Abstract
Antimicrobial resistance remains a global existential threat. Given that antimicrobial therapy
commonly starts before pathogen identification, rapid and scalable methods capable of determining
effective antimicrobial compounds are needed. In this paper , we demonstrate a 2 × 2 array of
parallelised microscope s that uses low numerical aperture (NA=0.25) detection optics and LED
excitation to determine bacterial viability based on their fluorescence response to an electrical
stimulus. Following a 2-hour incubation, the fluorescent viability readout requires less than one
minute. We use K-means clustering to classify pixels in a time lapse sequence of widefield fluorescence
images and extract changes seen within bacterial clusters. We demonstrate sufficient sensitivity to
measure fluorescence changes after electrical stimulation in a bacterial monolayer. To capture these
subtle fluorescence changes at high signal -to-background ratios, we place a limit on the minimum
optical density of the bacteria l sample . This novel approach is scalable to 96 -well formats using a
suitable consumable electrode array.
Introduction
Antimicrobial resistance (AMR) remains a major health threat to the global population . The annual
cost of AMR in Europe alone is currently estimated at €1.5 bn with AMR predicted to cause 10 million
deaths annually by 2050 1. Developing a rapid and scalable AMR test would reduce this burden by
ensuring patients receive the appropriate antibiotic at the right concentration sooner.
The current clinical gold standard for determining antibiotic resistance in a patient sample is to culture
the bacterial isolate with a panel of antibiotics across multiple concentrations for 24 to 48 hours. The
efficacy of an antibiotic and its minimum inhibitory concentration (MIC) are established by monitoring
bacterial growth (turbidity) in in each well. The limitation of this approach is that, in acute cases, results
are required within much shorter time frames (a few hours) for effective treatment. Th e long
incubation period also limits throughput and contributes to a backlog of testing.
In recent years, there has been a rapid expansion of antimicrobial susceptibility testing (AST) methods
available (see Reszetnik, et al. [2] for a review). Recent advances include direct-from-specimen testing,
single-cell imaging, miniaturized growth chambers, microfluidics, and novel detection modalities
including plasmonic and nucleic acid -based methods that shorten turnaround times by accelerating
growth detection or viability assessment under antimicrobial exposure. Several commercial platforms
now provide phenotypic AST directly from positive blood cultures or urine specimens significantly
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
2
faster than traditional methods (such as the electric field detection method employed by iFast
Diagnostics3). Emerging non-commercial approaches show promise for further reductions in time-to-
result, including methods that combine rapid imaging or metabolic sensing with deep learning for
enhanced signal interpretation. Challenges remain in extending these technolog ies to diverse
specimen types, standardising inoculum effects, and achieving broad clinical validation.
One recent technique developed by the Munehiro lab4 measures bacterial viability via electrical
stimulation of the bacteria under investigation, whilst monitoring changes in fluorescence intensity,
with positive and negative changes indicating viable and non-viable bacteria, respectively . After
incubation in a medium containing fluorescent dye thioflavin T (ThT), electrical stimulation
hyperpolarises the membrane of healthy cells by opening voltage -sensitive ion channels. Viable
bacteria rapidly accumulate surrounding (positively charged) ThT through these ion channels, leading
to increased fluorescence brightness. In contrast, non-viable bacteria with disrupted membranes and
a weak ened potassium ion gradient will undergo depolarisation and expel any intracellular ThT
through electrical stimulation. This translates into either no change or a reduction in fluorescence
brightness.
This technique has since been refined and commercialised 5. Whilst significant developments have
been made to improve sensitivity and repeatability , there is clear potential (as with many of the
techniques summarised in [2]) to greatly increase sample throughput . This could be achieved by
creating arrays of stimulation electrodes, each with their own bacterial sample. However, in studies
performing fluorescence imaging of multiple fields of view, 75% of the total acquisition time was spent
moving the sample 6, with similar estimates for fast axial scanning systems 7. Throughput can be
dramatically increased by eliminating the need to move samples by combining electrical stimulation
with parallelised microscopy.
