Keywords
NIOSH, COVID-19, SARS-CoV-2, Coronavirus, Novel Coronavirus, 3D Printing, N95,
Respirator, Mask, Qualitative, Fit Testing, Saccharin
Conflict of interest: None
Abstract
The 2019 Novel Coronavirus (COVID-19) has caused an acute shortage of personal protective equipment
(PPE) globally as well as shortage in the ability to test PPE such as respirator fit testing. This limits not
only the ability to fit PPE to medical practitioners, but also the ability to rapidly prototype and produce
alternative sources of PPE as it is difficult to validate fit. At the George Washington University, we
evaluated an easily sourced method of qualitative fit testing using a nebulizer or “atomizer” and a sodium
saccharin solution in water. If aerosolized saccharin entered candidate masks due to poor fit or inadequate
filtration, then a sweet taste was detected in the mouth of the user. This method was tested against
previously fit tested Milwaukee N95 and 3D Printed Reusable N95 Respirator as a positive control. A
Chinese sourced KN95, cotton cloth material, and surgical mask were tested as other masks of interest.
Sensitivity testing was done with no mask prior to fit test. A sweet taste was detected for both the surgical
mask and cotton cloth, demonstrating a lack of seal. However, there was no sweet taste detected for the
Milwaukee N95, 3D Printed Reusable N95 Respirator, or Chinese KN95. These results demonstrate this
could be a valuable methodology for rapid prototyping, evaluation, and validation of fit in a non-clinical
environment for use in creation of PPE. This method should be not be used without confirmation in a
formal qualitative or quantitative fit test but can be used to preserve those resources until developers are
confident that potential new N95 comparable respirators will pass. We strongly suggest validation of
masks and respirators with Occupational Safety and Health Administration (OSHA) approved fit testing
prior to use in a clinical environment.
1.0 Background:
The Occupational Safety and Health Administration (OSHA) defines N95 respirators as tight fitting
filtering face piece respirators or filter cartridges that are not resistant to oil, remove at least 95 percent of
the most-penetrating particles, and must be fit to the face.i Respirators classified as N95 quality face
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
NOTE: This preprint reports new research that has not been certified by peer review and should not be used to guide clinical practice.
pieces are therefore self-contained and thought to have a high level of protection from environmental
hazards. Because these masks must be air-tight around the face, they must also be fit tested. The 2019
Novel Coronavirus (COVID-19) has caused an unprecedented in the number of practitioners seeking N95
protection and corresponding fit testing. This has resulted in a shortage of not just N95 respirators, but
also a shortage in the materials used for fit testing respirators. A survey of 8,200 nurses across the USA
reported that only 55% has access to N95 respirators, and only 65% had been fit tested in the last year.ii
Commercially available fit tests like those produced by 3M cost $250-350 USD, and are currently on
global backorder, and can sometimes take 2 - 3 months to deliver.iii It is notable that a delay of 2-3
months in these resources make the use of this supply chain impractical given that peak resource use due
to COVID-19 in the United States is expected on April 15th, 2020, which is within several weeks of the
writing of this article.iv
Considering these shortages, OSHA has instituted temporary guidance to switch from quantitative
Methods
of testing N95 respirators to qualitative methods.v OSH recommends four substances for testing
qualitative fit of N95 respirators vi:
▪ Isoamyl acetate, which smells like bananas;
▪ Saccharin, which leaves a sweet taste in your mouth;
▪ Bitrex, which leaves a bitter taste in your mouth; and
▪ Irritant smoke, which can cause coughing.
Our group at the George Washington University and Hospital determined that there was a need for a
qualitative test substitute to continue to rapidly prototype, iterate, and test N95 respirators to compensate
for the shortage of PPE. We developed a protocol to mirror OSHA qualitative testing processes and detail
it in this report. Our method should be not be used without confirmation in a formal OSHA approved
qualitative or quantitative fit test, the method but can be used to preserve those resources until developers
are confident that potential new N95 comparable respirators will pass.
2.0 Methods:
Current OSHA Recommendations:
The United States Occupational Safety and Health Administration defines two official methods for fit
testing of N95 equivalent masks: qualitative using four substances that a user can smell or taste, and
quantitative that look for negative pressure and aerosol. (OSHA, regulation 1910.134)vii The four
qualitative methods as defined above are Isoamyl acetate, which smells like bananas; Saccharin, which
leaves a sweet taste in the mouth; Bitrex (denatonium benzoate), which leaves a bitter taste in the mouth;
and Irritant smoke, which can cause coughing. OSHA also defines three quantitative methods of
generated aerosol, ambient aerosol, and controlled negative pressure.
