Introduction
The worldwide global demand for N95 filtering face respirators (FFRs) for healthcare professionals will
quickly outpace supply as the global COVID19 pandemic continues. Several protocols for disinfecting and
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
NOTE: This preprint reports new research that has not been certified by peer review and should not be used to guide clinical practice.
reusing FFRs that are normally for one-time use only have been proposed,1 and the FDA has authorized
emergency use of vapor hydrogen peroxide (VHP) as a broad-spectrum sterilant for re-use of FFRs2
Because VHP requires centralized operations and specialized equipment, it may not be available in all
emergency situations.
In this study, we investigate the use of heat treatment at 75 oC as a potential method for recycling N95
FFRs. Although not a broad-spectrum sterilant, heat treatments may be easily adapted in a variety of
settings and filter efficiency of FFRs has been reported to be maintained after one hour at 90 oC under
dry heat for 1 hour.3 A very recent report found that filter efficiency of meltblown polypropylene filter
fabric was not affected by either dry or moist heat at ≤85 oC for up to 50 cycles.4 Therefore, the heating
schedule used is cool enough to maintain the filter efficiency of the FFR.
In addition to efficiently capturing aerosolized particles in the filter elements, the FFR must also form a
good seal around the nose and face of the wearer. Reports are now emerging on the quality of the seal
via quantitative fit testing of FFRs following decontamination protocols. A recent evaluation of typical
hospital decontamination protocols showed that quantitative fit was retained following multiple cycles
of ethylene oxide, VHP, and (for some FFR models) autoclaving.5 Another recent study, released during
the preparation of the present report, has reported that quantitative fit factor of FFRs is retained for
two cycles of dry heat at 70 oC but fail thereafter.6 In this study, we performed standard OSHA
quantitative fit testing on FFRs that underwent both one and ten heating cycles at 75 oC under both dry
and humid conditions.
Although heating at this temperature is not a broad-spectrum sterilant, previous studies in liquid media
have reported SARS-CoV-1 inactivation at 60 oC/30 minutes (≥5 log reduction in viral activity) 7 and at 75
oC/15 minutes (≥4 log reduction).8 Most recently, SARS-CoV-2 has been shown to be inactivated (≥6.8
1 See summary on the CDC website, https://www.cdc.gov/coronavirus/2019-ncov/hcp/ppe-
strategy/decontamination-reuse-respirators.html (retrieved 4/14/20)
2 For examples, see FDA letter to STERIS Corporation, https://www.fda.gov/media/136843/download and FDA
letter to Battelle Memorial Institute, https://www.fda.gov/media/136529/download (retrieved 4/14/20)
3 Viscusi, et. al., The Annals of Occupational Hygiene, 53, 815 (2009)
4 Liao, et. al., MedRxiv pre-print, https://doi.org/10.1101/2020.04.01.20050443
5 Kumar, et. al., MedRxiv pre-print, https://doi.org/10.1101/2020.04.05.20049346
6 Fischer, et. al., MedRxiv pre-print, https://doi.org/10.1101/2020.04.11.20062018
7 Rabenau, et. al., Medical Microbiology and Immunology, 194, 1 (2005)
8 Darnell, et. al., Journal of Virological Methods, 121, 85 (2004)
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
log reduction) by heating to 70 oC for 5 minutes in liquid media.9 While promising, confirmation that the
virus can be inactivated after deposition on FFR filter material is needed to investigate FFR re-use. A
study released during the preparation of the present report found that SARS-CoV-2 on FFR filter material
could be inactivated (≥4 log reduction) after dry heating at 70oC for 60 minutes.6 To simulate
decontamination under use conditions, we also measured the activity of a coronavirus (in this case,
mouse hepatitis virus (MHV) as a surrogate to SARS-CoV-2), initially inoculated onto FFR filter material
and titered using a TCID50 protocol before and after dry heat treatment at 75 oC.
Method
Heat treatments
Prior to heat treatment, a volunteer participant briefly fitted a new, unused FFR (3M Model 8210) to
their face and nose structure to simulate a first-time use. FFRs were then loaded into a sterilization
pouch (CrossTex Sure-Check, SCL12182).
