Abstract
Rapid SARS-COV-2 related serology testing can help identify and manage the spread of infection
in decentralized testing environments but the limitation in performance of existing tests in blood
has restricted implementation of testing at the point-of-care. Optimization of existing rapid tests
in whole blood will require significant effort in the short-term and there is a need for solutions to
help bridge the gap in performance between plasma and whole blood . We demonstrate here the
implementation of the H.E.R.M.E.S platform, a portable plasma separation system that can
enhance the performance of blood-based diagnostic testing, with a commercially available SARS-
COV-2 IgG/IgM serology rapid diagnostic test (RDT) in a blinded study with 61 human samples.
We compare the performance of the RDT in whole blood and separated plasma and highlight that
plasma yields a 39% increase in positivity agreement with PCR in samples collected from patients
with early infections. We further legitimize the increase in positivity agreement rate with the help
of an independent evalua tion by 10 previously untrained users. The H.E.R.M.E.S plasma
separation system circumvents the need for assay optimization in whole blood and furthers the
legitimacy of incorporating SARS-COV-2 serology RDTs at the point -of-care. The data
highlighted in this work makes a compelling case for the incorporation of the H.E.R.M.E.S system
in large scale efforts to perform SARS -COV-2 serology testing in decentralized testing
environments.
Introduction
As of September 15th 2020, the novel coronavirus (SARS -COV-2) has infected over 28 million
people worldwide1. Serology testing has emerged as an effective avenue to monitor the spread and
prevalence of infection in commun ities and healthcare settings 2–5 since IgG and IgM
seroconversion can be reliably observed in individuals infected with SARS -CoV-2 within 2 -17
days of initial symptom onset 6–10. As opposed to molecular testing, serology testing can be used
as an epidemiological tool to gather information on infection history across population clusters
and help enhance contact tracing efforts in the event of a localized outbreak 11,12. To expedite
availability of diagnostic tests to the public, t he Food and Drug Administration (FDA) issued an
Emergency Use Authorization (EUA) on Feb 21 2020 allowing immunoassay and serology tests
to be available commercially based on manufacturer-reported data without formal clearance and
approval13. To date, the FDA has issued EUAs for 31 different serology tests 14; a third of which
are further classified as lateral flow rapid diagnostic tests (RDT) which offer the benefit of ease of
operation, low sample volumes and quick results (10-15 minutes)15–17.
RDTs are the ideal platform to facilitate mass testing as they can be easily distributed and
incorporated in decentralized testing environments . However, RDTs are currently limited in
application for SARS -COV-2 serology testing at t he point-of-care because of issues with their
performance. While m anufacturers claim equivalent diagnostic accuracy in blood, serum and
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
NOTE: This preprint reports new research that has not been certified by peer review and should not be used to guide clinical practice.
plasma, the majority of self-reported validation data has been collected by testing serum or plasma
samples alone. Several lab studies have shown that the actual performance of these tests is much
lower than advertised 18–20 which has called the legitimacy of serology testing into question 21.
Additionally, under the current EUA, there are no serology tests approved for use in point-of-care
settings which wo uld further suggest that existing tests have not demonstrated reliable
performance in venous and fingerstick blood samples. RDTs are currently restricted for use in
moderate and high complexity labs22, which are diametrically opposite in function to point-of-care
testing environments and are likely have access to more robust lab-based tests. The limitation in
performance has ultimately bottlenecked the application of RDTs for use at the point-of-care and
a widescale rollout of RDT serology tests will require demonstration of robust and reliable
performance in blood.
The difficulty to demonstrate clinically acceptable performance in blood stems from the
interference introduced by red blood cells, which ca n make up to 5 5% of the total volume of the
sample23. While the cell themselves adversely affect traditional immunochemical diagnostic
methods24,25, red blood cell separation itself is a tedious process that is not compatible with
facilities and resources typically available in point-of-care settings. To address the gap in available
solutions for sample processing in decentralize d testing environments, w e previously
demonstrated the High Efficiency Rapid Magnetic Erythrocyte Separator (H.E.R.M.E.S) portable
plasma separation platform: a low-cost magnetic bead capture assay to remove up to 99.9% of all
red blood cells in a sample to deliver plasma that can enhance the diagnostic accuracy of lateral
flow tests that are designed for use with blood26. In this work, we demonstrate the implementation
of the H.E.R.M.E.S platform with a commercially available FDA EUA approved SARS-COV-2
IgG/IgM RDT in 31 PCR positive and 30 prospective negative samples. We compare the
performance of the RDT in venous whole blood and separated plasma in a randomized blinded
study and demonstrate that the tests show a significantly higher positivity agreement with PCR
when plasma is used as the sample input . We also illustrate that H.E.R.M.E.S p lasma is an
advantageous sample type for use in serology testing as it widens the specific IgG detection
window to within 2 weeks of infection. To further establish the validity of the study, we asked 10
previously untrained users to participate in a survey to assess the outcomes of the tests who also
noted a similar increase in positivity agreement between the two sample types . The increase in
performance as noted by trained professionals and untrained users with H.E.R.M.E.S plasma
suggests that it is possible to avoid the issue of direct testing in blood altogether which could
enhance the legitimacy of the use case for SARS-COV-2 serology RDTs in decentralized testing
environments and facilitate the implementation of mass screening efforts to help mitigate the
ongoing pandemic.
