Results
Surgical procedures were completed with the Auto-
Lap™ system in 64 cases (97%). The mean overall setup
time of the AutoLap™ system was 4 min (04:08 ± 0.10).
Procedure times were not prolonged due to the use of the
system when compared to literature average. The reported
user satisfaction was 3.85 and 3.96 on a scale of 1 to 5 in two
studies. More than 90% of the image-based movements were
accurate. No system-related adverse events were recorded
while using the system.
Conclusion
Safe and efficient use of the core technology
of the AutoLap™ system was demonstrated with high image
stability and good surgeon satisfaction. The results support
further clinical studies that will focus on usability, improved
ergonomics and additional image-based features.
Keywords
Robotic · Steering · Camera holder ·
Laparoscopy · Active camera control systems · Autolap™
Visualization in endoscopic surgery is a challenge for both
surgeon and camera assistant. The surgeon highly depends
on the camera assistant to provide a stable, centered, and
non-rotated image of the target area and has no control other
than verbal commands or manual correction, which requires
release of one of the instruments.
An often encountered pitfall is that the image is unsta-
ble due to tremor, unintended movements or rotation of the
camera by the surgical assistant. Furthermore, manual con-
trol can also be physically demanding leading to fatigue and
suboptimal control. Inexperienced or inattentive assistants
may displace the camera frequently, make jerky movements,
malposition the laparoscope in relation to the horizon, and
point the camera outside the focus of interest more often
Abstract
Background Robotic camera holders for endoscopic sur -
gery have been available for 20 years but market penetration
is low. The current camera holders are controlled by voice,
joystick, eyeball tracking, or head movements, and this
type of steering has proven to be successful but excessive
disturbance of surgical workflow has blocked widespread
introduction. The Autolap™ system (MST, Israel) uses a
radically different steering concept based on image analysis.
This may improve acceptance by smooth, interactive, and
fast steering. These two studies were conducted to prove safe
and efficient performance of the core technology.
Methods
A total of 66 various laparoscopic procedures
were performed with the AutoLap™ by nine experienced
surgeons, in two multi-center studies; 41 cholecystectomies,
13 fundoplications including hiatal hernia repair, 4 endome-
triosis surgeries, 2 inguinal hernia repairs, and 6 (bilateral)
salpingo-oophorectomies. The use of the AutoLap™ system
was evaluated in terms of safety, image stability, setup and
and O ther Inte rventional Techniques
* Paul J. M. Wijsman
[email protected]
1 Deparment of Surgery, Meander Medical Center, Maatweg
3, Amersfoort, The Netherlands
2 Department of Surgery, University Medical Center Utrecht,
Utrecht, The Netherlands
3 Assia Medical Group, Assuta Medical Center, Tel Aviv,
Israel
4 Department of Surgery, Niguarda Cà Granda Hospital,
Milan, Italy
5 Department of Gynecologic Oncology, UMC Utrecht Cancer
Center, Utrecht, The Netherlands
6 Department of Surgery, Assuta Medical Center, Haifa, Israel
2561Surg Endosc (2018) 32:2560–2566
1 3
than experienced assistants [1]. In addition, touching tissues
with the camera results in image blurring and a need for fre-
quent removal of the camera for cleaning. These parameters
can cause motion sickness and surgeon distraction that may
Result
in prolonged procedures and patient safety hazards [2].
Robotic devices have been introduced 20 years ago to
overcome some of these pitfalls [2]. A robotic laparoscopic
positioner can perform the task of the surgical assistant and
enables the surgeon to control camera movements person-
ally. Robotic camera holders provide a stable, non-rotated
field of view with a focus based on the surgeon’s preference.
In the last two decades, different robotic arms have been
developed, controlled by voice, joysticks, eyeball tracking,
and head movements. The various commercialized systems
have proven to be reliable and efficient but market penetra-
tion is low. This might be due to excessive hindrance of
surgical flow as a result of cumbersome control modalities
causing surgeons to decide to rely on human assistants.