Parallelised microscopy, in which an array of samples can be imaged individually without the need to
mechanically translate the sample, has gained popularity in recent years. Unlike macroscopic imaging
of the entire sample array, where each sample occupies a small fraction of the whole image, in parallel
microscopy, each sample is imaged by the entire detector area, greatly increasing the achievable
resolution from ~100 µm to less than 10 µm. Versions of parallelised microscopes include schemes
that have a separate excitation and detection path for each sample, using arrays of light sources,
lenses, and cameras to capture image data from each sample simultaneously8,9. This approach has the
advantage of true simultaneity for up to 24 samples , but the machine vision cameras used lack the
sensitivity required to detect the fluorescent emission from small bacterial microcolonies, which may
be less than 10 µm across. In addition, the large data rates associated with collecting data from 24
cameras at once (45 Gbps compared to 10 Gbps for USB 3.1 Gen 2.0) require custom data transfer
methods, limiting scalability and impact . An alternative method is to use a galvanometer mirror to
sweep a macroscopic image of all samples across a single detector to increase sampling at the desired
locations, but this low N.A. (i.e. 0.04) imaging method lacks spatial resolution10.
“Random Access Parallel” (RAP) microscopy 11 uses an array of LED light sources to illuminate each
sample individually, but uses a common detection path and therefore requires only one camera. This
sequential acquisition approach has recently been applied to the measurement of cardiomyocyte
monolayer contraction (implemented as the “Exeter Multiscope”)12. The utility of using one detection
camera becomes particularly apparent in fluorescence imaging, as the prospect of having an array of
low-noise, high quantum efficiency fluorescence cameras quickly becomes impractical in terms of
mechanical integration and cost. We have developed a proof-of-concept parallelised microscope with
the ability to distinguish between viable and non-viable bacteria using a custom-built UV-fluorescence
Multiscope that can read out a 2 × 2 array of bacterial samples, testing up to four different conditions,
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
3
in rapid succession, without any moving parts in less than 5 minutes . K-means clustering is used to
extract fluorescence intensity changes associated with bacterial clusters from time lapse data . We
demonstrate the fluorescence detection sensitivity required to scale the readout to larger numbers of
bacterial samples, including standard 96-well plate formats.
Results
Baseline results from an inverted microscope
To provide a baseline measurement of bacterial fluorescence changes in response to electrical
stimulation, we first performed the viable/non-viable assay on an inverted microscope using a multi-
electrode array composed of gold electrodes on a coverglass-bottom dish (Figure 1A-B). The electrodes
were stimulated individually (100 Hz, 4V peak -peak, 2.5 seconds) using an external control box
(‘CytePulse’, Cytecom Ltd.) connected to the electrode array via an IDE ribbon cable. Two of the six
electrodes were used to collect reference data (electrode #5 and #6, Figure 1B).
To measure bacterial viability, Bacillus subtilis was cultured to log-phase and an optical density (OD) of
1.5. 1 μL of this culture was pipetted onto one side of 5 × 5 mm agarose pad s containing ThT. The
inoculated side of each agarose pad was then placed in direct contact with the individual electrodes in
the array, ensuring close contact. The pads were then incubated at room temperature for 2 hours to
allow bacterial proliferation.
After loading the electrode array onto the microscope stage, a widefield fluorescence image of each
sample (electrodes #5 and #6) was taken to confirm the presence of bacteria colonies (Figure 1C).
Fluorescence time lapse images were acquired for each pad in sequence using a 10× 0.4 N.A. objective
lens and a UV LED source (400 nm, p300 ultra, CoolLED Ltd.). Each time lapse acquired 15 seconds of
images (1 image every second, 500ms exposure) before stimulation to provide a fluorescence
Reference
value and then for a further 30 seconds after stimulation to monitor the change (Figure 1F).
The first bacterial assay used proliferating, healthy B. subtilis cells. As t he process of electrical
stimulation is detrimental to the health of the cell membrane, the best measure of the non-viable cell
assay was to repeat the experiment on the same site whilst monitoring changes in cell fluorescence
after a second stimulation. This repeatable approach avoids complications associated with antibiotic-
based killing, such as incomplete loss of viability, while enabling interrogation of the same cell
population on the same pad.
