Impact of metallurgical and geometric characteristics on cyclic fatigue resistance of three thermally treated NiTi rotary instruments

preprint OA: closed
Full text JSON View at publisher

Abstract

This study aimed to evaluate the cyclic fatigue resistance and metallurgical characteristics of thermally treated NiTi rotary instruments: ProDesign Logic, Solla Purple, and Solla Colors. Cyclic fatigue testing was conducted at room temperature (20°C ± 1°C) and simulated body temperature (36°C ± 1°C) using a standardized artificial canal with a 60° curvature and 5 mm radius. Ten instruments per group (25.06) were tested at each temperature. Differential Scanning Calorimetry (DSC) and Scanning Electron Microscopy (SEM) were employed to assess phase transformation behavior and fracture topography, respectively. ProDesign Logic exhibited the highest cyclic fatigue resistance at room temperature compared to both Solla systems (p0.05). DSC analysis revealed distinct thermal behaviors. ProDesign Logic had the highest austenite finish temperature (Af = 53.0°C), indicating a predominant martensitic phase at clinical temperatures; Solla Purple showed a mixed-phase profile (Af = 43.5°C); and Solla Colors demonstrated characteristics of R-phase transformation with greater microstructural heterogeneity (Af = 43.1°C). SEM confirmed typical fatigue fracture patterns in all groups. ProDesign Logic demonstrated greater cyclic fatigue resistance, attributed to its high Af temperature and cross-sectional design. In contrast, the Solla systems demonstrated lower resistance to cyclic fatigue. Solla Purple exhibited conventional NiTi behavior, whereas Solla Colors displayed a heterogeneous microstructure with the presence of the R-phase.
Full text 37,669 characters · extracted from preprint-html · click to expand
Impact of metallurgical and geometric characteristics on cyclic fatigue resistance of three thermally treated NiTi rotary instruments | Authorea try { document.documentElement.classList.add('js'); } catch (e) { } var _gaq = _gaq || []; _gaq.push(['_setAccount', 'G-8VDV14Y67G']); _gaq.push(['_trackPageview']); (function() { var ga = document.createElement('script'); ga.type = 'text/javascript'; ga.async = true; ga.src = ('https:' == document.location.protocol ? 'https://ssl' : 'http://www') + '.google-analytics.com/ga.js'; var s = document.getElementsByTagName('script')[0]; s.parentNode.insertBefore(ga, s); })(); Skip to main content Preprints Collections Wiley Open Research IET Open Research Ecological Society of Japan All Collections About About Authorea FAQs Contact Us Quick Search anywhere Search for preprint articles, keywords, etc. Search Search ADVANCED SEARCH SCROLL This is a preprint and has not been peer reviewed. Data may be preliminary. 17 July 2025 V1 Latest version Share on Impact of metallurgical and geometric characteristics on cyclic fatigue resistance of three thermally treated NiTi rotary instruments Authors : Raimundo Sales de Oliveira Neto 0000-0002-4726-8106 [email protected] , Guilherme Ferreira da Silva , Diego Rafael Nespeque Correa , Rodrigo Vivan 0000-0002-0419-5699 , Murilo Priori Alcalde , and Marco Antonio Duarte 0000-0003-3051-737X Authors Info & Affiliations https://doi.org/10.22541/au.175273951.15422748/v1 Published Journal of Clinical and Experimental Dentistry Version of record Peer review timeline 164 views 141 downloads Contents Abstract Information & Authors Metrics & Citations View Options References Figures Tables Media Share Abstract This study aimed to evaluate the cyclic fatigue resistance and metallurgical characteristics of thermally treated NiTi rotary instruments: ProDesign Logic, Solla Purple, and Solla Colors. Cyclic fatigue testing was conducted at room temperature (20°C ± 1°C) and simulated body temperature (36°C ± 1°C) using a standardized artificial canal with a 60° curvature and 5 mm radius. Ten instruments per group (25.06) were tested at each temperature. Differential Scanning Calorimetry (DSC) and Scanning Electron Microscopy (SEM) were employed to assess phase transformation behavior and fracture topography, respectively. ProDesign Logic exhibited the highest cyclic fatigue resistance at room temperature compared to both Solla systems (p0.05). DSC analysis revealed distinct thermal behaviors. ProDesign Logic had the highest austenite finish temperature (Af = 53.0°C), indicating a predominant martensitic phase at clinical temperatures; Solla Purple showed a mixed-phase profile (Af = 43.5°C); and Solla Colors demonstrated characteristics of R-phase transformation with greater microstructural heterogeneity (Af = 43.1°C). SEM confirmed typical fatigue fracture patterns in all groups. ProDesign Logic demonstrated greater cyclic fatigue resistance, attributed to its high Af temperature and cross-sectional design. In contrast, the Solla systems demonstrated lower resistance to cyclic fatigue. Solla Purple exhibited conventional NiTi behavior, whereas Solla