The AutoLap™ system (MST, Israel) aims to overcome
unsuccessful camera control concepts by radically changing
laparoscope manipulation technology. The steering concept
is based on image analysis and computer-based instrument
recognition instead of direct electromechanical steering
of a holder that is unaware of the relation between camera
position and field of view. The aim of these studies were
to evaluate reliability and efficiency of the core electrome-
chanical elements of the Autolap™ system in clinical use.
The system was evaluated in terms of safety, image stability,
procedural and setup time, accuracy of image-based move-
ments and user satisfaction during general and gynecological
laparoscopic surgeries in two multi-center studies, which
were conducted in Israel, the Netherlands, and Italy.
Materials and methods
System’s description
The AutoLap™ (MST, Israel) is an image-guided system
that utilizes unique image processing algorithms to enable
the motorized movement of the laparoscope at any oblique
path by following the movement of a designated surgical
instrument. It can be mounted on either side of the operating
bed rail, providing flexibility for supporting various laparo-
scopic procedures and negates the need for calibration if the
patient position is altered. The system comprises a robotic
motion assembly unit and a processing unit, which holds the
AutoLap™ unique software and algorithms (Fig. 1). The
robotic motion assembly unit, which holds the laparoscope
(Fig. 2), is covered by a sterile sleeve during the procedure
(Fig. 3). Control over camera’s movements and operative
field is enabled by a small wireless sterile disposable button,
which transmits RF signals and is worn by the surgeon as a
ring or is attached to the surgical tool (Fig. 4).
There are three modes of operation, manual operation
with a force joystick and two modes initiated by the wireless
button: “joystick mode” and “follow-me mode.” In joystick
mode, the surgeon can move in four directions; up, down,
left, right, and zoom in/out. In follow-me mode, the system
movements are controlled by tracking the movements of a
Fig. 1 The main components of
the AutoLap™ system
2562 Surg Endosc (2018) 32:2560–2566
1 3
designated tool, which is selected by the surgeon. The sys-
tem is able to recognize any off-the-shelf surgical tool used
in laparoscopic surgery and does not need special marking
or brand.
Additionally, the use of the data gathered from the video
image enables additional image-based features such as per-
forming zoom in/out movement with an angled laparoscope
while maintaining the center of the displayed image, correc-
tion of directional movement in case of camera rotation, pro-
vide adaptive velocity according to the working zoom level
and enable digital zoom using the AutoLap™ controller.
Study design
Two multi-center studies have been performed in five cent-
ers in three countries [Assuta Medical Center Haifa (Israel),
Assuta Medical Center Tel Aviv (Israel), Meander Medi -
cal Center (the Netherlands), Niguarda Cà Granda Hospi-
tal (Italy), UMC Utrecht (the Netherlands)]. The first study
was conducted in 2013–2014 to evaluate safety and perfor -
mance of the AutoLap™ system (the “safety study”, Clini-
cal Trials.gov identifier NCT01828580). The second study
was conducted in 2014–2016 to evaluate the follow-me
mode (the “follow-me study”, Clinical Trials.gov identifier
NCT02326870). Patients who were scheduled for hiatal her-
nia repair/fundoplication, cholecystectomy, endometriosis
surgery, (bilateral) salpingo-oophorectomy and inguinal her-
nia repair were included in both studies. The clinical studies
were approved by the centers’s relevant Ethic Committees
and if required also by the Ministry of Health.
Inclusion criteria were age > 18 and signed informed
consent form for both studies. Exclusion criteria for the
safety study were previous abdominal surgery and con-
traindications to pneumoperitoneum, pregnancy, obesity
(BMI > 35 kg/m
2), generalized peritonitis, septic shock
from cholangitis, severe acute pancreatitis, uncorrected
coagulopathy, advanced cirrhosis with failure of hepatic
function, suspected gallbladder cancer, acute cholecystitis,
Fig. 2 Laparoscopic- and base unit of the AutoLap™ system
Fig. 3 Setup of the AutoLap™ system during a laparoscopic fun-
doplication
Fig. 4 The AutoLap™ Command Unit—ring configuration
2563Surg Endosc (2018) 32:2560–2566
1 3
presence of any medical or psychiatric condition or any
other condition that, in the opinion of the investigator,
could affect the successful participation of the patient in
the study and patient participates in any other clinical
study 60 days prior to the start of the study and through-
out the study duration. Exclusion criteria for the follow-me
study were pregnancy, American Society of Anesthesiolo-
gist (ASA) classification > 2, and extensive adhesions that
preclude the standard laparoscopic surgical technique.
