Material
failure modes often increase in likelihood as time progresses. The parylene-C
insulation commonly used for Utah style intracortical arrays can crack and delaminate, shunting
current to the biological tissues (Caldwell et al. 2020; Prasad et al. 2014; Schmidt et al. 1988;
Xie et al. 2014). The metal tips of Utah arrays, most commonly platinum or iridium oxide, are
generally stable in vitro, but may be eroded away by aggressive stimulation (Negi et al. 2010) or
the comparatively harsh in vivo environment (Negi et al. 2010). Furthermore, use of the
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
electrodes for stimulation can impact the rate of tip degradation (Cogan 2008; Gilgunn et al.
2013).
To establish stimulation limits for these clinical studies, experiments were performed in non-
human primates and showed that frequent microstimulation over six months did not cause more
loss of neurons around the electrode tips than insertion of the devices themselves and that
stimulation had no behavioral effe ct for tasks that r equired tactile feedback (Chen et al. 2014;
Kim et al. 2015). Using these established parameters, we would not expect stimulation to cause
further damage to the brain tissue after implantation or have deleterious effects on behavior. In
fact, stimulation over five years in a participant with these established parameters has not
resulted in significant differences in signal between stimulated and non-stimulated arrays and
detection thresholds have improved over time (Hughes et al. 2020). However, to our knowledge
there have been no post-implant examinations of the material properties of intracortical arrays
implanted in humans. Here we examine the extent to which any material degradation occurred
on explanted human intracortical electrodes, which will aid in the design and development of
robust BCIs for long-term clinical use.
In this work, electrodes explanted from two human participants were examined to determine the
extent of tissue encapsulation and material failure and to assess how these factors affected
chronic recording performance. These electrodes were implanted for different lengths of time
and were surgically explanted: 987 days for the two arrays in participant 1 (P1) and 182 days for
the four arrays in participant 2 (P2). Both arrays in P1 and two of the arrays in the P2 had
platinum tips and were used for recording only, while the other two of the arrays in P2 had
sputtered iridium oxide (IrOx) tips and were used for both stimulating and recording (Negi et al.
2010). First, the extent and nature of the tissue encapsulation of the arrays was investigated
using optical microscopy and two-photon microscopy (TPM). Following this, the electrode arrays
were examined with scanning electron microscopy (SEM) and energy-dispersive x-ray
spectroscopy (EDS) to evaluate the extent of material damage. Finally, we compared the results
of these analyses to endpoint recording performance of the devices and characterized the
relationship between electrical
stimulation and material degradation.
1. Methods
1.1 Participants
These studies (NCT01894802 and NCT01364480) were conducted under Investigational
Device Exemptions from the U.S. Food and Drug administration and were approved by the
Institutional Review Boards at the University of Pittsburgh (Pittsburgh, PA) and the Space and
Naval Warfare Systems Center Pacific (San Diego, CA). Informed consent was obtained before
any study procedures were conducted. Two participants were implanted with microelectrode
arrays in the brain. The first subject (P1) was implanted with two intracortical Pt microelectrode
arrays (4 mm × 4 mm, Blackrock Microsystems, Salt Lake City, UT, USA) each with 96 wired
electrode shanks (length 1.5 mm) in a 10x10 grid in the participant’s left motor cortex (Figure 1).
The second subject (P2) was implanted with two Pt microelectrode arrays (Blackrock
Microsystems, Salt Lake City, UT) in the left somatosensory cortex and two iridium oxide (IrOx)
microelectrode arrays in the left posterior parietal cortex. Each Pt array in the somatosensory
cortex consisted of 88 wired electrodes in a 10x10 grid while each IrOx array in the posterior
parietal cortex consisted of 32 wired electrodes distributed throughout a 6x10 grid (Figure 1).
Following implantation of the arrays into P2, it was discovered that the implant locations were
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
posterior to the intended sites. Following which, the pedestals were removed, and a second
implantation was performed two months later.
1.2 Neural recording and signal quality metrics
Neural data were collected for both P1 and P2 using Neuroport Neural Signal Processors
(Blackrock Microsystems, Salt Lake City, UT). At the beginning of each test session, a threshold
for all channels was set at -5.25 (P1 before day 565) and -4.5 (all other test sessions) times the
root-mean-square voltage. Data were collected across 287 sessions spanning 33 months for P1
and 40 sessions across four months for P2. No recordings were done for the final two months of
P2’s implant as the percutaneous pedestal connectors had been removed to prepare for the
reimplant.
One of the main goals of the clinical study was to provide the participants with high degree-of-
freedom control of a robotic arm. To accomplish this, participants performed a brain-computer
interface calibration paradigm at the beginning of a test session. We used three minutes of data
collected during this calibration procedure to run spike sorting analyses offline. The sorting
method, described in detail in Downey et al., 2018 (Downey et al. 2018) used principal
component analysis (PCA) to separate units, defined as threshold crossings from an individual
electrode, based on the similarity of their waveform shape. Characteristics for each unit were
then calculated. Peak-to-peak voltage (Vpp) was defined as the voltage difference between the
peak and the trough of the average waveform for each unit. Since there could be more than one
unit identified per electrode, the unit with the maximum Vpp was chosen to represent the signal
quality for the given electrode. Electrodes were considered to be viable if they contained
waveforms with a minimum Vpp of 30 µV and a minimum firing rate of 0.25 Hz.