To map out the temporal variations in fluorescence intensity over the field of view, a K-means
clustering approach was taken. The temporal variation in fluorescence intensity was extracted for each
individual pixel in the 256 × 256 pixel region of interest (i.e. over 65,000 fluorescence traces). The K-
means clustering algorithm then identified five fluorescence traces (or ‘clusters’) that best
characterised the majority of traces present (in a way not dissimilar to principal components analysis).
The clusters were ranked according to the magnitude of fluorescence change, with the largest changes
being the ‘top’ cluster (=5) and the smallest changes being the ‘bottom’ cluster (=1). The images in
Figure 1D and Figure 1E have been colour coded according to the cluster to which each pixel has been
assigned. A total of fi ve clusters were chosen to ensure that there were enough to describe the key
features anticipated in each field of view: bacteria in the focal plane, bacteria just beyond the focal
plane, agarose pad, electrodes, plus an addition al cluster for any unanticipated features. Specifying
substantially more than five clusters would result in redundant classifications with highly similar cluster
characteristics. The clusters were not constrained to be equally
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
4
populated, as the bacterial clusters occupy a relatively small fraction of the total pixel number.
After clustering, a heat map was produced describing the cluster assignment for each pixel (Figure 1D-
E). The cluster that was most strongly associated with the in-focus bacteria (‘top cluster’) was used as
the best measure of the change in flu orescence intensity over time (Figure 1F). The absolute
fluorescence change was obtained from this cluster by calculating the peak fluorescence difference
Figure 1: (A) Overview of the method used to electrically stimulate bacterial samples in an inverted microscope. (B)
Plan view of an array of six electrode positions. Only positions #5 and #6 were used. (C) Widefield fluorescence
image of a bacterial colony, imaged through the electrodes (black vertical bars) on an inverted microscope. One
region of interest (ROI) was selected from within this larger field of view for processing. (D) Cluster assignment maps
for pixels within each ROI. The top cluster (most strongly correlated to bacterial clusters) was used to calculate the
fluorescence change after stimulation. (E) Detail of cluster assignment map indicates the degree of heterogeneity
within each colony of B. subtilis. (F) Changes in fluorescence observed after first stimulation (viable bacteria average
response, black line) , and second stimulation (non -viable bacteria average response, red dashed line). Error bars
indicate standard deviation (N=2). Scale bars: (C) 500 µm (D) 100 µm (E) 50 µm.
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
5
between stimulation and end points. The peak change in fluorescence for electrode positions #5 and
#6 are recorded in Table 1. Table 1 also records the mean pixel value of the traces for the top and
bottom cluster from which the signal to background ratio is calculated.
Table 1: Summary of fluorescence changes detected using the inverted microscope. Bacteria at each electrode position were
stimulated twice before moving to the next electrode.
Electrode
position
Stimulation
number
Peak
fluorescence
change (Δf/f)
Top trace
avg. (a.u.)
Bottom
trace avg.
(a.u.)
Signal to
Background
5 1 +4.9% 622.3 4.4 141.4
5 2 -12.0% 585.4 4.3 136.1
6 1 +3.2% 449.8 3.8 118.4
6 2 -6.0% 331.6 4.0 82.9
Detection of bacterial viability using the Multiscope
The RAP/Multiscope approach to parallelised microscopy is outlined in Figure 2. The samples are
arranged on a 9 mm pitch (as in a 96 well plate) and are indicated by the electrode positions in Figure
1B. Each sample is illuminated by collimated light from a UV LED. Baffles reduce cross-talk between
each of the illumination paths, ensuring each LED only illuminates one sample. An objective lens below
each sample collects emitted light to form an image at infinity. The final element in this system has a
diameter large enough to cover all of the samples in the array and acts to both steer and focus the
light from each sample to a common exit pupil, which is in the same plane as the camera. T his final
element could be a transmissive lens, but for the purposes of maintaining a compact setup, a reflective
parabolic mirror was used. To maintain the camera at the focus of the parabolic mirror, whilst not
obscuring the light from the sample, the ax is of the mirror was offset from the centre of the array. A
photograph of the prototype system used in this project is shown in Figure 2(B).
