Description and Validation of Flow-Through Chambers of Respirometry for Measuring Gas Exchange in Animal Trials

preprint OA: closed CC-BY-4.0
🔓 Open OA copy View at publisher

Abstract

Indirect calorimetry (IC) is a widely used method to study animal energy metabolism by measuring gas exchange. The accuracy of IC depends on detecting variations in signals reflecting the metabolic response, which can be challenging due to measurement noise and external factors. In this study, we propose a methodology to validate IC systems, including an easy-to-use spreadsheet for data computing, to verify accuracy and detect whole-system leaks. We conducted a re-covery test using a simulation of CO2 dynamics in MS Excel and injecting a known CO2 concentration into four respirometry chambers. The clearance rate of CO2 was observed and compared to the expected clearance rate from the simulation. The results from the recovery test in our system show that the proposed methodology is accurate and precise. The proposed methodology and recovery test can be used to standardize the validation of IC systems, providing accurate measurements of animal energy metabolism in different environmental conditions and energy utilization from feeds.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00
unpaywall
last seen: 2026-05-30T02:00:01.510937+00:00
License: CC-BY-4.0