Results
In initial instrumental optimization,
the IDLs of the 69 halogenated analytes (PCBs, BDEs, OCPs, and PCDD/Fs)
were compared on two different GC column lengths (15 and 30 m; 0.25
mm × 0.25 μm DB-5MS, Agilent) using 1 μL injections
of standard mixtures between 0.0025 and 50 ng/mL ( Figure S1 , original data in Table S6 ); phthalates and PAHs were not considered for this aim because of
the low-level background contamination for certain analytes in these
classes, making IDLs more difficult to determine on either column.
For approximately half of the analytes ( n = 35),
the 30 m column resulted in better sensitivity ( Figure S1 , green), while for approximately one-third of the
analytes ( n = 22), the column length had no significant
effect on detection limits ( Figure S1 ,
gray), and only 10 target analytes had better sensitivity on the shorter
15 m column ( Figure S1 , red); two analytes
(pentachlorobenzene and PCB-3) had very low IDLs (i.e., ≪2.5
fg) that could not be adequately compared in the concentration ranges
tested. Those analytes with lower detection limits on the 15 m column
were mostly late-eluting (RI > 2570) substances with higher boiling
points, including one OCP, five BDEs, and four PCDD/Fs. The highly
brominated BDEs are known to degrade on longer or thicker GC columns
and thus perform better on shorter columns. 47 Nevertheless, for this multiclass chemical exposomics method, we
decided to use the 30 m column because of the 2–5 fold increased
sensitivity for a majority (51.5%) of the analytes examined and considering
that this may lead to greater sensitivity for nontarget molecular
discovery over most of the retention range.
For further
method development, commercial human serum was spiked with 103 priority
target analytes ( Table S4 ) representing
the wide chemical space of hydrophobic environmental contaminants
targeted by GC-based methodologies in national 48 or international biomonitoring programs. 49 These analytes belonged to 6 chemical classes (PCBs, BDEs,
OCPs, PCDD/Fs, PAHs, and phthalates), ranged in molecular weight from
128 (i.e., naphthalene) to 637 (i.e., BDE-154) Da, and had log K ow ranging from 1.7 (i.e., dimethyl phthalate)
to 11.2 (i.e., dechlorane 603). 50
For the traditional GC-MS sample preparation of blood plasma extracts,
it is a common strategy for major lipid interferences to be removed
by destructive techniques under acidic conditions, such as by addition
of concentrated sulfuric acid 51 , 52 or using chromatographic
cleanup on acidified silica gel. 53 For
the suite of multiclass target analytes here, we briefly tested their
recoveries in acidified silica gel columns by loading standards in
isohexane (10–13.5 ng/mL). However, this step was relatively
laborious and resulted in very low recoveries for a majority of PAHs
and phthalates (<5%, Figure S2 ), likely
due to degradation or hydrolysis under acidic conditions. 54 , 55 To avoid these issues, this step was abandoned, and we focused on
solvent extraction conditions that would minimize lipid coextractives.
Acidic conditions were also avoided in further development; thus,
prior to solvent extraction, we denatured and precipitated plasma
proteins by addition of acetonitrile, rather than by formic acid. 29 Plasma protein precipitation by acetonitrile
is a common technique in metabolomics protocols and is also compatible
with some LC-HRMS-based chemical exposomics methods, 31 thereby opening the future possibility to split deproteinized
supernatants for dual analysis by LC- and GC-based exposomics.
Following the protein precipitation and centrifugation step, further
method development focused on minimizing lipid coextractives from
the A-P supernatant. We first tested the effect of adding 40 mg of
dispersive solid-phase extraction material (Bond Elut EMR-Lipid, preconditioned
with 250 μL of water) on relative recoveries into isohexane
extracts (detailed in the SI). Although EMR-Lipid reduced major sterol
lipids in the chromatograms ( Figure S3a , RT = 24.5 min), 25 late-eluting analytes (i.e., RT > 20 min)
were
also lost (average 20% absolute loss, range 10–28%). Moreover,
the precision was much lower using EMR, and few improvements to the
relative recoveries were evident ( Figure S3b ). Subsequent tests focused solely on solvent types and their ratios
under liquid–liquid extraction conditions, including pure isohexane
and mixtures of isohexane with more polar solvents (isohexane: toluene
(9:1); isohexane: ethyl acetate (2:1); isohexane: chloroform (9:1))
( Figure S4 ). However, the addition of polar
solvents always resulted in decreased recoveries, in particular with
ethyl acetate, and especially for the more polar early-eluting analytes,
likely because the tested polar solvents partition significantly into
the acetonitrile phase. Moreover, the addition of toluene resulted
in three layers and was abandoned. In all cases, the isohexane-containing
extract layer (on the top) also contained more interference peaks
of lipids or fatty acids when polar solvents were included.
Therefore, it was concluded that 100% isohexane (H) was the optimum
extraction solvent for the A-P supernatant; in subsequent sections,
we refer to this as the HA-P method. It is germane for us to note
that a recent study investigating various extraction conditions for
plasma exposomics reported the best quantitative performance for a
hexane-acetonitrile-plasma extraction condition, but the authors chose
alternate methods due to the assumption that the hexane layer would
contain lipid interferences; 56 as discussed
later, the hexane layer in the HA-P method was in fact free from most
lipid interferences.
