Abstract
Textbooks of anaesthesia describe in detail the calculations associated with splitting ratios
created in variable bypass vapourisers. These calculations are based on assumptions of carrier
gas flow(eg:assume 150 ml) as practised with measured flow vaporizers[1,2]. Carrier gas flow
in a variable bypass vaporizer is agent specific,dynamic,unknown and operator
independent.Therefore assuming carrier gas flow in calculations related to variable bypass
vaporisers is without rationale[3]. In spite of technological advance,the amount of vapour
produced is calculated based on clinically impractical assumptions. A simple method and
formula is suggested for the quick estimate of vapour output, carrier gas flow and determine the
splitting ratio while operating a sevoflurane variable bypass vaporizer.
Authors
1.D Girijanandan Menon MD
2.Manjit George MD,FRCA
3.Salih V Salim MD
4.R.Sreekumar MSc PhD(Physics)
5.Sam Philip MD
6.Roshin K Mathew MD
7.Princymol Bju MD
8.Merin Helen Mathai MD
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
NOTE: This preprint reports new research that has not been certified by peer review and should not be used to guide clinical practice.
2
Author affiliations
Authors numbered 1, 5,6,7& 8 are affliliated to Malankara Orthodox Syrian Church Medical
college, Kolenchery, Kerala, India
Authors numbered 2 & 3 are affiliated to Al Azhar Medical college, Thodupuzha, Kerala, India
and Broomsfield Hospital, Chelmsford , UK respectively.
Author numbered 4 is affiliated to St.Joseph’s College, Devagiri, Kozhikode, Kerala
Contribution by authors
( author name with abbreviation of name is given in brackets)
Manjit George (MG)
Salih V Salim (SS)
D Girijanandan Menon (DG)
R.Sreekumar (RS)
Sam Philip (SP)
Roshin K Mathew (RM)
Princymol Bju (PB)
Merin Helen Mathai (MM)
DG is involved in concept development, derivation of formula, writing up the manuscript,
editing, concept validation and reliability of the concept as teaching aid in the department.
MG Mainlead in concept validation, manuscript editing, verified thereliability of the concept as
teaching aid in his instituion
SS concept development, concept proof verification, manuscript editing, verified the reliability
of the concept as teaching aid in his instituion
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
3
RS did the derivation styling, illustrations and supplementary file editing
SP involved in concept validation, and verification by department faculty, granted permission for
including the explanation in discussions on vapouriser, ultiisation of concept for post graduate
teaching in the Department of Anaesthesiology
RM,PB&MM are non author contributors for implementation of the concept for postgraduate
teaching and for analyzing the reliability of the concept as a better teaching aid in the
Department of Anaesthesiology.
Splitting ratio in sevoflurane vapouriser-a relook on calculations
Introduction
What is the current method to estimate the amount of vapour produced in vapourising
chamber?
Carrier gas flow is assumed to estimate the amount of vapour produced in vapourising chamber.
What is the new method suggested to calculate the amount of vapour produced in vapourisng
chamber?
Vapour produced in vaporising chamber is simply the product of the percentage of the
concentration control dial setting and the fresh gas flow in ml per minute. Instead of assuming
the carrier gas flow, it is calculated real time from the settings. An attempt is made to simplify
the calculations by logically relating the assumptions to the operator’s act of turning the
vaporizer dial to achieve the required output concentration in a variable bypass vapouriser. This
will enable judicious control of the concentration of the volatile requirements like achieving the
ED
95,uptake by circulation,tissues and the priming dose of vapour needed for ventilation etc in
low flow anaesthesia[4]. Therefore the volume of vapour produced in the vaporising chamber
of a modern variable bypass vaporiser can be estimated instantaneously and achieve the
targeted amount of vapour proportional to the fresh gas flow and dial setting.
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
4
Proposed explanation
The dial settings of the variable bypass vaporizers currently used are off, zero and the markings
in percentage. When the vapouriser dial is in the off position fresh gas flow does not enter the
vapouriser.It exits via the back bar of the anaesthesia machine. In the zero position all the fresh
gas flow pass through the bypass chamber of the vapouriser manifold. When the dial is set to
c%, the fresh gas flow (f) splits into two streams namely, the vapourising chamber(carrier gas)
flow(x) and bypass flow(f-x)as explained in the Figure1.