A parallelised fluorescence microscope was built following the design of the Exeter Multiscope12, which
in turn is based on the original RAP prototype11. Previous designs had been bright field only, detecting
transmitted light with a machine vision camera. The key difference in this work was that we were
detecting UV/VIS fluorescence and required UV excitation, and a scientific CMOS camera to achieve
the required fluorescence sensitivity. Due to the in -line nature of the fluorescence detection, there
was a need for strong filtering to block directly transmitted light, requiring the use of OD6 excitation
(390 ± 20 nm) and emission (494 ± 10 nm) filters. The objective lenses had a diameter of 9 mm (limited
by the sample pitch) and a focal length of 18 mm, providing an NA of 0.25 which was not sufficient to
resolve individual bacteria, but bacterial microcolonies could be clearly discerned (Figure 2E).
The electrode array was loaded onto the sample tray and time lapse fluorescence images were
acquired of B. subtilis samples at electrodes #1, #2, #5 and #6 (electrode positions #3 and #4 were not
used as they were not centred on the 9 mm pitch of the objective lens array, Figure 1B). As previously,
two sets of time lapse data were acquired for each electrode position, corresponding to two separate
electrical stimulations to compare viable/non-viable responses. The Multiscope provided all -optical
switching between the four fields of view by turning on the LED associated with each electrode
position ( Figure 2A). Only one field of view was imaged at any one time. An example widefield
fluorescence image of electrode #2 is shown in Figure 2 C. Within each 2048 x 2048 raw image , a
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
6
smaller 256 x 256 pixel ROI was selected to reduce processing times for the K-means clustering
analysis. ROIs were located where the bacterial colonies appeared in sharpest focus. To reduce the
high level of background fluorescence, these ROI images were first processed using a rolling ball
Figure 2: (A) Overview of the method used by the Multiscope to parallelise imaging of multiple samples . Each sample is
illuminated one at a time and the images are relayed to the same sCMOS camera. ExF = excitation filter. EmF = emission filter.
(B) A photograph of the Multiscope system constructed to image bacterial fluorescence (see Methods for full description of
individual components (i) – (viii)). (C) Widefield fluorescence images of the bacteria sample mounted on agarose. Regions of
interest which appeared in best focus were selected for processing. (D) Processed ROIs after K-means clustering. (E) Magnified
image of ( D). (F) Averaged fluorescence profiles for the top cluster in each K-means analysis. Profiles corresponding to first
stimulation (black line) and second stimulation (red dashed line). Scale bars: (C) 500 µm (D) 100 µm (E) 50 µm.
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
7
subtraction (50 pixel diameter) in FIJI 13. The K-means clustering algorithm was then applied to the
processed ROI data and each pixel assigned to a cluster (Figure 2D-E).
The time sequence of each pixel value in the ROI was then analysed using K-means clustering as for
the inverted microscope data . The ‘top’ cluster (i.e. that most closely associated with the bacterial
microcolony pixels) was then extracted. The peak difference in fluorescence intensity after stimulation
was determined and the results are summarised in Table 2. The data set for electrode #6 was
overwritten by the second stimulation and could not be analysed. To provide an indication of the signal
to background ratio, Table 2Table 2 also includes the mean value of each of the top cluster traces
(‘signal’) and the bottom cluster traces (‘background’) where the bottom cluster corresponds to the
pixels obscured by the opaque electrodes. The absolute change in fluorescence brightness appears to
be approximately correlated with the signal to background ratio.
Table 2: Summary of fluorescence changes detected using the Multiscope at a bacteria density of
OD=2.5. Bacteria at each electrode position were stimulated twice before moving to the next
electrode.
Electrode
position
Stimulation
number
Peak
fluorescence
change (Δf/f)
Top trace
avg. (a.u.)
Bottom
trace av g.
(a.u.)