Materials and methods
Sample Preparation: A total of 61 samples were used in this study (31 PCR positive and 30
presumed negative samples). PCR positive samples were acquired from two separate commercial
collection services: CantorBio (San Diego, CA) and AMSBio (Cambridge, MA). Of the 31 PCR
positive samples, 16 were collected with additional inclusion criteria of “Patients who tested PCR
positive within 14 days prior to collection”. The 30 negative samples were acquired from
Innovative Research (Novi, MI) and were further split into frozen plasma samples acquired prior
to the pandemic and fresh whole blood samples collected prospectively from donors who had not
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
yet tested PCR positive. For the purposes of equivalency, the frozen negative samples were mixed
with pooled washed red blood cells (Innovative Research, Novi, MI) to mimic the consistency of
a blood sample wi th a 50% hematocrit. Once the samples were received from their respective
vendors, they were anonymized and assigned a random 6-digit alphanumeric code to facilitate a
blinded study.
Lateral Flow Rapid Diagnostic Tests : COVID 19 Ra pid IgG/IgM tests (Healgen Scientific, San
Diego, USA) were acquired from Confirm Biosciences (San Diego, CA). We compared the
performance of the test in two different sample types: whole blood and plasma separated by the
H.E.R.M.E.S platform. Briefly, the protocol for the whole blood test involved dispensing 10L of
sample into the input p ortion of the test strip. This was immediately followed by the application
of 2 drops of buffer provided by the manufacturer. For the separated plasma samples , 20L of
whole blood was dispensed into a PCR tube containing the magnetic bead separation assay and
placed in a small benchtop device for sepa ration as outlined in Vemulapati et al (ref) . After the
plasma was separated, a capillary pipette was used to extract 5l of plasma which was dispensed
onto the sample input portion of the test. The test was then initiated by dispensing 2 drops of buffer.
For either sample type, t he test was allowed to run for 10 minutes before a positive or negative
Result
was noted. Individual IgG and IgM responses were recorded based on the presence of red
bands in the IgG and IgM “Test” line. Positive agreement with PCR was recorded if a red band
was observed in either the IgG or IgM section of the test. A faint test line was also considered as
a positive sample. Results were recorded by a trained laboratory professional. After the results of
each test were recorded, the test strip was imaged by a rudimentary smartphone imaging setup
using an Iphone 7 (Apple Inc, Cupertino CA) for use in the untrained user survey.
Untrained User Survey: The smartphone images collected during testing were cataloged and input
into a survey (Google Forms) to use in a short study to gather blinded performance data from
previously untrained users. A total of 122 images (61 images for each sample type) were used for
the evaluation. 10 participants were recruited using Amazon Mechanical Turk to participate in the
study. The participants were required to have a high school degree at the very minimum in order
to meet the suggested education requirement for laboratory personnel in a Certificate for
Laboratory Improvement Amendment (CLIA) approved lab. There was no explicit mention that
the images were collected from SARS-COV-2 serology tests.
To start the survey, participants were provided with brief instructions and background information
necessary to classify the tests as positive, negative or invalid. The participants were then asked to
complete a six-question training survey to help familiarize themselves with the selection criteria .
Each question in the survey contained an image of a rapid test with either a sample positive for
IgG alone, positive for IgM alone, positive for both IgG and IgM and negative for both IgG and
IgM. For each training question, participants were asked to determine if the test was ‘Positive’,
‘Negative’ or ‘Indeterminate’. If a participant answered a question wrong in the training survey,
they were informed of their error and given another chance to get the question right. The training
survey was primarily designed to ensure that the participants had an accurate understanding of the
classification criteria and was not used to judge results from the full survey. Once the training
survey was completed, participants were advanced to the main section of the survey which
included evaluating the full set of 122 images. These images were further divided into 6 sections
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
in order for participants to accurately track their own progress. The order that images were
displayed were also randomized for each participant.
