Background
42
Serological tests are widely used in various medical disciplines for diagnostic and 43
monitoring purposes. Unfortunately, the sensitivity and specificity of test systems is often 44
poor, leaving room for false positive and false negative results. However, conventional 45
Methods
used to increase specificity decrease sensitivity and vice versa. Using SARS-46
CoV-2 serology as an example, we propose here a novel testing strategy: the 47
"Sensitivity Improved Two-Test" or "SIT²" algorithm. 48
Methods
49
SIT² involves confirmatory re-testing of samples with results falling in a predefined 50
retesting-zone of an initial screening test, with adjusted cut-offs to increase sensitivity. 51
We verified and compared the performance of SIT² to single tests and orthogonal testing 52
(OTA) in an Austrian cohort (1,117 negative, 64 post-COVID positive samples) and 53
validated the algorithm in an independent British cohort (976 negatives, 536 positives). 54
Results
55
The specificity of SIT² was superior to single tests and non-inferior to OTA. The 56
sensitivity was maintained or even improved using SIT² when compared to single tests 57
or OTA. SIT² allowed correct identification of infected individuals even when a live virus 58
neutralization assay could not detect antibodies. Compared to single testing or OTA, 59
SIT² significantly reduced total test errors to 0.46% (0.24-0.65) or 1.60% (0.94-2.38) at 60
both 5% or 20% seroprevalence. 61
Conclusion
62
For SARS-CoV-2 serology, SIT² proved to be the best diagnostic choice at both 5% and 63
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
4
20% seroprevalence in all tested scenarios. It is an easy to apply algorithm and can 64
potentially be helpful for the serology of other infectious diseases. 65
Running Head 66
Sensitivity-improved two-test serology 67
Abbreviations 68
Confidence interval (CI) 69
Coronavirus disease 2019 (COVID-19) 70
Enzyme-linked immunosorbent assay (ELISA) 71
Intensive Care Unit (ICU) 72
Neutralization titer (NT) 73
Nucleocapsid (NC) 74
Nucleoprotein (NP) 75
Orthogonal testing algorithms (OTA) 76
Polymerase chain reaction (PCR) 77
Receptor binding domain (RBD) 78
Sensitivity-Improved Two-Test (SIT2) 79
Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) 80
Spike protein (S) 81
Virus neutralization test (VNT) 82
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
5
Introduction
83
Serological tests are commonly used diagnostic tools in a broad medical field, spanning 84
from infectiology (1) to autoimmunity (2), oncology (3) and transplantation medicine (4). 85
They also play a critical role in animal disease surveillance (5). However, many 86
serological tests come with acceptable but imperfect sensitivities and specificities. Tests 87
with specificities slightly above 90% are considered good (6) or even highly specific (3). 88
However, at low seroprevalence rates, every single percent counts: if the frequency of a 89
given disease in the tested population is only 5%, a specificity of 90% would mean that - 90
even at a sensitivity of 100% - 5 true positives would be matched by ten false positives. 91
Thus, the probability of an individual with a positive test (positive predictive value, PPV) 92
to be a true positive would be only 33%. 93
During the early phase of the SARS-CoV-2 pandemic, seroprevalences were far below 94
1% (7). Therefore, highly specific test systems were necessary (>99,5%) to provide 95
good positive predictive values (8). Sensitivity and specificity can be seen as 96
communicating vessels – the improvement of one is usually at the expense of the other 97
(9). Consequently, test systems adjusted by the manufacturers to very high specificities 98
(>99%) showed moderate sensitivity. This problem was particularly evident when non-99
hospitalized patients were included in the cohorts studied (10-12). To further increase 100
specificity at very low seroprevalence levels, various methods have been proposed, e.g., 101
raising the thresholds for positivity or confirming a positive result with a second test 102
(orthogonal testing) (9, 13, 14). Unfortunately, these specificity improvement strategies 103
inevitably lead to a further reduction of the previously moderate sensitivities. 104
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
6
As the pandemic progressed, the problem became more pronounced as antibodies 105
declined, and sophisticated statistical models were required to compensate for the 106
waning sensitivity (15). In the case of SARS-CoV-2, as with any evolving pandemic, 107
increasing seroprevalence rates worldwide have attenuated the need for higher 108