Signal to
Background
1 1 4.0% 61.1 7.9 7.7
1 2 -6.7% 60.3 7.9 7.6
2 1 13.9% 74.0 8.7 8.5
2 2 -10.7% 74.5 8.7 8.8
5 1 9.4% 126.4 9.2 13.7
5 2 -10.6% 120.3 8.9 13.5
6 2 -4.7% 99.5 9.3 10.6
Multiscope data at lower cell densities
To further explore the sensitivity limits of the UV fluorescence Multiscope, we repeated the
experiments at a lower density preparation of B. subtills (OD=1.5) for agarose pad inoculation .
Reducing the bacterial density used for agarose pad inoculation offers additional clinical benefit by
enabling earlier identification of effective antimicrobial compounds. Experiments were repeated as for
the OD=2.5 samples. Multiscope widefield fluorescence images from each of the electrode positions
are shown in Figure 3A, together with the selected ROI (red square). The processed data indicate that
in some cases, such as electrode #1 (Figure 3B-C), the bacteria were very sparse.
The top cluster profiles are shown in Figure 3D. None of the traces indicate a significant change in
fluorescence brightness. As shown in Table 3, the changes are all small with slightly negative changes
in fluorescence. It is possible that, even though the signal to background levels are similar to that at
high cell densities, there are simply not enough cells to register a significant change in fluorescence.
Doubling the number of clusters from five to ten did not significantly affect the results. The more
defocused images of the sample led to poorer rolling ball background subtraction leaving higher
baseline levels of fluorescence for some electrode positions.
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
8
Table 3: Measurements of peak fluorescence changes for each of the Multiscope experiments for a
low bacteria density (OD=1.5).
Discussion
Combining electrical stimulation with Multiscope readout enables the clear determination of viable
from non-viable bacteria (Figure 2F). Differences in the amplitude of the fluorescence change (Figure
3E) are linked to both (i) contrast in the sample image and (ii) the strength of the fluorescence signal.
Image contrast is maintained by keeping the sample in focus and minimising aberrations. Variations in
sample focusing is thought to be largely due to the mounting of the electrode array in the Multiscope.
Tension in the IDC cable produced a small uplift of electrode positions #2 and #5 relative to #1 and #6
(Figure 2B). Focusing the objective array on electrodes #2 and #5 led to reduced contrast in sample
images at positions #1 and #6, as reflected in Table 2.
Electrode
position
Stimulation
number
Peak
fluorescence
change (Δf/f)
Top trace avg.
(a.u.)
Bottom trace
avg. (a.u.)
Signal to
Background
2 1 -9.0% 507.4 42.8 11.8
2 2 -3.5% 65.3 6.5 10.0
5 1 -2.1% 67.2 6.3 10.7
5 2 -4.2% 64.2 6.4 10.0
6 1 -2.4% 48.9 5.2 9.4
6 2 -3.5% 48.9 5.2 9.4
Figure 3: (A) Widefield fluorescence images of the bacteria samples mounted on agarose imaged through the opaque
CytePulse electrodes. Regions of interest (red squares) were selected for processing. Example shown is a live cell response
after the first stimulation. (B) Processed ROIs after K-means clustering. (C) Magnified view of (B). (D) Change in the
fluorescence brightness for the top cluster averaged over three electrode positions after the first ( black line) and second
(red dashed line) stimulation. (E) Change in fluorescence brightness after first (S1) and second (S2) stimulation for the three
conditions considered above (Multi=Multiscope, Invert=Inverted microscope).
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
9
Aberrations from the parabolic reflector will also contribute to the reduction in image contrast.
Comatic aberrations increase as (𝐵 𝑓⁄ )2 where 𝐵 is the lateral distance of the sample from the optic
axis of the mirror and 𝑓 is the focal length of the mirror. This can be improved greatly by increasing the
focal length of the mirror from the current value of 100 mm to around 300 mm and thereby reducing
the comatic aberrations by a factor of nine. A longer parabolic mirror focal length would also increase
the standoff distance of the sCMOS camera, reducing the number of sample positions that might be
obscured by the camera itself.