Results
Comparison of performance in whole blood and H.E.R.M.E.S plasma: Due to the lack of a
commercially available gold standard serology test for SARS -COV-2, we used a Positive
Percentage Agreement (PPA) and Negative Percentage agreement (NPA) against PCR to evaluate
diagnostic accuracy of the antibody test. PPA/NPA for the whole blood and H.E.R.M.E.S can be
seen in Table 1 and Table 2. For the 31 positive samples, we note an increase in PPA from 83.9%
in whole blood to 96.7% in separated plasma. The difference in PPA is further highlighted in the
set of samples collected from patients within the first 14 days of PCR positive diagnosis (6 8.8%
in whole blood and 93.8% in H.E.R.M.E.S plasma). The difference in PPA can be largely attributed
to the false negatives from whole blood samples collected from patients with early stage infections,
an example of which is shown in Figure 1A. The test correctly identified IgG in all 15 blood
samples that were collected from patients within 1 -2 months of infection . We also observe d a
significant increase in the IgM detection rate from 64.5% in whole blood to 80.6% in H.E.R.M.E.S
plasma. The NPA remained at 100% for both sample types.
Untrained User Evaluation: Table 3 shows the average PPA and NPA as determined by the 10
previously untrained users. We note a slightly diminished increase in total PPA from whole blood
to H.E.R.M.E.S (80.7% to 89.4%) and a similar increase in PPA for the early infections sample
subset (62.6% to 9 3.1%). The NPA for both sample types were comparable (98.3% vs 99.7 %).
The Fleiss’ kappa for inter rater reliability27 was evaluated to be significantly greater than 0.81 in
both cases demonstrating near perfect agreement between the evaluators across both sample types.
Discussion
Rapid serological screening at the point of care can enable local governments, communities and
workplaces to effectively monitor and track the spread of SARS -COV-2 in the ir respective
populations28,29. For successful implementation, it is critical that serology RDTs exhibit a high
diagnostic accuracy (>95%) in venous or fingerstick blood, as they are likely the only accessible
sample type at the point -of-care. In this work, we demonstrate the validity of a portable plasma
separation system as an alternative approach to improve the performance of SARS -COV-2
serology RDTs for use at the point-of-care. By enabling plasma separation in decentralized testing
environments, the H.E.R.M.E.S platform bypasses the need for arduous optimization of exi sting
assays for use in blood. We offer evidence to suggest that plasma shows a significant improvement
in positivity agreement against PCR when compared to whole blood and further demonstrate that
the use of plasma can enhance the detection window for specific IgG to within 2 weeks of infection.
We also demonstrate that the difference in observed performance is similar when the tests are
evaluated by a laboratory professional and an untrained user, further validating the use case for
implementing H.E.R.M.E .S with RDTs in decentralized testing environments . The ability to
facilitate earlier detection and the capability to make plasma accessible at the point-of-care make
a compelling case for the incorporation of the H.E.R.M.E.S plasma separation system for a ny
widescale serology testing pursuit.
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
A higher positivity agreement rate with PCR would likely translate to a higher sensitivity for these
tests in clinical settings. A highly sensitive serology RDT for SARS-COV-2 could be impactful in
a variety of clin ical situations. In addition to detecting a higher number of infections in patients
who are within the first few days of infection, an enhanced sensitivity would also help inform
infection history in those individuals who were infected during the onset of the pandemic. With an
expectation of lowered levels of specific antibody in those who are several months removed from
exposure to SARS -COV-2, a highly sensitive test could help accurately diagnose past infection
and prevent more false negatives. Additional ly, a point-of-care serology test could be extremely
beneficial when an effective vaccine has been verified and deployed across large masses of the
population. The ability to reliably verify antibody presence at the point -of-care could help state,
county and city governments track the progress of their respective vaccination efforts. In turn, the
means to quickly verify successful vaccination could lead to more robust pathways to help
workplaces and business gradually return to optimal operating conditions.
