Comparison of Two Novel Swept-Source Optical Coherence Tomography Devices to A Partial Coherence Interferometry-Based Biometer

preprint OA: closed CC-BY-4.0
📄 Open PDF View at publisher

Abstract

Purpose: To evaluate the repeatability and agreement of cornea and biometry measurements obtained with two swept-source optical coherence tomography (SSOCT) and a partial coherence interferometry-based device. METHODS This is a cross-sectional study. Forty-eight eyes of 48 patients had three consecutive measurements for ANTERION (Heidelberg Engineering, Germany), CASIAII (Tomey, Japan) and IOLMaster500 (Carl Zeiss Meditec, USA) on the same visit. Mean keratometry (Km), central cornea thickness (CCT), anterior chamber depth (ACD) and axial length (AL) were recorded. Corneal astigmatic measurements were converted into vector components – J0 and J45. Intra-device repeatability and agreements of measurements amongst the devices were evaluated using repeatability coefficients (RCs) and Bland-Altman plots, respectively. RESULTS All devices demonstrated comparable repeatability for Km (p≥0.138). ANTERION had the lowest RC for J0 amongst the devices (p≤0.039). Systematic difference was found for the Km and J0 obtained with IOLMaster500 compared to either SSOCTs (p≤0.010). The ACD and AL measured by IOLMaster500 showed a higher RC compared with either SSOCTs (p<0.002). Systematic difference was found in CCT and ACD between the two SSOCTs (p<0.001), and in AL between ANTERION and IOLMaster500 (p<0.001), with a mean difference of 1.6 µm, 0.022mm and 0.021 mm, respectively. CONCLUSIONS Both SSOCTs demonstrated smaller test-retest variability for measuring ACD and AL compared with IOLMaster500. There were significant disagreement in keratometry and AL measurements between the SSOCTs and PCI-based device; their measurements should not be considered as interchangeable but the disagreement could be refined by readjustment of intraocular lens constant in clinical practice.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00
unpaywall
last seen: 2026-06-06T02:00:05.402940+00:00
License: CC-BY-4.0