Colors displayed a heterogeneous microstructure with the presence of the R-phase. Impact of metallurgical and geometric characteristics on cyclic fatigue resistance of three thermally treated NiTi rotary instruments Raimundo Sales de Oliveira Neto 1 , Guilherme Ferreira da Silva 1 , Diego Rafael Nespeque Correa 2 , Rodrigo Ricci Vivan 1 , Murilo Priori Alcalde 1 , Marco Antonio Hungaro Duarte 1 1 Department of Operative Dentistry, Endodontics and Dental Materials, Bauru School of Dentistry, University of São Paulo – USP, Bauru, Brazil 2 São Paulo State University (UNESP), School of Sciences, Campus Bauru, Laboratory of Anelasticity and Biomaterials, Bauru, Brazil Corresponding author address: Raimundo Sales de Oliveira Neto (Contact information: +55-14-3235-8344; [email protected] . Department of Dentistry, Endodontics and Dental Materials- University of São Paulo, Al. Octávio Pinheiro Brizolla 9-75, CEP: 17012-901 Vila Universitária Bauru, SP, Brazil) ABSTRACT This study aimed to evaluate the cyclic fatigue resistance and metallurgical characteristics of thermally treated NiTi rotary instruments: ProDesign Logic, Solla Purple, and Solla Colors. Cyclic fatigue testing was conducted at room temperature (20°C ± 1°C) and simulated body temperature (36°C ± 1°C) using a standardized artificial canal with a 60° curvature and 5 mm radius. Ten instruments per group (25.06) were tested at each temperature. Differential Scanning Calorimetry (DSC) and Scanning Electron Microscopy (SEM) were employed to assess phase transformation behavior and fracture topography, respectively. ProDesign Logic exhibited the highest cyclic fatigue resistance at room temperature compared to both Solla systems (p0.05). DSC analysis revealed distinct thermal behaviors. ProDesign Logic had the highest austenite finish temperature (Af = 53.0°C), indicating a predominant martensitic phase at clinical temperatures; Solla Purple showed a mixed-phase profile (Af = 43.5°C); and Solla Colors demonstrated characteristics of R-phase transformation with greater microstructural heterogeneity (Af = 43.1°C). SEM confirmed typical fatigue fracture patterns in all groups. ProDesign Logic demonstrated greater cyclic fatigue resistance, attributed to its high Af temperature and cross-sectional design. In contrast, the Solla systems demonstrated lower resistance to cyclic fatigue. Solla Purple exhibited conventional NiTi behavior, whereas Solla Colors displayed a heterogeneous microstructure with the presence of the R-phase. Keywords: cyclic fatigue; differential scanning calorimetry; nickel-titanium alloy INTRODUCTION Nickel-titanium (NiTi) endodontic instruments are widely used in root canal preparation due to their flexibility and ability to shape curved canals. However, instrument separation can occur during canal instrumentation, with cyclic fatigue being a primary cause of fracture [1, 2]. Cyclic fatigue arises when a NiTi instrument rotates within a curved canal, subjecting it to repeated cycles of tension and compression. The constant stress concentration at the point of maximum curvature leads to progressive material fatigue, ultimately resulting in instrument fracture [3, 4, 5]. To enhance the fracture resistance of NiTi files, manufacturers have developed new instruments through design modifications, surface and thermal treatments [6]. These thermal treatments have enabled the production of controlled-memory (CM) instruments, which primarily exhibit a martensitic crystalline structure at room temperature. This characteristic allows the instrument to return to its original shape upon heating, as a consequence of the phase transition (martensitic crystallographic phase to austenitic phase) [7, 8]. Compared to conventional/superelastic instruments, CM files demonstrate superior resistance to both cyclic and torsional fatigue [9, 10]. In addition to mechanical testing (e.g., cyclic and torsional fatigue assessments), metallurgical characterization enables a more comprehensive evaluation and a deeper understanding of the mechanical properties of new instruments. This approach provides a physicochemical explanation for their mechanical behavior [1, 11, 12]. According to a previously published literature review, differential scanning calorimetry (DSC) is the most frequently reported metallurgical analysis in the literature [13]. DSC is primarily used to determine enthalpy changes and phase transformation temperatures of NiTi alloys. The combination of DSC analysis and mechanical testing enables a more accurate and clinically relevant assessment of the properties of new NiTi instruments [11]. The ProDesign Logic instruments (Easy Bassi, Belo Horizonte, Minas Gerais, Brazil) are manufactured from thermally treated CM NiTi. These instruments are available with various tip sizes and tapers, featuring distinct cross-sectional designs: a double-helix configuration for instruments with .03, .05, and .06 tapers; a triple-helix design for .04 