In final HA-P method optimization, we compared
target analyte recoveries
with different volumes of isohexane for the primary extraction (600
μL, 1.2, and 2 mL). Although the larger volume (2 mL) slightly
improved recovery for some analytes (average 3%), precision was lower
(i.e., see higher standard deviations in Figure S5a ). Therefore, 1.2 mL was determined to be the optimum volume
of isohexane. Moreover, we tested the effect of performing a second
follow-up extraction with the addition of isohexane. The addition
of this secondary isohexane extraction (400 μL) after the primary
extraction (0.6 mL) increased the absolute recoveries on average by
8.5% (range 1.1–14.4%, Figure S5b ), particularly for later-eluting analytes. Thus, this method was
included in the optimized HA-P method for validation tests.
In order to increase the sensitivity
for trace levels of contaminants, an LVI method was applied and optimized.
The initial parameters (temperature gradient ramp, split flow rates,
etc.) were first optimized with pure standards ( Table S4 ), and the optimal injection volume was selected based
on performance with a commercial serum extract, prepared by the optimized
HA-P method and spiked at 10 ng/mL. Peak areas increased with larger
injection volumes between 1 and 40 μL ( Figure S6 ), but at 30 and 40 μL, the peak areas did not further
increase linearly, and some analytes had greater standard deviations
and split peaks, suggesting overloading at these injection volumes.
Thus, 25 μL injections were chosen as optimal for method validation
tests. Under these conditions, method precision and robustness were
tested over 4 days by 60 continuous injections of the same spiked
sample extract (2 ng/mL). Although peak areas for many target analytes
decreased slowly over the entire sequence (mean of 28%), this minor
effect was adequately controlled by internal standard correction.
The median RSD was 5.1% for target analytes over 4 days (range 1.6–24.5%, Figure S7 ).
Using the optimized HA-P method with
LVI (25 μL), we performed method validation for all target analytes,
including determination of matrix effects, internal standard corrected
recoveries, calibration linearity, and MLOQs ( Table S7 ). The internal standard for correcting each native
analyte was selected based on similar retention, absolute recovery,
and matrix effect. The median MLOQ was 0.088 ng/mL (range 0.005–4.83
ng/mL; Figure 1 a),
and for most analytes (99 out of 103), calibration curve linearity
was good ( R 2 > 0.99) between the MLOQ
and 5 ng/mL. Two phthalates (diethyl phthalate and dibutyl phthalate)
had relatively lower R 2 values (0.98 each),
and diisobutyl phthalate and di(2-ethylhexyl) phthalate were semiquantified
by single-point calibration because of background interference at
lower contamination (detailed in Table S7 ). When all analytes were spiked to native serum at 1 ng/mL, among
the 91 detectable analytes (out of 103), the mean matrix effect was
null (i.e., mean 100% response; range 76–140%) and the median
internal standard recovery was 104% (range 22–132%; Figure 1 b). For 22 analytes
with interfering background levels in method blanks or commercial
serum, spiking experiments at 10 or 75 ng/mL showed a mean recovery
of 96% and a median matrix effect of 110% ( Table S7 ). The lowest recoveries (22–41%) and worst matrix
effect (513%) were found for three phthalates, specifically dimethyl
phthalate, diethyl phthalate, and diethoxyethyl phthalate, likely
owing to their lower hydrophobicity (log K ow = 1.6, 2.5, and 2.1) 57 and their presumed
preferential partitioning to the plasma-acetonitrile layer during
extraction. All other spiked target analytes have log K ow > 3; thus, we suggest that the HA-P method is only
appropriate for analytes in this more hydrophobic range. This limitation
of the HA-P method is not a limitation for chemical exposomics in
general, for example, dimethyl phthalate and diethyl phthalate are
more sensitive and quantitative by LC-HRMS-based chemical exposomics, 58 and it is understood that GC-HRMS will need
to be applied together with LC-HRMS for comprehensive analysis of
the exposome. 9 Finally, it is noteworthy
that the relatively low recovery of β-HCH and relatively high
recovery of α-HCH ( Figure 1 b) are likely due to interconversion of these isomers
in the presence of water (i.e., plasma) 59 and therefore not necessarily a limitation of the analytical method.
Method
validation results for multiclass target analytes, arranged
by classes, spiked to 200 μL of human serum and analyzed by
the HA-P method with LVI, showing (a) MLOQ ( n = 4)
and (b) internal standard corrected analyte recoveries ( n = 3, 1 ng/mL spiking levels). Panel (a): MLOQ of BDE-154 was >10
ng/mL and is not plotted. Panel (b): 13 analyte recoveries were calculated
at 10 or 75 ng/mL (detailed in Table S7 ).
The performance of the HA-P method
with LVI was also examined by
analysis of NIST SRM 1958 (i.e., a freeze-dried fortified human serum),
which has certified or noncertified reference values, for 33 of the
analytes targeted here. For most of these ( n = 23),
quantified concentrations by the HA-P method were within 75–125%
of certified/noncertified values (i.e., ratio 0.75–1.25, Figure S8 ), and the mean ratio for all analytes
was 0.98 (range 0.32–1.43). Notable outliers were again HCH
isomers (i.e., low concentrations of β-HCH and high concentrations
of γ-HCH), which have noncertified values in the SRM and are
known to interconvert as described above. The authors of the literature
method reported that the ratios of 20 target analytes were within
75–125% of certified/noncertified values and that the mean
ratio for all analytes was 0.9 (range 0.3–1.79). 29 Overall, the current chemical exposomics method
performs similarly as well as the literature chemical exposomics method
for target analysis based on analysis of the same SRM (see Figure S8 for comparison). 29
During
method development, it was noted that the total ion chromatograms
of HA-P method extracts were relatively clean (visually) and free
from large interfering peaks in GC-HRMS analysis ( Figure S3a ). This suggested that few lipid species had been
coextracted and may explain why dispersive solid-phase extraction
by EMR-Lipid had no beneficial effect in our method development tests.