The enhanced performance of RDTs in separated plasma can be largely attributed to the absence
of red blood cells. In addition to interfering with the underlying immunochemistry, red cells make
visual readout based on colorimetric bands difficult due to the increased background and noise due
to the pigment of the cells. This also makes it harder to visually identify faint signal responses in
samples that might have relatively low level of analyte. This is further exemplified in the higher
IgM positivity rate in plasma over whole blood as noted in the results of the study . While faint
lines can be difficult to recognize, they can be used to trigger a follow -up confirmatory test, a
common practice employed in decentralized testing environments. The faint response could likely
be solved by using a test from a different manufacturer or dispensing 30% more sample volume
than recommended. We also expect that an automated imaging system could be a viable alternative
to avoid the issue of human error altogether. In this study, all faint responses we re classified as
positive which could help explain the diminished increase in PPA determined by the untrained
users in comparison to the laboratory professional.
The overall performance of the test in whole blood as determined in this study is much lower than
described by the manufacturer and the National Cancer Institute’s independent evaluation in serum
samples30. The difference in performance is more noticeable in the case of IgM. The diminished
sensitivity in detection of specific IgM is expected in whole blood due to a combination of lower
overall performance and the smaller amount of IgM generated by the body during an immune
response as compared to IgG 31. However, we also suspect that the lower performance in IgM is a
function of the design of multiplexed lateral flow assays; the architecture of the test strip tends to
favor the detection of the analyte that is captured further away from the point of sample input due
to the dependence of reaction time on flow rate32. As such, we expect the IgM performance could
be imp roved with the design of a standalone IgM test or by employing a different test strip
architecture. Finetuning the IgM responses could play a bigger role in enabling earlier detection,
particularly when used in combination with a PCR test33.
In this work, we demonstrate the compatibility of the H.E.R.M.E.S platform with a single
commercially available EUA approved SARS-COV-2 serology rapid test and we expect the
performance difference to be similar in other serology tests that are compatible with plasma and
whole blood. The difference in performance between blood and plasma will likely also depend on
the recommended protocol outlined by the manufacturers in the “Instruction for Use” (IFU)
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
documents. Specifically, the recommend sample volumes will dictate the difference in
performance of the test as immunochemical tests are sensitive to changes in total analyte34. While
some manufacturers suggest a specific amount for a given sample type (typically the volume of
recommended plasma is half of the amount of blood) , others recommend the use of the same
amount of volume regardless of sample type. In the case of the latter, we expect the differences in
performance between whole blood and plasma to be even greater than demonstrate here due to the
higher total amount of analyte in the plasma sample.
The number of positive samples we chose to evaluate in this study is in line with the number of
positives expected to be tested by the FDA as outlined in their EUA requirements. Although the
EUA recommends testing a total of 75 negative samples for each new antibody test , we focused
our efforts in this study to evaluate the differences in positive percentage agreement that would
arise from a difference in sample type. Next steps will involve further validation for use at the
point of care by conducting a fingerstick study with volunteers in a point-of-care clinic of a similar
size to evaluate the applicability and of our device in settings outside of the lab. Future studies will
also include evaluation of the test from samples with known underlying conditions that are known
to cross react with the COVID test (e.g., H.I.V).
Conclusion
We demonstrate here the clear value that H.E.R.M.E.S plasma presents as a sample type for use
with existing SARS-COV-2 serology testing RDTs as opposed to whole blood. Incorporation of
the separation platform into existing testing protocols can immediately enhance the legitimacy of
the use case for rapid serology testing and circumvent the arduous need to demonstrate equivalent
performance in blood. We expect the H.E.R.M.E.S platform to play an integral role in any large-
scale testing efforts that will require the deployment of serology RDTs to communities across the
country.
Acknowledgments
SV would like to thank Serhat Pala at Confirm Biosciences for providing the Healgen antibody
tests for the study.
Competing interests
SV is founder of Hermes Life Sciences and is pursuing the commercialization of the plasma
separation technology described in this work.
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
References
1. Johns Hopkins University. COVID-19 Dashboard by the Center for Systems Science and
Engineering (CSSE) at Johns Hopkins University (JHU).
https://coronavirus.jhu.edu/map.html.
2. Peeling, R. W. et al. Serology testing in the COVID-19 pandemic response. The Lancet
Infectious Diseases 20, e245–e249 (2020).
3. Ong, D. S. Y., de Man, S. J., Lindeboom, F. A. & Koeleman, J. G. M. Comparison of
diagnostic accuracies of rapid serological tests and ELISA to molecular diagnostics in
patients with suspected coronavirus disease 2019 presenting to the hospital. Clin. Microbiol.
Infect. (2020) doi:10.1016/j.cmi.2020.05.028.