Primary outcomes of the both studies were adverse
events and performance evaluation [operation time (skin
to skin), number of successful movements]. The procedure
time for each type of operation was measured; however,
no comparison was made with a control group. The aver -
age percentage of successful movements was calculated by
the number of successful movements divided by the total
number of movements during the procedure × 100%. A
movement is deemed successful if the laparoscope reached
the desired position, which was verbally verified with the
surgeon after each movement.
Secondary outcomes were system setup time (including
installation, draping and positioning), the number of times
the laparoscope was removed for cleaning and a usability
evaluation (questionnaire). Usability questionnaires were
completed postoperatively by the surgeon for subjective
evaluation of usability aspects, specifically regarding sys-
tem handling, image stability, effort, satisfaction, and the
efficiency to perform the procedure with the AutoLap™
system. Answers were given on a scale ranging from one
(disagree completely) to five (agree completely, the most
positive response). Different usability questionnaires were
used for both studies comprising of 15 questions for the
safety study and 18 questions for the follow-me study.
Data analysis
In addition to the primary and secondary outcomes, demo-
graphic data were recorded including age, gender, ASA
score, weight, height, and BMI. The data were analyzed
with Excel (Microsoft, Redmond, Washington, 2016).
Results
From January 2013 to October 2015, 66 patients were
enrolled to participate in two multi-center studies in five
centers. A total of nine surgeons participated in the stud-
ies. The demographic data of the patients are shown in
Table 1. The number of procedures performed per center
is shown in Table 2.
Setup and operation time
The mean setup time was 4 min (4:08 ± 0.1). In most cases,
the system setup was performed while the patient was being
prepared for surgery. Draping of the system was mostly done
during insufflation of the abdomen, integrating the setup of
the system in the normal OR setup time. The mean operative
time in minutes for the different procedures are descripted
in Table 3.
Laparoscope cleaning
The laparoscope was not removed at all in most of the pro-
cedures. An average of one (0.72 range 0–5) removal per
procedure was recorded. In some cases, the need to clean
the scope was due to fogging that resulted from tempera-
ture differences and not due to smearing of the lens of the
laparoscope.
Number of successful movements
On average, 99 joystick- and 12.8 follow-me movements
were made during a procedure. More than 90% of these
movements were successful in both studies. Most of the
surgeons used the joystick mode for small movements close
to the tissue and the follow-me mode for larger movements,
further away from the tissue.
Safety and usability evaluation
No system-induced complications occurred during all the
performed procedures. The usability questionnaires dem-
onstrated adequate satisfaction of all surgeons in all of the
procedures with a median score of four (on a scale of 1–5)
Table 1 Demographic data of the patients
a Recorded only in the safety study
b Recorded only in the follow-me study
All patients (n = 66)
Age in years (range) 48.9 (23–73)
Gender
Male 17 (25.75%)
Female 49 (74.25%)
BMIa (n = 33) 25.7 (18.7–36.2)
ASAb (n = 32) 1.72 (1–3)
Procedure type
Cholecystectomy n = 41
Hiatal hernia/fundoplication n = 13
Endometriosis surgery n = 4
Salpingo-oophorectomy n = 6
Inguinal hernia repair n = 2
2564 Surg Endosc (2018) 32:2560–2566
1 3
in both studies. The average usability score of the safety
study was 3.85 (± 0.37) and 3.96 (± 0.54) for the follow-me
study. The satisfaction specifically related to the usage of the
follow-me mode was 3.92. All surgeons expressed satisfac-
tion from the high image stability provided by the system
(4.41) as well as the advantage of full control over the field
of view (3.72), intuitive use (3.76), and anticipated short
learning curve (3.97). Less satisfaction was expressed from
the release of the laparoscope for cleaning, which had the
lowest average usability score of 3.1.