The major lipids in human plasma include glycerolipids, glycerophospholipids,
and sterol lipids such as cholesterol esters. 26 Nonpolar bulk lipids (i.e., glycerolipids and some sterol esters)
generally have very low solubility in polar solvents such as acetonitrile
and are known to be removed with proteins during protein precipitation. 60 Most phospholipids remain in the plasma-acetonitrile
phase during extraction, and here, we found no traces of phospholipids
in the isohexane extract ( Figure S9 , cyan).
To fully understand the relative extent of lipid interferences,
and the impact of these on chemical exposomics, we compared plasma
extracts from the HA-P method (in 100 μL of isohexane) to extracts
of the same plasma by a literature method (in 200 μL of ethyl
acetate). 29 The relatively clear appearance
of extracts from the HA-P method was evident, relative to yellow-colored
extracts by the literature method ( Figure 2 ). Consistent with visual appearances, after
injecting 2 μL of each extract ( Figure 2 a,b), the corresponding total ion chromatograms
revealed a comparably complex matrix for the literature method, with
many abundant coextracted substances. The relatively clean total ion
chromatogram of the HA-P extract is noteworthy considering that twice
as much plasma equivalents were injected on-column. The major chromatographic
peaks for the literature method extract included long-chain fatty
acids (RT = 11.19, 13.77, 16.46 min) and sterol lipids (RT = 16.06,
20.96–24.61 min, Figure S10b ), which
were either absent or much lower in the HA-P method, even with 10-fold
more plasma equivalents injected on-column ( Figure S10a ).
Total ion chromatograms and photos of extracts from pooled
Swedish
plasma (200 μL, n = 3 each) prepared by the
HA-P method (cyan, left panels) and a literature method (brown, right
panels) injected with various volumes to GC-HRMS. Chromatograms for
the HA-P method extract are shown for a 2 μL injection (a) and
a 25 μL injection (c), corresponding to 4 and 50 μL of
plasma equivalents on-column, respectively, and these can be contrasted
with chromatograms for the literature method extract for a 2 μL
injection (b) and a 5 μL injection (d), corresponding to 2 and
5 μL of plasma equivalents on-column. Photos of the associated
solvent extracts are also shown for (a) HA-P method in 100 μL
of isohexane and (b) from the literature method in 200 μL of
ethyl acetate, with 100 μL taken for photography.
For the extracts from the literature method, the MS response
was
saturated for many of the largest peaks, as maximum peak heights were
in the range of 2 × 10 10 and did not increase with
increasing injection volumes from 2 to 5 μL. The largest peaks
in TICs of both extracts were for sterol lipids (RT = 24.5 min); thus,
both methods are still similarly prone to interference from these
major plasma metabolites. When the injection volume of the plasma
extract was increased from the literature method ( Figure 2 d), the major chromatographic
peaks became visibly broader, suggesting that the GC column was overloaded,
and larger injection volumes were not attempted. In comparison, a
larger injection volume of the HA-P extract (25 μL, corresponding
to 50 μL plasma equivalents on-column, Figure 2 c) was nevertheless applied, and the total
ion chromatogram was still relatively clean but did show elevated
baseline at later retention times (>25 min). This increased background
overlaps with the retention range of 18 (17.5%) of the target analytes;
thus, its potential as a minor interference cannot be discounted.
As shown in extracted ion chromatograms of example target analytes
in the HA-P method ( Figure 3 a,b), the increased injection volume resulted in larger target
analyte peaks and also revealed new detectable analyte peaks that
were not evident at lower injection volumes (e.g., trans-nonachlor, m / z 408.7840 at 16.04 min, and PCB-105, m / z 325.8810 at 18.25 min), demonstrating
enhanced method sensitivity by LVI for low-abundance analytes in plasma.
Using 200 μL plasma aliquots of pooled Swedish plasma, 16 target
PCBs/OCPs (0.02–3.3 ng/mL) were consistently detected in triplicate
( n = 3/3) samples by the HA-P method (25 μL
injection), compared to only 5 target analytes (0.17–2.8 ng/mL)
by the literature sample preparation method (2 μL injection)
and the same instrumental settings ( Figure 3 e,f). Among the 11 PCBs/OCPs not consistently
detected by the literature method, most had plasma concentrations
below 0.2 ng/mL, according to the HA-P method, except for PCB-180
(0.45 ng/mL), which was inconsistently detected (2 of 3 replicates)
by the literature method, despite relatively higher concentrations.
Example
extracted ion chromatograms (EICs, a–d), as well
as concentrations and detection frequencies (e, f) of analytes in
triplicate extracts of pooled Swedish plasma (200 μL aliquots)
prepared by the HA-P method and literature method and injected to
GC-HRMS with various injection volumes. Panels (a) and (b) present
the HA-P method and injection volumes of 2 and 25 μL. Panels
(c) and (d) present the literature method and injection volumes of
2 and 5 μL, respectively. [Top row (panels a–d) is the
EIC of ion m / z 221.9998 for PCB
13 (Level 4 annotation, RT = 11.39 min), PCB-15 (RT = 11.89 min),
and other dechlorinated PCBs. Middle row (panels a–d) is the
EIC of m / z 408.7840, corresponding
to trans-nonachlor (RT = 16.03 min). Bottom row (panels a–d)
is the EIC of ion m / z 325.8810 for
PCB-118 (RT = 17.54 min) and PCB-105 (RT = 18.25 min).] Panel (e)
shows general agreement of quantified targeted analyte concentrations
by the two methods above 0.2 ng/mL, but below this approximate threshold,
the literature method resulted in nondetection for most analytes;
mean concentrations and detection frequencies are listed in panel
(f).