For dry heating, the FFRs were loaded into an oven (Cascade Tek TFO-1) pre-warmed to 75 oC and
operating at ambient humidity. We estimate that ambient humidity at 75 oC is <5% relative humidity.
The oven door was held open for <30 seconds during loading and experienced a temperature loss of <2
oC. Some FFRs were instrumented with a thermocouple as shown in Figure 1a to monitor the thermal
history. Figure 1b shows the thermal history of several FFRs from both the top and bottom shelf of the
oven. The FFRs heated to 75 oC in approximately 5 minutes and this temperature was maintained to ±3
oC throughout the 30-minute treatment. FFRs that were subjected to multiple cycles were allowed to
cool to room temperature but were not removed from the pouch or re-donned prior to the next heating
cycle.
For humid heating, FFRs were loaded into an environmental chamber (Espec SH-242). We visually
confirmed that the pouch allowed steam to permeate to the surface of the FFR during experimental
process development. The chamber was programmed to ramp over 15 minutes to 75 oC and 90%
relative humidity, hold those conditions for 30 minutes, and ramp back down to room temperature and
humidity over 15 minutes. For FFRs treated for multiple cycles, these conditions were held for seven
hours before the next heat and humidity cycle began. The FFRs were not re-donned between cycles.
9 Chin, et. al., The Lancet Microbe, in press (2020). https://doi.org/10.1016/S2666-5247(20)30003-3
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
Neither oven used has exposed heating elements, as the infrared radiation emitted from exposed
heating elements may be absorbed by the polymer components of the FFR and cause rapid heating and
damage to the FFR (see Results and Dscussion).
Figure 1: a) Thermocouple attached to an FFR with Kapton tape. This picture shows a 3M Model 8511
N95 FFR used for validating the process. All fit tests were performed with a 3M Model 8210 N95 FFR.
b) Thermal history of FFRs on top (blue) and bottom (red) shelf of oven over multiple cycles.
Quantitative fit test
Quantitative fit tests were performed per OSHA fit test protocol 1910.134, Appendix A using a TSI
PortaCount Respirator Fit Tester 8038 (TSI Instruments, Shoreview MN).10 This instrument does not test
the effectiveness of the filter, which previous studies have validated up to 90 oC.3,4 The tests performed
in this study quantify changes to the sealing surface insofar as their effect on fit. The passing criteria for
the fit test was a quantitative fit factor of 100, which is the OSHA criteria for new, unused FFRs.
A sodium chloride (NaCl) aerosol generator and two humidifiers with tunable droplet size set at the
smallest droplet size setting were used to achieve background levels of aerosol. The PortaCount Fit
Tester 8038 uses a pre-selector to ensure that the detected aerosols reflect those penetrating the
sealing surfaces and not the filter itself. Aerosolized particle counts were compared inside and outside
10 https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.134AppA (retrieved 4/14/20)
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
the FFR while the participant performed a series of seven 60-second and one 15-second breathing,
movement, and speaking exercises including:
• Normal breathing
• Deep breathing
• Head side-to-side
• Head up and down
• Talking
• Grimacing (15 seconds)
• Bending over/reaching down
• Normal breathing
Samples were only fit-tested on the volunteer who originally donned and doffed the FFR prior to the
heat treatment. Fit tests on samples 01-08 (Volunteer A) were performed sequentially, but the order of
fit tests on samples 09-16 (Volunteer B) was randomized.
A total of 18 FFRs were treated in this study as shown in Table 1. For each heating schedule tested, the
samples were cycled either once or ten times. Because the fit test requires inserting a metal rivet into
the FFR, we considered the test destructive. Therefore, pre- and post-treatment measurements were
not possible on the same FFR and control measurements were instead made for each volunteer with
new, unused FFRs without heat treatment.
Table 1: Sample set of FFRs and conditions tested in this study.
