Methods
From March 2020 to May 2021, we conducted a case-control study (IUD in adequate position vs. displaced) tracing MC exposure since IUD insertion. Participation in the study was systematically offered to women who consulted for systematic follow-up of their IUD, or pain/symptoms related to IUD, pregnancy on IUD or IUD expulsion in two primary care medical facilities in Paris. General practitioners and midwives recruited the patients. Patients with abnormal uterine cavity such as a fibroid or adenomyosis and patients with pregnancy on displaced IUD were excluded.
IUD position was assessed by standard, 2D, transvaginal ultrasound (TVU) for all patients except for the one reporting an IUD expulsion. There is no specific recommendation on the ultrasound criteria for a well-positioned IUD. In this context, results were classified as “IUD in adequate position in the uterine cavity” (including pregnancy on an IUD seemingly in proper position i.e. non-cervically displaced) versus “IUD in a non-adequate position” (including expelled IUD and IUD in the lower uterine segment or cervix).
Data were collected by the investigators, by interviewing the patient, regarding known and potential risk factors of IUD displacement: age, parity, past pregnancies, BMI, type of IUD, menorrhagia, dysmenorrhea, history of IUD expulsion and surgical abortion within 6 weeks before IUD insertion.
MC use and exposure was collected using a standardized investigator-administered questionnaire. The use of MC was defined by the answer “yes” to the question: “Has the patient used a MC since IUD insertion?“. To describe the different uses of MC that women could have, its utilization was classified as “always” (no use of other menstrual protection than the cup), “almost daily” (almost every day of menstruation but can occasionally have used other protections), “regularly” (at least 1 day/cycle or not every cycle but several days), “occasionally” (e.g.: for an activity), “anecdotally” (e.g.: try once) or “never”. The type of use could vary with time, so women could describe their type of use for several periods of time. For each period of time, the type of use, the number of cycles, and the approximate number of MC withdrawal per cycle were collected. An exposure score was calculated by multiplying the number of cycles since the IUD insertion by the type of exposure for each MC period of use, semi-quantitatively estimated as follows: “Always” worth 5 pts, “almost daily” 4 pts, “regularly” 3 pts, “occasionally” 2 pts, “anecdotally” 1 pts and “never” 0 pts. All periods scores were summed up to get a final score.
Assuming an IUD displacement of 15% among MC users and 5% among non-users and, a MC use by at least 10% of women, the required sample size was estimated at 740 to obtain a power of 80% and an alpha risk of 5%. Statistical analyses were performed by the Clinical Research Department of the Adolphe de Rothschild Foundation. Wilcoxon tests and Chi2 tests (or Fisher exact tests) were used for univariate comparisons of continuous parameters and qualitative parameters, respectively. Univariate logistic regressions were performed to assess the association between IUD position (adequate or not) and the other variables. For multivariate analysis, regression was adjusted for factors selected with a stepwise variable selection. Odds Ratios were calculated with a 95% confidence interval. Analysis was performed on available data. All statistical analyses were performed using R software version 4.0.3. ( www.r-project.org ).
Results
Data from 769 consultations were collected for this study (Fig. 1 ). Among them, 38 were not included in the analysis: 13 because of abnormal uterine cavity, 3 for pregnancy on displaced IUD, 2 for absence of TVU realisation, 1 for missing data regarding US result, 1 for presence of two IUDs in the uterine cavity, 2 for missing data regarding MC use and 16 for consultations checking for the same IUD and for which only data of the most recent consultation were kept for the analysis. Thus, the analysis included 731 IUD follow-up consultations.
Fig. 1 IUD follow-up consultations inclusion flowchart
IUD follow-up consultations inclusion flowchart
Table 1 provides the main characteristics of the study population. Median age was 27 years old, nulliparity was reported in 89.9% of the cases and never having been pregnant in 81.4%. Copper IUDs were in the vast majority in our population (76.6%). The most frequent copper IUD was the UT 380 ® Short (Mona Lisa NT Cu 380 mini, CCD UT 380 short, 7 MED USHA, EUROMEDIAL 380 CU mini, GYNEAS T 380 CU Plus mini): (93% of copper IUDs). Hormonal IUDs were Kyleena ® (levonorgestrel 19,5 mg) for 47%, Jaydess ® (levonorgestrel 13,5 mg) for 43% and Mirena ® (levonorgestrel 52 mg) for 9%. The median time since IUD insertion was 7.7 months, with a wide range of situations as the minimum was 6 days and the maximum 5 years and 8 months.
On the 731 consultations, 18.6% of women had used a MC. Women using a MC had been using their IUD for a longer time (median of 15.4 months vs. 5.5 months for women not using a MC, p < 0.001), were more likely to have never been pregnant (88.2% vs. 79.8%, p = 0.023), were more often using a copper IUD (94.1% vs. 72.6%, p < 0.001) and were more likely to report heavy menstrual bleeding since IUD insertion (47% vs. 35.7%, p = 0.016).