specificity. 109
However, the problem persists in non-epidemic diseases where seroprevalence remains 110
low. Moreover, each new pandemic begins with extremely low seroprevalence rates as 111
well, and in the future, we should have better diagnostic strategies in infectious serology 112
ready here. 113
In the present work, we propose for the first time an orthogonal test algorithm based on 114
real-life data for the SARS-CoV-2 antibody tests of Roche, Abbott, DiaSorin, and two 115
commercial SARS-CoV-2 ELISAs (16) intending to maximize both specificity and 116
sensitivity at the same time. Although our algorithm follows a general principle, it was 117
developed based on SARS-CoV-2 antibody tests. The SARS-CoV-2 pandemic provided 118
a unique opportunity in this regard, as historical samples from before the pandemic are 119
negative by definition (thus allowing accurate specificity testing). In addition, sufficient 120
PCR-confirmed positive cases were available quickly, ensuring a sophisticated and 121
reliable sensitivity verification. Thus, for SARS-CoV-2 - in contrast to most other 122
circulating microorganisms - a realistic and unusually accurate estimation of specificities 123
and sensitivities of serological tests was possible. We benefited from this advantage to 124
develop our sophisticated diagnostic algorithm. 125
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
7
Methods
126
Study design and cohorts 127
Sera used in this non-blinded prospective cross-sectional study were either residual 128
clinical specimens or samples stored in the MedUni Wien Biobank (n=1,181), a facility 129
specialized in the preservation and storage of human biomaterial, which operates within 130
a certified quality management system (ISO 9001:2015) (17). 131
For derivation of the SIT2 algorithm, sample sets from individuals known to be negative 132
and positive were established for testing. As previously described (18), samples 133
collected before 01.01.2020 (i.e., assumed SARS-CoV-2 negative) were used as a 134
specificity cohort (n=1117): a cross-section of the Viennese population (the LEAD 135
study)(19), preselected for samples collected between November and April to enrich for 136
seasonal infections (n=494); a collection of healthy voluntary donors (n= 265); a 137
disease-specific collection of samples from patients with rheumatic diseases (n=358); 138
(see also Tables S1 and S2). 139
The SARS-CoV-2 positive cohort (n=64 samples from n=64 individuals) included 140
patients testing positive with RT-PCR during the first wave and their close, symptomatic 141
contacts. Of this cohort five individuals were asymptomatic, 42 had mild-moderate 142
symptoms, four reported severe symptoms, and 13 were admitted to the Intensive Care 143
Unit (ICU). The timing of symptom onset was determined by a questionnaire for 144
convalescent donors and by reviewing individual health records for patients and was in 145
median 41 [26,25-49] days. For asymptomatic donors (n=5), SARS-CoV-2 RT-PCR 146
confirmation time was used instead (for more details, see Tables S1 and S3). All 147
included participants gave written informed consent to donate their samples for research 148
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
8
purposes. The overall evaluation plan conformed with the Declaration of Helsinki as well 149
as relevant regulatory requirements. It was reviewed and approved by the ethics 150
committee of the Medical University of Vienna (1424/2020). 151
For validation of the SIT² algorithm, we used data from an independent United Kingdom 152
cohort (20), including 1,512 serum/plasma samples (536 PCR confirmed SARS-CoV-2 153
positive cases; 976 negative cases, collected earlier than 2017). 154
Antibody testing 155
For the derivation analyses, SARS-CoV-2 antibodies were either measured according to 156
the manufacturers' instructions on three different commercially available automated 157
platforms (Roche Elecsys® SARS-CoV-2 [nuclecapsid total antibody assay, further 158
referred to as Roche NC], Abbott SARS-CoV-2-IgG assay [nucleocapsid IgG assay, 159
Abbott NC], DiaSorin LIASION® SARS-CoV-2 S1/S2 assay [S1/S2 combination antigen 160
IgG assay, DiaSorin S1/S2]) or using 96-well enzyme-linked immunosorbent assays 161
(ELISAs) (Technoclone Technozym® RBD and Technozym® NP) yielding quantitative 162
results(16) (for details see Supplement, Supplemental Methods). The antibody assays 163
used in the validation cohort were Abbott NC, DiaSorin S1/S2, Roche NC, Siemens RBD 164
total antibody, and a novel 384-well trimeric spike protein ELISA (Oxford Immunoassay) 165