4. Serre-Miranda, C. et al. Performance assessment of 11 commercial serological tests for
SARS-CoV-2 on hospitalized COVID-19 patients.
http://medrxiv.org/lookup/doi/10.1101/2020.08.06.20168856 (2020)
doi:10.1101/2020.08.06.20168856.
5. Ghaffari, A., Meurant, R. & Ardakani, A. COVID-19 Serological Tests: How Well Do They
Actually Perform? https://www.preprints.org/manuscript/202006.0278/v1 (2020)
doi:10.20944/preprints202006.0278.v1.
6. Long, Q.-X. et al. Antibody responses to SARS-CoV-2 in patients with COVID-19. Nature
Medicine 26, 845–848 (2020).
7. Zhao, J. et al. Antibody responses to SARS-CoV-2 in patients of novel coronavirus disease
2019. Clinical Infectious Diseases (2020) doi:10.1093/cid/ciaa344.
8. To, K. K.-W. et al. Temporal profiles of viral load in posterior oropharyngeal saliva samples
and serum antibody responses during infection by SARS-CoV-2: an observational cohort
study. The Lancet Infectious Diseases 20, 565–574 (2020).
9. Wu, F. et al. Neutralizing Antibody Responses to SARS-CoV-2 in a COVID-19 Recovered
Patient Cohort and Their Implications. SSRN Electronic Journal (2020)
doi:10.2139/ssrn.3566211.
10. Okba, N. M. A. et al. Severe Acute Respiratory Syndrome Coronavirus 2−Specific Antibody
Responses in Coronavirus Disease Patients. Emerging Infectious Diseases 26, 1478–1488
(2020).
11. Kubina, R. & Dziedzic, A. Molecular and Serological Tests for COVID-19 a Comparative
Review of SARS-CoV-2 Coronavirus Laboratory and Point-of-Care Diagnostics.
Diagnostics (Basel) 10, (2020).
12. Winter, A. K. & Hegde, S. T. The important role of serology for COVID-19 control. The
Lancet Infectious Diseases 20, 758–759 (2020).
13. Secretary of Health and Human Services. Determination of a public health emergency and
declaration that circumstances exist justifying authorizations pursuant to section 564(b) of
the federal food, drug, and cosmetic act, 21 U.S.C. § 360bbb-3.
https://www.fda.gov/media/135010/download.
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
14. U.S Food and Drug Administration. Individual EUAs for Serology Tests for SARS-CoV-2.
https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-
authorizations-medical-devices/vitro-diagnostics-euas#individual-serological.
15. Posthuma-Trumpie, G. A., Korf, J. & van Amerongen, A. Lateral flow (immuno)assay: its
strengths, weaknesses, opportunities and threats. A literature survey. Anal Bioanal Chem
393, 569–582 (2009).
16. Sajid, M., Kawde, A.-N. & Daud, M. Designs, formats and applications of lateral flow assay:
A literature review. Journal of Saudi Chemical Society 19, 689–705 (2015).
17. Hoffman, T. et al. Evaluation of a COVID-19 IgM and IgG rapid test; an efficient tool for
assessment of past exposure to SARS-CoV-2. Infection Ecology & Epidemiology 10,
1754538 (2020).
18. Whitman, J. D. et al. Test performance evaluation of SARS-CoV-2 serological assays.
http://medrxiv.org/lookup/doi/10.1101/2020.04.25.20074856 (2020)
doi:10.1101/2020.04.25.20074856.
19. Velay, A. et al. Evaluation of the performance of SARS-CoV-2 serological tools and their
positioning in COVID-19 diagnostic strategies.
http://biorxiv.org/lookup/doi/10.1101/2020.06.16.156166 (2020)
doi:10.1101/2020.06.16.156166.
20. Lisboa Bastos, M. et al. Diagnostic accuracy of serological tests for covid-19: systematic
review and meta-analysis. BMJ m2516 (2020) doi:10.1136/bmj.m2516.
21. Duong, Y. T., Wright, C. G. & Justman, J. Antibody testing for coronavirus disease 2019:
not ready for prime time. BMJ m2655 (2020) doi:10.1136/bmj.m2655.
22. Shah, A. & Shuren, J. US Food and Drug Administration. Insight into FDA’s Revised Policy
on Antibody Tests: Prioritizing Access and Accuracy. https://www.fda.gov/news-events/fda-
voices/insight-fdas-revised-policy-antibody-tests-prioritizing-access-and-accuracy (2020).
23. Al-Soud, W. A. & Radstrom, P. Purification and Characterization of PCR-Inhibitory
Components in Blood Cells. Journal of Clinical Microbiology 39, 485–493 (2001).