taper instruments; and a quadruple-helix geometry for .01 taper instruments. Previous studies have demonstrated that these instruments exhibit superior cyclic and torsional fatigue resistance compared to other commercially available systems [3, 14, 15]. Although their mechanical properties have been extensively evaluated, no studies to date have investigated the metallurgical characteristics of the ProDesign Logic system. Recently, two new NiTi instrument systems have been introduced to the market: Solla Purple and Solla Colors (Shenzhen Forever Medical, Longgang District, Shenzhen, China). The rotary Solla Purple system features a triangular cross-section (inactive tip) for its smaller instruments (15.04, 20.05, and 25.06) and a trapezoidal cross-section for larger files (35.04). The Solla Purple system undergoes a proprietary ’Purple’ manufacturing process, while the Solla Colors instruments feature a ’Colors’ treatment. However, the manufacturer’s specifications do not clearly distinguish whether these distinctive colorations result from thermal or surface treatments. Despite their commercial availability, no studies have yet evaluated their metallurgical properties and mechanical performance, highlighting a significant gap in the current literature. Given the critical importance of mechanical and metallurgical properties in ensuring clinical safety and case-specific instrument selection, this study aims to comprehensively evaluate the performance characteristics of ProDesign Logic, and the newly introduced and commercially available Solla Colors and Solla Purple files. MATERIAL AND METHODS jabbrv-ltwa-all.ldf jabbrv-ltwa-en.ldf Sample size calculation The sample size was determined through power analysis using G*Power 3.1 (Heinrich Heine University, Düsseldorf, Germany), which indicated that 20 instruments per group (10 per temperature condition) were needed to reliably detect differences in cyclic fatigue resistance, based on an effect size of 1.8 from previous data [5] with 95% power and α = 0.05. Cyclic fatigue Cyclic fatigue testing was conducted at room temperature (20°C ± 1°C) and simulated body temperature (36°C ± 1°C). For maintaining precise temperature control, a histology water bath (Leica HI 1210; Leica Biosystems, Nussloch, Germany) was used to regulate body temperature. The container was filled with 600 mL of water to ensure complete submersion of the simulated canal. Temperature was monitored throughout the test using both the water bath’s digital thermometer and an external infrared thermometer [5, 11, 16]. Six instruments of each type (n = 6 per group) were tested at each temperature, including Solla Purple 25/0.06, Solla Colors 25/0.06, and ProDesign Logic 25/0.06. Before testing, all instruments were examined for defects or deformities under a stereomicroscope (Stemi 2000C; Carl Zeiss, Jena, Germany) at 16× magnification to ensure structural integrity. The cyclic fatigue test was conducted using an apparatus designed to simulate the curvature of an artificial stainless-steel canal, as detailed in prior studies [3, 4, 5, 17]. This apparatus featured a 60° curvature and a 5 mm radius. The curvature adjustment was achieved using a guide cylinder (5 mm radius) and an outer arc with a 1 mm deep groove, facilitating the rotation of instruments while maintaining them within the curvature. An electric motor, VDW Reciproc Silver (VDW, Munich, Germany), was coupled to this apparatus, and instruments were operated with reciprocation motion (Reciproc all function). The time taken for instrument fracture was measured using a digital stopwatch and corroborated by simultaneous filming. Differential Scanning Calorimetry (DSC) analysis Differential scanning calorimetry (DSC) analysis (Mettler Toledo, Barueri, São Paulo, Brazil) was performed to determine the phase transformation temperatures of the NiTi alloy, by ASTM standards [18]. The austenite start (As) and austenite finish (Af) temperatures were assessed, as these transitions are clinically relevant when instruments move from a cooler environment to the warmer root canal, mirroring the thermal process simulated by DSC [19]. For analysis, segments approximately 3–4 mm in length were sectioned using a water-cooled, slow-speed diamond saw. Samples weighing 2–3 mg were placed in open aluminum pans and subjected to the following thermal cycle: heating from room temperature to 100°C, cooling to –30°C (to record the cooling curve) and reheating to 150°C (to obtain the second heating curve). Scanning Electron Microscopy (SEM) analysis Transverse images of the fractured instruments were obtained using a scanning electron microscope (SEM - Aspex Express Fei Europe, Eindhoven, the Netherlands) to determine the topographic features of the fractured surface. Before the SEM evaluation, the instruments were ultrasonically cleaned (L100, Schuster, Santa Maria, RS, Brazil) in a saline solution for 3 minutes. The instruments were examined at ×200 