The strength of the fluorescence signal has been shown to be dependent on both how well the sample
was focused, and the abundance of bacterial colonies present in the sample, with ODs of 1.5
generating insufficient fluorescence signal. This could lead to the top cluster expanding to include more
dim pixels from defocused bacterial microcolonies , thereby “averaging out” the strength of the
fluorescence response. This is indicated by the number of top cluster pixels remaining approximately
the same for the OD=1.5 samples (Figure 3B), despite the approximately two-fold reduction in bacterial
abundance.
Future improvements in system throughput could be enabled by reducing the current 2.9 second
integration time required to accumulate sufficient fluorescence signal from the bacterial monolayer.
These long integration times undermine the benefits of all-optical switching between samples. Using
UV sources that are brighter than the current 35 mW LEDs is one option. As only 1 mW of excitation
light currently reaches the sample , there is significant potential to increase the optical coupling
between source and sample by removing the plastic bulb covering the LED emitter and using higher
NA collection lenses with higher transmission UV coatings. Transitioning from an in-line to an epi -
fluorescent imaging architecture would also greatly increase contrast in the acquired images. The
increase in signal to background ratio would also open up the possibility of obtaining reliable measures
of viability at lower bacterial concentrations, even enabling imaging direct from biofluid, bringing the
technology in line with competing AST approaches2.
If the camera integration time can be reduced from 2.9 seconds to around 500 ms (as used on the
inverted microscope), then it would easily allow all four samples to be imaged in a single 45 second
exposure by optically moving from one sample to the next between camera exposures. The raw image
sequence could then be de-interlaced into four time-lapse sequences. If the integration time could be
reduced further to a more typical 50 ms exposure , this would allow over 50 wells to be imaged
simultaneously, including additional time budget for streaming images to disk. This is a significant
target as it would allow for bacterial viability measurements in the presence of seven concentrations
(plus controls) of the six main antibiotics administered clinically in acute infections (i.e. Piperacillin–
tazobactam, Ceftriaxone, Gentamicin, Metronidazole, Vancomycin, Meropenem).
Finally, it can be seen in Figure 1C that the bacterial clusters were rarely distributed evenly across the
agarose pad but would instead accumulate in a ‘coffee ring’ pattern left after the inoculation droplet
had dried on the agarose pad. One potential modification to the protocol would be inoculating the
agarose pad via spraying, as demonstrated in [14] to achieve a more uniform distribution across the
pad.
Conclusion
With the rapid emergence of AMR there is a growing and urgent need for rapid diagnostic methods
capable of identifying antibiotic efficacy. Established ASTs are culture-based and inherently slow: even
after a pure bacterial isolate has been obtained, standard AST methods typically require 24-48 hours
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
10
of incubation to generate results. In contrast, bacterial optical electrophysiology can provide viability-
dependent fluorescence readouts in under one minute , offering a fundamentally faster route for
assessing antimicrobial effectiveness. To date, all previous implementations of the fluorescent
measure of bacterial viability using electrical stimulation have used a standalone research grade
microscope and require a high degree of user intervention to obtain bacterial viability data. In this
work, we have demonstrated the viability of using an automated Multiscope approach to parallelise
optical excitation and readout of fluorescence changes in bacterial monolayers. This provides an
important stepping stone towards high throughput screening of bacterial populations using rapid
optical electrophysiology methods. Whilst the initial test used a relatively modest number of testing
sites (a 2 x 2 array) there is clear potential for scaling the number of tests to 50 or more wells using the
same architecture, providing measures of antibiotic efficacy in one to two minutes. By matching rapid
testing with high throughput, the goal of reducing bottlenecks in clinical AST can be achieved.