1.2.1 Impedances
Electrode impedances were measured for both participants using the NeuroPort patient cable
data acquisition system (Blackrock Microsystems, Salt Lake City, UT). For P1, impedances
were measured at the beginning of a test session once a month. Impedances values for P2
were measured at the beginning of each test session. The system delivered a 1 kHz, 10 nA
peak-to-peak sinusoidal current to each implanted electrode for one second.
1.2.2 Intracortical stimulation and calculated metrics
Seven test sessions across approximately one month involved microstimulation on the IrOx
arrays. Stimulation was delivered using a CereStim R96 multichannel microstimulation system
(Blackrock Microsystems, Salt Lake City, UT). Pulse trains consisted of cathodal phase first,
current-controlled, charge-balanced pulses delivered at frequencies from 20-300 Hz and at
amplitudes from 1-100
μ A. The cathodal phase was 200 μ s long, the anodal phase was 400 μ s
long, and the anodal phase was set to half the amplitude of the cathodal phase. The phases
were separated by a 100
μ s interphase period. Stimulus pulse trains were varied in terms of
amplitude, frequency, and train duration.
The voltage transients associated with each stimulus pulse were recorded using National
Instruments data acquisition modules. Voltage traces were displayed in real time using LabView
and saved to disk for analysis. Interphase voltage was measured as the voltage at the end of
the interphase period immediately prior to the anodal phase for a given stimulation pulse. The
total charge delivered to each electrode was calculated across all stimulation experiments using
the charge delivered during the cathodal phase.
1.3 Explanted array handling before imaging
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
The two Pt arrays in P1 were explanted on day 987 post-implant and the four arrays in P2 were
explanted on day 182. Following explantation, all arrays were removed from their wire bundles
by clipping the wires proximal to the probe and were washed with saline. The P1 arrays were
immediately fixed in formalin and then transferred to PBS bath for storage.
Immunohistochemical staining procedure was performed on these two arrays with the goal of
identifying neuron (NeuN) and microglia/macrophage (Iba-1). The staining process involves
incubation of the arrays with primary antibodies solutions overnight, with secondary antibodies
for 4 hours followed by Hoescht solution for 20 min for nuclei staining. The antibody staining
was unsuccessful, and only nuclei stain was used for the tissue analysis. The P2 arrays were
fixed 2 months post-implant, and one of the Pt arrays had visible tissue encapsulation and was
imaged using TPM. Because these arrays were not immediately fixed, we did not perform
immunostaining, and only characterized the collagen structure, which can be stable without the
fixation.
After optical and TPM imaging, two arrays explanted from P1 were sent to the FDA for initial
analysis. The arrays were initially imaged with an environmental SEM, then enzymatically
cleaned with Asepti-Zyme neutral pH enzymatic instrument presoak/cleaner (4ml in 250ml
saline) at 37°C for 90 minutes, followed by Getinge Clean Enzymatic detergent (1ml in 250ml
saline) at 37°C for 90 minutes, and then by MetriZyme detergent (1ml in 250ml saline) 37°C for
90 minutes. Samples were then thoroughly washed with water and air dried, ready for SEM
imaging. This process was effective at removing some of the tissue and revealing the electrode
tip/shank for material analysis. Arrays from P2 did not undergo the enzymatic cleaning
procedure. All arrays were stored adhered to copper tape, tips up.
1.4 Electrode Imaging
Explanted electrodes were first characterized by optical and two-photon microscopy to assess
the degree of tissue encapsulation. For TPM, we used a two-photon laser scanning microscope
with a Bruker scan head (Prairie Technologies, Madison, WI), TI:sapphire laser tuned to 920 nm
(Mai Tai DS; Spectra-Physics, Menlo Park, CA), light collection through non-descanned
photomultiplier tubes (Hamamatsu Photonics KK, Hamamatsu, Shizuoka, Japan), and a 10x or
16x, 0.8 numerical aperture wate r immersion objective (Nikon In c., Milville, NY). Laser power
was maintained between 20-40 mW. For each electrode tip, Z-stacks were collected with filters
to resolve second harmonic generation (SHG) at half the laser wavelength (~460nm), which
enabled intrinsic imaging of collagen-I representing the meningeal encapsulation. Images along
the length of the electrode shanks were collected as Z-stacks. Z-stack images were either
collected at specific regions of interest, or in a grid at all locations across the face of the
electrode array. Grid images were automated by the Prairie software with a 10% overlap
between images. All image stitching and subsequent image processing was conducted with
ImageJ software (NIH). Electrode integrity was characterized by scanning electron microscopy
(SEM) and energy-dispersive x-ray spectroscopy (EDS). Samples were washed, dried under
alcohol, and sputter-coated with 4nm Au/Pd. Images were taken by JSM 6335F electron
microscope. EDS was taken by Zeiss Sigma 500VP, excluding Au and Pd from quantification.
Using the SEM and optical images, a qualitative category of ‘non-degraded/unencapsulated’ or
‘degraded/encapsulated’ was assigned to each electrode based on the degree of damage to the
tip or shank, or the level of encapsulation around the electrode ( Figure S1). Arrays explanted
from P1 were more extensively cleaned prior to imaging, and the encapsulation score was
based on optical images of the explanted arrays. Encapsulation on arrays from P2 was
determined by examining the SEM images. Degraded electrode tips were defined as having
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
obvious and substantial surface defects in the metal coating, including pitting of the metal,
flaking of the metal, and ex posure of the underlyi ng silicon. Degraded shanks were defined
relative to the parylene insulation, with defects including insulation cracking along the shank,
peeling of the insulation away from the shank near the tip, and other obvious defects in or below
the insulation. These categories were compared to EDS images, confirming the
presence/absence of metalation at the tip (Pt/IrOx). Electrodes which could not be quantified,
due to breakage during removal or gross encapsulation, were assigned a null score and
excluded from analysis.