IUD position in the uterus cavity on TVU was adequate in 93.6% of cases. Adequate position without pregnancy ( n = 680, 93%) was the most common but 4 (0.5%) pregnancies with IUD in seemingly proper position i.e. non-cervically displaced were observed. We noted 37 (5%) displaced IUD including 4 with pregnancies. IUD expulsion accounted for 13 cases (1.8%).
Table 1 Characteristics of the study population Median (IQR) or N (%) Total sample ( N = 731) Use of MC ( N = 136) No use of MC ( N = 595) p value Age (years) 27 (25–29) 27 (25–30) 27 (25–29) 0.108 BMI (kg/m 2 ) 21.3 (19.8–23.4) 21.3 [20.0-23.3] 21.2 [19.7–23.4] 0.915 NA
6
0
6
Nulliparity 657 (89.9%) 124 (91.2%) 533 (89.6%) 0.578 Never been pregnant 595 (81.4%) 120 (88.2%) 475 (79.8%) 0.023 Surgical abortion within 6 weeks of IUD insertion 26 (3.6%) 5 (3.7%) 21 (3.5%) 0.933 History of IUD expulsion 56 (7.7%) 15 (11.0%) 41 (6.9%) 0.102 IUD type : Hormonal Copper 171 (23.4%) 8 (5.9%) 163 (27.4%) < 0.001 560 (76.6%) 128 (94.1%) 432 (72.6%) Time since IUD insertion (months) 7.7 (2.6–23.5) 15.4 (4.2–29.8) 5.2 (2.5–20.9) < 0.001 Dysmenorrhea since IUD insertion 254 (34.9%) 51 (37.5%) 203 (34.3%) 0.479 NA
3
0
3
Heavy menstruation since IUD insertion: 270 (37.8%) 62 (47.0%) 208 (35.7%) 0.016 NA
16
4
12
BMI: Body mass index; IUD: Intrauterine device; MC: menstrual cup
Characteristics of the study population
Hormonal
Copper
BMI: Body mass index; IUD: Intrauterine device; MC: menstrual cup
The proportion of MC use in the adequate IUD position group was 17.1% compared to 40.4% in the non-adequate IUD position group ( p < 0.001) which corresponds to a proportion of displaced IUDs of 4.7% in patients not using MC and of 13.9% in patients using it ( p < 0.001). After adjustment, the use of MC appeared independently associated with IUD displacement (aOR [95CI] = 3.13 [1.55–6.25]). Among other risk factors, only previous IUD expulsion was independently associated with IUD displacement (aOR [95CI] = 6.57 [3.01–14.00]). Details are presented in Table 2 . Regarding IUD expulsion, it was up to 5.9% in MC users and 0.5% in non-users ( p 75%) in our population, subgroup analyses were made regarding the type of IUD. In the copper IUD group, the association between MC use and IUD displacement was still significant (aOR [95CI] = 3.20 [1.57–6.44]), but it was not in the hormonal IUD group (OR could not be calculated as only 3 hormonal IUDs were in non-adequate position, none of them in women using MC).
Table 2 IUD displacement and associated risk factors Median (IQR) or N (%) Adequate IUD position ( N = 684) Non-adequate IUD position ( N = 47) Crude OR [95CI] Adjusted* OR [95CI] ( N = 709) Age (years) 27 (25–29) 27 (25-29.5) 0.99 [0.92–1.06] - BMI (kg/m 2 ) 21.3 (19.8–23.4) 21.3 (19.9–23.7) 1.06 [0.97–1.14] 1.06 [0.96–1.16] NA
6
0
- Nulliparity 614 (89.8%) 43 (91.5%) 1.23 [0.48–4.16] 2.14 [0.70–9.40] Never been pregnant 556 (81.3%) 39 (83%) 1.12 [0.54–2.64] - Surgical abortion within 6 weeks of IUD insertion 26 (3.8%) 0 (0%) - - History of IUD expulsion 42 (6.1%) 14 (29.8%)
6.48 [3.15–12.86]
6.57 [3.01–14.00]
IUD type: Copper 516 (75.4%) 44 (93.6%)
4.78 [1.72–19.87]
3.06 [1.00–13.29] Hormonal 168 (24.6%) 3 (6.4%) Time since IUD insertion (months) 7,7 (2,6–23,6) 7.3 (2,7–17) 0.98 [0.96–1.01]
0.97 [0.94–0.99]
Dysmenorrhea since IUD insertion 235 (34.5%) 19 (40.4%) 1.29 [0.69–2.34] 1.27 [0.63–2.50] NA
3
0
Heavy menstruations since IUD insertion 248 (37%) 22 (48.9%) 1.63 [0.88–2.99] 0.98 [0.49–1.95]
NA
14
2
Use of a MC since IUD insertion 117 (17.1%) 19 (40.4%)
3.29 [1.75–6.05]
3.13 [1.55–6.25]
MC: menstrual cup; IUD: Intrauterine device; BMI: Body mass index * Variables not mentioned for adjusted OR were not included in the model after stepwise variable selection
IUD displacement and associated risk factors
MC: menstrual cup; IUD: Intrauterine device; BMI: Body mass index
* Variables not mentioned for adjusted OR were not included in the model after stepwise variable selection
In 136 consultations, patients declared having used MC at least once since IUD insertion (details can be found in Table 3 ). Among them, regarding the last period of utilisation, 56 (43%) used it always, 38 (29%) used it almost daily, 16 (12%) used it regularly, 3 (2%) used it occasionally and 17 (13%) had used it anecdotally. We can consider that 94 (72%) were using the MC as main menstrual protection (always or almost daily). Regarding the evolution of MC utilisation, the great majority of patients ( N = 115, 85%) reported that they had not changed their frequency of use of MC since they started using it (= one period of use).