Discussion
Minimal invasive surgery has become a leading surgical
modality. Every year, the rate of procedures performed is
growing. However, laparoscopy requires a high demand
of skill, concentration, and maneuverability leading to
increased fatigue and physical discomfort for primary sur -
geons and assistants [3 ]. Operating in the small pelvis and
the upper abdomen is especially physically challenging and
burdensome. These complex surgeries require the assistant
to hold the camera for a long period of time at the same, fre-
quently angled, position. The resulting uncomfortable static
working postures can lead to increased strain to different
parts of the musculoskeletal system [ 4–6]. By improving
ergonomics, primary surgeons and assistants may greatly
benefit, allowing them to perform more complex and lengthy
procedures with ease.
Different robotic active camera holders have been intro-
duced in order to improve ergonomics and overcome some
disadvantages of laparoscopy. One of the first systems on the
market was the AESOP™ system (Intuitive Surgical, Inc.),
which was controlled by voice or manually. The LapMan™
(Medsys S.A, Belgium) uses a joystick that is connected
to the surgical tool to steer and the EndoAssist™ (Arm-
strong Healthcare LLC, UK), which is the first version of
FreeHand™, uses a combination of a headset with motion
sensor and a foot pedal to control the laparoscope’s move-
ments. Smaller systems include the ViKY™ system (Endo-
control, France), FreeHand™ system (OR Productivity plc,
UK) and the Soloassist™ (Actormed, Barbing, Germany).
The Soloassist™ controls the scope with a joystick and the
ViKY™ system with voice control in combination with a
foot pedal to ensure that movement is only possible when
the pedal is pressed.
Market penetration of active robotic camera holders is
low. Despite the fact that several types of movement controls
have proven to be successful, excessive disturbance of the
surgical workflow has blocked widespread introduction. The
main difference of the AutoLap™ system with other systems
on the market is the usage of a different steering mechanism
by image processing software. The AutoLap™ system is not
limited to perpendicular movements alone, as can only be
done by other systems in the market, but enables the surgeon
to move the laparoscope in any direction and path. Acquir -
ing the video images and using image processing algorithms
enables the AutoLap™ to continuously detect the instru-
ments within the field of view and control the laparoscope’s
movements according to the instruments’ movements, as
commanded by the surgeon.
In these studies, the accuracy of the image processing
algorithms was 90%. Failure of detecting the tip of the tool
is caused by several factors. First of all, the image quality
needs to be sufficient in order for the algorithm to perform
properly. Factors that adversely affect the image quality are
fogging, blurring, and smearing of the lens. The amount of
light in the field of view is also an important factor. A dark
image due to poor equipment or a narrow and deep working
space like the pelvis can contribute to poorer detection of
the tip. Additionally, user errors were made. For instance,
Table 2 The number of
procedures performed per center Number of procedures performed Center
Amersfoort Utrecht Milan Haifa Tel Aviv
Cholecystectomy 19 9 7 0 6
Hiatal hernia/fundoplication 8 0 0 0 5
Endometriosis surgery 0 0 0 4 0
Salpingo-oophorectomy 0 5 0 1 0
Inguinal hernia repair 0 0 2 0 0
Total 27 14 9 5 11
Table 3 Operation time
a Both procedures started with the AutoLap™ system but were not
completed with the system
Time in minutes
Cholecystectomy (n = 41) 53 (22–87)
Hiatal hernia/fundoplication (n = 13) 69 (50–90)
Endometriosis surgery (n = 4) 58 (40–80)
Salpingo-oophorectomy (n = 6) 59 (49–70)
Inguinal hernia repaira (n = 2) 108 (74–142)
2565Surg Endosc (2018) 32:2560–2566
1 3
activating the detection algorithm when the tip of the instru-
ment is not within the field of view. Lastly, this was the first
clinical experience with the image processing algorithms.
Small bugs and glitches were detected and corrected during
the clinical work, resulting in a new software version.