Sample Humidity Heating cycles Volunteer
01, 02, 03, 04 Dry 1x A
05, 06, 07, 08 Dry 10x A
17 (control) N/A None A
09, 10, 11, 12 90% RH 1x B
13, 14, 15, 16 90% RH 10x B
18 (control) N/A None B
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
Viral activity measurements
Twenty microliters of viral stock diluted in Dulbecco’s Modified Eagle Medium (DMEM) media (Thermo
Scientific, Waltham MA) with 10% Fetal Bovine Serum (FBS) and antibiotics was inoculated onto 12
replicate ½ cm2 coupons cut from an FFR (3M Model 8210). The inocula were dried inside sterile petri
plates within a biosafety cabinet for 1 hour, and 2 sets of 3 coupons were placed in 2 autoclave bags
(CrossTex Sure-Check, SCL12182), sealed, and placed in the oven pre-heated to 75 oC (Cascade Tek TFO-
1) . At 30 minutes one bag containing 3 coupons were was removed from the oven and allowed to cool
for 30 minutes, and the second bag containing 3 coupons was removed from the oven at 60 minutes and
allowed to cool for 30 minutes. Control coupons were stored at room temperature and were processed
at the same timepoints as when the corresponding 30- and 60-minute coupons were processed. Thirty
minutes after removal from the oven, each coupon was immersed in 2 ml of media (DMEM, 10% FBS,
and antibiotics) and vortexed intermittently for 10 minutes to dislodge the viral particles from the
coupon into the media. The virus was then titered by TCID50 assays using 17CL-1 cells. Cytopathic effect
for each well was recorded day 3 post-inoculation and TCID50 titer was calculated using the Spearman
and Karber method.11
Results
and discussion
Quantitative fit test results
Results
of the quantitative fit tests are shown in Table 2. A passing score is 100, and the maximum
possible score is 200.
None of the processed samples showed any qualitative change in feel or appearance. One of the elastic
straps on sample 05 snapped upon doffing the FFR after passing the quantitative fit test.
All samples subjected to dry heat cycles passed the quantitative fit tests with a fit factor of >100. After
one heating cycle there was no significant change to the quantitative fit test result for the four tested
samples. After 10 cycles all four tested samples had the maximum score upon fit testing. Due to FFRs 01-
08 being tested sequentially, we are unable to disambiguate whether the improvement in quantitative
fit test results for FFRs treated for ten cycles versus one cycle is due to the heat treatment or an
improvement in donning procedure over time.
11 Hierholzer and Killington, “Virus isolation and quantitation” in Virology Methods Manual (Mahy and Kangro,
eds), 25–47 (1996)
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
Likewise, all samples subjected to moist heat cycles passed the quantitative fit tests with a fit factor
>100. There was no significant change to the quantitative fit test result for the eight samples tested, nor
a correlation between score and number of treatment cycles. No significant difference was observed
between the two volunteers.
Table 2: Results of quantitative fit tests
Sample Humidity Heating cycles Volunteer Fit factor
01 Ambient 1x A 107 (pass)
02 Ambient 1x A 195 (pass)
03 Ambient 1x A 165 (pass)
04 Ambient 1x A 149 (pass)
05 Ambient 10x A 200+ (pass)*
06 Ambient 10x A 200+ (pass)
07 Ambient 10x A 200+ (pass)
08 Ambient 10x A 200+ (pass)
17 (control) N/A None A 181 (pass)
09 90% RH 1x B 200+ (pass)
10 90% RH 1x B 163 (pass)
11 90% RH 1x B 200+ (pass)
12 90% RH 10x B 200+ (pass)
13 90% RH 10x B 200+ (pass)
14 90% RH 10x B 200+ (pass)
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
15 90% RH 10x B 200+ (pass)
16 90% RH 10x B 161 (pass)
18 (control) N/A None B 200+ (pass)
* Strap broke upon doffing
In our initial experiments, we attempted to heat an FFR in an oven with an exposed heating element.
We found that the FFR heated rapidly and showed visible signs of softening and melting. We believe this
is due to the infrared radiation emitted by heating elements, which typically operate at temperatures of
~800-1000 oC. Polymers strongly absorb blackbody radiation (2-3 μm wavelength) emitted by the
heating elements at this temperature. We therefore caution against using any heating method which
exposes the FFR directly to radiation from the heat source.