Figure 1 showing line diagram of cross section of temperature compensated variable bypass vapouriser
with baffles.Valves are denoted by the X shape. Cones represent the concentration control and the
thermostat,c -dialled concentration of the volatile agent , f- fresh gas flow entering vapouriser inlet , x- the
carrier gas flow and y- the volume of vapour produced inside the vapourising chamber, v- the volume &
SVP- the saturated vapour pressure.Black dots represent the vapour molecules and arrows represent the
dirction of flow of fresh gas.
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
5
The carrier gas flowing above the volatile liquid in the vapourising chamber is enriched by the
saturated vapour volume(y ml). This is equal to the dialled in percentage concentration of the
vaporizer multiplied by the fresh gas flow(f ml/min)entering the vaporizer inlet. The vapour so
produced has a saturated vapour pressure of 160 mm of Hg at 20
oC at atmospheric pressure.
Therefore as per Dalton's law of partial pressure,concentration of saturated vapour produced is
160 mmHg/760mmHg=0.21(21% volume) which is lethal for humans. Carrier gas forms the
remaining 79%(600/760) volume(x ml) of the mandatory atmosphere above the volatile
liquid.When the fresh gas flow(f) is 1000 ml per minute with a vaporizer dial setting(c) of one
percent, the volume of vapour added to the carrier gas flow is simply 10 ml(y ml). The
vapourising chamber output concentration of 21% sevoflurane vapour (in a volume of ten ml) in
the carrier gas flow of unknown volume (x ml) is reduced to the clinically safe concentration of
one percent vapour at the vapouriser output. This occurs on dilution with bypass gas flow(f-x) at
the vaporizer outlet. The carrier gas flow (x ml/min) may be calculated from the equation
x/y=(600/760)/(160/760)=x(ml/min)/ 10 (ml/min) ie x=3.75*10=38 ml. Therefore the bypass
flow is 1000-38=962 ml and the splitting ratio is 962/38=25, ie 25:1. The carrier gas flows for a
set of dial settings and fresh gas flows are given in the Table 1 below.
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
6
Table1 Calculation of vapour output, bypass & carrier gas flow,vaporiser outlet concentration
and splitting ratio of a variable bypass sevoflurane vapouriser.3.75 is the ratio of partial pressure
of carrier gas(600mm Hg)at atmospheric pressure(760mm Hg) to the saturated vapour pressure of
sevoflurane(160mmHg)at 20oC produced in the vapourising chamber.
Fresh
Gas
Flow
(f ml)
Vapouriser
Dial setting
(c%)
Volume of Saturated
vapour
(160/760=0.21(21%)
produced in vaporising
c h am be r o u t p u t (y
ml) (cf=y)
Volume of carrier gas
entering vaporising
chamber (x ml)=
3.75*vapour
produced (y ml/min)
‘*‘denote to multiply
Bypass flow/
Carrier Gas flow
(f-x)/x
Vaporizer
outlet
concentration
(%)= y / (f+y)
Splitting
ratio
1000 1 (1/100)*1000=10 3.75*10(=38) 962/38(=25) 10/1010=1 25:1
1000 2 (2/100)*1000=20 3.75*20(=75) 925/75(=12) 20/1020=2 12:1
2000 1 (1/100)*2000=20 3.75*20 (=75) 1925/75(=25.6) 20/2020=1 25.6:1
2000 2 (2/100)*2000=40 3.75*40 (=150) 1850/150(=12.3) 40/2040=2 12.3:1
3000 3 (3/100)3000=90 3.75*90(=338) 2662/338(=7.8) 90/3090=3 7.8:1
3000 4 (4/100)3000=120 3.75*120(=450) 2550/450(=5.6) 120/3120=4 5.6:1
The splitting ratio may be verified using the formula R={(S/F)-1}/(1-S) where R =splitting ratio,
S= fraction of saturated vapour concentration at 200C for sevoflurane being 0.21, F is the fraction
of dial setting, 0.01 for 1 %, 0.02 for 2% etc[5].
Simplified formula for calculating the splitting ratio in a sevoflurane variable bypass vaporizer
is given below. Please see the appendix for derivation of the formula. S=(1/3.75c)-1, where S is
splitting ratio,c is fraction of the dialled concentration of sevoflurane (c = 0.01 for dial setting of
1 %). Constant 3.75 is the ratio of partial pressure of carrier gas at atmospheric pressure to the
saturated vapour pressure of sevoflurane vapor produced in the vapourisng chamber at 20
o C.
