A Simple Way to Minimize Error on Humidity Measure with a Psychrometer

preprint OA: closed CC-BY-4.0
🔓 Open OA copy View at publisher

Abstract

This work demonstrates that simple calibration procedures can reduce the error in de-termining relative humidity with a psychrometer by up to 75 times. The psychrometer uses two temperature transducers to measure relative humidity. The calibration consists of comparing both transducers before assembling the psychrometer. A psychrometer assembly is presented, built from readily available parts, which provides temperature data for comparison with a theoretical model. The results are compared against a refer-ence psychrometer to assess errors with and without calibration. A twenty-two-fold de-crease in error was observed for a relative humidity of 62%.