Methods
Multiscope construction
The prototype system in shown in Figure 2B. The light path starts with a 4 × 4 array of LEDs (RS
Components #903-3771) which output up to 35 mW in the range 390-425 nm. As the light was emitted
over a 30 degree cone angle, baffles were used to avoid light spilling over into adjacent well positions
and creating cross-talk. The baffles were 3D printed as a 4 × 4 array of opaque tubes. These tubes were
6 mm in diameter and had a length which corresponded to the 9 mm focal length of the collimation
lenses. The baffle was mounted onto the underside of the custom LED array and together occupied
the first tray position (Figure 2B(i)). The second tray position was occupied by a 3 × 3 array of 9 mm
diameter, 9 mm focal length collimation lenses (Edmund Optics, #65-549-INK). The short focal length
was chosen to maximise the light transmitted onto the bacterial sample. By moving the lens array
fractionally further away from the LED array, it was possible to focus the excitation light onto only the
region of the bacterial sample that lies within the field of view of the objective lenses, further
increasing the light intensity at the sample. A 50 mm diameter, OD6 excitation filter with a 390 ± 20
nm pass band (Edmund Optics #86-359) was placed after the collimation lenses. The filter was aligned
to the well positions of the 4 × 4 LED array within a 3D printed mount that occupied the third well plate
holder position (Figure 2B(iii)).
The CytePulse electrode array occupied the fourth tray position (Figure 2B(iv)). Directly underneath
this layer was a 2 × 2 array of imaging objective lenses, centred on the samples (Figure 2B(v)). The four
positions corresponded to electrode positions 1, 2, 5 and 6 (Figure 1(B)). These lenses (Edmund Optics
#47-653, 9 mm diameter, 18 mm focal length, NA=0.25) collected the fluorescence emission from the
bacteria and relayed this to the sCMOS camera (Hamamatsu ORCA Flash4.0 V2) via a parabolic mirror
(100 mm focal length, 220 mm diameter, Edmund Optics #68-793). A 1” OD6 emission filter with pass
band 494±10nm (Edmund Optics #84-096) was fixed immediately in front of the fluorescence camera
to further suppress excitation light.
The 0.25 NA objective lenses provided both better light collection, higher spatial resolution and smaller
depth of field to improve the sensitivity to fluorescence changes within the thin bacterial monolayer.
The theoretical resolution of 1.2 µm combined with residual aberrations from the parabolic mirror
meant it was not possible to resolve individual bacteria. The shallow depth of field required a finer
degree of control over the objective lens position. This was provided by a single axis piezo ( Figure
2B(vi), Mad City Labs MMP1) which achieved a minimum step size of 95 nm over a 25 mm range.
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
11
Fluorescence then reflected off the parabolic mirror (2B(vii)) and was collected by the sCMOS camera
(Figure 2B(viii)). The magnification of the system is 100 mm / 18 mm = 5.5 ×. The maximum field of
view of the system is therefore given by the size of the imaging chip (2048 × 6.5 um = 13.3 mm) divided
by the magnification (13.3 mm / 5.5 = 2.4 mm square). To minimise stray light incident upon the
camera, emanating from outside of the sample field of view, a 3D printed plate was used. This 2 mm
thick plate contained an array of 3 mm diameter holes that acted as a field stop when centred on the
CytePulse electrode locations and was aligned with the camera to avoid any clipping of the field of
view.
Data acquisition and hardware control
The Multi-dimensional acquisition toolbox in MicroManager15 was used to trigger the sCMOS camera
and define the integration time. For the inverted microscope (IX73, Olympus), images were acquired
on an sCMOS camera (Zyla 4.2, Andor) every second (500 ms exposure), on a 10× 0.4 N.A. dry objective.
For the acquisition of Multiscope image data, a 2.9 second camera integration time on a 3 second
interval was used for a total of 20 exposures. Electrical stimulation was manually triggered vi a the
CytePulse graphical user interface 15 seconds after initiating image acquisition (i.e. after 15 frames on
the inverted system and 5 frames on the Multiscope). Electrical stimulation of 4V amplitude and 100
Hz frequency for 2.5 seconds was used in all cases. All images in the sequence were saved directly to
disk.
Sample preparation
B. Subtilis 168 streak plates were prepared from cyrostock by streaking onto lysogeny broth (LB) agar
plates and incubating overnight at 37 °C. Stationary-phase cultures were prepared by inoculating fresh
LB with a single colony and incubating for 17 h at 37 °C with shaking. Log-phase cultures were prepared
by diluting overnight cultures into fresh LB and incubating until mid-log phase (OD600 ≈ 1.5 or 2.5). Log-
phase cells were used immediately for preparation of imaging samples.