All target analytes detected and
quantified in SRM 1958 ( n = 33 analytes) had similar
accuracies by the HA-P method
and literature method, including those analytes present at moderate
to high concentrations (i.e., 293–1250 ng/kg, certified values)
and at lower concentrations, such as for PCB-123 and PCB-114 (0.0525,
0.0466 ng/kg, noncertified values) ( Figure S8 ). When analyzing the pooled Swedish plasma in triplicate by the
HA-P and literature methods, the detection and quantification were
similar by both methods when concentrations were above 0.2 ng/mL ( Figure 3 e,f). The major discrepancies
were at lower concentrations, where most analytes were nondetectable
by the literature method, thereby demonstrating enhanced sensitivity
by the HA-P extraction method and LVI.
An additional observation
in the analysis of the extracts produced
by the literature method was that, when increasing injection volumes
from 2 to 5 μL, some analyte peaks disappeared (e.g., PCB-13
at 11.39 min, m / z 221.9998) ( Figure 3 c,d). This may be
due to abundant interferences ( Figure 4 a), and the auto gain control function of the Orbitrap
mass spectrometer, which applies a dynamic ion injection time (i.e.,
to the C-trap and subsequently to the Orbitrap analyzer) throughout
the analytical run to balance sensitivity and mass spectral resolving
power. 35 , 36 For a very clean injection, as shown for
analysis of instrumental blanks composed only of isohexane ( Figure 4 b, black), the ion
injection time was initially maximal (i.e., 112 ms), thereby allowing
maximum signal to the Orbitrap analyzer, but declined to approximately
20 ms after 22 min due to increasing background signal from column-bleed
at higher temperatures (300 °C at 22 min); overall, the mean
ion injection time throughout the analysis of isohexane instrumental
blanks was high (mean = 73 ms; 97 ms before 25 min). In comparison,
lower ion injection times were observed with LVIs of the HA-P extract
(mean: 38.6 ms, 56 ms before 25 min, Figure 4 b, cyan). Nevertheless, these ion injection
times were still substantially higher than that for the extract produced
by the literature method (mean: 3.22 ms, 3.85 before 25 min, Figure 4 b, brown), even considering
a 10× more plasma-equivalent volume injected on-column from the
HA-P extract ( Figure 4 b; 50 μL plasma equivalents by HA-P (25 μL injection)
and 5 μL plasma equivalents by the literature method (5 μL
injection)). Lower ion injection times predictably corresponded to
regions of the chromatograms with a higher total ion signal ( Figure 4 a). For analytes
that are still detectable, a correction factor is applied by the software
to maintain quantitative analysis, 61 but
for trace analytes near detection limits, the signal can become nondetectable
due to the lower ion injection time, as shown for PCB-13 in the extract
from the literature method ( Figure 3 d).
Comparison of (a) total ion chromatograms and (b) ion
injection
times for extracts of pooled Swedish plasma prepared by the HA-P method
(cyan, 50 μL plasma equivalents on-column) and by a literature
method (brown, 5 μL plasma equivalents on-column), both injected
to the same GC-HRMS. In both plots, the instrumental blank is also
shown (black, 25 μL isohexane) for comparison.
The HA-P method with LVI was applied to 32 individual plasma samples
of Swedish adults as well as to pooled Swedish plasma for reference
and quality assurance. Among all samples, 51 (out of 103) target analytes
were detected in at least one individual ( Table S8 ). The detected analytes included 7 dioxin-like PCBs (#105,
#114, #118, #123, #156, #157, and #167), 14 nondioxin-like PCBs (#1,
#3, #4, #19, #15, #28, #52, #37, #101, #138, #153, #202, #180, and
#205), 9 PAHs, 12 OCPs, 1 BDE, and 8 phthalates. Among these, 28 analytes
had detection frequencies above 20%, the distributions of which are
shown by sex ( Figure 5 ), and 9 analytes (β-HCH, DEHP, PCB-156, pyrene, p,p’-DDT,
DiBP, PCB-167, HCB, and acenaphthene) showed gender differences ( p < 0.05), but they were not significant anymore after
the Bonferroni correction. Contrasting the current results to previous
target analyses of the same plasma samples (separate aliquots) 6 showed linear associations between the two methods
( Figure S11 ).
Violin plots showing
concentrations and distributions for detected
target analytes in 32 individual Swedish adult plasma samples by sex
(female in purple, male in green). Panel (a) shows 28 analytes with
detection frequencies >20%, while panel (b) shows 23 analytes detected
at lower frequencies (i.e., in 1–6 samples). Black “+”
symbols indicate the target analyte MLOQ. For data visualization,
detectable signals below the MLOQ are plotted as MLOQ/2, and nondetectable
signals are plotted as MLOQ/4. For analytes marked with black “*”
symbols, the MLOQs were calculated by methods other than the native
standard in the matrix-matched calibration curve, detailed in Table S7 . Values of DBP were extrapolated from
the calibration curves, and DiBP and DEHP were semiquantified using
one point.
Correlations between concentration
and sampling year were tested
by the Spearman method as most data were not normally distributed.