magnification. Statical Analysis Normality of the data was verified using the Shapiro-Wilk test. For intergroup comparisons under the same temperature condition, one-way ANOVA followed by Tukey’s post hoc test was applied. Pairwise comparisons between instruments tested at room temperature and simulated body temperature were performed using unpaired two-tailed Student’s t-tests. A significance level of 5% was adopted for all analyses. Statistical procedures were conducted using GraphPad Prism 9.3 (GraphPad Software Inc., San Diego, CA, USA). RESULTS Table 1 presents the times in seconds and the number of cycles to failure (NCF) of the examined instruments. In addition, the percentage of reduction in cyclic fatigue at body temperature was calculated, and these data are also shown in Figure 1. At room temperature, ProDesign Logic exhibited significantly higher cyclic fatigue resistance (time and NCF) compared to Solla Purple and Solla Colors (p0.05). At body temperature, no significant differences were observed among instruments (p>0.05). Temperature did not significantly alter the performance of any individual instrument (all p>0.05). Figure 2 displays SEM images of the fractured instruments from each system. These images depict characteristic features indicative of cyclic fatigue (crack initiation zones with fatigue striations, progressive crack propagation areas showing plastic deformation of lateral cutting edges, and overload zones containing numerous microvoid coalescence dimples) behavior observed under SEM. The DSC analysis of the second heating cycle revealed distinct phase transformation characteristics among the endodontic instruments. Solla Purple exhibited an Af temperature of 43.5°C, indicating a mixed martensite-austenite phase at body temperature (37°C) with moderate superelasticity (ΔH = -5.06 J/g), consistent with conventional NiTi or M-Wire alloys. Solla Collors showed a low transformation start temperature (As = 31.1°C) and broader peak width (6.79°C), suggesting the presence of R-phase and microstructural heterogeneity characteristic of CM-Wire alloys. ProDesign Logic demonstrated the highest Af temperature (53.0°C), maintaining pure martensite at 37°C with, a characteristic typical of high-Af thermally treated alloys, such as CM-Wire. All data derived from the second heating cycle DSC (ASTM F2004-17), representing stabilized material behavior post-thermal cycling. Figure 3 presents the DSC curves comparing the second heating cycle of Solla Purple, Solla Colors, and ProDesign Logic endodontic instruments. The normalized heat flow (mW/mg) is plotted against temperature (°C), showing distinct phase transformation behaviors from martensite to austenite. Key transformation temperatures (onset, peak, and endset) are annotated for each instrument. DISCUSSION While endodontic NiTi file systems are continuously developed and marketed to enhance clinical efficiency and procedural safety, many newly introduced systems, including the Solla files (Purple and Colors), enter clinical use without a comprehensive evaluation of their mechanical and metallurgical properties. This study provides a critical characterization of these instruments by systematically assessing both their phase transformation behavior and cyclic fatigue resistance. Also, this study provides the first metallurgical characterization of ProDesign Logic instruments. The DSC results provide a metallurgical basis for the observed cyclic fatigue performance [11, 20]. ProDesign Logic’s superior fatigue resistance at room temperature aligns with its high austenite finish temperature (Af = 52.99°C), which enables it to maintain a stable martensitic structure during instrumentation. This phase stability, coupled with narrow thermal hysteresis (6.3°C), minimizes energy dissipation during cyclic loading, delaying crack initiation [21]. In contrast, Solla Colors and Solla Purple exhibited lower NCF values, consistent with their earlier phase transformations (Af = 43.14°C and 43.5°C). While Solla Colors’ R-phase presence (evidenced by low As = 31.1°C and broad DSC peaks) can enhance stress distribution, its heterogeneous microstructure likely introduced localized stress concentrators [8, 22]. Solla Purple’s conventional NiTi behavior (ΔH = -5.06 J/g) and mixed-phase composition at 37°C further limited its cyclic fatigue resistance, as the absence of advanced thermal treatments may reduce energy absorption capacity during cyclic deformation [6, 14, 17, 19]. Temperature variations did not significantly alter the cyclic fatigue resistance of the tested 25/0.06 instruments (p>0.05), despite DSC-predicted phase differences. These findings suggest that instrument-specific design characteristics - particularly the different cross-section designs of ProDesign Logic (modified S-shaped), Solla Purple (triangular), and Solla Colors (trapezoidal) - combined with their distinct thermal treatments may have a greater impact on fatigue