Successful vapor-based endometrial ablation: in vivo peri-hysterectomy study

Other OA: closed CC0 ⤵ 1 in-corpus citation
View on OpenAlex View on PubMed View at publisher
⚙ AI-generated summary by qwen3.7-flash, 2026-09-11 ⓘ

In a peri-hysterectomy model of nine women, the AEGEA vapor-based endometrial ablation system achieved complete endometrial coverage with no uterine perforation or significant thermal injury, demonstrating safety and efficacy for future clinical trials.

One-sentence paraphrase of the abstract; not a substitute for reading it. No clinical advice. How this works

Abstract

STUDY OBJECTIVE: To evaluate the AEGEA vapor-based endometrial ablation system using an in vivo peri-hysterectomy model. DESIGN: Single-site feasibility study (Canadian Task Force classification II-2). SETTING: University medical center. PATIENTS: Nine women consented to undergo AEGEA endometrial ablation before previously scheduled abdominal hysterectomy to treat abnormal uterine bleeding. INTERVENTIONS: In vivo AEGEA endometrial ablation was performed using a 90-second vapor treatment cycle. After hysterectomy, the uteri were examined for the extent and location of endomyometrial ablation (macroscopic triphenyltetrazolium chloride staining) and fallopian tube injury (microscopic nitroblue tetrazolium staining). MEASUREMENTS AND MAIN RESULTS: The mean (SD) posttreatment measurements of the 9 uteri were as follows: weight, 143 (40) g; length, 10.3 (1.3 cm); thickness, 4.4 (0.6) cm; and width, 6.2 (0.7) cm. The endometrial thickness was 1.1 (0.7) mm. Three uteri had myomas that measured less than 2 cm; and 2 uteri demonstrated focal adenomyosis. No myometrial perforation or thermal serosal injury was identified. The median corpus, lower uterine cavity and bilateral cornua percentages of TTC-negative surface endometrial treatment were 100% (range: 100-100%), 100% (range: 80-100%), and 100% (range: 95-100%), respectively. The closest distance between the ablation and serosa was 11.5 (3.2) mm. No lower endocervical or exocervical thermal injury was identified. Minimal fallopian tube thermal injury was identified in 18% of interstitial segments evaluated, and measured 0.6 to 0.8 mm in maximal depth and extended to within 6.3 to 9.5 mm of the serosa. No thermal injury was identified in the extrauterine fallopian tube segments. CONCLUSION: The AEGEA vapor-based endometrial ablation system has the potential to provide excellent cavity coverage with full-thickness endometrial ablation. The study results further support an acceptable in vivo safety profile for future clinical efficacy trials.

My notes (saved in your browser only)

MeSH descriptors

Endometrial Ablation Techniques Hysterectomy Hysterectomy Preoperative Care Preoperative Care Endometrial Ablation Techniques Endometrial Ablation Techniques Feasibility Studies Feasibility Studies Female Female Humans Humans Uterus Uterus Uterus

Citation neighborhood (sparse)

Too few in-corpus citations on either side for a chart; here are the lists.

Cited by (1)

References (6)

Cited by (1)

Source provenance

europepmc
last seen: 2026-10-16T06:09:50.462052+00:00
openalex
last seen: 2026-06-04T00:00:01.174412+00:00
pubmed
last seen: 2026-10-08T21:09:10.644269+00:00
unpaywall
last seen: 2026-10-16T06:24:35.758740+00:00
License: CC0 · commercial use OK