Twelve target analytes showed statistically significant temporal associations
between 1990 and 2013 (test method and statistics in Table S8 ), six of which are shown in Figure 6 . Particularly, PCBs and OCPs were negatively
associated with sampling year, consistent with bans and restrictions
that started in the 1970s. 62 , 63 To the contrary, DEHP,
a commonly used phthalate plasticizer, showed a positive association
with sampling year, from below 10 ng/L in the early 1990s to >40
ng/L
by 2005, and possibly lower concentrations thereafter. We acknowledge
that no field blanks were available in this cohort to rule out phthalate
contamination from medical sampling equipment, but similar or higher
levels of DEHP have been reported in other studies. For example, mean
DEHP was 180 ng/mL (maximum 1030 ng/mL) in healthy women in the 2000s 64 and 65 ng/mL in the serum of women diagnosed
with endometriosis in the 2010s. 65
Concentrations
of example target analytes, with statistically significant
negative associations with sampling year; the Spearman correlation
coefficient (ρ) and statistical significance ( p ) are shown in each plot.
We found no significant associations between the concentrations
of these target analytes and individual metadata such as birth year,
sampling age, meat consumption (Spearman correlation test), or smoking
status (two-tailed Wilcoxon mean value test). Nevertheless, concentrations
of several PCBs (#15, #156, and #180), as well as β-HCH and
HCB, were significantly higher in snuff users than nonsnuff users
(means: 0.008 vs 0.002 ( p < 0.01), 0.07 vs 0.05
( p = 0.02), 0.37 vs 0.05 ( p = 0.02),
0.06 vs 0.05 ( p = 0.03), and 0.15 vs 0.11 ng/mL ( p = 0.05), respectively, for PCB-15, 156, 180, β-HCH,
and HCB), whereas DEHP was lower in snuff users (mean: 8.5 vs 21.4
ng/mL; p < 0.01) by the two-tailed Wilcoxon mean
value test. However, snuff users were not equally distributed over
sampling years (with fewer snuff users in later years); thus, these
results are likely confounded by the associated temporal trends.
We
additionally evaluated the suitability of the HA-P method with LVI
for the discovery of unexpected substances in a nontarget exposomics
workflow. After data processing by MS-DIAL and blank filtration, a
total of 875 molecular features were detectable among all individual
plasma samples (see Table S9 , including
relative responses). Among these features, 112 were matched to reference
library spectra, corresponding to a relatively high annotation rate
of 12.8%. The annotations included 30 of the target analytes (confirmed
Level 1, 42 ΔRT < 1%), as well
as 82 new annotations (Level 2 confidence, 42 ΔRI < 50) that included 28 prospective environmental substances.
Authentic standards for 8 of these environmental chemicals were purchased,
resulting in 7 confirmed identifications (Level 1, ΔRT <
1%, chromatograms and spectra in Figures S12–18 ). These included the related analytes 2,4-di- tert -butylphenol ( Figure S12 ) and tris(2,4-di- tert -butylphenyl) phosphite ( Figure S18 ), which have been used as antioxidants and UV stabilizers
in rubber and plastics; the co-occurrence of the two was reported
in indoor dust in 2018. 66 Although 2,4-di- tert -butylphenol, which is a degradation product of the
latter 67 and other substances, has been
detected in human blood and urine previously, 68 , 69 we are not aware that tris(2,4-di- tert -butylphenyl)
phosphite has been reported previously in human biomonitoring. This
latter substance was previously reported in chemical migration tests
from polymeric materials to water and simulated foods. 66 , 67 , 70 , 71 This analyte has a high boiling point and very late elution time
in our method (26.14 min) and in some methods may suffer from high
background interferences. The discovery of this substance and, in
general, the high nontarget annotation and confirmation rates are
likely due to a combination of compounding factors, including higher
sensitivity from LVI and lower matrix interference in the HA-P extracts.
The low matrix interference by the HA-P method may not only result
in higher sensitivity (i.e., due to higher ion injection times) but
could also improve the performance of the in silico spectral deconvolution (i.e., in MS-DIAL), resulting in higher quality
spectra that will have better matches to spectral libraries.
In this study, a comprehensive multiclass target and nontarget chemical
exposomics method was developed and validated for human plasma. With
the combination of lipid removal at the extraction stage and LVI,
the method sensitivity was significantly improved in comparison to
a literature GC-HRMS method without these steps. These developments
resulted in more target analytes being consistently detected and a
higher rate of molecular annotation and confirmation in nontarget
analysis. An important reason for the enhanced method performance
was the longer ion injection times achieved in the Orbitrap instrument
throughout the GC run time owing to a much lower signal from coextracted
lipids. The method was rugged and can be applied to larger studies
of the chemical exposome in combination with LC-HRMS-based analysis.
Materials
A representative
list of 103 target analytes from 6 chemical classes was selected based
on higher detection frequencies in major historical biomonitoring
initiatives (e.g., HBM4 EU, US NHANES), and also considering a broad
range of physiochemical properties for use in method development,
and their native and isotopic labeled standards were acquired from
commercial suppliers ( Table S1 ). These
included 33 polychlorinated biphenyls (PCBs), 8 polybrominated diphenyl
ethers (BDEs), 10 polychlorinated dibenzodioxins/dibenzofurans (PCDD/Fs),
16 polycyclic aromatic hydrocarbons (PAHs), 24 organochlorine pesticides
(OCPs), and 12 phthalates. Toluene, formic acid, n -hexane, and silica gel (high-purity grade, pore size 60 Å,
230–400 mesh particle size) were from Merck (Germany). Isohexane,
ethyl acetate, acetonitrile, dichloromethane, chloroform, and water
were from Fisher Scientific (USA); isohexane was used instead of n- hexane because it has similar physiochemical properties
but lower toxicity. 37 , 38 MgSO 4 and a dispersive
solid-phase extraction powder, Bond Elut EMR-Lipid, were purchased
from Agilent (USA).
