“Similar performances but markedly different triaging thresholds in three CAD4TB versions risk systematic errors in TB screening programs”
preprint
OA: closed
Abstract
Tuberculosis (TB) screening programs may apply computer-aided diagnostic (CAD) tools on chest radiographs to select people for microbiological sputum examination using a pre-selected triaging threshold. CAD software-updates are frequently introduced and it is unknown whether their use requires adjustment of triaging thresholds. In a community-based screening program in South Africa, we compared the scores between the three recent CAD4TB versions (v5, v6, and v7) and assessed their performance to identify microbiologically-confirmed TB. The performance of all versions was similar (v5: AUC 0.78, v6: AUC 0.79, v7: AUC 0.80; p-values>0.05), but along a 0-100 point scale, each had markedly different score distributions and optimal triaging thresholds (v5: 40, v6: 47, v7: 20). This has the potential to cause confusion within TB screening programs as these tools are increasingly adopted and new versions released. Independent guidance for adapting CAD triaging thresholds for frequently released software updates is needed.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.
Source provenance
- europepmc
- last seen: 2026-05-19T01:45:01.086888+00:00