Abstract Speed and efficiency of data collection and image processing in cryo electron microscopy have increased over the last decade. However, cryo specimen preparation techniques have lagged behind and faster, more reproducible specimen preparation devices are needed. Here we present a new vitrification device with highly automated sample handling, requiring only limited user interaction. Moreover, the device allows inspection of thin films using light microscopy, since excess liquid is removed through suction by tubes, not blotting paper. In combination with dew-point control, this enables thin film preparation in a controlled and reproducible manner. The advantage is that quality of the prepared cryo specimen is characterized prior to electron microscopy data acquisition. Practicality and performance of the device are illustrated by experimental results obtained by vitrification of protein suspensions, lipid vesicles, bacterial and human cells, followed by imaged using single particle analysis, cryo electron tomography and cryo correlated light and electron microscopy.