24. Bhamla, M. S. et al. Hand-powered ultralow-cost paper centrifuge. Nature Biomedical
Engineering 1, 0009 (2017).
25. Mabey, D., Peeling, R. W., Ustianowski, A. & Perkins, M. D. Diagnostics for the developing
world: Tropical infectious diseases. Nature Reviews Microbiology 2, 231–240 (2004).
26. Vemulapati, S. & Erickson, D. H.E.R.M.E.S: rapid blood-plasma separation at the point-of-
need. Lab on a Chip 18, 3285–3292 (2018).
27. Falotico, R. & Quatto, P. Fleiss’ kappa statistic without paradoxes. Qual Quant 49, 463–470
(2015).
28. Cheng, M. P. et al. Diagnostic Testing for Severe Acute Respiratory Syndrome–Related
Coronavirus 2: A Narrative Review. Annals of Internal Medicine 172, 726–734 (2020).
29. Weinstein, M. C., Freedberg, K. A., Hyle, E. P. & Paltiel, A. D. Waiting for Certainty on
Covid-19 Antibody Tests — At What Cost? N Engl J Med 383, e37 (2020).
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
30. National Cancer Institute. Serology Test Evaluation Report for “COVID-19 IgG/IgM Rapid
Test Cassette” from Healgen.
https://www.accessdata.fda.gov/cdrh_docs/presentations/maf/maf3247-a001.pdf (2020).
31. Gonzalez-Quintela, A. et al. Serum levels of immunoglobulins (IgG, IgA, IgM) in a general
adult population and their relationship with alcohol consumption, smoking and common
metabolic abnormalities: Serum immunoglobulin levels in adults. Clinical & Experimental
Immunology 151, 42–50 (2007).
32. Bahadır, E. B. & Sezgintürk, M. K. Lateral flow assays: Principles, designs and labels. TrAC
Trends in Analytical Chemistry 82, 286–306 (2016).
33. Liu, R. et al. The comparative superiority of IgM-IgG antibody test to real-time reverse
transcriptase PCR detection for SARS-CoV-2 infection diagnosis.
http://medrxiv.org/lookup/doi/10.1101/2020.03.28.20045765 (2020)
doi:10.1101/2020.03.28.20045765.
34. Hong, L. et al. High performance immunochromatographic assay for simultaneous
quantitative detection of multiplex cardiac markers based on magnetic nanobeads.
Theranostics 8, 6121–6131 (2018).
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
Tables and Figures
Table 1: Summary of results for comparison between Whole blood and H.E.R.M.E.S plasma
Comparator Method (PCR) Prospective Negative
IgM+/IgG+ IgM-/IgG-
SARS COV 2 IgM/IgG
Antibody Test Whole Blood H.E.R.M.E.S Whole Blood H.E.R.M.E.S
IgM+/IgG+ 20 25 0 0
IgM+/IgG- 0 0 0 0
IgM-/IgG+ 6 5 0 0
IgM-, IgG - 5 1 30 30
Total 31 30
Table 2: Comparison of PPA and NPA between Whole Blood and H.E.R.M.E.S. % represents
agreement with PCR in the case of positives. C.I. represents the 95% Confidence interval.
Whole Blood H.E.R.M.E.S
% C.I. % C.I.
PPA 83.9% (26/31) (70.7%-97%) 96.7% (30/31) (90.5%-100%)
PPA
(early infections) 68.8% (11/16) (45.2%-92.2%) 93.8% (15/16) (81.5%-100%)
NPA 100% (30/30) 100% (30/30)
Table 3: Comparison of PPA/NPA as determined by 10 untrained users. Fleiss’ Kappa for
interrater reliability is shown in the last row.
Whole blood H.E.R.M.E.S
% C.I. % C.I.
PPA 80.7 (66.5% - 94.8%) 89.4 (78.9% - 99.8%)
PPA (early
infections) 62.6 (38.1% - 87.1%) 93.1 (80.9% - 100%)
NPA 98.3 (96.4% - 100%) 99.7 (99.1% - 100%)
Fleiss’ Kappa 0.97 0.94
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
Figure 1: Example of an incorrectly assessed False Negative test in whole blood. A) Test result
from using whole blood and B) test result using H.E.R.M.E.S plasma from the same sample. A
faint IgG response is noted by the arrow.
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted October 2, 2020. ; https://doi.org/10.1101/2020.10.02.20199083doi: medRxiv preprint
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.