(20), resulting in 20 evaluable combinations. All samples from the Austrian SARS-CoV-166
2-positive cohort also underwent live virus neutralization testing (VNT), and 167
neutralization titers (NT) were calculated, as is described in detail in the Supplemental 168
Methods. 169
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
9
Sensitivity improved two-test method (SIT²) 170
Our newly developed sensitivity improved two-test (SIT²) method consists of the 171
following key components: i) sensitivity improvement by cut-off modification and ii) 172
specificity rescue by a second, confirmatory test (Fig. 1A). 173
For the first component of the SIT² algorithm, positivity thresholds were optimized for 174
sensitivity according to the first published alternative thresholds for the respective 175
assays, calculated e.g. by ROC-analysis (21-23). Additionally, a high cut-off, above 176
which a result can be reliably regarded as true positive without the need for further 177
confirmation, was defined. These levels were based on in-house observations(18) and 178
represent those values (including a safety margin) above which no more false positives 179
were found. The highest results seen in false-positives were 1.800 COI, 2.86 Index, and 180
104.0 AU/mL, respectively. Hence, we defined the high cut-off for Roche and Abbott as 181
3.00 COI/Index and for DiaSorin as 150.0 AU/mL. The lowering of positivity thresholds 182
improves sensitivity; the high cut-off prevents unnecessary re-testing of clearly positive 183
samples. Moreover, the high cut-off avoids possible erroneous exclusion by the 184
confirmatory test. The newly defined interval between the reduced threshold for positivity 185
and the high cut-off is the re-testing zone (Fig. 1A). The initial antibody test (screening 186
test) is then followed by a confirmatory test, whereby positive samples from the re-187
testing zone of the screening test are re-tested. Also, for the confirmatory test, 188
sensitivity-adapted assay thresholds are needed (Figs.1A, 1B). As false-positive 189
samples are usually only positive in one test system (Fig. S1), false positives can be 190
identified, and specificity markedly restored with minimal additional testing as most 191
samples do not fall within the re-testing zone (18, 24). A flowchart of the testing strategy 192
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
10
and the applied cut-off levels and their associated quality criteria are presented in Figs. 193
1B, 1C. 194
Test strategy evaluation 195
On the derivation cohort, we compared the overall performance of the following SARS-196
CoV-2 antibody testing strategies: i) testing using single assays; ii) simple lowering of 197
thresholds; iii) classical orthogonal testing (OTA), and iv) our newly developed SIT2 198
algorithm at assumed seroprevalences of 5% and 20%. As part of the derivation, we 199
then compared the performance of OTAs and SIT2 against the results of a virus 200
neutralization assay. On the validation cohort, we then compared the performance of 201
OTAs and SIT2. Finally, we used data from this cohort to evaluate the performance of 202
SIT2 versus single tests at seroprevalences of 5%, 10%, 20%, and 50% if the Abbott and 203
DiaSorin assays (i.e., assays with varying degrees of discrepancies in sensitivity and 204
specificity) were used. 205
Statistical analysis 206
Unless otherwise indicated, categorical data are given as counts (percentages), and 207
continuous data are presented as median (interquartile range). Total test errors were 208
compared by Mann-Whitney tests or, in case they were paired, by Wilcoxon tests. 95% 209
confidence intervals (CI) for sensitivities and specificities were calculated according to 210
Wilson, 95% CI for predictive values were computed according to Mercaldo-Wald unless 211
otherwise indicated. Sensitivities and specificities were compared using z-scores. P 212
values <0.05 were considered statistically significant. All calculations were performed 213
using Analyse-it 5.66 (Analyse-it Software, Leeds, UK) and MedCalc 19.6 (MedCalc 214
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
11
bvba, Ostend, Belgium). Graphs were drawn using Microsoft Visio (Armonk, USA) and 215
GraphPad Prism 7.0 (La Jolla, USA). 216
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
12
Results
217
In the derivation cohort of 1,117 pre-pandemic sera and 64 sera from convalescent 218
COVID-19 patients (80% non-hospitalized, 20% hospitalized), the Roche NC, Abbott 219
NC, and DiaSorin S1/S2 antibody assays gave rise to 7/64, 10/64, and 11/64 false-220
negative, as well as to 3/1,117, 9/1,117, and 20/1,117 false-positive results. Assuming a 221