1.5 Statistics
Changes in signal and impedances over time were assessed using linear regression. For
impedances, data were log-transformed because data did not follow a linear trend. For P2
impedances and Vpp, we excluded data prior to day 30 for regression because the impedances
measured in this range were highly variable.
Total charge delivered, minimum interphase voltages, and charge delivered after exceeding an
interphase voltage of -0.6V were compared between the two electrode arrays that had received
stimulation using Mann-Whitney tests. We used a non-parametric test because the data was
determined to not be normally distributed using an Anderson-Darling test. We used a Fisher
exact test to determine if there was a significant relationship between an electrode’s material
properties (undamaged or damaged) and the length of implantation (Pt arrays in P1 vs. P2) or if
it received stimulation (P2 IrOx arrays, yes or no). We further quantified if there was a
relationship between both total charge injected and charge injected with interphase voltages
below -0.6V on stimulated electrodes and their material properties (undamaged or damaged)
using logistic regression. Electrode categories were compared to impedances and Vpp using
Mann-Whitney tests. We used a non-parametric test because the variances between groups
were not the same.
2. Results
2.1 Signal amplitude and impedances decreased over time
Changes in the impedances and peak-to-peak voltages over time were observed on implanted
electrodes in both participants (Figure 2). Impedances decreased over time on electrodes
implanted in P1 (p<0.001, log-transformed linear regression, Figure 2A). For P2, the starting
impedances of IrOx electrodes were lower than the platinum electrodes, which is consistent
with the manufacturer’s specification (Negi et al. 2010). From day 1 to 20 we observed an
increase in impedances. The initial increase in impedance reversed after one month (30 days),
and a significant downward trend in impedances was observed until the end of recording for
both the IrOx (p<0.001, linear regression) and platinum arrays (p<0.001, linear regression).
Impedances gathered from P1 eventually stabilized after approximately two years. The
difference between the final impedance values recorded in P1 and P2 can be explained by the
difference in length of implantation. Previous studies have determined that impedance values of
stimulated and non-stimulated intracortical electrodes decrease dramatically over the first
couple of years after implantation in humans (Hughes et al. 2020) and monkeys (Chestek et al.
2011; Suner et al. 2005). Since P2 was implanted for a significantly shorter period, we would
expect the electrode impedance values to be larger and more variable, which the data supports.
In the same manner as the impedance measurements, an initial increase in Vpp was observed
for both P1 and P2. However, after an increase in the first 30 days, the measured Vpp from P1
and P2 exhibited a downward trend (p<0.001, linear regression, Figure 2C,D). The rates of
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
decrease in the Vpp between day 30 and 120 for P1 and P2 were -4.0 µV/month and -4.86
µV/month, respectively. Median Vpp decreased by 52% across 550 days in P1 and by 14%
across 90 days in P2. The median Vpp for P1 leveled off at approximately 25-30 µV.
2.2 Encapsulating tissues were apparent on multiple arrays
Based on the gross optical micrographs, both P1 arrays had a significant degree of adherent
tissue on the electrode base and shanks. For the P2 arrays, one of the Pt arrays and one of the
IrOx arrays showed some tissue deposits while the other arrays were clean. The nature of the
encapsulating tissue was examined with TPM, measuring the second harmonic signal
characteristic of collagen. For the more heavily encapsulated P1 arrays, the encapsulation
sheet was found both along the shanks of the array (Figure 3A-D) and at the base (Figure 3E-H
(P1)). Strong second-harmonic signal within the tissue sheet confirmed that it was primarily
composed of collagen-I fibers (Figure 3C,D,G,H). After further examining the indicated
electrodes and staining for cell nuclei, we observed that the encapsulation was highly
cellularized (Figure 3C,D,G,H). In addition, the encapsulation continued down the shank of the
electrode, with cellular and collagenous material detected along the shanks and tips of the
array. On the underside of the array at the base of the shanks the encapsulation was not
homogenous, instead exhibiting greater second harmonic signals nearer to the edges (Figure
3G) while having greater cell density nearer the center (Figure 3H). SHG imaging is also a good
tool for detecting blood vessels because of the strong presence of collagen in the vessel wall,
however we did not observe clear blood vessel structure in the P1 explants.
For the posterior Pt array in P2, the encapsulation tissue covers the whole array (Figure 4B) and
the TPM imaging from the side revealed significant tissue covering the majority of the electrode
tips. Here, we identified clear vascular architecture in the encapsulation tissue (Figure 4D,
highlighted in blue). The blood vessel in the encapsulation tissue was traced and super-imposed
to the image of pial vasculature observed pre-implantation (Figure 4E). As can be seen in
Figure 4F, the blood vessel traces match the pia vasculature. This indicates that the blood
vessels observed to be at the tip of this array were pial blood vessels. Two mechanisms may
lead to this: 1) the array did not fully penetrate the pia at the time of implantation; 2) the array
was successfully implanted in the brain parenchyma and the pia membrane was pulled out with
the array. Since we were able to obtain high quality single unit recordings from this array even
from the affected region, the first potential mechanism was ruled out. Therefore, we conclude
that at least some of the tissue on this explant is pia membrane that was pulled out with the
device, not fibrotic scar tissue formed as the result of foreign body reaction.
