Validation of the GE Carescape V100 and Omron HEM-9210T by the AAMI/ESH/ISO 81060-2:2018 in multi-ethnic South-East Asian adult population
preprint
OA: closed
Abstract
ABSTRACT Objective To ascertain the accuracy of 2 upper-arm blood pressure monitors, GE Carescape V100 Vital Signs and Omron HEM-9210T, used in a large, South-East Asian epidemiological cohort by the AAMI/ESH/ISO 81060-2:2018. Methods 149 participants were recruited from the ongoing Health for Life in Singapore (HELIOS) Study in multi-ethnic Singapore. Results 110 datasets were analysed. For criterion 1, the mean ± SD of differences for GE Carescape V100 was -3.5 ± 7.4/ -3.5 ± 7.4 mmHg (systolic/diastolic), and -4.2 ± 6.9/ -4.2 ± 6.9 mmHg (systolic/diastolic) for Omron HEM-9210T. For criterion 2, GE Carescape V100was -3.5 ± 5.6 / -3.5 ± 5.6 mmHg (systolic/diastolic), and -4.2 ± 4.5 / - 4.2 ± 4.6 mmHg (systolic/diastolic) for Omron HEM-9210T. Conclusion Both GE Carescape V100 Vital Signs and Omron HEM-9210T monitor passed the AAMI/ESH/ISO 81060-2:2018 validation standard.