seroprevalence of 20%, this led to 2180, 3120, and 3440 false-negative results per 222
100,000 tests, and 240, 650 and 1,440 false-positive results per 100,000 tests 223
respectively (Fig. 2A, right panel). 224
Effects of threshold lowering on Sensitivity and Specificity 225
Lowering the positivity thresholds for the Roche NC, Abbott NC, and Diasorin S1/S2 to 226
0.165 COI, 0.55 Index and 9 AU/mL increased the sensitivity significantly and reduced 227
false-negative results to 63/64, 62/64, and 57/64 (320, 620, and 2,180 per 100,000 tests 228
at a seroprevalence of 20%), but substantially increased false-positive results to 229
18/1,117, 27/1,117, and 31/1,117, respectively (1,280, 1,920 and 2,240 per 100,000 230
tests, an assumed seroprevalence of 20%; Table S4, Fig. 2A, right panel). 231
Classical OTA compared to SIT2 232
Subsequently, we evaluated 12 OTA combinations using the fully automated SARS-233
CoV-2 antibody tests from Roche NC, Abbott NC, and DiaSorin S1/S2 as screening 234
tests, each combined with one of the other fully automated assays or a commercially 235
available NC or RBD-specific ELISA as a confirmation test. Combining these tests as 236
classical OTAs significantly increased specificity and reduced false positives to 0 (0-237
1)/1,117. However, the rate of false negatives was 14 (12-16)/64 (1,095 [955-1,230] per 238
100,000 tests at 20% seroprevalence), and therefore considerably higher than for single 239
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
13
testing strategies. In contrast, the SIT2 algorithm minimized false positives to 0 (0-240
2)/1,117 (0 [0-140) per 100,000 tests at 20% seroprevalence) while also reducing false 241
negatives to 5 (3-8)/64 (1,560 [940-2420] per 100,000 tests at 20% seroprevalence, Fig. 242
2A right panel; Table S5). 243
Reduction of total error rates by the Sensitivity-Improved Two-Test 244
Of all the methods assessed, SIT2 reached the lowest total error rates per 100,000 tests 245
under both 5% and 20% assumed seroprevalence (455 [235-685] and 1,600 [940-2,490] 246
per 100,000 tests) (Fig. 2B). At a seroprevalence of 5 %, OTA on average performed 247
better than individual tests, and the total error rates of the single tests were higher for 248
the Abbott NC and DiaSorin S1/S2 assay (OTA 1,095 [955-1,325] vs. 830 [Roche NC], 249
1,540 [Abbott NC] and 2,570 [DiaSorin S1/S2] per 100,000 tests). But with a 250
seroprevalence of 20 %, performance of OTAs, worsened compared to single tests 251
(OTA 4,380 [3,820-5,000] vs 1,600 [Roche], 2,540 [Abbott] and 4,420 [DiaSorin] per 252
100,000 tests) (Fig. 2B). Therefore, at both 5% and 20% seroprevalence, SIT2 resulted 253
in the lowest overall errors. Compared to OTAs, SIT2 yielded a similar improvement in 254
specificity while not suffering from the significant sensitivity reduction (Fig. S2). Since the 255
better overall performance of SIT2 compared to OTAs was not due to increased 256
specificity but improved sensitivity, we subsequently set out to examine these 257
differences in more detail. 258
Sensitivities of single tests, OTA and SIT2 in relation to Neutralization Testing 259
Next, we compared the sensitivities of the three screening tests as single tests and in 260
both two-test methods (OTA and SIT2), benchmarking them using the Austrian 261
sensitivity cohort (n=64) simultaneously evaluated with an authentic SARS-CoV-2 virus 262
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
14
neutralization test (VNT). Regardless of the screening test used (Roche NC, Abbott NC, 263
or DiaSorin S1/S2), OTAs had lower sensitivities than single tests (80.5% [78.5-83.6], 264
78.1% [75.8-82.8], or 75.8% [71.5-78.9] vs. 89.1%, 84.4%, or 82.8% respectively), and 265
SIT2 showed the best sensitivities of all methods (95.3% [93.0-96.5], 93.8% [92.2-96.5], 266
or 87.5% [85.1-88.7]) (Fig. 3). SIT2 algorithms, including the Roche NC and Abbott NC 267
assays, achieved similar or even higher sensitivities than VNT (Fig. 3, VNT reference 268
line), made possible by the unique re-testing zone of SIT2 (Fig. S3). 269
Validation of the Sensitivity-Improved Two-Test using an independent cohort 270
To confirm the improved sensitivity of SIT 2 compared to OTA, we analyzed the 271
sensitivities of OTAs and SIT2 in an independent validation cohort of 976 pre-pandemic 272
samples and 536 post-COVID samples. Out of 20 combinations using the assays Roche 273
NC (total antibody), Abbott NC (IgG), DiaSorin S1/S2 (IgG), Siemens RBD (total 274
antibody), and Oxford trimeric-S (IgG), a statistically significant improvement in 275
