrl CrossRef ↵ Tustison , N. J. , Avants , B. B. , Cook , P. A. , Yuanjie Zheng , Egan , A. , Yushkevich , P. A. , & Gee , J. C. ( 2010 ). N4ITK: Improved N3 Bias Correction . IEEE Transactions on Medical Imaging , 29 ( 6 ), 1310 – 1320 . doi: 10.1109/TMI.2010.2046908 OpenUrl CrossRef PubMed Web of Science ↵ Urushibata , Y. , Kuribayashi , H. , Fujimoto , K. , Kober , T. , Grinstead , J. W. , Isa , T. , & Okada , T . ( 2019 ). Advantages of fluid and white matter suppression (FLAWS) with MP2RAGE compared with double inversion recovery turbo spin echo (DIR-TSE) at 7T . European Journal of Radiology , 116 , 160 – 164 . doi: 10.1016/j.ejrad.2019.04.019 OpenUrl CrossRef ↵ Van de Moortele , P.-F. , Auerbach , E. J. , Olman , C. , Yacoub , E. , Uğurbil , K. , & Moeller , S. ( 2009 ). T1 weighted brain images at 7 Tesla unbiased for Proton Density, T2⁎ contrast and RF coil receive B1 sensitivity with simultaneous vessel visualization . NeuroImage , 46 ( 2 ), 432 – 446 . doi: 10.1016/j.neuroimage.2009.02.009 OpenUrl CrossRef PubMed Web of Science ↵ Vasanawala , S. S. , Alley , M. T. , Hargreaves , B. A. , Barth , R. A. , Pauly , J. M. , & Lustig , M . ( 2010 ). Improved Pediatric MR Imaging with Compressed Sensing . Radiology , 256 ( 2 ), 607 – 616 . doi: 10.1148/radiol.10091218 OpenUrl CrossRef PubMed ↵ Weiskopf , N. , Suckling , J. , Williams , G. , Correia , M. M. , Inkster , B. , Tait , R. , Ooi , C. , Bullmore , E. T. , & Lutti , A . ( 2013 ). Quantitative multi-parameter mapping of R1, PD*, MT, and R2* at 3T: a multi-center validation . Frontiers in Neuroscience , 7 . doi: 10.3389/fnins.2013.00095 OpenUrl CrossRef PubMed ↵ Wright , P. J. , Mougin , O. E. , Totman , J. J. , Peters , A. M. , Brookes , M. J. , Coxon , R. , Morris , P. E. , Clemence , M. , Francis , S. T. , Bowtell , R. W. , & Gowland , P. A . ( 2008 ). Water proton T1 measurements in brain tissue at 7, 3, and 1.5T using IR-EPI, IR-TSE, and MPRAGE: results and optimization . Magnetic Resonance Materials in Physics, Biology and Medicine , 21 ( 1 ), 121 . doi: 10.1007/s10334-008-0104-8 OpenUrl CrossRef PubMed ↵ Xiaowei He ( 2024 ). Positive-zero-negative colorbar ( https://www.mathworks.com/matlabcentral/fileexchange/110190-positive-zero-negative-colorbar ), MATLAB Central File Exchange. Retrieved January 26, 2024. (n.d.). ↵ Yarnykh , V. L . ( 2007 ). Actual flip-angle imaging in the pulsed steady state: A method for rapid three- dimensional mapping of the transmitted radiofrequency field . Magnetic Resonance in Medicine , 57 ( 1 ), 192 – 200 . doi: 10.1002/mrm.21120 OpenUrl CrossRef PubMed View the discussion thread. Back to top Previous Next Posted July 01, 2024. Download PDF Supplementary Material Data/Code Email Thank you for your interest in spreading the word about medRxiv. NOTE: Your email address is requested solely to identify you as the sender of this article. Your Email * Your Name * Send To * Enter multiple addresses on separate lines or separate them with commas. You are going to email the following Quantitative T1 and Effective Proton Density (PD*) mapping in children and adults at 7T from an MP2RAGE sequence optimised for uniform T1-weighted (UNI) and FLuid And White matter Suppression (FLAWS) contrasts Message Subject (Your Name) has forwarded a page to you from medRxiv Message Body (Your Name) thought you would like to see this page from the medRxiv website. Your Personal Message CAPTCHA This question is for testing whether or not you are a human visitor and to prevent automated spam submissions. Share Quantitative T1 and Effective Proton Density (PD*) mapping in children and adults at 7T from an MP2RAGE