Trend-Based Intermittent Neuromonitoring in Thyroid and Parathyroid Surgery: A Prospective Preliminary Observational Study
preprint
OA: closed
CC-BY-4.0
Abstract
Background: /Objectives: Intraoperative neuromonitoring (IONM) has improved safety in thyroid and parathyroid surgery, yet intermittent IONM (I-IONM) may miss traction injuries developing between stimulations. We evaluated the feasibility and clinical utility of a trend-based intermittent monitoring mode (NIM Vital NerveTrend®) that records closely spaced stimulations and plots amplitude and latency over time. Methods: We conducted a prospective observational study at a high-volume endocrine surgery unit (January–September 2025). Forty-four consecutive patients undergoing thyroidectomy and/or parathyroidectomy with NerveTrend® were enrolled. EMG responses were categorized as Green (amplitude >50% of baseline and latency < 110%), Yellow (amplitude 110%), Red (amplitude 110%), and Loss of Signal (LOS: amplitude < 100 µV). Primary outcomes included LOS prevalence and the association between stimulation frequency and the appearance of Yellow trends. Ethical approval: AVEN protocol 486/2024/OSS/AOUPR; informed consent obtained. Results: Of 71 nerves at risk (NAR), 55 had a valid baseline and were analyzed; LOS occurred in 3/55 NAR (5.5%). The mean number of stimulations per NAR was 4.5 (range 1–9). Cases with both Green and Yellow points had a significantly higher mean number of stimulations than cases with only Green points (5.1 vs. 3.8; Student's t-test p = 0.0059). One Red measurement occurred in a case that progressed to LOS. Conclusions: NerveTrend® provided near real-time functional feedback while maintaining the simplicity of I-IONM. Increased stimulation frequency was associated with early Yellow trend alerts, potentially signaling traction stress and enabling timely surgical adjustments. Larger multicenter studies and protocol standardization are warranted.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2025) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.
Source provenance
- europepmc
- last seen: 2026-05-20T01:45:00.602351+00:00
- unpaywall
- last seen: 2026-05-29T02:00:03.542394+00:00
License: CC-BY-4.0