Comparative Analysis of Nebulizers in Clinical use for Pressurized Intraperitoneal Aerosol Chemotherapy (PIPAC)

preprint OA: closed
📄 Open PDF View at publisher

Abstract

Objective Technical ex-vivo comparison of commercial nebulizer nozzles used for Pressurized Intraperitoneal Aerosol Chemotherapy (PIPAC). Methods The performance of four different commercial nebulizer nozzles (Nebulizer; HurriChem™; MCR-4 TOPOL ® ; QuattroJet) was analysed by comparing: i) technical design and principle of operation, ii) operational pressure as function of the liquid flow rate, iii) droplet size distribution via laser diffraction spectrometry, iv) spray cone angle, spray cone form, and horizontal drug deposition through image-metric analyses, and v) chemical resistance via exposing to a cytostatic solution and metallurgic composition by means of spark optical emission spectral analysis. Results The Nebulizer exhibits a nearly identical technical design, implying a comparable performance (e.g., mass median droplet size of 29 μm) as the original PIPAC nozzles (MIP/ CapnoPen). The other three nozzles demonstrate varying degrees of performance deviation from the original PIPAC nozzles. The HurriChem™ shares a similar design and principle of operation as the Nebulizer, but produces a finer aerosol with a particle size of 22 μm. The operating principles of MCR-4 TOPOL ® and QuattroJet significantly differ from that of the original PIPAC nozzle technology. The MCR-4 TOPOL ® nebulizer has a hollow spray cone that leads to the production of significantly larger aerosol droplets (50 μm) compared to the original PIPAC nozzles. The QuattroJet generates an aerosol droplet (22 μm) similar in size to the HurriChem™ and exhibits improved spatial drug distribution. Conclusion While the introduction of new PIPAC nozzles is a welcome development, differences in performance and efficacy were noted. Therefore, it is recommended that PIPAC nozzles that deviate from the current standard undergo bioequivalence testing and be implemented in accordance with the IDEAL-D framework prior to routine clinical use.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00