For method
development and validation,
sterile-filtered human serum was obtained commercially (Merck, Germany,
human male AB plasma-derived, origin USA). Standard reference material
(SRM 1958, fortified and lyophilized human serum) was purchased from
NIST (USA). The optimized exposomics workflow was applied to 32 individual
human plasma samples (100 μL) from the Västerbotten Intervention
Programme (VIP) cohort, 39 a cardiovascular
study cohort launched in 1985 in north Sweden. The samples (16 men
and 16 women), which were collected between 1991 and 2013 ( Table S2 ) and stored at −80 °C, were
randomly selected among participants whose samples (separate aliquots)
had previously been analyzed for persistent organic pollutants. 6 Participant smoking and moisture snuff consumption
as well as dietary intake of meat and fish were self-reported. As
described previously, 31 a Swedish pooled
plasma reference sample was also prepared in-house from residual heparinized
plasma of 953 adults (male and female) from the VIP cohort. Approval
from the Swedish Ethical Review Authority [Dnr 2020-03301] was granted
for work with these plasma samples.
All glassware
was newly furnaced to minimize background contamination, and sample
preparation was in a positive-pressure clean laboratory with high-efficiency
particulate air filtration. Plastic containers and materials (e.g.,
disposable pipet tips) were avoided during sample preparation and
analysis to minimize contamination of the samples and associated extracts.
Method development included various tests of lipid removal techniques,
including liquid–liquid and solid-phase extraction (see the Supporting Information , SI). The optimized method
for human plasma and serum can be used for a range of small volumes
typically available in cohort studies (100–200 μL). The
plasma/serum was added to a conical borosilicate-glass centrifuge
tube (10 mL, Pyrex, Corning, USA), and each sample was then spiked
with 8 μL of a 100 ng/mL surrogate internal standard mixture
containing 26 isotopically labeled chemicals ( Table S1 ). For method development, small volumes of the internal
standard solution were spiked onto each sample using precise Hamilton
syringes (10 μL). For application to cohort plasma samples,
the internal standard solution was indirectly added to larger volumes
of acetonitrile and later added for protein precipitation using glass
tip pipettes (200–2000 μL, Socorex). Procedural blanks
composed of LC water, substituted for plasma, were prepared with all
experiments and batches. For protein precipitation, a volume of acetonitrile
was added that was 4 × sample volume (e.g., 400 μL of acetonitrile
added to a 100 μL sample or 800 μL of acetonitrile added
to a 200 μL sample), vortexed at 1400 rpm for 1 min (multitube
vortex mixer, Ohaus, USA) and then centrifuged (5804R, Eppendorf)
at 4400 g for 5 min at room temperature. The supernatant
was transferred to a new borosilicate-glass tube, and 1.2 mL of isohexane
was added, followed by vortexing at 1400 rpm (1 min) and centrifugation
at 3000 g (1 min). The upper isohexane layer was transferred
to a new borosilicate-glass tube; another 400 μL of isohexane
was added to the extraction test tube, and the extraction step was
repeated. The two isohexane layers were combined and evaporated to
100 μL by gentle nitrogen flow at room temperature (TurboVap
LV, Biotage, Sweden), followed by adding 10 μL of 20 ng/mL volumetric
internal standard (methoxychlor-D14) by a Hamilton syringe (700 series,
100 μL), vortexing, and transferring to an amber glass
vial with an insert (0.3 mL, Thermo Scientific) for GC-HRMS analysis.
All 32 individual samples (VIP cohort) were extracted in one daily
batch along with 3 procedural blanks and 3 Swedish pooled plasma reference
samples.
For comparison, plasma samples were also prepared by
a leading literature chemical exposomics method designed for GC-Orbitrap
HRMS. 29 Briefly, 8 μL of 100 ng/mL
surrogate internal standard was added to 200 μL of Swedish pooled
plasma ( n = 3) in conical glass tubes, followed by
adding 50 μL of formic acid and 200 μL of n -hexane: ethyl acetate (v:v = 2:1). This was vortexed for 1 h on
ice and then centrifuged at 4400 g at 4 °C (10
min). The organic supernatant was transferred to a new tube with 25
mg of MgSO 4 , vortexed, and centrifuged at 4400 g (10 min). The final supernatant was transferred to an injection
vial with 6 μL of 100 ng/mL volumetric internal standard (methoxychlor-D14)
for instrumental analysis.
Plasma extracts
were analyzed by GC
(TRACE 1300 Series, Thermo Fisher Scientific, US) interfaced to an
Orbitrap HRMS (Q-Exactive, Thermo Fisher Scientific, US) operating
in full scan (34–750 m / z )
in electron ionization mode. Nominal resolution was set to 60,000,
with AGC Target at 1 × 10 6 , and Maximum IT set to
auto. The Orbitrap was calibrated with the internal calibration gas
and evaluated before every sequence, typically every 1–3 days.