The AutoLap™ system demonstrated to be an effective
and safe system for robotic camera control during different
types of abdominal procedures in our studies. High image
stability leading to high surgeons’ satisfaction was reported
in the questionnaires. No system-induced complications
occurred and in most procedures the laparoscope did not
need to be cleaned due to unintended contact with tissue.
The average setup time of the AutoLap™ system is found
to be comparable with other active camera holders. Wagner
et al. reported an average setup time for the Endoassist™ and
the AESOP™ of 2 ± 0.8 and 5.3 ± 2.4 min, respectively [7 ].
Kommu et al. also measured the setup time for the Endoas-
sist™, varying from 5.1 ± 1.2 to 6.8 ± 2.3 min depending
on the type of procedure [8]. The average setup time for the
SoloAssist™ was reported to be 7 min by Beckmeier et al.
[9] There was a significant learning curve of 20 procedures
regarding the mounting of the system [9 ]. Maheswari et al.
reported an average setup time of 3–5 min using the ViKY™
system in a small study of three patients [10].
During conventional laparoscopy, about 7% of the OR
time is spent on cleaning the scope according to a recent
article of Yong et al. [11]. Besides the fact that this is a
considerable amount of valuable OR time, cleaning of the
scope can disrupt the flow of the operation and be very
frustrating. In our AutoLap™ trials, the scope was cleaned
only once per operation on average. Other trials on robotic
active camera holders also demonstrated less frequent clean-
ing of the scope and a reduction or equal total OR time [1 ,
12–14]. However, a longer OR time was demonstrated in a
study by Gillen et al. using the Soloassist™ system, favor -
ing the human assistant [15]. Kommu et al. reported more
frequent cleaning of the scope during nephrectomies using
the EndoAssist™, which was probably related to the product
design [8].
Although there is no comparison with human assistance
in our studies, the measured operational times when using
the AutoLap™ system are in line with the literature [16–20].
Comparative studies between human- and robotic camera
assistance demonstrated similarity in operational procedure
times.[12, 21–23] However, the transabdominal preperito-
neal hernia procedure (TAPP) took longer compared to lit-
erature due to hindrance of system, limiting the movement
range of the surgeon’s tool. A new curved design of the bar
solves this problem, and successful TAPP procedures are
already performed with the system.
The follow-me mode is based on instrument recogni-
tion, followed by dragging the camera to the new focus
point by a smooth instrument movement. An alternate
mode of the follow-me, which has been developed fol-
lowing these studies, is the go-to mode. This mode enables
the surgeon to use an instrument to tag the new desired
center field of view. The surgeon moves the tagging tool
to the new desired center field of view and releases the
button of the remote controller. The field of view will then
be centered around the virtually marked new position, at a
comparable distance from the tissue. Thus, without losing
grip of the surgical instrument, the surgeon only needs to
point and click to adjust his desired field of view.
The Autolap™ system may prove economic value by
reducing the number of surgical team members. Several
procedures in these studies were performed as solo sur -
gery, i.e., without the need of a surgical assistant. The
operation was performed with a surgeon and scrub nurse
alone. Greater involvement of the scrub nurse may facili-
tate fewer OR personnel—a factor that may have an impor -
tant economic impact where skilled staff shortage is an
obstacle to the performance of laparoscopic surgery.
The main limitations regarding the performed studies
must be noted. There is a potential risk of bias due to het-
erogeneity; a total of nine surgeons, in five different cent-
ers, performed six different types of procedures. Moreover,
the learning curve was not measured. It is possible that
some surgeons were still in their learning process, which
can influence the length of the procedures.
There are also some limitations regarding the Auto-
Lap™ system. Correct positioning of the system on the
bedrail is very important to have maximum movement
range. Sometimes, adjustments to the positioning have
to be made by a non-sterile OR personnel during the
procedure. Although the system main compartments are
rather large in appearance, positioning of the system is
quite flexible. Various types of complex procedures have
been performed successfully with the system outside the
studies, including colectomies, surgery in the pelvis and
gastric surgery. Also, using the wireless disposable button
without losing grip of the instrument takes some practice.