sensitivities over OTAs was shown for SIT2 in 18 combinations (Fig. 4). The 276
performance was comparable for the remaining two combinations (Siemens RBD with 277
Oxford trimeric-S and vice versa). Still, no statistically significant improvement could be 278
achieved due to the high pre-existing sensitivities of these assays on this particular 279
sample cohort. 280
To further illustrate the effect of SIT2 on the outcome of SARS-CoV-2 antibody testing, 281
we compared single testing versus SIT2 with the Abbott and DiaSorin assays at varying 282
assumed seroprevalences (5, 10, 20, and 50%), given that the Abbott NC assay is a 283
highly specific (99.9%), but moderately sensitive test (92.7%), and the DiaSorin S1/S2 284
assay has the most limited specificity (98.7%) of all evaluated assays but an acceptable 285
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
15
sensitivity (96.3%). Regardless of whether a lack of specificity (DiaSorin S1/S2) or 286
sensitivity (Abbott NC) had to be compensated for, SIT2 improved the overall error rate 287
compared to the individual tests in all four combinations and at all four assumed 288
seroprevalence levels (Fig. 5). 289
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
16
Discussion
290
Serology is a commonly used, multi-purpose analytical method (1-4). However, not all 291
serologic assays have appropriate sensitivities and specificities, especially in low-292
prevalence settings. The SARS-CoV-2 pandemic prompted the simultaneous 293
development of several antibody tests and, which is rare otherwise, allowed to evaluate 294
these tests with both confirmed positive and negative cases, the latter derived from 295
biobank collections established before the virus emerged. In the case of SARS-CoV-2, 296
false-positive samples are usually not simultaneously reactive in different test systems 297
(14, 25). This led to the hypothesis that reducing the threshold for positivity in screening 298
and confirmation tests would increase the specificity without impairing the sensitivity. A 299
further improvement in sensitivity was possible by defining a high cut-off for the 300
screening test, above which, due to the excellent reliability of high test results, no further 301
confirmation (and, thereby, a possible false-negative result in the confirmation test) was 302
necessary. 303
In the early waves of the SARS-CoV-2 pandemic, many commercially available SARS-304
CoV-2 antibody tests did not provide sufficient specificity to achieve acceptable positive 305
predictive values (PPVs), for example, at a seroprevalence rate of 1-5% (13, 25). 306
Lowering positivity thresholds might improve test sensitivity (21-23) and conventional 307
orthogonal testing can maximize specificity (26-28). The latter might increase the 308
positive predictive value, but PPV will only be relevantly increased at low 309
seroprevalences. However, since seroprevalence is often neither known and varies 310
widely from region to region, it is difficult to judge whether a less specific or less 311
sensitive test is the lesser of two evils. 312
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
17
However, the reported imperfect specificities and sensitivities of SARS-CoV-2 antibody 313
assays are still considerably higher than those of many other serological tests used in 314
clinical routine. Based on actual data related to SARS-CoV-2, which are in principle 315
generalizable for all types of serological tests, we, therefore, we propose a new, 316
universally adaptable two-test system that could, in the case of SARS-CoV-2, perform 317
better than any other known approach regardless of the actual seroprevalence: the 318
sensitivity-improved Two-Test or SIT2. Its generalizability can be inferred from the 319
following features: i) the adapted cut-offs used to optimize sensitivity were determined in 320
various independent studies and were not explicitly calculated for our cohort (21-23); ii) 321
SIT2 was effective, albeit with different efficiencies, in a total of 32 different test 322
combinations; and iii) SIT2 was successfully validated in an independent cohort which 323
was profoundly different from the derivation cohort. Therefore, SIT2 does not require a 324
particular infrastructure or the availability of high-performance individual test systems but 325
achieves the best performance out of an available test. 326
Our SIT2 strategy can rescue the specificity with minimal repeat testing required (see 327
Table S6). For example, when applying the Roche NC as a screening test to our cohort, 328
only 27 out of 1,181 samples needed confirmation testing with the Abbott NC test to 329
