Optical Biosensor Based on Porous Silicon and Tamm Plasmon Polariton for Detection of CagA Antigen of Helicobacter Pylori
preprint
OA: closed
Abstract
Helicobacter Pylori (H. Pylori) is a common pathogen with high prevalence of infection in human populations. Diagnosis of H. Pylori infection is critical for its treatment, eradication, and prognosis. Biosensors have been demonstrated to be powerful for rapid onsite detection of pathogens, particularly for point-of-care test (POCT) scenarios. In this work, we propose a novel optical biosensor, based on nanomaterial porous silicon (PSi) and photonic surface state Tamm Plasmon Polariton (TPP), for detection of cytotoxin-associated antigen A (CagA) antigen of H. Pylori bacterium. We fabricated the PSi TPP biosensor, analysed its optical characteristics and demonstrated through experiments towards sensing of CagA antigen, that TPP biosensor has sensitivity of 100 pm/(ng/ml), limit of detection of 0.01 ng/ml, and specificity in terms of positive-to-negative ratio of more than six. From these performance factors, it can be concluded that TPP biosensor can serve as an effective tool for diagnosis of H. Pylori infection, either in analytical labs or in POCT applications.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2024) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.
Source provenance
- europepmc
- last seen: 2026-05-20T01:45:00.602351+00:00