performance than phase transformation behavior alone [3, 16, 23]. The interaction between these geometric factors and proprietary manufacturing processes can influence the mechanical response [24], potentially explaining the similar fatigue resistance observed between the Solla instruments, despite their differing DSC profiles. The greater reduction in cyclic fatigue resistance at body temperature observed for ProDesign Logic (26.1%) compared to Solla Colors (3.2%) and Solla Purple (8.4%) may be attributed to distinct phase distributions resulting from their proprietary manufacturing processes. While ProDesign Logic’s significant fatigue reduction (26.1%) at body temperature aligns with its high As temperature (45.5°C) and predominantly martensitic structure, the more stable performance of Solla instruments (3.2-8.4% reduction) suggests they likely contain lower martensite fractions and greater austenite stability at clinical temperatures. However, the exact phase composition and distribution in these commercially available systems remain undefined. X-ray diffraction (XRD) analysis is therefore essential to conclusively determine their crystalline characteristics, including precise quantification of phase ratios (austenite/martensite/R-phase) and identification of potential microstructural heterogeneities that may influence mechanical performance [13, 19]. This study implemented a static testing protocol to assess cyclic fatigue resistance, aligning with previous studies [3-5, 15-17, 19]. Static testing effectively eliminates torsional stress artifacts present in dynamic testing configurations that utilize bending tubes [25,26], while simultaneously overcoming the technical limitations associated with achieving pure axial motion in dynamic systems [27]. Although alternative dynamic testing approaches are available, their heightened susceptibility to extraneous factors may mask fundamental differences in instrument performance. Consequently, static testing was chosen to ensure that the collected data accurately represents the inherent mechanical behavior of each file design, independent of the influences of the testing apparatus. Our results demonstrate that the cross-sectional design can also significantly influence cyclic fatigue resistance, with ProDesign Logic’s superior performance attributable to its S-shaped geometry, which features a reduced cross-sectional area and optimized stress distribution. This finding aligns with finite element analyses, which show that instruments with smaller cross-sectional diameters exhibit lower stress concentrations during cyclic loading [28]. Consistent with previous findings that S-cross-section designs enhance fatigue resistance [29-31], ProDesign Logic’s modified S-shape showed approximately 25% greater cycles to failure than triangular/trapezoidal designs. The absence of significant differences in performance between Solla Colors and Solla Purple may be attributed to their similar core metal mass, despite their distinct geometric configurations. SEM examination revealed characteristic ductile fracture patterns across all instrument systems following cyclic fatigue testing. The fractured surfaces consistently exhibited crack initiation zones with fatigue striations, progressive crack propagation areas showing plastic deformation of lateral cutting edges, and overload zones containing numerous microvoid coalescence dimples. These observations align with failure patterns reported in previous studies of NiTi endodontic instruments [4, 16, 17, 19]. As the first study to evaluate both the metallurgical properties of ProDesign Logic and the mechanical characteristics of Solla Purple and Solla Colors systems, our findings provide critical benchmarks for clinical decision-making. The superior performance of ProDesign Logic recommends its use in technically demanding cases involving prolonged instrumentation of curved canals. In contrast, the Solla files exhibited less predictable fracture behavior, suggesting they may be better suited for lower-stress clinical scenarios. The results should be interpreted considering the inherent limitations of in vitro models, which cannot fully replicate the clinical environment or cumulative thermomechanical stresses (e.g., repeated sterilization cycles) [3]. Future studies should expand this research by conducting clinical validation of shaping ability and torsional fatigue testing, complemented by advanced material characterization techniques such as XRD, to provide a more comprehensive analysis of these NiTi instruments. CONCLUSIONS Within the limitations of this study, the metallurgical properties and instrument design significantly influence cyclic fatigue resistance. The ProDesign Logic system demonstrated superior cyclic fatigue resistance, attributable to its specific metallurgical characteristics and cross-sectional geometry. In contrast, the Solla instruments (Colors and Purple) showed comparable but lower fatigue