Single-Point Left Ventricular Pacing Optimized by ECG Results in Better Resynchronization than Multipoint Pacing
preprint
OA: closed
Abstract
Background: Multipoint pacing (MPP) in cardiac resynchronization therapy (CRT) activates the left ventricle from two locations, thereby shortening the QRS duration and enabling better resynchronization; however, compared with conventional CRT, MPP reduces battery longevity. On the other hand, electrocardiogram-based optimization using the fusion-optimized intervals (FOI) method achieves more significant reverse remodeling than nominal CRT programming. Our study aimed to determine whether MPP could attain better resynchronization than single-point pacing (SPP) optimized by FOI. Methods: This prospective study included 32 consecutive patients who successfully received CRT devices with MPP capabilities. After implantation, the QRS duration was measured during intrinsic rhythm and with three pacing configurations: MPP, SPP-FOI, and MPP-FOI. In 14 patients, biventricular activation times (by electrocardiographic imaging, ECGI) were obtained during intrinsic rhythm and for each pacing configuration to validate the findings. Device battery longevity was estimated at the 45-day follow-up. Results: The SPP-FOI method achieved greater QRS shortening than MPP (-56±16 vs. -42±17 ms, P<0.001). Adding MPP to the best FOI programming did not result in further shortening (MPP-FOI: - 58±14 ms, P=0.69). Furthermore, the biventricular activation time according to ECGI was also shorter with SPP-FOI than with MPP. The estimated battery longevity was longer with SPP than with MPP (8.1±2.3 vs. 6.3±2.0 years, P=0.03). Conclusions: SPP optimized by FOI resulted in better resynchronization and longer battery duration than MPP.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.
Source provenance
- europepmc
- last seen: 2026-05-19T01:45:01.086888+00:00