sequence optimised for uniform T1-weighted (UNI) and FLuid And White matter Suppression (FLAWS) contrasts Ayşe Sıla Dokumacı , Katy Vecchiato , Raphael Tomi-Tricot , Michael Eyre , Philippa Bridgen , Pierluigi Di Cio , Chiara Casella , Tobias C. Wood , Jan Sedlacik , Tom Wilkinson , Sharon L. Giles , Joseph V. Hajnal , Jonathan O’Muircheartaigh , Shaihan J. Malik , David W. Carmichael medRxiv 2024.06.28.24307535; doi: https://doi.org/10.1101/2024.06.28.24307535 Share This Article: Copy Citation Tools Quantitative T1 and Effective Proton Density (PD*) mapping in children and adults at 7T from an MP2RAGE sequence optimised for uniform T1-weighted (UNI) and FLuid And White matter Suppression (FLAWS) contrasts Ayşe Sıla Dokumacı , Katy Vecchiato , Raphael Tomi-Tricot , Michael Eyre , Philippa Bridgen , Pierluigi Di Cio , Chiara Casella , Tobias C. Wood , Jan Sedlacik , Tom Wilkinson , Sharon L. Giles , Joseph V. Hajnal , Jonathan O’Muircheartaigh , Shaihan J. Malik , David W. Carmichael medRxiv 2024.06.28.24307535; doi: https://doi.org/10.1101/2024.06.28.24307535 Citation Manager Formats BibTeX Bookends EasyBib EndNote (tagged) EndNote 8 (xml) Medlars Mendeley Papers RefWorks Tagged Ref Manager RIS Zotero Tweet Widget Facebook Like Google Plus One Subject Area Radiology and Imaging Subject Areas All Articles Addiction Medicine (573) Allergy and Immunology (865) Anesthesia (303) Cardiovascular Medicine (4457) Dentistry and Oral Medicine (445) Dermatology (383) Emergency Medicine (610) Endocrinology (including Diabetes Mellitus and Metabolic Disease) (1517) Epidemiology (15244) Forensic Medicine (30) Gastroenterology (1132) Genetic and Genomic Medicine (6620) Geriatric Medicine (669) Health Economics (1002) Health Informatics (4557) Health Policy (1372) Health Systems and Quality Improvement (1615) Hematology (543) HIV/AIDS (1272) Infectious Diseases (except HIV/AIDS) (15936) Intensive Care and Critical Care Medicine (1106) Medical Education (624) Medical Ethics (147) Nephrology (670) Neurology (6634) Nursing (346) Nutrition (999) Obstetrics and Gynecology (1148) Occupational and Environmental Health (957) Oncology (3348) Ophthalmology (980) Orthopedics (369) Otolaryngology (421) Pain Medicine (436) Palliative Medicine (130) Pathology (665) Pediatrics (1696) Pharmacology and Therapeutics (693) Primary Care Research (714) Psychiatry and Clinical Psychology (5463) Public and Global Health (9257) Radiology and Imaging (2210) Rehabilitation Medicine and Physical Therapy (1371) Respiratory Medicine (1198) Rheumatology (598) Sexual and Reproductive Health (716) Sports Medicine (532) Surgery (714) Toxicology (99) Transplantation (289) Urology (265) (function(){function c(){var b=a.contentDocument||a.contentWindow.document;if(b){var d=b.createElement('script');d.innerHTML="window.__CF$cv$params={r:'a0313f5abd688a34',t:'MTc4MDAxMzEyNg=='};var a=document.createElement('script');a.src='/cdn-cgi/challenge-platform/scripts/jsd/main.js';document.getElementsByTagName('head')[0].appendChild(a);";b.getElementsByTagName('head')[0].appendChild(d)}}if(document.body){var a=document.createElement('iframe');a.height=1;a.width=1;a.style.position='absolute';a.style.top=0;a.style.left=0;a.style.border='none';a.style.visibility='hidden';document.body.appendChild(a);if('loading'!==document.readyState)c();else if(window.addEventListener)document.addEventListener('DOMContentLoaded',c);else{var e=document.onreadystatechange||function(){};document.onreadystatechange=function(b){e(b);'loading'!==document.readyState&&(document.onreadystatechange=e,c())}}}})();
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.