The ion source and transfer line temperature were 300 °C. The
optimized method utilized a DB-5MS column (30 m × 0.25 mm ×
0.25 μm, Agilent) and temperature gradient program starting
at 30 °C for 1 min, increasing to 50 °C at 20 °C/min,
then ramping to 170 °C at 25 °C/min, to 250 °C at 6
°C/min, and then increasing to 315 °C/min at 25 °C/min
with a 12 min hold. The carrier gas was helium at a constant flow
of 1.3 mL/min. Injections of 25 μL of extract were made to a
programmable temperature vaporizer (PTV) injector with a baffle liner
in large-volume mode ( Table S3a ). The temperature
programs on the injection port and GC oven were coordinated. The starting
temperature was held at 30 °C to evaporate the injection solvent
(isohexane) while retaining the semivolatile analytes. During method
development, a shorter DB-5MS column (15 m) was compared, based on
instrumental detection limits (IDLs) of the target analytes, and 1
μL of each solution was injected into each system in PTV splitless
mode, with injection temperature increasing from 30 to 315 °C
at 7.3 °C/s, and oven and MS conditions were the same as in the
optimized method. The IDL was defined as the lowest concentration
in an 11-point calibration curve (range 0.0025–50 ng/mL, triplicate)
with the signal-to-noise ratio >3, and at least 3 data points across
the peak. For LVI, an optimal injection volume was decided by comparing
responses for various injection conditions: 1 and 5 μL injections
with a small syringe (10 μL) and 5, 10, 25, 30, and 40 μL
injections with a larger syringe (100 μL); parameters for autosampler
are shown in Table S3b . For investigation
of lipid removal effects, the extracts prepared by the literature
method were analyzed by the same instrumental method as above, but
only with 2 and 5 μL injection volumes.
For application
to cohort samples, calibration curves of target analytes were run
3 times during the injection sequence (beginning, middle, and end).
Kovats retention index (RI) was applied, consisting of n -alkane standards (C7–40 mixture, Merck, 25 ng/mL in isohexane)
injected separately at the beginning and end of the sequence. The
entire sequence, including samples, calibration solutions, procedural
and instrumental blanks, and Swedish pooled reference samples, lasted
approximately 71 h (35 min gradient plus 15 min re-equilibration per
injection), while all samples were maintained at 10 °C in the
autosampler. The instrumental blanks consisted of a clean solvent
and were run multiple times in the sequence to monitor for carryover.
Method robustness was examined by
60 continuous injections of the same spiked (2 ng/mL) serum extract
(split into 12 vials), with an additional 12 injections of isohexane
spread throughout the sequence, with one solvent injection after every
5 sample injections. The whole sequence lasted 60 h, with no interruptions
to the sequence. Target analyte recoveries, precision, and matrix
effects were evaluated in triplicate spiked samples at 0.1, 1, and
10 ng/mL for all analytes in 200 μL of the commercial serum.
An additional spike recovery experiment was also conducted at 75 ng/mL
( n = 3) for those analytes with higher background
concentrations. During method development, absolute recoveries (without
internal standard correction) were calculated from the peak areas
of native analytes spiked to serum before sample preparation compared
to the same spike added to processed serum extracts immediately before
GC-HRMS analysis. For method validation, final recoveries of spiked
native analytes were calculated by the relative response to internal
standards spiked before extraction. Method precision was evaluated
by relative standard deviations of the triplicate recovery experiments
at 1 ng/mL (if not available because of the native serum background,
a value at a higher concentration was taken). For evaluating matrix
effects, peak areas of spiked native analytes were compared between
(i) extracts of serum and (ii) extracts of blank water. For analytes
not present in method blanks, the method limit of quantification (MLOQ)
was defined as the lowest concentration spiked to commercial serum
(8 points between 0.005 and 10 ng/mL, n = 4), in
which the peak area of the analytes had relative standard deviations
below 20% with no internal standard correction but corrected for the
volumetric internal standard. In the cases where background levels
of the analytes in commercial serum interfered, the response of the
corresponding isotopic labeled standard (spiked at 4 points between
0.005 and 5 ng/mL, n = 4) was used to calculate the
MLOQ when possible; otherwise, solvent-based calibration curves were
used (for 3 analytes). In the cases where the analyte was present
in method blanks, the MLOQ was alternatively calculated from the average
blank response plus 10 standard deviations. The linearity was evaluated
by the R 2 of standard solvent calibration
curves (MLOQ–5 ng/mL, n = 3). Method accuracy
was examined in two ways, including by comparison of calculated target
analyte concentrations to a previous target analysis of the same VIP
plasma samples (different aliquots) 6 and
by comparison of target analyte concentrations in NIST SRM 1958 with
certified reference values.
External
calibration curves
(7 points, 0.005–5 ng/mL, n = 3) were used
to quantify the 103 target analytes. TraceFinder (v.5.0, Thermo Scientific)
was used for peak detection, identification, and quantification of
the target analytes. Isotopically labeled surrogate internal standards
( n = 26) were used to correct the recoveries and
variations during sample preparation ( Table S4 ). Target analytes were generally absent in procedural blanks; however,
peaks were detectable for certain phthalates and PAHs, and in this
case, analytes were only considered detected if sample peak areas
were 3 times higher than the corresponding procedural blank; concentrations
were also blank-subtracted in these cases.
For nontarget analysis,
MS-DIAL 40 (v.4.9.221218, parameters in Table S5 ) was used to align features and deconvolute
corresponding electron ionization spectra. Each nontarget feature
had a corresponding retention time (RT), RI, deconvoluted spectrum,
and peak area. Mass spectral library matching was performed from the
combination of an in-house GC-HRMS Orbitrap library (244 chemicals),
MassBankEU, 41 and NIST20 (version 2.4,
purchased from Thermo Fisher Scientific, USA). Annotation confidence
levels were applied and defined according to the annotation scoring
framework for GC-HRMS, 42 with RI matching
scores considered.
