Oxygen Sensor-Guided Fine Needle Biopsy Studies of Human Cancer Xenografts in Mice

preprint OA: closed
📄 Open PDF View at publisher

Abstract

An oxygen sensor-mounted fine-needle biopsy tool was used for in vivo measurement of oxygen levels in tumor xenografts. The system provides a means of measuring the oxygen content in harvested tumor tissue from specific locations. Oxygen in human tumor xenografts in a murine model was observed for over 1 min. Tissues were mapped in relation to oxygen tension (pO 2 ) readings and sampled for conventional cytological examination. Careful modeling of the pO 2 readings over 60 seconds yielded a diffusion coefficient for oxygen at the sensor tip, providing additional diagnostic information about the tissue before sampling. Oxygen level measurement may provide a useful adjunct to the use of biomarkers in tumor diagnosis.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2024) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.

Source provenance

europepmc
last seen: 2026-05-20T01:45:00.602351+00:00