Video-rate remote refocusing through continuous oscillation of a membrane deformable mirror

preprint OA: closed
📄 Open PDF View at publisher

Abstract

This paper presents the use of a deformable mirror (DM) configured to rapidly refocus a microscope employing a high numerical aperture objective lens. An Alpao DM97-15 membrane DM was used to refocus a 40×/0.80 NA water-immersion objective through a defocus range of −50 to 50 μm at 26.3 sweeps per second. We achieved imaging with a mean Strehl metric of > 0.6 over a field of view in the sample of 200×200 μm 2 over a defocus range of 77 μm. We describe an optimisation procedure where the mirror is swept continuously in order to avoid known problems of hysteresis associated with the membrane DM employed. This work demonstrates that a DM-based refocusing system could in the future be used in light-sheet fluorescence microscopes to achieve video-rate volumetric imaging.