resistance despite their distinct phase transformation profiles. Data availability statement: Data available on request from the authors. Funding statement: The first author would like to thank São Paulo Research Foundation (FAPESP- Process number: 2024/01063-0) for providing financial support. Conflict of interest disclosure: The authors deny any conflicts of interest related to this study. Ethics approval statement: Not applicable. Patient consent statement: Not applicable. Permission to reproduce material from other sources: Not applicable Clinical trial registration: Not applicable REFERENCES 1. Arias A, Peters OA. Present status and future directions: Canal shaping. Int Endod J. 2022 May;55 Suppl 3(Suppl 3):637-655. doi: 10.1111/iej.13698. 2. Fernández-Pazos G, Martín-Biedma B, Varela-Patiño P, Ruíz-Piñón M, Castelo-Baz P. Fracture and deformation of ProTaper Next instruments after clinical use. J Clin Exp Dent. 2018 Nov 1;10(11):e1091-e1095. doi: 10.4317/jced.54910. 3. Sales de Oliveira Neto R, da Rocha Tavares Duarte R, Hungaro Duarte MA, Ferreira da Silva G, Priori Alcalde M, Rigoldi Bonjardim L. Impact of Repeated Use on Cyclic and Torsional Fatigue of 3 Rotary Files: Implications for Clinical Safety. J Endod. 2025 Apr 21:S0099-2399(25)00194-3. doi: 10.1016/j.joen.2025.04.008. 4. Duque JA, Bramante CM, Duarte MAH, Alcalde MP, Silva EJNL, Vivan RR. Cyclic Fatigue Resistance of Nickel-Titanium Reciprocating Instruments after Simulated Clinical Use. J Endod. 2020 Nov;46(11):1771-1775. doi: 10.1016/j.joen.2020.08.010. 5. Klymus ME, Alcalde MP, Vivan RR, Só MVR, de Vasconselos BC, Duarte MAH. Effect of temperature on the cyclic fatigue resistance of thermally treated reciprocating instruments. Clin Oral Investig. 2019 Jul;23(7):3047-3052. doi: 10.1007/s00784-018-2718-1. 6. Gavini G, Santos MD, Caldeira CL, Machado MEL, Freire LG, Iglecias EF, Peters OA, Candeiro GTM. Nickel-titanium instruments in endodontics: a concise review of the state of the art. Braz Oral Res. 2018 Oct 18;32(suppl 1):e67. doi: 10.1590/1807-3107bor-2018.vol32.0067. 7. Shen Y, Zhou HM, Zheng YF, Campbell L, Peng B, Haapasalo M. Metallurgical characterization of controlled memory wire nickel-titanium rotary instruments. J Endod. 2011 Nov;37(11):1566-71. doi: 10.1016/j.joen.2011.08.005. 8. Zupanc J, Vahdat-Pajouh N, Schäfer E. New thermomechanically treated NiTi alloys - a review. Int Endod J. 2018 Oct;51(10):1088-1103. doi: 10.1111/iej.12924. 9. Braga LC, Faria Silva AC, Buono VT, de Azevedo Bahia MG. Impact of heat treatments on the fatigue resistance of different rotary nickel-titanium instruments. J Endod. 2014 Sep;40(9):1494-7. doi: 10.1016/j.joen.2014.03.007. 10. Campbell L, Shen Y, Zhou HM, Haapasalo M. Effect of fatigue on torsional failure of nickel-titanium controlled memory instruments. J Endod. 2014 Apr;40(4):562-5. doi: 10.1016/j.joen.2013.12.035. 11. Silva EJNL, Martins JNR, Alcalde MP, Versiani MA. Assessing the fatigue resistance of NiTi instruments: A DSC-based approach to understanding temperature effects. Int Endod J. 2025 May 23. doi: 10.1111/iej.14257. 12. Silva EJNL, Martins JNR, Versiani MA. Advancing insights in contemporary endodontics: Proposing a multimethod approach for comprehensive evaluation of NiTi instruments. Int Endod J. 2024 Feb;57(2):116-118. doi: 10.1111/iej.14009. 13. Zanza A, Seracchiani M, Reda R, Miccoli G, Testarelli L, Di Nardo D. Metallurgical Tests in Endodontics: A Narrative Review. Bioengineering (Basel). 2022 Jan 12;9(1):30. doi: 10.3390/bioengineering9010030. 14. de Menezes SEAC, Batista SM, Lira JOP, de Melo Monteiro GQ. Cyclic Fatigue Resistance of WaveOne Gold, ProDesign R and ProDesign Logic Files in Curved Canals In Vitro. Iran Endod J. 2017 Fall;12(4):468-473. doi: 10.22037/iej.v12i4.17494. 15. Só GB, Siocheta G, Calefi P, Alcalde M, Vivan RR, Duarte MAH, Só MVR, da Rosa RA. Cyclic and torsional fatigue resistance of a new rotary file on a rotary and reciprocating motion. Microsc Res Tech. 2023 Dec;86(12):1635-1641. doi: 10.1002/jemt.24399. 16. de Oliveira Neto RS, Alcalde MP, Titato PCG, Calefi PHS, Ramos CAS, da Silva GF, Vivan RR, Duarte MAH. Shaping ability and cyclic fatigue resistance between Genius ProFlex, ZenFlex, and TruNatomy rotary systems: an experimental study. Restor Dent Endod. 2025 Feb;50(1):e9. doi: 10.5395/rde.2025.50.e9. 17. Alcalde MP, Tanomaru-Filho M, Bramante CM, Duarte MAH, Guerreiro-Tanomaru JM, Camilo-Pinto J, Só MVR, Vivan RR. Cyclic and Torsional Fatigue Resistance of Reciprocating Single Files Manufactured by Different Nickel-titanium Alloys. J Endod. 2017 Jul;43(7):1186-1191. doi: 10.1016/j.joen.2017.03.008. 18. ASTM:F2004-17. Standard test method for transformation temperature of nickel- titanium alloys by thermal analysis. 2004. p. 1–5. 19. Teves Cordova AVI, Alcalde MP, Vivan RR, Zevallos Quiroz CA, Guzmán HP, Calefi PS, Pinto BO, Correa DRN, Grandini CR, Duarte MAH. Impact of Metallurgical Features on the Cyclic and Torsional Properties of Five Reciprocating Instruments Manufactured With Blue Thermal Treatment. Microsc Res Tech. 2025 Feb 28. doi: 10.1002/jemt.24840. 