The fit test involves a sensitivity screening following by a mask fit testing with aerosolized solution as
defined by the four parameters above. The sensitivity screening involves placing the subject under a hood
without a respirator and aerosolizing one of the four qualitative substances. The subject is asked if they
can smell the bananas for Isoamyl acetate, taste the sweet taste for Saccharin, bitter taste for Bitrix, or
cough for the smoke. If the fit is positive the subject then puts on the mask and is routinely aerosolized
within the hood. While the subject is wearing the mask, they are then instructed to test several conditions
including: a) normal breathing b) deep breathing c) turning head side to side d) moving head up and down
e) reciting a passage f) jogging in place and e) normal breathing. The test is finished and considered
successful if the subject can not taste or smell the substance under any of the aforementioned conditions.
If at any point the subject can sense the substance, the test is failed and the fit is judged to be inadequate.
Aerosolization method:
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
We reviewed the four qualitative methods for the OSHA recommended fit test. Of the four, Saccharin was
determined to be low-cost, readily available, and easily detectable. For these reasons Saccharin was
selected for candidacy in developing an alternative fit test. We used pellets of sodium saccharin made by
NectaSweet, and purchased from a commercial vendor (Amazon, etc.). The tablets contain approximately
30 mg of sodium saccharin each, with a binder of sodium bicarbonate and cellulose. An alternative
source may also include “Sweet n Low” packets, which contains approximately 36 mg of sodium
saccharin per packet. However, we used NectaSweet tablets because it contained less binder material.
NectaSweet tablets were dissolved into a warm neutral water solvent. To form this solution 60 ml water
was first warmed to near boiling and then 10 pellets (each containing 30 mg of sodium saccharin) were
added to the water and stirred until fully dissolved. The solution was then cooled to room temperature
before use. The final solution had a concentration of 5 mg/ ml of sodium saccharin.
Commercial OSHA test typically uses a nebulizer to aerosolize particles. We repurposed a nebulizer in
the form of an atomizer (a cosmetic spray product available at commercial retailers like Amazon for $10-
30 USD) to test fit. A clinical nebulizer used for aerosolizing inhaled medications is also acceptable, if it
is available. The subject was instructed to stand within a confined space to mimic the hood. The
saccharin solution was loaded into the nebulizer then sprayed first next to the subject, but not directly at
the subject, without respirator protection. The subject had to verify they were able to taste the substance
before moving to the next step. The subject was instructed to clean their palate of the saccharin with a
neutral substance like a glass of water. Next the substance was sprayed around the subject, but not
directly at the subject, to verify the subject could not taste the substance through the respirator. Various
movements were instructed similar to the commercial fit test such as normal breathing, deep breathing,
turning head side to side, moving head up and down, speaking, jogging in place, and normal breathing
once again. If the subject could not taste the substance under any of these conditions, the fit of the mask
was deemed to be adequate and to pass the test.
Filtration methods tested:
Five filtration and respirator designs were tested using the above saccharin and nebulizer combination.
First a subject was instructed to step into a confined space and was sprayed with saccharine solution from
the nebulizer. Every subject for every test was able to taste saccharine within fifteen seconds of initial
exposure. Next the subject was instructed to drink water to neutralize their palate, then was fit with the
respirator. Each mask was tested multiple times with no regard for test order (surgical mask was test
before and after the N95 comparable respirators for example to ensure negative result was not false
negative).
The following five methodologies were tested (Figure 1):
1. 3D Printed Reusable N95 Respiratorviii
2. Kimberly Clark Surgical Mask
3. Powecom KN95 (Sourced from China)
4. Milwaukee N95
5. Hanes Cotton cloth
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
Figure 1: In order from left to right: 3D Printed Reusable N95
Respirator, surgical mask, KN95, N95, cotton cloth.
The 3D Printed Reusable N95 Respirator, KN95, and Milwaukee N95 were all considered N95 or
comparable and were first tested for fit and suction prior to test. Subjects placed mask and then performed
a qualitative check for any leaks around the surface. The 3D printed Reusable N95 Respirator and
Milwaukee N95 respirator had previously passed a qualitative Bitrix fit test at GWUH employee health
and functioned as a positive control for the experiment.
Cotton cloth and surgical mask were secured prior to test with elastic, however were not considered N95
equivalent and were included as negative controls.
3.0 Results:
Qualitative Pre-Assessment of N95 or comparable respirator fit:
All three of the N95 or comparable respirators were first visually examined and determined to have no
leaks. The 3D printed reusable N95 respirator had the tightest seal, where as the Powecom KN95
appeared to have the least tight seal. Visual inspection of the KN95 revealed perforation along the nose,
although this did not seem to ultimately impact the filtration capabilities.
Figure 2: From left to right: Milwaukee N95 Respirator, Powecom
KN95, 3D Printed Reusable Respirator. Perforations are visible at
the top of middle mask (KN95).
Qualitative fit test results:
Two subjects wore each of five masks two times to have the qualitative saccharin fit test in a confined
area.
