Preventing Laboratory-acquired Brucellosis in the Era of MALDI-TOF Technology and Molecular Tests
preprint
OA: closed
Abstract
Brucellosis is one of the most common etiologies of laboratory-acquired infections worldwide, and handling of living brucellae should be performed in a Class II biological safety cabinet. The low infecting dose, multiple portals of entry to the body, the great variety of potentially contaminated specimens, and the unspecific clinical manifestations of human infections facilitate the unintentional transmission of brucellae to laboratory personnel. Work accidents such as spillage of culture media cause only a small minority of exposures, whereas >80% of events result from unfamiliarity with the phenotypic features of the genus, misidentification of isolates, and unsafe laboratory practices such as aerosolization of bacteria and working on an open bench without protective goggles or gloves. Although the bacteriological diagnosis of brucellae by traditional methods is simple, the Gram stain and the biochemical profile of the organism, as determined by commercial kits, can be misleading, resulting in inadvertent exposure and contagion. The use of novel identification technologies is not hazard-free. The MALDI-TOF technology requires an initial bacterial inactivation step, while the instruments’ reference database may misidentify Brucella as belonging to other Gram-negative species. The rapid identification by the FISH method mistakes brucellar isolates for members of the closely related Ochrobactrum genus.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.
Source provenance
- europepmc
- last seen: 2026-05-19T01:45:01.086888+00:00