Improvements of these limitations are being made and will
be implemented soon. Examples are a new command unit
design, easy release of the laparoscope and the possibil-
ity to reposition the system by single button activation
by a sterile person. Future perspectives of the system are
development of the next generation system including a
smaller footprint and the expansion of the image-based
properties, in search for the most effective image-based
steering technology without disturbance of surgical flow.
In conclusion, the AutoLap™ system is an effective,
safe and easy to use system for robotic camera driving dur -
ing a variety of abdominal procedures. Future studies will
focus on evaluating additional image-based features, ergo-
nomic and economic advantages while using the AutoLap™
system.
2566 Surg Endosc (2018) 32:2560–2566
1 3
Acknowledgements
The study was sponsored by Medical Surgery
Technologies ltd (M.S.T). The company paid the required fees to the
ethics committee and all other relevant study-related expenses. No
other benefits were received by participating in this study.
Compliance with ethical standards
Disclosures Paul Wijsman is a Clinical Field Engineer of Medical
Surgery Technologies ltd (M.S.T) since 2016. Ivo Broeders, Amir
Szold, and Yuval Kaufman are the paid members of the Clinical Advi-
sory Board of M.S.T. Hylke Brenkman, Antonello Forgione, Henk
Schreuder, Esther Consten, Werner Draaisma, Paul Verheijen, and Jelle
Ruurda have no conflicts of interest or financial ties to disclose.
References
1. Dunlap KD, Wanzer L (1998) Is the robotic arm a cost-effective
surgical tool? AORN J 68:265–272
2. Ballantyne GH (2002) The pitfalls of laparoscopic sur -
gery: challenges for robotics and telerobotic surgery.
Surg Laparosc Endosc Percutan Tech 12:1–5. https://doi.
org/10.1097/00129689-200202000-00001
3. Berguer R (1999) Surgery and ergonomics. Arch Surg 134:1011–
1016. https://doi.org/10.1001/archsurg.134.9.1011
4. Hu C-L, Yang C-Y, Lin Z-S, Yang S-Y, Kuo C-H, Lin M-T (2013)
An interactive method for achieving ergonomically optimum
conditions during laparoscopic surgery. J Robot Surg 7:125–130.
https://doi.org/10.1007/s11701-012-0353-4
5. Miller K, Benden M, Pickens A, Shipp E, Zheng Q (2012) Ergo-
nomics principles associated with laparoscopic surgeon Injury/
Illness. Hum Factors J Hum Factors Ergon Soc 54:1087–1092.
https://doi.org/10.1177/0018720812451046
6. Van Der Schatte Olivier RH, Van’t Hullenaar CDP, Ruurda JP,
Broeders IAMJ (2009) Ergonomics, user comfort, and perfor -
mance in standard and robot-assisted laparoscopic surgery. Surg
Endosc Other Interv Tech 23:1365–1371. https://doi.org/10.1007/
s00464-008-0184-6
7. Wagner AA, Varkarakis IM, Link RE, Sullivan W, Su LM
(2006) Comparison of surgical performance during laparoscopic
radical prostatectomy of two robotic camera holders, EndoAs-
sist and AESOP: a pilot study. Urology 68:70–74. https://doi.
org/10.1016/j.urology.2006.02.003
8. Kommu SS, Rimington P, Anderson C, Rané A (2007) Initial
experience with the EndoAssist camera-holding robot in laparo-
scopic urological surgery. J Robot Surg 1:133–137. https://doi.
org/10.1007/s11701-007-0010-5
9. Beckmeier L, Klapdor R, Soergel P, Kundu S, Hillemanns P,
Hertel H (2014) Evaluation of active camera control systems in
gynecological surgery: construction, handling, comfort, surger -
ies and results. Arch Gynecol Obstet 289:341–348. https://doi.
org/10.1007/s00404-013-3004-8
10. Maheshwari M, Ind T (2015) Concurrent use of a robotic uterine
manipulator and a robotic laparoscope holder to achieve assistant-
less solo laparoscopy: the double ViKY. J Robot Surg 9:211–213.
