Safety of Bubble Nasal Intermittent Positive Pressure Ventilation (NIPPV) versus Bubble Nasal Continuous Positive Airway Pressure (NCPAP) in preterm infants with respiratory distress
preprint
OA: closed
Abstract
Abstract Objective Nasal Intermittent Positive Pressure Ventilation (NIPPV) is an effective therapy for infants in respiratory distress. We here report the safety of a novel, low-cost, non-electric bubble NIPPV device in comparison with bubble NCPAP.Study Design: At Paramitha Children’s Hospital (Hyderabad, India), preterm (n = 60) neonates with moderate respiratory distress were pragmatically allocated to bubble NCPAP (5–8 cm H2O) or bubble NIPPV (Phigh 8–12 cm H2O/Plow 5–8 cm H2O) based on staff and equipment availability. Primary outcomes to assess safety included clinically relevant pneumothorax, nasal septal necrosis, or abdominal distention.Results One patient in each arm developed minor nasal septal injury (grade 3 on NCPAP, grade 2 on NIPPV); no patients in either arm developed a clinically significant pneumothorax or abdominal distention.Conclusion The similar rates of nasal septal injury, pneumothorax and abdominal distention suggest that bubble NIPPV has a similar safety profile as bubble NCPAP for preterm infants in respiratory distress.