A Green, Versatile, and Facile Strategy for Anti-biofouling Surface with Ultra-high Graft Density Polyethylene Glycol
preprint
OA: closed
Abstract
Abstract Implantable catheters are susceptible to severe complications due to non-specific protein adhesion on their surfaces. Polyethylene glycol (PEG) coatings, the gold standard for resistance to non-specific protein adhesion, present a challenge in achieving high-density grafting, which significantly restricts their use as anti-biofouling coatings. Herein, we exploited the strong interaction between polyphenols (PCs) and polycations (K6-PEG) to graft PEG onto the surface of PC-Cu (A network of metal polyphenols composed of proanthocyanidins and metal copper ions, with expectation for the coating with excellent resistance to non-specific protein adhesion (PC-Cu@K6-PEG). The introduction of K6-PEG resulted in enhanced stability and modulus of PC-Cu, as well as a reduction in the surface adhesion energy and contact angle of PC-Cu. In contrast to previously reported PEG coatings, PC-Cu@K6-PEG exhibited a markedly elevated grafting density of PEG (4.06 chains/nm²), which was more than double the highest value previously documented (1.9 chains/nm²), due to the diffusing ability of K6-PEG throughout the PC-Cu networks. PC-Cu@K6-PEG displays robust resistance to a variety of proteins, microbials, and platelet attachment, thereby preventing thrombosis. The coating ability of PC-Cu onto diverse substrates, coupled with the simple, straightforward and environmentally benign process of fabricating PC-Cu@K6-PEG, suggests that this strategy has significant potential for use in anti-biofouling surfaces.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2024) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.
Source provenance
- europepmc
- last seen: 2026-05-20T01:45:00.602351+00:00