correctly identify 62/64 true positives. Simultaneously, all false-positive results were 330
eliminated, including those added by lowering the cut-offs (Table S4 and Fig. S1). 331
Additionally, it was more sensitive than virus neutralization testing, which identified only 332
60/64 clinical positives (Fig. 3). This result is not completely surprising as it is known that 333
not all patients who recovered from COVID-19 show detectable levels of neutralizing 334
antibodies (29). Nevertheless, it should be noted that although antibody binding assays 335
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
18
may have a higher sensitivity than neutralization assays, they only partially reflect the 336
functional activity of SARS-CoV-2-reactive antibodies(30). 337
Our study has both strengths and limitations. One strength is the size of the cohorts 338
examined, both in deriving the SIT2 algorithm (N=1,181) and validating it (N=1,512). The 339
composition of our specificity cohort is also unique: it consists of three sub-cohorts with 340
selection criteria to further challenge analytical specificity. The lower cut-offs used to 341
increase sensitivity were not modeled within our datasets but were derived from ROC-342
analyses data of independent studies (21-23). Furthermore, we were able to test the 343
performance of the two-test systems in a total of 32 combinations, 12 in the derivation 344
cohort and another 20 combinations in the validation cohort. As a limitation, in the 345
Austrian cohort, only samples ≥ 14 days after symptom onset were included. Therefore, 346
no conclusions on the sensitivity of the early seroconversion phase can be made from 347
these data. Furthermore, mild and asymptomatic cases were underrepresented in the 348
British cohort, perhaps leading to an observed higher sensitivity of the test systems. 349
In conclusion, we describe the novel two-test algorithm SIT2, which makes it possible to 350
maintain or even significantly improve sensitivity while approaching 100% specificity. 351
Acknowledgments 352
We sincerely thank Marika Gerdov, Susanne Keim, Karin Mildner, Elisabeth Ponweiser, 353
Manuela Repl, Ilse Steiner, Christine Thun, and Martina Trella for excellent technical 354
assistance. Finally, we want to thank all the donors of the various study cohorts. The 355
MedUni Wien Biobank is funded to participate in the biobank consortium BBMRI.at 356
(www.bbmri.at) by the Austrian Federal Ministry of Science, Research and Technology. 357
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
19
There was no external funding received for the work presented. However, test kits for 358
the Technoclone ELISAs were kindly provided by the manufacturer. All authors have 359
read the journal’s policy on disclosure of potential conflicts of interest and the following 360
has to be disclosed: NP-N received a travel grant from DiaSorin. DWE reports lecture 361
fees from Gilead outside the submitted work. OCB reports grants from GSK, grants from 362
Menarini, grants from Boehringer Ingelheim, grants from Astra, grants from MSD, grants 363
from Pfizer, and grants from Chiesi, outside the submitted work. SH does receive 364
unrestricted research grants (GSK, Boehringer, Menarini, Chiesi, Astra Zeneca, MSD, 365
Novartis, Air Liquide, Vivisol, Pfizer, TEVA) for the Ludwig Boltzmann Institute of COPD 366
and Respiratory Epidemiology, and is on advisory boards for G. SK, Boehringer 367
Ingelheim, Novartis, Menarini, Chiesi, Astra Zeneca, MSD, Roche, Abbvie, Takeda, and 368
TEVA for respiratory oncology and COPD. PQ is an advisory board member for Roche 369
Austria and reports personal fees from Takeda outside the submitted work. The Dept. of 370
Laboratory Medicine (Head: OWF) received compensations for advertisement on 371
scientific symposia from Roche, DiaSorin, and Abbott and holds a grant for evaluating 372
an in-vitro diagnostic device from Roche. CJB is a Board Member of Technoclone. HH 373
receives compensations for biobank services from Glock Health Science and Research 374
and BlueSky immunotherapies. 375
References
376
1. Lass-Flörl C, Samardzic E, Knoll M. Serology anno 2021-fungal infections: from 377
invasive to chronic. Clin Microbiol Infect. 2021;27(9):1230-41. 378
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
20
2. Popp A, Kivelä L, Fuchs V, Kurppa K. Diagnosing Celiac Disease: Towards Wide-379
Scale Screening and Serology-Based Criteria? Gastroenterol Res Pract. 380
2019;2019:2916024. 381
3. Hibbert J, Halec G, Baaken D, Waterboer T, Brenner N. Sensitivity and Specificity 382
of Human Papillomavirus (HPV) 16 Early Antigen Serology for HPV-Driven 383