2.3 Length of implantation impacts the degree of material degradation and fibrous
encapsulation.
Based on the optical, TPM and SEM images, electrodes were assigned a binary score for the
tip, shank, and degree of fibrous encapsulation (Figure 5). Electrodes that appeared to be
broken or damaged by implantation/explantation were excluded from analysis. Tips and shanks
were evaluated separately to examine the effects of both tip metallization and electrode
insulation on device performance. The number of electrodes for each group are displayed in
Table 1, excluding electrodes which were not wired or used for recording or stimulation.
Differences in the total number of electrodes receiving a tip category (n=387), shank category
(n=413), and encapsulation (n=380) are due to damage to the electrodes or excess
encapsulation preventing the assignment of a proper category (Figures S1-3).
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Categories assigned to P1 and P2 platinum arrays were compared to identify any potential
changes in material deterioration or encapsulation which may be attributed to the length of
implantation (Table 2). We found that both measures of material degradation (tip and shank
damage) were more prominent for longer implantation times (27.8% tip damage for P1 and
9.9% for P2, 15.2% shank damage for P1 and 1.7% for P2, p<0.001 for both). We also found
that the degree of encapsulation is more significant for longer implants with 72.4% for the P1
arrays and 49.4% for the P2 arrays (p<0.001).
2.4 Stimulation resulted in electrode damage on one stimulating array but not the other
Two IrOx arrays implanted in P2 received a low amount of total charge (<160 µC per electrode
site). Each of the two stimulated IrOx arrays had 60 electrodes, half of which electrically
connected and used for stimulation. Preimplant optical images of the arrays did not show any
variation between arrays. The stimulated electrode sites are arranged primarily in a
checkerboard fashion. SEM shows that the lateral array had a high degree of tip and shank
degradation (Figure 6A). Interestingly, tips and shanks showing visible damage appeared to
coincide with the electrodes that were used for stimulation. Furthermore, EDS revealed that
stimulated tips had lower iridium content than non-stimulated tips (Figure 6B). The loss of
metallization for the lateral stimulating a rray occurred only on th e electrodes used for
stimulation. The medial array did not show this pattern (Figure 6C). The damage scores for
each electrode tip and shank are summarized in the Figure 6D,E. The checkerboard pattern of
damages of the lateral array is clearly seen, which correspond very well with arrangement of the
stimulation electrodes. The medial stimulating array has overall much less observable material
damage but more tissue encapsulation. Of the 62 electrodes used for stimulation on both
arrays, 56 were analyzed, of which 23 had notable tip degradation, 21 of which were located on
the lateral electrode array. Metal loss, and the corresponding decrease in iridium signal, was not
observed on any non-stimulated electrodes. These results are summarized in Table 3.
Delivered charge and measured interphase voltages were compared to the material
degradation. The amount of stimulation provided was quantified by both the total charge
delivered and number of pulses delivered. Although the mean amount of charge injected on the
lateral array was greater, it was not significantly different than the mean charge injected on the
medial array (Mann-Whitney test, p = 0.22). The medial array contained the three electrodes
with the most charge delivered, none of which displayed observable material degradation.
However, we examined the minimum voltage during the interphase period (Figure 6F) and
found that the lateral array electrodes experienced higher voltage excursions on average than
the medial array electrodes (mean minimum voltage was = -1.7 V for the lateral, and -1.1 V for
the medial array). Furthermore, there was a significant relationship between the total charge
injected at voltages more negative than -0.6V (Figure 6G) and the tip score (p = 0.025, crit-p =
0.034, logistic regression) and shank category (p = 0.023, crit-p = 0.034, logistic regression) on
the lateral stimulating array. There was no relationship between total charge injected at voltages
less than -0.6V and tip category (p = 0.60, logistic regression) or shank category (p = 1, logistic
regression) on the medial array.
We found no significant differences in recording quality (Vpp) (Figure 6H) between the damaged
and non-damaged electrodes on the two stimulation arrays, after excluding the encapsulated
electrodes.
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
3. Discussion
In order for BCIs to become a viable therapy, the longevity of the devices and mechanisms of
failure must be well understood. Effective electrode design requires knowledge of the stability of
the materials in the harsh in vivo environment and the effects of gradually accumulating damage
to the device. However, the relationships between chronic material degradation and device
performance are poorly understood. The effects of material degradation on performance in
human subjects is further complicated by the limited number of human subjects and the even
smaller amount of explanted human BCI arrays. In this work, the in vivo performance of human
neural electrode arrays was compared to the material integrity after explant. We have found
signs of material degradation on all electrode arrays, with longer implantation times correlating
with an increased number of degraded electrodes (Table 2). Additionally, biological tissue
encapsulation on the explanted device was also documented as another potential factor to
influence recording quality. The biological encapsulation tissues were highly collagenous and
also highly cellularized, and appear to form in a time dependent manner, increasing with the
length of implantation. A similar form of tissue response has been observed in a post mortem
analysis of tissue surrounding a MEA implanted for seven months.(Szymanski et al. 2021)
Further, the nature of the encapsulation at the periphery of the array and the center is different.
Together, these results suggest that the encapsulation originated from the meninges, as
opposed to the CNS.