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
7
Splitting ratio=bypass gas flow /carrier gas flow = (f-x) / x ……. Equation1
x=3.75fc………………….Equation 2
On applying Equation 2, Equation 1 becomes
Splitting ratio=(f-3.75fc) / 3.75f=f(1-3.75c) / 3.75fc =(1-3.75c) / 3.75c
Splitting ratio= (1/3.75c) -1
When dialled concentration is 1%, c=0.01,splitting ratio=(1/3.75*0.01)-1=26.66 -1=25.66
Therefore splitting ratio is approximately 25:1
Conclusion
Reliable estimation of the amount of anaesthetic vapour produced in the vapourising chamber
andthat delivered at the the vaporizer outlet is clinically more important for the operator while
providing inhalational general anaesthesia. Calculating the same based on assumptions of carrier
gas flow is not logical. The above explanation helps in the safer use of the volatile agent during
low flow anaesthesia. Vapour out put is the product of fresh gas flow and set dial concentration.
Carrier gas flow for sevoflurane vaporizer is 3.75 times the vapour output. Splitting ratio derived
above is independent of fresh gas flow but dependent on the dialled concentration of the agent
used.
Reference
1. Miller RD, Michael P. Bokoch, Stephen D. Weston. MIller’s Anaesthesia. 9th ed. 2020.
2. Jan Ehrenwerth, Ehrenwerth JB, Berry JM. Anesthesia Vaporizers ,Chapter3, page 68-70. In:
Anaesthesia Equipment, Principles & Practice. 2nd ed. Elsevier; p. 68–70.
3. Dhulkhed V, Shetti A, Naik S, Dhulkhed P. Vapourisers: Physical Principles and
Classification. Indian J Anaesth [Internet]. 2013 [cited 2024 Apr 5];57(5):455–63. Available
from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3821262/
4. Jacobs D. Closed circle anesthesia [Internet]. NYSORA. 2023 [cited 2024 Mar 6]. Available
from: https://www.nysora.com/closed-circle-anesthesia/
5. Leigh JM. Variations on a theme: splitting ratio. Anaesthesia. 1985 Jan;40(1):70–2.
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
8
Appendix
Derivation of splitting ratio of a modern sevoflurane vapouriser
Saturated vapour pressure (SVP) of sevoflurane at 20°C = 160 mm Hg
Atmospheric Pressure= 760mm Hg
Partial pressure of carrier gas in vapourising chamber =760-160 =600mm Hg
Concentration of sevoflurane vapour in vapourising chamber= 160760=0.21(21%)
Concentration of carrier gas(C
cg)in vapourising chamber =600760=0.79(79%)
f = fresh gas flow entering the vapouriser inlet (ml/min)
c = concentration in percentage or fraction dialed on the concentration control dial (1%=0.01)
x = carrier gas volume entering the vapouring chamber (ml/min)
y = vapour produced in the vapourising chamber (ml/min)
f-x = bypass gas flow (ml/min)
x + y = sum of the volumes of carrier gas and vapour (ml/min)
At 20°C, y is 21 % of vapourising chamber atmosphere and x is 79% in a modern sevoflurane
vapouriser. Therefore the ratio between carrier gas and vapour produced is given by
x / y=(79% ) / (21%)= (600/760) / (160/760)=600 / 160=3.75
So x=3.75 *y and y=f*c
Therefore x=3.75*f*c
When the flow split in the variable bypass vapouriser occurs at the vapouriser in let and outlet
the splitting ratio of the fresh gas flow is given by S=(f-x) / x & S= (f-x) / (x+y) respectively. By
applying the values of x & y as given above the splitting ratio formulae may be derived as given
below. The splitting ratio(s) for effective gas split at vapouriser inlet is given by s = (1/3.75*c)-1
and the exit flow split occuring at the vapourising chamber outlet is s=(1/4.75*c) – Ccg , where
c is the concentration control dial setting and Ccg
g is the carrier gas concentration in percentage.
Thus it is proof that splitting ratio is independent of fresh gas flow. It depends on the dialled
concentration and the constant specific for the volatile anaesthetic (sevoflurane). This constant
is the ratio of the partial pressure of carrier gas at atmospheric pressure to the saturated vapour
pressure of the vapour produced in the vapourising chamber generally at 20
oC. This constant
depends on the concentration of the vapour produced inside the vapourising chamber. For
halothane this constant is (760-243) / 243=2.1.
. CC-BY-NC-ND 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted April 12, 2024. ; https://doi.org/10.1101/2024.04.11.24305499doi: medRxiv preprint
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.