Oropharyngeal Cancer: A Systematic Literature Review and Meta-Analysis. Cancers 384
(Basel). 2021;13(12). 385
4. Smatti MK, Al-Sadeq DW, Ali NH, Pintus G, Abou-Saleh H, Nasrallah GK. 386
Epstein-Barr Virus Epidemiology, Serology, and Genetic Variability of LMP-1 Oncogene 387
Among Healthy Population: An Update. Front Oncol. 2018;8:211. 388
5. Gaudino M, Moreno A, Snoeck CJ, Zohari S, Saegerman C, O'Donovan T, et al. 389
Emerging Influenza D virus infection in European livestock as determined in serology 390
studies: Are we underestimating its spread over the continent? Transbound Emerg Dis. 391
2021;68(3):1125-35. 392
6. Shibata Y, Horita N, Yamamoto M, Tsukahara T, Nagakura H, Tashiro K, et al. 393
Diagnostic test accuracy of anti-glycopeptidolipid-core IgA antibodies for Mycobacterium 394
avium complex pulmonary disease: systematic review and meta-analysis. Sci Rep. 395
2016;6:29325. 396
7. Arora RK, Joseph A, Van Wyk J, Rocco S, Atmaja A, May E, et al. SeroTracker: a 397
global SARS-CoV-2 seroprevalence dashboard. Lancet Infect Dis. 2020. 398
8. Bailey D, Konforte D, Barakauskas VE, Yip PM, Kulasingam V, Abou El Hassan 399
M, et al. Canadian society of clinical chemists (CSCC) interim consensus guidance for 400
testing and reporting of SARS-CoV-2 serology. Clin Biochem. 2020. 401
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
21
9. Turbett SE, Anahtar M, Dighe AS, Garcia Beltran W, Miller T, Scott H, et al. 402
Evaluation of Three Commercial SARS-CoV-2 Serologic Assays and their Performance 403
in Two-Test Algorithms. J Clin Microbiol. 2020. 404
10. Gattinger P, Borochova K, Dorofeeva Y, Henning R, Kiss R, Kratzer B, et al. 405
Antibodies in serum of convalescent patients following mild COVID-19 do not always 406
prevent virus-receptor binding. Allergy. 2020. 407
11. Long QX, Tang XJ, Shi QL, Li Q, Deng HJ, Yuan J, et al. Clinical and 408
immunological assessment of asymptomatic SARS-CoV-2 infections. Nat Med. 409
2020;26(8):1200-4. 410
12. Baron RC, Risch L, Weber M, Thiel S, Grossmann K, Wohlwend N, et al. 411
Frequency of serological non-responders and false-negative RT-PCR results in SARS-412
CoV-2 testing: a population-based study. Clin Chem Lab Med. 2020. 413
13. Manthei DM, Whalen JF, Schroeder LF, Sinay AM, Li SH, Valdez R, et al. 414
Differences in Performance Characteristics Among Four High-Throughput Assays for the 415
Detection of Antibodies Against SARS-CoV-2 Using a Common Set of Patient Samples. 416
Am J Clin Pathol. 2020. 417
14. Pfluger LS, Bannasch JH, Brehm TT, Pfefferle S, Hoffmann A, Norz D, et al. 418
Clinical evaluation of five different automated SARS-CoV-2 serology assays in a cohort 419
of hospitalized COVID-19 patients. J Clin Virol. 2020;130:104549. 420
15. Buss LF, Prete CA, Jr., Abrahim CMM, Mendrone A, Jr., Salomon T, de Almeida-421
Neto C, et al. Three-quarters attack rate of SARS-CoV-2 in the Brazilian Amazon during 422
a largely unmitigated epidemic. Science. 2021;371(6526):288-92. 423
16. Klausberger M, Duerkop M, Haslacher H, Wozniak-Knopp G, Cserjan-424
Puschmann M, Perkmann T, et al. A comprehensive antigen production and 425
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
22
characterisation study for easy-to-implement, specific and quantitative SARS-CoV-2 426
serotests. EBioMedicine. 2021;67:103348. 427
17. Eskandary F, Regele H, Baumann L, Bond G, Kozakowski N, Wahrmann M, et al. 428
A Randomized Trial of Bortezomib in Late Antibody-Mediated Kidney Transplant 429
Rejection. J Am Soc Nephrol. 2018;29(2):591-605. 430
18. Perkmann T, Perkmann-Nagele N, Breyer MK, Breyer-Kohansal R, Burghuber 431
OC, Hartl S, et al. Side-by-Side Comparison of Three Fully Automated SARS-CoV-2 432
Antibody Assays with a Focus on Specificity. Clin Chem. 2020;66(11):1405-13. 433
19. Breyer-Kohansal R, Hartl S, Burghuber OC, Urban M, Schrott A, Agusti A, et al. 434
The LEAD (Lung, Heart, Social, Body) Study: Objectives, Methodology, and External 435
Validity of the Population-Based Cohort Study. J Epidemiol. 2019;29(8):315-24. 436
20. National S-C-SAEG. Performance characteristics of five immunoassays for 437
SARS-CoV-2: a head-to-head benchmark comparison. Lancet Infect Dis. 438
2020;20(12):1390-400. 439
21. Favresse J, Eucher C, Elsen M, Tre-Hardy M, Dogne JM, Douxfils J. Clinical 440
Performance of the Elecsys Electrochemiluminescent Immunoassay for the Detection of 441
SARS-CoV-2 Total Antibodies. Clin Chem. 2020;66(8):1104-6. 442
22. Lau CS, Oh HML, Hoo SP, Liang YL, Phua SK, Aw TC. Performance of an 443
automated chemiluminescence SARS-CoV-2 IG-G assay. Clin Chim Acta. 444
2020;510:760-6. 445