3.1 Encapsulation and material degradation were both related to the length of implantation
References
Ajiboye, A. B ., Will et t, F. R. , Young, D. R., Memberg, W . D., Murp hy, B. A ., Mille r, J . P., et al. (2017).
Restor atio n of reaching and grasping mo vements thr ough brain-cont rolle d muscle stimulati on in a
person with t et raplegia : a proof-of-concept demons tra tion . The Lan cet 389, 1821 -1830.
doi:ht tps://doi.o rg/10.1016/S0140-6736 (17)30601-3
Arment a Salas, M ., Bashfo rd, L., Kellis, S ., Jafari, M ., Jo , H., Krame r, D., e t al. (2018). Proprioceptive a nd
cutaneo us sensatio ns in humans elicit ed by intracor tical microstimula tion . eLife 7, e32904.
doi:10.7554/eLife.32904
Barr ese, J . C., Ace ros, J . and Donoghue, J. P. (2016). Scanning electron microscopy of chronically
implanted in traco rtical micro elect rode a rrays in non-human primat es. J our nal of Neural En gine ering 13,
026003. doi:10.1088/1741-2560/13/2/0 26003
Bouton , C. E., Shaikhouni , A., Anne tta , N. V., Bockbrade r, M. A ., Fri edenb erg, D. A . , Nielson, D. M ., e t al.
(2016). Restoring cortical con trol of funct ional movement in a human wit h quadri plegia. Na t u re 533,
247-250. doi:10.1038/natu re17435
Bullard, A . J ., Hutchison , B. C., Le e, J ., Chestek, C. A . and Patil, P. G . (2020). Estimating Risk for Future
Intr acranial , Fully Implant ed, Mod ular N europr osth etic Systems: A Syst ematic Re view of Hardware
Complications in Clinical Deep Brain S timulation and E xperim enta l Human Int raco rtical Ar rays.
Neuro mod ula tion : Tech nol ogy at t he Ne ural In terfac e 23, 411-426.
doi:ht tps://doi.o rg/10.1111/ner.13069
Buzsáki, G. (2004). Large-scale r ecording of neuronal e nsembles. Na ture Ne urosci ence 7, 446-451.
doi:10.1038/nn1233
Caldwell, R., St ree t, M . G. , Sharma , R., Ta kmakov, P., Baker, B. an d Riet h, L. (2020) . Characte riza tion of
Parylene-C degradat ion mechanisms: In vitro reac tive accele rat ed aging model co mpared to mul tiyear in
vivo implantation . Biom ateri als 232, 119731. doi: ht tps://doi.o rg/10.1016/j.biomaterials .2019.119731
Chen, K. H., Dammann, J . F., B oback, J. L. , Tenore, F . V., Ot to, K. J., Gaun t, R. A ., e t al. (2014). The effect
of chronic intracor tical microstimul ation on the el ectr ode –tissue in terfac e. J ourn a l of Neural Eng ineeri ng
11, 026004. doi:10.1088/1741-2560/11/ 2/026004
Chestek, C. A. , Gilja, V., N uyujukian, P., Foster , J. D., Fa n, J . M., Kaufman, M . T., e t al. (2011). Long-term
stability of neur al prost hetic con trol sign als from silicon cortical arr ays in rhesus macaque moto r cort ex .
Jour nal of Ne ural En gineeri ng 8, 045005. doi:10.1088/1741-2560/8/4/045005
Cody, P. A., Eles, J. R ., Lagenau r, C. F., Ko zai, T. D. Y. and Cui, X. T. (2018). Unique electr ophysiological
and impedance signa tures b etwe en enca psulation typ es: An an alysis of biological Utah ar ray failure an d
benefit of a biomime tic coating in a ra t model. B i om at e r i a l s 161, 117-128.
doi:ht tps://doi.o rg/10.1016/j.biomateri a ls.2018.01.025
Cogan, S. F. (2008). Neural S timulati on a nd Recording Elect rodes . An nu al Review of Biomedi cal
Engineeri ng 10, 275-309. doi :10.1146/an nurev.bio eng.10.061807 .160518
Cogan, S. F., Troyk, P. R., Ehrlich , J. an d Plante, T. D. (2005). In vitro compa rison of the charge-injec tion
limits of activated iridi um oxide (AI ROF) and platinum-iridium micro elect rodes . IE EE Transactio ns on
Biome dical En gin eering 52, 1612-1614. d oi:10.1109/TBME.2005.851503
Collinger, J. L. , Wodlinge r, B. , Downey, J. E., Wang, W ., Tyler-Kabar a, E. C., We ber , D. J., et al. (2013).
High-performance neu ropr osthe tic cont r ol by an individual with tet rapl egia. The L ance t 381, 557-564.
doi:ht tps://doi.o rg/10.1016/S0140-6736 (12)61816-9
Degenhart , A. D., Eles , J., Dum, R. , Misch el, J. L. , Smalianchuk, I ., Endler , B., et al . (2016). Histological
evaluation of a chronically-implanted ele ctrocor ticographic el ectr ode grid in a no n-human primate .