While we found that the quantitative fit factor of FFRs was not affected by up to ten heating cycles to 75
oC, a recent report by Fischer et. al. found that the quantitative fit factor of FFRs was only retained for 2
cycles of 70 oC. The key difference in the two studies was the treatment of samples between heating
cycles. In the present study, we donned and doffed the FFR a single time but did not simulate donning
and doffing in between heating cycles. In Fischer et. al., the FFR was donned and worn for two hours in
between cycles. This suggests that use duration and number of donning/doffing cycles of the FFR,
perhaps even independent of heat treatment (or other decontamination protocol such as VHP), may
play an important role in the quantitative fit factor of FFRs that are utilized beyond the recommended
single-use. Additional studies are needed to disambiguate the effects of total use duration,
donning/doffing cycles, and decontamination protocol on the quantitative fit factor of FFRs and in
particular on the elastic head and neck straps.
Viral activity results
Dry heat inactivation of MHV, a coronavirus previously used as a SARS-CoV surrogate virus for validating
decontamination protocols,12 was used to confirm that heating to 75oC for 30 min inactivates high titers
of coronavirus on FFR filter material. Diluting MHV in media containing 10% FBS prior to inoculating the
FFR filter material simulated the presence of proteins found in respiratory secretions that may increase
12 Casanova, et. al., Appl. and Environ. Microbiology, 76, 2712 (2010)
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
viral resistance to inactivation. 12 No viral activity was detected after the heat treatment. TCID50
measurements on the coupons used in these studies show ≥ 6 log reduction after heating to 75 oC for
either 30 or 60 minutes as compared to room temperature activity for similar time periods as seen in
Figure 2. These results are generally consistent with a recent report that 70 oC dry heat for 60 minutes
inactivates SARS-CoV-2 on FFR filter material (≥ 4 log reduction)6 and provide further evidence that dry
heat provides relatively rapid inactivation of coronaviruses on FFR filter material, especially as compared
to untreated FFRs stored for the same amount of time at room temperature.
Figure 2: Viral activity after 30 or 60 minutes of dry heat (75 oC) and after 30 or 60 minutes at room
temperature (RT). Heated samples showed a ≥ 6 log reduction in viral activity.
Conclusion
We subjected N95 FFRs to one and ten heating cycles up to 75 oC under dry and humid (90% relative
humidity) conditions. Quantitative fit testing did not show any degradation in the fit factor, showing that
the form and fit of these FFRs was retained following the heat treatments. We also found that dry
heating to 75 oC reduced the viral activity of MHV on an FFR filter element by ≥ 6 log. These
temperatures and times have already been shown to inactivate other coronaviruses (including SARS-
CoV-2) in liquid media7,8,9 and FFR filter material,6 and our study provides further evidence that virus
dried on FFR filter material can also be inactivated at by heating. Previous studies have shown that
these temperatures have do not to negatively impact filter efficiency and airflow of melt-blown
propylene filter elements found in N95 FFRs. 3,4
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
The emerging evidence shows that heat treatments may be used as an effective method for re-using
N95 FFRs. It should be noted that heat treatment is not a broad-spectrum sterilant and that N95 FFRs
are normally meant for one-time use. However, in emergency situations heat treatments specifically to
inactivate coronaviruses may be developed using commonly available equipment (incubators, blanket
warmers, clothes dryers, ovens, etc.). Heat treatments may therefore serve as a rapid method for re-
use of FFRs in areas where FFRs are in critically short supply, specialized decontamination equipment
(e.g., VHP) is not available, and surface sterilization (e.g., ultraviolet germicidal irradiation (UVGI)) is
insufficient. However, important questions remain on the retention of fit factor after long-term use and
repeated donning/doffing cycles to help resolve conflicting data on the number of cycles for which
quantitative fit factor can be maintained.
Acknowledgment
This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore
National Laboratory under Contract DE-AC52-07NA27344. We kindly thank Dr. Susan Weiss at University
of Pennsylvania Perelman School of Medicine for providing MHV strain A59.
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 22, 2020. ; https://doi.org/10.1101/2020.04.15.20065755doi: medRxiv preprint
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.