23. Bonelli F, Sarasini A, Zierold C, Calleri M, Bonetti A, Vismara C, et al. Clinical and 446
Analytical Performance of an Automated Serological Test That Identifies S1/S2-447
Neutralizing IgG in COVID-19 Patients Semiquantitatively. J Clin Microbiol. 2020;58(9). 448
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
23
24. Pfluger LS, Bannasch JH, Brehm TT, Pfefferle S, Hoffmann A, Norz D, et al. 449
Clinical evaluation of five different automated SARS-CoV-2 serology assays in a cohort 450
of hospitalized COVID-19 patients. J Clin Virol. 2020;130:104549. 451
25. Perkmann T, Perkmann-Nagele N, Breyer MK, Breyer-Kohansal R, Burghuber 452
OC, Hartl S, et al. Side by side comparison of three fully automated SARS-CoV-2 453
antibody assays with a focus on specificity. Clin Chem. 2020. 454
26. Turbett SE, Anahtar M, Dighe AS, Garcia Beltran W, Miller T, Scott H, et al. 455
Evaluation of Three Commercial SARS-CoV-2 Serologic Assays and Their Performance 456
in Two-Test Algorithms. J Clin Microbiol. 2020;59(1). 457
27. Bolotin S, Tran V, Osman S, Brown KA, Buchan SA, Joh E, et al. SARS-CoV-2 458
seroprevalence survey estimates are affected by anti-nucleocapsid antibody decline. J 459
Infect Dis. 2021. 460
28. Irsara C, Egger AE, Prokop W, Nairz M, Loacker L, Sahanic S, et al. Evaluation of 461
four commercial, fully automated SARS-CoV-2 antibody tests suggests a revision of the 462
Siemens SARS-CoV-2 IgG assay. Clin Chem Lab Med. 2021;0(0). 463
29. Kalkan Yazici M, Koc MM, Cetin NS, Karaaslan E, Okay G, Durdu B, et al. 464
Discordance between Serum Neutralizing Antibody Titers and the Recovery from 465
COVID-19. J Immunol. 2020;205(10):2719-25. 466
30. Padoan A, Bonfante F, Pagliari M, Bortolami A, Negrini D, Zuin S, et al. Analytical 467
and clinical performances of five immunoassays for the detection of SARS-CoV-2 468
antibodies in comparison with neutralization activity. EBioMedicine. 2020;62:103101. 469
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
24
Figure Legends 470
Fig. 1. A) The Sensitivity Improved Two-Test (SIT2) algorithm includes sensitivity 471
improvement by adapted cut-offs and a subsequent specificity rescue by re-testing all 472
samples within the re-testing zone of the screening test by a confirmatory test. B) 473
Testing algorithm for SIT2 utilizing a screening test on an automated platform 474
(ECLIA/Roche, CMIA/Abbott, CLIA/DiaSorin) and a confirmation test, either on one of 475
the remaining platforms or tested by means of ELISA (Technozym RBD, NP). C) All test 476
Results
between a reduced cut-off suggested by the literature, and a higher cut-off, 477
above which no more false-positives were observed, were subject to confirmation 478
testing. **… results between 12.0 and 15.0, which are according to the manufacturer 479
considered borderline, were treated as positives; ***… suggested as a cut-off for 480
seroprevalence testing; ****… determined by in-house modeling; 1… see (21); 2… see 481
(22); 3… see (23). 482
Fig. 2. False-positives (FP)/false-negatives (FN) (A) and total error (B) of single tests, 483
tests with reduced thresholds according to (21-23), orthogonal testing algorithms (OTAs) 484
and the Sensitivity Improved Two-Test (SIT2) algorithm at 5 and 20% estimated 485
seroprevalence. Data in (B) were compared by Mann-Whitney tests (unpaired) or 486
Wicoxon tests (paired). *… P<0.05; **...P<0.01; ***…P<0.001. 487
Fig. 3. Sensitivities of single tests, orthogonal testing algorithms (OTAs) and the 488
Sensitivity Improved Two-Test (SIT2) algorithm. The dotted line indicates the sensitivity 489
of virus neutralization test (VNT). 490
Fig. 4. Differences in sensitivity and specificity (mean±95% confidence interval) between 491
the Sensitivity Improved Two-Test (SIT2) algorithm and standard orthogonal testing 492
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
25
algorithms (OTAs) within the UK validation cohort. *… P<0.05; **...P<0.01; 493
***…P<0.001; ****...P<0.0001 494
Fig. 5. Comparing false-positives (FP), false-negatives (FN), and total error (TE) for two 495
selected test systems, A) Abbott, B) DiaSorin, between different Sensitivity Improved 496
Two-Test (SIT2) combinations and the respective single test within the UK validation 497
cohort for different estimated seroprevalences. 498
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
All rights reserved. No reuse allowed without permission.
(which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.
The copyright holder for this preprintthis version posted January 12, 2022. ; https://doi.org/10.1101/2020.11.05.20226449doi: medRxiv preprint
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.