Among MC users 117 (86%) had an IUD in adequate position and 19 (14%) a displaced IUD.
Removing suction before withdrawal, MC exposition score or estimated number of withdrawal were not significantly associated to IUD displacement in univariate analysis (Table 3 ).
Table 3 Details of MC utilisation Median (IQR) or N (%) MC use sample ( N = 136) Adequate IUD position ( N = 117) Non-adequate IUD position ( N = 19) Crude OR [95CI] Estimated number of menstrual cycles with MC use 8.5 (2–19) 9 (2-21.5) 3 (1.5–15.5) 0.96 [0.91–1.01] NA
14
14
0
Estimated number of MC withdrawals since IUD insertion 49 (11.2-160.5) 50 (15–156) 32 (1-180) 1.00 [1.00–1.00] NA
62
56
6
Type of MC: With stem 115 (84%) 99 (85%) 16 (84%) ref Without stem 16(12%) 13 (11%) 3 (16%) 1.43 [0.30–5.06] Unkown 5 (4%) 5 (4%) 0 (0%) - Break of the MC suction before removal 115 (85%) 99 (85%) 16 (84%) 0.97 [0.29–4.47] MC as main menstrual protection 94 (72%) 78 (70%) 16 (84%) 2.26 [0.69–10.16] NA
6
6
0
MC exposition score 35 (6.5–76.5) 35 (7–75) 15 (7-67.5) 0.99 [0.98-1.00] NA
14
14
0
MC: menstrual cup; IUD: Intrauterine device
Details of MC utilisation
MC: menstrual cup; IUD: Intrauterine device
Conclusion
Our results show that regardless of other known risk factors for displacement, MC use was an independent risk factor for IUD displacement. Clinicians should be aware of this risk to offer patients the opportunity to choose the most appropriate menstrual protection when using an IUD. The mechanisms behind MC-induced IUD displacement remains to be explored. Further studies will be needed to determine whether it is indeed related to a suction effect and whether this effect could be reversed by therapeutic education sessions on MC use.
Discussion
We conducted a case-control study enrolling 731 IUD follow-up consultations, addressing IUD displacement with respect to MC use and other known or suspected displacement factors. We found that the use of MC appears to be an independent risk factor for IUD displacement (aOR 3.13). However, this conclusion can only be drawn for the patients with copper IUD, representing the vast majority of our population. In the MC users, neither the level of exposure to MC since IUD insertion (combining frequency and length of use) the number of MC withdrawals since IUD insertion, nor the fact that the patient broke the MC suction or not before removal, appeared to be predictive of IUD displacement. The lack of significance of the latter results is probably related to a lack of power, the number of women using MC in our population being small (only 11 women in the displaced IUD group).
Our study was limited by a memory bias as the patients had to remember how they had been using the MC since IUD insertion. Another limitation is the way the questionnaire was completed regarding MC use. Indeed, the questionnaire was not filled by the patient herself but by the practitioner who interviewed her. This method may have influenced the patients’ answers, especially those concerning the proper use of MC. It is recommended in France for all women with IUD to have an annual follow-up or in case of experiencing any symptoms so we hope to have a limited selection bias.
Our young and urban population of patients, mainly nulliparous and with copper IUDs might not be representative of how MC can be used in other populations. Regarding the high proportion of nulliparous women in pour IUD population, the more recent French data on IUD use are from 2016, yet at that time IUD use in 25–29 years old had increased since 2010 from 6.9–19% [ 1 ]. It is likely that this increase has been maintained since 2016, as more and more gynaecologists are agreeing to insert IUDs in nulliparous women, as the discourse of health professionals has changed, particularly in connection with the crisis of the contraceptive pill, which has led to a shift in use to other contraceptive methods.