https://doi.org/10.1007/s11701-015-0518-z
11. Yong N, Grange P, Eldred-Evans D (2016) Impact of laparo -
scopic lens contamination in operating theaters: a study on the
frequency and duration of lens contamination and commonly
utilized techniques to maintain clear vision. Surg Laparosc
Endosc Percutan Tech 26:286–289. https://doi.org/10.1097/
SLE.0000000000000289
12. Stolzenburg JU, Franz T, Kallidonis P, Minh D, Dietel A, Hicks J,
Nicolaus M, Al-Aown A, Liatsikos E (2011) Comparison of the
FreeHand?? Robotic camera holder with human assistants dur -
ing endoscopic extraperitoneal radical prostatectomy. BJU Int
107:970–974. https://doi.org/10.1111/j.1464-410X.2010.09656.x
13. Aiono S, Gilbert JM, Soin B, Finlay PA, Gordan A (2002) Con-
trolled trial of the introduction of a robotic camera assistant
(EndoAssist) for laparoscopic cholecystectomy. Surg Endosc
Other Interv Tech 16:1267–1270. https://doi.org/10.1007/
s00464-001-9174-7
14. Geis WP, Kim HC, Brennan EJ, McAfee PC, Wang Y (1996)
Robotic arm enhancement to accommodate improved efficiency
and decreased resource utilization in complex minimally invasive
surgical procedures. Stud Health Technol Inform 29:471–481
15. Gillen S, Pletzer B, Heiligensetzer A, Wolf P, Kleeff J, Feussner
H, Fürst A (2014) Solo-surgical laparoscopic cholecystectomy
with a joystick-guided camera device: a case-control study. Surg
Endosc Other Interv Tech 28:164–170. https://doi.org/10.1007/
s00464-013-3142-x
16. Shushan A, Mohamed H, Magos AL (1999) How long does lapa-
roscopic surgery really take? Lessons learned from 1000 operative
laparoscopies. Hum Reprod 14:39–43. https://doi.org/10.1093/
humrep/14.1.39
17. Minutolo V, Licciardello A, Arena M, Nicosia A, Stefano BDI
(2014) Laparoscopic cholecystectomy in the treatment of acute
cholecystitis: comparison of outcomes and costs between early
and delayed cholecystectomy. Eur Rev Med Pharmacol Sci
18:40–46
18. Morino M, Pellegrino L, Giaccone C, Garrone C, Rebecchi F
(2006) Randomized clinical trial of robot-assisted versus laparo-
scopic Nissen fundoplication. Br J Surg 93:553–558. https://doi.
org/10.1002/bjs.5325
19. Keus F, de Jong J, Gooszen HG, Laarhoven CJ (2006) Laparo-
scopic versus open cholecystectomy for patients with sympto-
matic cholecystolithiasis. Cochrane Database Syst Rev. https://
doi.org/10.1002/14651858.CD006231
20. Memon MA, Subramanya MS, Hossain MB, Yunus RM, Khan
S, Memon B (2015) Laparoscopic anterior versus posterior fun-
doplication for gastro-esophageal reflux disease: a meta-analysis
and systematic review. World J Surg 39:981–996. https://doi.
org/10.1007/s00268-014-2889-0
21. Proske JM, Dagher I, Franco D (2004) Comparative study of
human and robotic camera control in laparoscopic biliary and
colon surgery. J Laparoendosc Adv Surg Tech A 14:345–348.
https://doi.org/10.1089/lap.2004.14.345
22. Kalteis M, Pistrich R, Schimetta W, Polz W (2007) Laparoscopic
cholecystectomy as solo surgery with the aid of a robotic camera
holder: a case-control study. Surg Laparosc Endosc Percutan Tech
17:277–282. https://doi.org/10.1097/SLE.0b013e31806030ae
23. Kavoussi LR, Moore RG, Adams JB, Partin AW (1995) Urologists
at work comparison of robotic versus human laparoscopic camera
control. J Urol 154:2134–2136