Jour nal of Ne ural En gineeri ng 13, 046019 . doi:10.1088/1741-2560/13/4/046019
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Downey, J. E., Schwed, N ., Chase, S . M., S chwartz, A . B. and Collinge r, J . L. (2018). Intracor tical r ecording
stability in human br ain–compu ter in terf ace users. Jour nal of Ne ural En gineeri ng 15, 046016.
doi:10.1088/1741-2552/aab7a0
Dunlap, C. F., Colachis, S. C., M eyers, E. C ., Bockbrad er, M . A. a nd Fried enbe rg, D. A. (2020). Classifying
Intr acortica l Brain-Machi ne In terfac e Signal Disruptions Bas ed on System Perfor mance and Applicabl e
Compensatory St rat egies: A R eview. Fro ntiers in Ne uroro bo tics 14, doi:10.3389/fnbot.20 20.558987
Fifer, M. S. , McMulle n, D. P., Thomas, T. M., Osbo rn, L. E., Nickl, R. W. , Candre a, D. N., et al . (2020).
Intr acortica l Microstimul ation Elicits Hu man Fingertip S ensati ons. me dRxi v 2020. 05.29.20117374.
doi:10.1101/2020.05.29 .20117374
Flesher, S . N. , Collinger, J . L., Fold es, S. T. , Weiss, J . M., Downey, J . E., Tyler-Kabar a , E. C., et al . (2016).
Intr acortica l microstimula tion of human somatosenso ry corte x. Scie nce Tra nslati o nal Me dicin e 8,
361ra141.
Flesher, S . N. , Downey, J. E., Weiss, J . M. , Hughes, C. L., Herr era, A. J ., Tyler-Kabar a, E. C., et al . (2019).
Restor ed t actile sens ation imp roves neur oprosth etic arm con trol . bioR xiv 653428. doi:10.1101/653428
Flesher, S . N. , Downey, J. E., Weiss, J . M. , Hughes, C. L., Herr era, A. J ., Tyler-Kabar a, E. C., et al . (2021). A
brain-compute r int erface th at evokes tac tile sensa tions improves r obotic a rm control. Scie nc e 372, 831.
doi:10.1126/science. abd0380
Hochberg, L. R. , Ser ruya, M. D., F riehs, G. M., Mukand, J. A ., Sal eh, M. , Caplan, A . H., et al. (2006).
Neuro nal ensembl e cont rol of prosth etic devices by a human with tetr aplegia . Na ture 442, 164-171.
doi:10.1038/natu re04970
Hughes, C. L., Fleshe r, S. N ., W eiss, J. M ., Downey, J. E., Collinger, J. L. and Gaun t, R. A. (2020). Neur al
stimulation a nd reco rding performa nce i n human somatosens ory cort ex over 1500 days. medR xiv
2020.01.21.20018341. doi :10.1101/2020.01.21.20018341
James, C. B. , Nave en, R. , Kaivon, P., Core y, T., Carlos, V.-I., Lachlan , F., e t al. (2013 ). Failure mode an alysis
of silicon-based intraco rtical micro elect r ode arr ays in non-human primat es. J ourn al of Neural
Engineeri ng 10, 066014.
Kim, S., Callier, T., Tabo t, G. A ., Ga unt, R . A., Teno re, F . V. and Bensmai a, S. J . (2015). Behavioral
assessment of sensitivity t o intr acortic al microstimulatio n of primate soma tose nsory corte x. Proce edin gs
of the Na tio nal A ca demy of Sciences 112, 15202. doi:10.1073/pnas.1509265112
Kozai, T. D. Y., Catt , K., Li, X., Gugel , Z. V., Olafsson, V. T., Vazque z, A. L. , et al . (2015). Mechanical failure
modes of chronically implanted pl anar sil icon-based neur al prob es for laminar r ecording. B i om at e r i a l s
37, 25-39. doi: ht tps://doi.org/10.1016/j. biomate rials.2014 .10.040
Lago, N., Cest er, A ., W rachien , N., Nat ali, M., Quiroga , S. D., Bon et ti, S., et al . (2016). A physical-based
equivalent circui t model for an o rganic/e lectrolyt e int erface. Or gani c Elec troni cs 35, 176-185.
doi:ht tps://doi.o rg/10.1016/j.orgel.2016. 05.018
Lee, I.-S ., Whang, C.-N ., Choi, K., Choo , M .-S. and Lee, Y.-H . (2002). Characteriz atio n of iridium film as a
stimulating neu ral el ectro de. Bi om ateri al s 23, 2375-2380. doi: https://doi. org/10.1 016/S0142-
9612(01)00373 -8
Negi, S., B handa ri, R. , Riet h, L. and Solzb acher, F. (2010). In vitro compa rison of sputt ered ir idium oxid e
and platinum-coa ted n eural implan tabl e microelect rode a rrays. Bi ome dica l Mat er ials 5, 015007.
doi:10.1088/1748-6041/5/1/015007
Negi, S., B handa ri, R. , Riet h, L., Van Wag enen, R . and Solzb acher, F . (2010). Neura l electro de degr adati on
from continuous el ectrical stimul ation : Comparison of sputt er ed and activa ted iri dium oxide. Jo urnal of
Neurosci en ce Met ho ds 186, 8-17. doi: ht t ps://doi.org/10.1016/j.jneumeth .2009.1 0.016
Polikov, V. S., Tresco, P. A. and R eicher t, W. M. (2005). Response of br ain tissue t o chronically implante d
neural el ectr odes. Jo urnal of Neurosci enc e Meth ods 148, 1-18.