Our results corroborate those of the D-COUPE study [ 28 ] conducted on 207 patients, which found a displacement 4 times more frequent in MC users than in non-users, as well as the results of a survey of 638 responses [ 29 ] showing a 2.75-fold increase in the risk of IUD displacement in patients using MC. However, our results differ from those of Wiebe et al., investigating 743 women, in whom the risk of IUD displacement did not differ with respect to their menstrual protections. This study did not adjust the results for other IUD displacement risk factors and the lack of homogeneity between the groups compared, especially in terms of age, was described as a major limitation of the study by the authors [ 27 ]. However, a literature review in 2023 on 7 studies concluded that there is a possible association between menstrual cup use despite scarce evidence [ 30 ].
We chose to assess IUD displacements and not only expulsions, as the majority of our patients were using copper IUDs. Indeed, literature data do not show a significant difference between patients with a hormonal IUD in place or not for the risk of pregnancy. However, there is a significant difference in the risk of pregnancy in patients with a copper-bearing IUD [ 26 , 31 , 32 ].
There is no specific recommendation on the ultrasound criteria for a well-positioned IUD and this subject is controversial. Some studies use the IUD - uterine fundus distance (from 4 to 30 mm) [ 11 , 33 – 36 ], others the IUD - endometrium distance (from 5 to 10 mm) [ 9 , 35 – 39 ], rarer studies suggest that the IUD - myometrium distance(< 10 mm) is more reliable than the IUD - endometrium distance because the endometrium thickness varies according to the cycle [ 34 , 36 , 38 ], and finally, many studies do not use a measurement criterion but note whether the IUD is in place, in the lower part of the uterus, partially or entirely in the cervix [ 6 , 12 , 18 , 33 , 40 ]. Regarding the lack of recommendations and consistency among studies, we decided to use the latter method and that each investigator would judge the proper position of the IUD as in his usual practice.
Introduction
In France, about 26% of contracepting women had chosen intrauterine devices (IUD) as a birth control method, and since the contraceptive pill crisis, more and more young nulliparous women choose IUD as a birth control method [ 1 ]. This crisis built upon earlier debates from 2012 to 2013 when several lawsuits against pharmaceutical companies highlighted risks of venous thromboembolism associated with newer-generation pills and lead to many women seeking alternative contraceptive methods [ 2 ]. Contraceptive failures with IUDs are mainly related to displacement or expulsion [ 3 ]. Young age (< 25 years) [ 3 – 6 ], menorrhagia, dysmenorrhoea [ 3 , 7 , 8 ], use of a copper IUD rather than a levonorgestrel releasing one [ 6 , 9 , 10 ], history of IUD expulsion or uterine abnormality (fibroid, adenomyosis) [ 3 , 7 , 11 – 13 ], obesity, and higher parity have been identified as risk factors for IUD displacement or expulsion [ 14 – 19 ]. There is no consensus in the literature regarding other potential risk factors such IUD insertion within 6 weeks of an abortion [ 5 , 20 – 22 ].
Menstrual cups (MC) are being increasingly used worldwide as an alternative to tampons and sanitary pads and are a safe option for menstruation management, especially in younger generations of women [ 23 ]. Recently, concerns have been raised regarding possible IUD displacement linked to MC use [ 24 – 26 ]. A meta-analysis published in 2019 pointed out dislodgement of IUD in 13 women who used MC (8 in case reports, and 5 in a study) within 1 week to 13 months after insertion of the IUD [ 24 , 25 ]. A possible mechanism underlying MC-induced IUD displacement would be a suction effect when removing the MC, another mechanism is that strings are pulled [ 25 ]. To date, only 3 studies were published addressing the risk of IUD displacement associated to MC use. A 2012 Canadian retrospective chart survey on 743 women reported that IUD displacement rate was not different between women using tampons, pads or MC [ 27 ]. A 2018 French cohort study identified an association between declared IUD expulsion and MC use, the risk being higher for MC use inferior to 3 menstrual cycles [ 28 , 29 ]. A 2019 American Internet-based survey identified positive association between MC use and IUD expulsion (OR: 2.75, 95% CI: 1.40–5.42, p = 0.002) [ 28 , 29 ]. None of these studies adjusted for risk factors. Moreover, focus was on IUD expulsion but intrauterine displacement of the IUD, which is associated to a loss in contraceptive efficiency, was not addressed.
The objective of the present study was to measure the association between IUD displacement (as diagnosed by ultrasound measure of IUD position) and MC use in patients consulting for IUD follow-up. The study was designed to consider potential risk factors.