Characteristics and Incidence of Patients With Tricuspid Valve Prolapse After Micra Transcatheter Pacing System Implantation

preprint OA: closed
📄 Open PDF View at publisher

Abstract

Background: The characteristics and incidence of patients with tricuspid valve (TV) prolapse after leadless pacemaker implantation are unknown. Methods and Results: We retrospectively identified 35 of 85 patients with sufficient echocardiographic TV imaging before and after Micra transcatheter pacing system (Micra TPS) implantation. The post-procedure incidence of TV prolapse was 8.6%, and the cause of prolapse was chordae tendineae rupture. Patients with TV prolapse had significantly longer procedure times and more deployments than patients without TV prolapse. Conclusions: : TV prolapse after Micra TPS implantation is not a rare complication and is accompanied by frequent deployments and prolonged procedure times.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00
unpaywall
last seen: 2026-06-05T02:00:03.366016+00:00