Improving amplification bandwidth by combining Josephson Parametric Amplifiers for Active Axion Search Experiments at IBS/CAPP

preprint OA: closed
View at publisher

Abstract

Abstract IBS/CAPP in South Korea is home to multiple active axion search experiments using cavity haloscopes that operate within the frequency range of 1-6 GHz. The haloscopes convert axions to photons, resulting in an output power of about 10-24 to 10-22 W. To detect such a small signal amidst noise, quantum-limited noise amplifiers and ultra-low-temperature environment (∼10 mK) are required for all critical read-out components to minimize noise from all of active and passive lossy components. Additionally, operating at low dissipation levels is necessary to maintain the haloscope and amplifier at a temperature of 25-40 mK. Our primary objective is to achieve the highest possible scanning-frequency speed, which includes the time for maintenance and system calibration. For the CAPP-MAX flagship experiment, we achieved a record speed of 3 MHz/day for the DFSZ-level axion sensitivity during uninterrupting measurements. This paper presents the development and operation of low-noise amplifiers for five haloscope experiments targeting different frequency ranges and provides design, operational, and performance details of the amplifiers.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00