doi:ht tps://doi.o rg/10.1016/j.jneumeth. 2005.08.015
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Prasad, A., Xu e, Q.-S. , Dieme, R. , Sankar, V., Mayrand, R ., Nishid a, T., e t al. (2014). Abiotic-biotic
characte riza tion of Pt/Ir microel ectr ode arrays in chronic implants . Fron tiers in N euroen gin eering 7 ,
doi:10.3389/fneng.2014.00002
Rousche, P. J . and No rmann, R . A. (1998). Chronic recording capabili ty of the Ut ah Intr acortica l Electro de
Array in cat sens ory cort ex. Jour nal of Ne uroscie nce Me th ods 82, 1-15.
doi:ht tp://dx.d oi.o rg/10.1016/S0165-02 70(98)00031 -4
Salatino , J. W., Ludwig, K. A., Koz ai, T. D. Y. and Purcell, E. K. (2017). Glial respo nses to implant ed
electr odes in th e brain . Na ture Bi ome dic al Engin eerin g 1, 862-877. doi:10.1038/s 41551-017-0154-1
Santhan am, G. , Ryu, S. I ., Yu, B. M ., Afsha r, A. and Sh enoy, K. V. (2006). A high-per formance brain –
computer in terfac e. Na t u re 442, 195-198. doi:10.1038/na ture04968
Schmidt, E. M., McInt osh, J. S . and Bak, M. J. (1988). Long-term implants of Paryl ene-C coated
microelect rodes . Medi cal a nd Bi olog ical Engineeri ng a nd Co mp utin g 26, 96-101.
doi:10.1007/BF02441836
Schwartz, A . B., Cui, X. T., Webe r, Dougla s J. and Mor an, D. W . (2006). Brain-Contr olled In terfaces :
Movement R esto ratio n with Neu ral Pros thetics . Neur on 52, 205-220.
doi:ht tps://doi.o rg/10.1016/j.neuron.20 06.09.019
Simeral, J . D., Kim, S. P., Black, M . J. , Donoghue, J . P. and Hochberg, L. R . (2011). Neural con trol of cursor
traject ory and click by a human with tet r aplegia 1000 days after implan t of an intr acortical
microelect rode a rray. Jour nal of Ne ural E ngine ering 8, 025027 . doi:10.1088/1741-2560/8/2/025 027
Suner, S ., Fellows, M . R., Vargas-I rwin, C. , Nakat a, G . K. and Donoghue, J. P. (2005). Reliability of signals
from a chronically implanted , silicon-bas ed elect rode array in non-human p rimat e primary moto r cort ex.
IEEE Transacti ons o n Neur al Systems a nd Rehabili ta tio n Engi neerin g 13, 524-541.
doi:10.1109/TNSRE.2005 .857687
Szymanski, L. J., Kellis, S. , Liu, C. Y., Jon es, K. T., Ande rsen, R . A., Commins, D., e t a l. (2021).
Neuro path ological effects of chronically i mplanted , intrac ortical micro elec trod es in a tet raplegic pa tien t.
Jour nal of Ne ural En gineeri ng 18, 0460b 9. doi:10.1088/1741-2552/ac127e
Thakore, V., M olnar, P. an d Hickman, J. J. (2012). An Optimization-B ased Stu dy of Equivalent Circuit
Models for Rep resen ting Reco rdings at t he Neu ron–El ectro de In terfac e. IEEE Tra nsact ions o n Biom edic al
Engineeri ng 59, 2338-2347. doi:10 .1109/TBME.2012.2203820
Velliste, M ., Perel , S., Spal ding, M. C., W h itford, A . S. and Schwar tz, A . B. (2008). Cortical con trol of a
prosthe tic arm for self-feeding. Nat ure 4 53, 1098-1101. doi:10.1038/natur e0699 6
Wodlinger , B., Downey, J . E., Tyler-Kabar a, E. C., Schwartz , A. B ., Boning er, M . L. a nd Collinger, J. L.
(2014). Ten-dimensional anthro pomorph ic arm control in a human br ain−machin e interface : difficulties,
solutions, and limi tati ons. J ourn al of Ne u ral Engine ering 12, 016011 . doi:10.1088/ 1741-
2560/12/1/01 6011
Woepp el, K., Yang, Q. a nd Cui, X. T. (2017). Recent advanc es in neural elect rode – tissue int erfaces.
Current Op inio n in Bio medi cal En gineeri n g 4, 21-31. doi: ht tps://doi.o rg/10.1016/j.cobme.2017.09.003
Woolley, A . J. , Desai, H. A . and O tt o, K. J . (2013). Chronic intracortical micro elect r ode arr ays induce non-
uniform, depth-r ela ted tissu e respo nses. Journ al of Neur al Engi neeri ng 10, 02600 7. doi:10.1088/1741-
2560/10/2/02 6007
Xie, X., Riet h, L., Willi ams, L., Negi, S ., Bh andari, R ., Caldwell, R ., et al. (2014). Long-term reliabili ty of
Al2O3and Parylen e C bilayer encapsul ate d Utah el ectro de ar ray based ne ural in te rfaces for chronic
implanta tion. Jour nal of Ne ural En gineeri ng 11, 026016. doi:10 .1088/1741-2560/1 1/2/026016
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Table 1. Number of electrically connected electrodes that were classified as
undamaged/unencapsulated or damaged/encapsulated based on tip degradation, shank
degradation, and tissue encapsulation.
