Enzyme Immunoassays May Be Inaccurate: Familiar Correlates of Male Testosterone (T) Replicate in Britons when T is Measured by Mass Spectrometry but Do Not Replicate in Americans When T is Measured by Enzyme Immunoassay
preprint
OA: closed
AI-generated summary
Salivary testosterone correlations with age, time of day, and obesity replicated in Britons using mass spectrometry but failed in Americans using enzyme immunoassay, suggesting EIA inaccuracy.
One-sentence paraphrase of the abstract; not a substitute for reading it. No clinical advice. How this works
Abstract
ABSTRACT Several correlations have been reported between men’s testosterone (T) and other characteristics, e.g., T declines during the day, declines with obesity, and declines with advancing age. We asked if these relationships hold in older men when T is assayed from saliva. Here seven familiar correlations are tested among older American men, their salivary T measured by enzyme immunoassay (EIA). Some correlations can also be tested among older British men, their salivary T assayed by liquid chromatography-tandem mass spectroscopy (LC-MS/MS), a technique noted for its specificity. Most of our attempts at replication failed in the American data but succeeded in the British data. We conclude that failure to replicate in Americans is likely due to inaccuracy of EIA values for T, and that most T correlations hold true in older men when salivary T is accurately measured by LC-MS/MS.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.
Source provenance
- europepmc