A defined CytePulse medium (see Supplementary information) was prepared to support bacterial
viability during imaging. MOPS and potassium phosphate buffers were adjusted to pH 7 and
autoclaved, whereas trace-ion and carbon/nitrogen supplements (including amino acids, glucose,
metal salts, and thiamine), and ThT were filter sterilised to avoid thermal degradation. ThT was
included to enable fluorescence -based monitoring. Agarose pads (2% w/v) were prepared by
combining molten agarose with the CyteCom medium.
Agarose pads were cast under aseptic conditions by pipetting molten agarose onto 22 × 22 mm
coverslips and gently placing a second coverslip on top to form a uniform layer. After solidification, the
upper coverslip was carefully removed, and each pad was cut into 16 equal sections with a sterile
scalpel. Log-phase bacteria were spotted onto individual sections with a pipette and allowed to adsorb.
Each agarose section was then inverted onto a CytePulse electrode such that the bacterial layer
contacted the electrode surface while avoiding the central ground electrode. The pads were then
incubated at room temperature for 2 hours. Loaded electrodes were placed on a custom 3D-printed
holders and transferred to either the inverted microscope or Multiscope for electrical stimulation and
time-lapse imaging. This preparation yielded consistent single -layer bacterial samples suitable for
combined CytePulse stimulation and fluorescence imaging.
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
12
References
1. Interagency Coordination Group on Antimicrobial Resistance. No Time to Wait: Securing the
Future from Drug-Resistant Infections. https://www.who.int/docs/default-source/documents/no-
time-to-wait-securing-the-future-from-drug-resistant-infections-en.pdf (2019).
2. Reszetnik, G. et al. Next-generation rapid phenotypic antimicrobial susceptibility testing. Nat.
Commun. 15, 9719 (2024).
3. Spencer, D. C. et al. A fast impedance-based antimicrobial susceptibility test. Nat. Commun. 11,
5328 (2020).
4. Stratford, J. P . et al. Electrically induced bacterial membrane-potential dynamics correspond to
cellular proliferation capacity. Proc. Natl. Acad. Sci. 116, 9552–9557 (2019).
5. Cytecom Limited. https://www.cytecom.co.uk/.
6. Matsumoto, K. et al. Advanced CUBIC tissue clearing for whole-organ cell profiling. Nat. Protoc.
14, 3506–3537 (2019).
7. Gintoli, M. et al. Spinning disk-remote focusing microscopy. Biomed. Opt. Express 11, 2874 (2020).
8. Harfouche, M. et al. Imaging across multiple spatial scales with the multi-camera array
microscope. Optica 10, 471 (2023).
9. Thomson, E. E. et al. Gigapixel imaging with a novel multi-camera array microscope. eLife 11,
e74988 (2022).
10. Symvoulidis, P . et al. NeuBtracker—imaging neurobehavioral dynamics in freely behaving
fish. Nat. Methods 14, 1079–1082 (2017).
11. Ashraf, M. et al. Random access parallel microscopy. eLife 10, e56426 (2021).
12. Mohanan, S. et al. Automated measurement of cardiomyocyte monolayer contraction using
the Exeter Multiscope. Biomed. Opt. Express 16, 4716–4729 (2025).
13. Schneider, C. A., Rasband, W. S. & Eliceiri, K. W. NIH Image to ImageJ: 25 years of image
analysis. Nat. Methods 9, 671–675 (2012).
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
13
14. Cano, Á. et al. Multiparametric quantification of bacterial cells using digital holographic
microscopy. Sci. Rep. 15, 41051 (2025).
15. D. Edelstein, A. et al. Advanced methods of microscope control using μManager software. J.
Biol. Methods 1, 1 (2014).
.CC-BY 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted March 11, 2026. ; https://doi.org/10.64898/2026.03.10.710830doi: bioRxiv preprint
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.