Tip Degradation Shank Degradation Encapsulation
Low (%) High (%) Low (%) High (%) Low (%) High (%)
P1 122 (72.2) 47 (27.8) 151 (84.8) 25 (15.2) 51 (27.6) 134 (72.4)
P2 Pt 146 (90.1) 16 (9.9) 172 (98.3) 3 (1.7) 90 (50.6) 88 (49.4)
P2 IrOx
Medial 26 (92.9) 2 (7.1) 32 (100) 0 (0) 6 (18.8) 26 (81.2)
P2 IrOx
Lateral 7 (25) 21 (75) 9 (32.1) 19 (67.9) 30 (100) 0 (0)
Total 301 (77.7) 86 (22.3) 364 (88.8) 46 (11.2) 177 (41.6) 248 (58.4)
Excluded 45 22 7
Table 2. Differences observed between patients with different length of implant (980 days for P1
and 182 days for P2) on material degradation and encapsulation for electrically connected
platinum recording electrode arrays
P1 (%) P2 Pt (%) Fisher exact p-value
Degraded Tips 27.8 9.9 <0.001
Degraded Shank 15.2 1.7 <0.001
Encapsulated 72.4 49.4 <0.001
Table 3. Effect of stimulation on material degradation and encapsulation for IrOx arrays. Non-
stimulated tips were not electrically connected.
Stimulated (%) Non-stimulated (%) Χ 2 Statistic p-value
Damaged Tips 41.1 5.6 14.7 <0.001
Damaged Shank 33.3 0.0 19.3 0.05
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
E
G
F
H
A B
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Supplemental
Low Degradation/Unencapsulated
High Degradation/Encapsulated
Excluded from analysis
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
A
B
C
D
E
F
G
H
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Figure 1. Six electrode arrays were implanted in two participants: two recording arrays in
P1 motor cortex, two recording arrays in P2 somatosensory cortex, and two stimulating
arrays in P2 medial parietal cortex. Intraoperative images of implanted arrays in P1 (A) and
P2 (B).
Figure 2 : Impedances and peak-to-peak voltages decreased over time. Data points
represent the median across electrodes for a given test date. The shaded regions show the
interquartile ranges smoothed with a nine-point moving average filter with a triangular kernel.
Median impedances recorded on (A) P1 electrodes and (B) P2 electrodes across the length of
implant. Impedance measurements on P1 were not conducted with the same temporal
resolution as P2. Different colors represent platinum or IrOx for P2 as indicated in the legend.
Vpp recorded on (C) P1 electrodes and (D) P2 electrodes across the length of implant. For P1,
there was a discontinuity in the Vpp at day post-implant 550 due to a change in the RMS
threshold from -5.25 to -4.5. Overlayed impedances and Vpp for P1 and P2 are shown in (E)
and (F), respectively.
Figure 3: Characterization of the encapsulation of the electrodes. Arrays were imaged with
an optical microscope in air. Both arrays are from P1. The encapsulation of array (A) was further
examined with TPM. (B) The location of 2P imaging along the Z axis and select electrode
shanks. The array was stained for cell nuclei and zoomed-in images were taken of the green (C)
and red (D) regions. In both regions there is prominent second harmonic signal, indicating the
presence of collagen. The array in (E) was chosen to display the lack of homogeneity of the
encapsulating tissues. (F) Location along the z-axis (blue box) and 2 selected areas further
imaged. 3D rotation images were generated displaying the tissue encapsulation and nuclei
staining from the regions highlighted in green (G) or red (H). The outer image (G) displays high
second harmonic signal while the inner image (H) has elevated cell counts, demonstrating the
heterogeneity of the encapsulation.
Figure 4: Brain vascularization can be visualized on one of the explanted arrays from P2.
The pre-implant location is indicated with a yellow box (A). (B) Optical image of the array
showing tissue coverage. (C) TPM of the shanks of the electrode, with green denoting second
harmonic signal from collagen and red denoting the autofluorescence of the device. Each
electrode was imaged and separated by row (side view). Most of the electrode tips are covered
by collagenous tissue. (D) TPM image of the array looking from the tips downward, with a
portion of vasculature marked in blue. (E) zoomed in image from (A) where electrode shanks
are superimposed on the underlying vasculature. (F) The vasculature visualized in (D) is
superimposed on (E), showing similar trajectory, demonstrating that the vasculature structure
identified in the tissue on the explanted array is likely of pia origin.
Figure 5 . Tip and shank damage occurred on some implanted electrodes and
encapsulation occurred on four implanted arrays. Representative high magnification images
of undamaged/unencapsulated and damage/encapsulated electrodes. Tip images were taken
from P1 array 1, with the degraded tip showing demetallation and biologic fouling (scale bar is
10µm). Shank images were taken from P2 lateral stimulating array (scale bare is 100µm). The
degraded shank shows multiple surface and subsurface irregularities including pitting and
delamination from the tip. Encapsulation images were from P2 medial stimulating array (scale
bare is 100µm).
. CC-BY 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted August 31, 2021. ; https://doi.org/10.1101/2021.08.28.21262765doi: medRxiv preprint
Figure 6. Stimulation-induced material damage on one of the two arrays. (A) SEM image of
four shanks of the lateral array, tip damages are found on the stimulated electrodes marked with
white arrows. (B) EDS of the stimulating electrodes for the lateral array showing reduced
presence of iridium (magenta) on most of the stimulated sites (white arrows). (C) SEM image of
the medial stimulating array tips. No differences were observed between the non-stimulated and
stimulated tips on this array. Scale bars are 100µm. (D,E) Arrays showing the measured