20. Brantley WA, Svec TA, Iijima M, Powers JM, Grentzer TH. Differential scanning calorimetric studies of nickel titanium rotary endodontic instruments. J Endod. 2002 Aug;28(8):567-72. doi: 10.1097/00004770-200208000-00001. 21. Sánchez P, Vidi B, Mena-Alvarez J, Gil J, Rico C, Aragoneses JM. Effect of Stabilized Martensite on the Long-Term Performance of Superelastic NiTi Endodontic Files. Materials (Basel). 2023 May 31;16(11):4089. doi: 10.3390/ma16114089. 22. Olbricht J, Yawny A, Pelegrina JL, Eggeler G, Yardley VA. Characteristics of the stress-induced formation of R-phase in ultrafine-grained NiTi shape memory wire. Journal of Alloys and Compounds. 2013 Dec ;579:249–52. doi: 10.1016/j.jallcom.2013.06.056 23. Bürklein S, Zupanc L, Donnermeyer D, Tegtmeyer K, Schäfer E. Effect of Core Mass and Alloy on Cyclic Fatigue Resistance of Different Nickel-Titanium Endodontic Instruments in Matching Artificial Canals. Materials (Basel). 2021 Oct 1;14(19):5734. doi: 10.3390/ma14195734. 24. Grande NM, Plotino G, Pecci R, Bedini R, Malagnino VA, Somma F. Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems. Int Endod J. 2006 Oct;39(10):755-63. doi: 10.1111/j.1365-2591.2006.01143.x. 25. Dederich DN, Zakariasen KL. The effects of cyclical axial motion on rotary endodontic instrument fatigue. Oral Surg Oral Med Oral Pathol. 1986 Feb;61(2):192-6. doi: 10.1016/0030-4220(86)90186-6. 26. Hülsmann M, Donnermeyer D, Schäfer E. A critical appraisal of studies on cyclic fatigue resistance of engine-driven endodontic instruments. Int Endod J. 2019 Oct;52(10):1427-1445. doi: 10.1111/iej.13182. 27. Pedullà E, Lizio A, Scibilia M, Grande NM, Plotino G, Boninelli S, Rapisarda E, Lo Giudice G. Cyclic fatigue resistance of two nickel-titanium rotary instruments in interrupted rotation. Int Endod J. 2017 Feb;50(2):194-201. doi: 10.1111/iej.12609. 28. Xu X, Eng M, Zheng Y, Eng D. Comparative study of torsional and bending properties for six models of nickel-titanium root canal instruments with different cross-sections. J Endod. 2006 Apr;32(4):372-5. doi: 10.1016/j.joen.2005.08.012. 29. Rubio J, Zarzosa JI, Pallarés A. A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study. Acta Stomatol Croat. 2019 Mar;53(1):28-36. doi: 10.15644/asc53/1/3. 30. Faus-Llácer V, Hamoud-Kharrat N, Marhuenda Ramos MT, Faus-Matoses I, Zubizarreta-Macho Á, Ruiz Sánchez C, Faus-Matoses V. Influence of the Geometrical Cross-Section Design on the Dynamic Cyclic Fatigue Resistance of NiTi Endodontic Rotary Files-An In Vitro Study. J Clin Med. 2021 Oct 14;10(20):4713. doi: 10.3390/jcm10204713. 31. Kaval ME, Capar ID, Ertas H, Sen BH. Comparative evaluation of cyclic fatigue resistance of four different nickel-titanium rotary files with different cross-sectional designs and alloy properties. Clin Oral Investig. 2017 Jun;21(5):1527-1530. doi: 10.1007/s00784-016-1917-x. TABLES: Table 1: Time (seconds) and number of cycles to fracture (NCF) of the instruments. 20°C 36°C 20°C 36°C Logic 368.5±37.07 Aa 272.5±142.5 Aa 2457±247.2 Aa 1817±950.0 Aa Solla Colors 170.8±50.88 Ab 165.3±21.49 Aa 1139±339.2 Ab 1102±143.3 Aa Solla Purple 170.7±53.90 Ab 156.3±9.60 Aa 1138±359.3 Ab 1042±64.0 Aa Values are presented as mean ± standard deviation. Different lowercase letters indicate statistically significant differences between instruments within the same temperature condition (p<0.05). Different capital letters indicate significant differences between the same instrument tested at different temperatures (room temperature vs. simulated body temperature, p<0.05). FIGURES: Figure 1: Percentage reduction of cyclic fatigue resistance of the instruments when exposed at simulated body temperature (36°C ± 1°C). Figure 2: Scanning electron microscopy (SEM) images (200× magnification) of fractured surfaces after cyclic fatigue testing. (A) ProDesign Logic at room temperature (20°C), (B) ProDesign Logic at simulated body temperature (36°C), (C) Solla Colors at room temperature, (D) Solla Colors at simulated body temperature, (E) Solla Purple at room temperature, and (F) Solla Purple at simulated body temperature. Red arrows identify the crack origins. All images exhibit characteristic fatigue striations and microstructural features associated with cyclic failure. Figure 3: Differential scanning calorimetry (DSC) heating curves of the tested endodontic instruments (read from left to right). The blue curve represents Solla Colors, the purple curve represents Solla Purple, and the green curve represents ProDesign Logic. Critical phase transformation temperatures are annotated: onset (As, austenite start), peak (Ap, austenite peak), and endset (Af, austenite finish). Information & Authors Information Version history V1 Version 1 17 July 2025 Peer review timeline Published Journal of Clinical and Experimental Dentistry Version of Record 