Agreement between a mobile applet-based visual acuity self-test program and the conventional method for distance and near visual acuity test

preprint OA: closed
View at publisher

Abstract

Abstract PURPOSE: To investigate the agreement between a mobile applet-based visual acuity (VA) self-test program and the conventional VA tests.METHODS: This consecutive case series study included 121 children and adults (242 eyes). Patients were classified into three groups according to age (children, adolescents, and adults). They underwent uncorrected distance visual acuity (UDVA) testing, distance visual acuity with frame spectacle (DVA with FS) testing at 2.5-m distance, uncorrected near visual acuity (UNVA) testing, and near visual acuity with frame spectacle (NVA with FS) testing at a 0.4-m distance using a mobile applet-based VA self-test program and conventional VA tests in two eyes. RESULTS: Correlations among UDVA, DVA with FS, UNVA, and NVA with FS between the two methods were significant in all subjects (all P<0.001). The intraclass correlation coefficient were 0.960. 0.845, 0.960, and 0.669, respectively (all P<0.001). The proportions of bias outside the 95% confidence interval limit of agreements were 6.20%, 4.82%, 7.08%, and 6.10%, respectively. There were significant differences in NVA with FS between the two methods (P<0.05) in the adolescent group, but no differences in measured parameters were found among children and adult groups.CONCLUSIONS: There is good agreement between the mobile applet-based VA self-test program and the conventional VA tests. The VA self-test has good practical value especially in the current pandemic, allowing self-screening of visual acuity, myopia control, and remote management of visual impairment in oculopathy.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00