Analyzed data were further processed and
visualized in Excel (Microsoft
Office 2019), Python (v.3.7.3), 43 Jupyter
Notebook (v.5.7.8), 44 and R (v.4.3.2) 45 and RStudio (v.2023.12.1 + 402) 46 with ggplot2 (v.3.4.4).
Statistical tests were performed
in R (v.4.4.1) using the rstatix
package (v.0.7.2). Normality was evaluated by the Shapiro–Wilk
test, equal variance was evaluated by the F-test (Data Analysis Tool,
Excel), and group differences were evaluated by two-tailed t tests, with p -values adjusted for multiple
testing by the Bonferroni method, if the data were normally distributed
and had equal variance. Otherwise, statistical differences were evaluated
by the two-tailed Wilcoxon–Mann–Whitney test, with the p -value adjusted by the Bonferroni method. For tests of
correlation, the normality of the data was first tested by the Shapiro–Wilk
method. Spearman's correlations were tested in R (stats package
v.4.4.1).
Introduction
The concept of the
exposome was introduced in 2005 to encourage
more research on the environmental determinants of disease. 1 , 2 Environmental chemicals have long been known to be human disease
risk factors, 3 and relevant exposures today
include ambient air pollution and mixtures of contaminants ingested
from food, water, and dust. 4 − 7 Analytical methods are now evolving to measure the
“chemical exposome”, broadly defined as all environmental
chemical exposures throughout the life course. 8 − 10 With this ambitious
scope of chemical exposomics come several practical methodological
challenges, including to comprehensively quantify a broad range of
priority target analytes, to discover and identify novel exposures
by nontarget workflows, to scale up sample throughput, and to achieve
high sensitivity in the small volumes of biofluids available in typical
cohort studies. 11
Traditional analytical
methods for human biomonitoring are mostly
targeted and mass spectrometry-based, employing either liquid chromatography
(LC) or gas chromatography (GC). 12 − 19 Such methods are very sensitive and quantitative but can require
large sample volumes and laborious sample processing steps and ultimately
reveal only a limited fraction of the chemical exposome, defined a priori by the investigators. Today, there are over 350,000
chemicals in global commerce, 20 but only
5% of these have ever been analyzed in environmental media, 21 and likely fewer have been biomonitored in humans.
In this context, modern instrumental advances in LC- and GC-high-resolution
mass spectrometry (HRMS) provide great potential for chemical exposomics,
as several commercial instruments now combine high mass spectral resolving
power, high mass accuracy, high scanning frequency, good sensitivity,
and a wide dynamic range in full scan mode. 10 , 22 , 23 These instruments are therefore well suited
to perform parallel target and nontarget data acquisition, but methodological
challenges remain for sample preparation. One specific challenge is
to quantitatively extract diverse analytes from complex biological
samples while minimizing interferences. In human blood, known organic
environmental contaminants vary greatly in their hydrophobicity, spanning
17 orders of magnitude in octanol–water partition coefficient
( K ow ), and their concentrations range
over 11 orders of magnitude (i.e., 160 fM–140 mM) 24 , 25 in a matrix dominated by major lipid classes and complex mixtures
of endogenous metabolites 26 that may overshadow
small signals from environmental chemicals. Multiclass target exposomes
are now being reported, as well as methods for combined target and
nontarget analysis, but the sample preparation methods are often adapted
directly from metabolomics, 27 − 29 and not specifically designed
or optimized for chemical exposomics, which strives to profile small
molecules present at 1000× lower concentrations than endogenous
substances. 25
Most reported chemical
exposomics methods for blood have so far
been LC-based and thus focused on the polar environmental chemical
fraction. 22 , 30 , 31 However, the
commercial availability of GC-Orbitrap HRMS instruments 32 has enabled recent development of methods for
low-polarity and semivolatile analytes. 29 , 33 , 34 The associated sample preparation methods for blood
serum and plasma are simple, rapid, and potentially scalable, including
liquid–liquid extraction 29 and liquid–liquid
extraction with dispersive powders 34 (i.e.,
QuEChERS). Nevertheless, the performance of existing methods to minimize
major blood lipid coextractives has not been directly evaluated, and
there is concern that abundant lipids may negatively influence method
robustness, lower method sensitivity, and interfere with molecular
discovery. Of particular relevance is that the resolving power of
Orbitrap HRMS analyzers is adversely impacted by the abundant matrix
signal, such that the auto gain control function dynamically lowers
the ion injection time to minimize space-charging when an abundant
signal is detected. 35 , 36 Therefore, we hypothesized that
by minimizing lipid coextractives, trace analytes would be more easily
detected by GC-Orbitrap HRMS; moreover, larger volumes could be injected
to improve method sensitivity.
Building on existing GC-based
chemical exposomics methods for human
blood, here, we explore methods to achieve improved sensitivity for
plasma chemical exposomics by a GC-Orbitrap HRMS workflow. Specifically,
we aimed to minimize coextracted plasma lipids at the extraction step
to achieve dual benefits to method performance from (i) minimizing
matrix interference and (ii) enabling large-volume injections (LVIs)
of plasma extracts. A simple and potentially scalable protocol, which
uses isohexane (H) to liquid–liquid extract acetonitrile-plasma
(A-P), was developed and validated for small samples of human plasma
(100–200 μL) while optimizing for lipid removal and method
sensitivity for 103 priority target analytes. The validated method,
termed the HA-P method, was applied to a subset of adult plasma samples
( n = 32) in a combined target and nontarget analysis,
and method performance was examined with respect to target analytes
and molecular discoveries.