1 Jan 2026 Published Copyright This work is licensed under a Non Exclusive No Reuse License. Keywords cyclic fatigue differential scanning calorimetry nickel-titanium alloy Authors Affiliations Raimundo Sales de Oliveira Neto 0000-0002-4726-8106 [email protected] Universidade de Sao Paulo Faculdade de Odontologia de Bauru View all articles by this author Guilherme Ferreira da Silva Universidade de Sao Paulo Faculdade de Odontologia de Bauru View all articles by this author Diego Rafael Nespeque Correa Universidade Estadual Paulista Julio de Mesquita Filho - Campus de Bauru View all articles by this author Rodrigo Vivan 0000-0002-0419-5699 Universidade de Sao Paulo Faculdade de Odontologia de Bauru View all articles by this author Murilo Priori Alcalde Universidade de Sao Paulo Faculdade de Odontologia de Bauru View all articles by this author Marco Antonio Duarte 0000-0003-3051-737X Universidade de Sao Paulo Faculdade de Odontologia de Bauru View all articles by this author Metrics & Citations Metrics Article Usage 164 views 141 downloads .FvxKWukQNSOunydq8rnd { width: 100px; } Citations Download citation Raimundo Sales de Oliveira Neto, Guilherme Ferreira da Silva, Diego Rafael Nespeque Correa, et al. Impact of metallurgical and geometric characteristics on cyclic fatigue resistance of three thermally treated NiTi rotary instruments. Authorea . 17 July 2025. DOI: https://doi.org/10.22541/au.175273951.15422748/v1 If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. Simply select your manager software from the list below and click Download. For more information or tips please see 'Downloading to a citation manager' in the Help menu . Format Please select one from the list RIS (ProCite, Reference Manager) EndNote BibTex Medlars RefWorks Direct import Tips for downloading citations document.getElementById('citMgrHelpLink').addEventListener('click', function() { popupHelp(this.href); return false; }); $(".js__slcInclude").on("change", function(e){ if ($(this).val() == 'refworks') $('#direct').prop("checked", false); $('#direct').prop("disabled", ($(this).val() == 'refworks')); }); View Options View options PDF View PDF Figures Tables Media Share Share Share article link Copy Link Copied! Copying failed. Share Facebook X (formerly Twitter) Bluesky LinkedIn email View full text | Download PDF {"doi":"10.22541/au.175273951.15422748/v1","type":"Article"} Now Reading: Share Figures Tables Close figure viewer Back to article Figure title goes here Change zoom level Go to figure location within the article Download figure Toggle share panel Toggle share panel Share Toggle information panel Toggle information panel Go to previous graphic Go to next graphic Go to previous table Go to next table All figures All tables View all material View all material xrefBack.goTo xrefBack.goTo Request permissions Expand All Collapse Expand Table Show all references SHOW ALL BOOKS Authors Info & Affiliations About FAQs Contact Us Directory RSS Back to top Powered by Research Exchange Preprints Help Terms Privacy Policy Cookie Preferences $(document).ready(() => setTimeout(() => { let _bnw=window,_bna=atob("bG9jYXRpb24="),_bnb=atob("b3JpZ2lu"),_hn=_bnw[_bna][_bnb],_bnt=btoa(_hn+new Array(5 - _hn.length % 4).join(" ")); $.get("/resource/lodash?t="+_bnt); },4000)); (function(){function c(){var b=a.contentDocument||a.contentWindow.document;if(b){var d=b.createElement('script');d.innerHTML="window.__CF$cv$params={r:'a0286113980f300f',t:'MTc3OTkyMDEzNg=='};var a=document.createElement('script');a.src='/cdn-cgi/challenge-platform/scripts/jsd/main.js';document.getElementsByTagName('head')[0].appendChild(a);";b.getElementsByTagName('head')[0].appendChild(d)}}if(document.body){var a=document.createElement('iframe');a.height=1;a.width=1;a.style.position='absolute';a.style.top=0;a.style.left=0;a.style.border='none';a.style.visibility='hidden';document.body.appendChild(a);if('loading'!==document.readyState)c();else if(window.addEventListener)document.addEventListener('DOMContentLoaded',c);else{var e=document.onreadystatechange||function(){};document.onreadystatechange=function(b){e(b);'loading'!==document.readyState&&(document.onreadystatechange=e,c())}}}})();

Text is read by the "Ask this paper" AI Q&A widget below. Extraction quality varies by source — PMC NXML preserves structure cleanly, OA-HTML may include some navigation residue, and OA-PDF can have broken hyphenation. The publisher copy (via DOI) is the canonical version.

My notes (saved in your browser only)

Ask this paper AI returns verbatim quotes from the full text · source: preprint-html

Answers must be backed by verbatim quotes from this paper's full text. Hallucinated quotes are dropped automatically; if no verbatim passage answers the question, we say so. How this works

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2025) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.

Source provenance

europepmc
last seen: 2026-05-20T01:45:00.602351+00:00