Both subjects were able to taste the saccharin solution immediately (< 5 seconds of exposure) as a
sensitivity test and the test was continued for each respirator. Cotton cloth was fit loosely over subjects’
face and surgical mask was fit according to manufacture instructions to hook elastic around the back of
the ears. Neither the cotton cloth nor the surgical mask was able to obtain a qualitative fit (Figure 3):
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
Figure 3: Example lack of
suction with Kimberly Clark
surgical mask.
Both cloth and surgical mask failed qualitative saccharin solution fit test within 15 seconds of initial
exposure and no additional testing was taken. Surgical mask was tested repeatedly to ensure this was not a
false result. On average, sweet taste was detected within 45 seconds with a surgical mask or cotton cloth.
All three of the remaining respirators resulted in no detectable taste of the saccharin solution by the
wearer. The KN95 mask despite initial pre-assessment qualitative appearances, withstood multiple tests
and movements and did not allow any particulates through for taste. It should be noted that the N95 as
well as the 3D printed reusable N95 respirator had previously passed a qualitative Bitrix fit test at GWUH
employee health (Results not shown).
The 3D Printed Reusable Respirator was tested with MERV 16 filter wrapped in MERV 13, which
previously passed the Bitrix fit test. Subject reported no ability to taste saccharine while the respirator and
filter were in place. Once the cartridge was removed, the subject was able to rapidly discern the saccharin.
4.0 Discussion:
The process described above demonstrated that it is possible to emulate a clinical qualitative fit test using
easily available materials. The sensitivity test was detectable for all tests of all masks and all subjects,
suggesting the atomizer and saccharin solution functioned comparable to a nebulizer for this purpose
similar to clinically used Saccharin or Bitrix in qualitative fit tests. The N95 respirator has been approved
by OSHA and US NIOSH, functioning as high efficiency particulate filter against coronavirus, and a
positive control for this study. (OSHA 29CFR1910.134) The 3D Printed Reusable N95 Respirator had
been previously tested in employee health, so it also served as a positive control for this purpose.
Most interesting was the KN95. The KN95 mask is outsourced from China and was approved for use in
the USA by the CDC in the setting of resource shortage caused by COVID-19.ix (CDC, GB/T 18664—
2002). This test passed the qualitative fit test described in this report, but has not yet been evaluated with
a OSHA certified fit test at GWUH.
The cloth mask provided the least protection against detection of saccharin aerosols and can not be
expected to be a tight sealing mask. This result was expected, but is of particular interest due to
widespread creation of cloth masks in the community. Additionally the Kimberly Clark surgical mask
also did not pass the seal test and also demonstrates that it is not adequate protection to aerosol
particulates. However, both masks resulted in some delay in the detection of saccharin aerosol compared
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
to no mask at all. Although it cannot be assumed that either cotton cloth nor surgical mask would provide
complete protection against aerosols like the 2019 Novel Coronavirus, they are likely better than not
wearing any mask at all.
These results suggest similar tests could be used to rapidly prototype and fit N95 comparable respirators
outside of a clinical environment due to the shortage of PPE, personnel, and access to clinical grade fit
testing. It should be emphasized that after passing the protocol described in this study, clinical fit testing
should still be performed and passed before N95 equivalent masks or respirators are put into use with
COVID 19 positive patients. However, this study does allow for rapid testing and cycling of designs and
may speed development of PPE during the COVID-19 pandemic in the setting of resource limitations.
5.0 Conclusion:
Aerosols created using a nebulizer and an easily sourced saccharin solution were able to be used to test
the fit of several different N95 equivalent and non-equivalent masks and respirators. This could provide a
valuable non-clinical grade evaluation of N95 respirators to rapidly develop PPE during the COVID-19
pandemic. We strongly suggest validation of masks and respirators with OSHA approved fit testing prior
to use in a clinical environment.
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
6.0 References:
i https://www.osha.gov/video/respiratory_protection/resptypes_transcript.html
ii National Nurses Association: “NNU COVID-19 Survey Results.” Survey. March 31, 2020.
iii https://www.envirosafetyproducts.com/3m-bitrex-qualitative-fit-test-kit.html
iv https://covid19.healthdata.org/projections
v https://www.osha.gov/news/newsreleases/national/03142020
vi https://www.osha.gov/video/respiratory_protection/fittesting_transcript.html
vii https://www.osha.gov/video/respiratory_protection/fittesting_transcript.html
viii Provenzano, D.; Rao, Y.J.; Mitic, K.; Obaid, S.N.; Pierce, D.; Huckenpahler, J.; Berger, J.; Goyal, S.; Loew,
M.H. Rapid Prototyping of Reusable 3D-Printed N95 Equivalent Respirators at the George Washington
University. Preprints 2020, 2020030444 (doi: 10.20944/preprints202003.0444.v1).
ix https://www.cdc.gov/coronavirus/2019-ncov/hcp/respirators-strategy/crisis-alternate-strategies.html
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint
. CC-BY-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 11, 2020. ; https://doi.org/10.1101/2020.04.06.20055368doi: medRxiv preprint