Repeatability of i.Profiler for Measuring Wavefront Aberrations in Healthy Eyes
preprint
OA: closed
Abstract
Abstract Purpose To evaluate the repeatability of wavefront aberration measurements and the correlation between corneal aberration and pupil size in normal eyes using a wavefront-based autorefractor (i.ProfilerPlus; Carl Zeiss Vision, Germany). Methods A prospective cross-sectional study. Wavefront aberrations, including SA (Z40), coma (Z3 − 1, Z31), trefoil (Z3 − 3, Z33) and total higher-order aberrations (tHOA), were measured at different pupil diameters. The repeatability was evaluated using one-way ANOVA method, and statistical indicators including within-subject standard deviation (Sw), test-retest repeatability (TRT), and intra-class correlation coefficient (ICC). The correlations between corneal aberrations and pupil sizes were evaluated by Pearson correlation analysis. Results A total of 96 healthy young volunteers were enrolled. Corneal and ocular higher-order aberrations (HOA) measured by i.Profiler showed Sw<0.01µm, TRT < 0.10µm, ICC > 0.90. There was a linear positive correlation between the corneal HOA and pupil size. The correlation coefficient between SA and tHOA was the largest (r = 0.996, P < 0.001). Conclusions The measurements of wavefront aberrations by i.Profiler are highly repeatable. Corneal HOA is significantly dependent on pupil size. SA is the most influential aberration for visual quality.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.
Source provenance
- europepmc
- last seen: 2026-05-19T01:45:01.086888+00:00