Methods
A search of the online databases PubMed, Scopus and Google Scholar was performed from inception to 30 October 2022 to ascertain whether a suitable scoring system for TVS IQ assessment exists. No such tool was found, and subsequently, a review of the literature was conducted to identify any other published IQ assessment tools for ultrasound imaging. The same online databases were searched again from inception to 30 October 2022 using the terms ‘ultrasound’, ‘scoring’ OR ‘grading’, ‘assessment’ and ‘quality’. Returned items were screened by title and abstract for relevance. Those considered likely to be relevant were then read in full. This ultimately revealed 15 studies detailing scoring systems related to ultrasound imaging (Table 1 )
4
,
5
,
6
,
7
,
8
,
9
,
10
,
11
,
12
,
13
,
14
,
15
,
16
,
17
,
18
. Based on these identified systems published previously, a new draft scoring system based on anatomy demonstration and technical optimization of images was created to quantify IQ assessment of gynecological TVS images. This system was then reviewed by four experts in gynecological ultrasound (G.C., M.L., S.K. and C.P.) with discussion around the suitability of each score until agreement was reached on score descriptions. The draft scoring system was then reviewed by an expert in the development of AI tools for medical imaging (H.‐T.C.) to ensure its suitability for use in the design of AI tools. A final review from all experts was then conducted to create the new scoring system. Details of the various iterations of the scoring system throughout this process are provided in Appendix S1 .
Summary of 15 published studies on ultrasound image‐quality (IQ) scoring systems identified through literature search
Only the first author is given for each study.
eFAST, extended focused assessment with sonography for trauma; POCUS, point‐of‐care ultrasound.
The proposed scoring system is given in Table 2 . If an image was deemed incorrect/inaccurate (e.g. blank image or transabdominal rather than transvaginal image required), the image was rejected (score of 1). All other images were assessed against the five factors listed in Table 2 relating to the following: correct depiction of anatomy; view of the anatomical structure in the field of view; image optimization (depth, focus and gain); ability to interpret image for diagnosis of pathology; and overall clarity of the image. A score of 1–4 that, in the opinion of the rater, best represented the image, was allocated for each image, where 1 = reject (image incorrect), 2 = poor IQ, 3 = suboptimal IQ and 4 = optimal IQ. A selection of images representing scores 2–4 are shown in Figure 1 .
Proposed transvaginal ultrasound image‐quality (IQ) scoring system
Images are assessed by the five factors listed and assigned a score of 1–4 that best represents the image; a score of 4 represents optimal IQ while a score of 2 represents poor IQ.
N/A, not applicable.
Example of ovarian ultrasound images with poor image quality (IQ) (score 2) (a), suboptimal IQ (score 3) (b) and optimal IQ (score 4) (c). A score of 1 was allocated to images deemed incorrect; these images were then rejected. (a) It is difficult to use this image for diagnosis as the ovary cannot be identified confidently, with the entire ovary not seen in the image, and optimization is poor. As the overall IQ is poor, a score of 2 is assigned. (b) The ovary can be seen. With the entire ovary visualized, optimization is satisfactory, but it could be improved, and diagnosis should be possible from the image. As the overall IQ is satisfactory, a score of 3 is assigned. (c) The entire ovary is documented clearly, optimization is good, and this image can be interpreted easily for diagnosis. As the overall IQ is good, a score of 4 is assigned. Lt, left.
To assess the intra‐ and interobserver reliability, IQ scoring was performed independently by six imaging professionals, paired by specialization: two gynecologists with a subspecialization in ultrasound (gynecological sonologists) (G.C. and M.L.), two sonographers (C.P. and S.M.) and two radiologists with subspecialization in gynecological imaging (S.K. and G.L.). All professionals have a high level of expertise in the performance and/or offline interpretation of TVS images, with a minimum experience of 6 years.
There were 50 cases contributing three still images each (no cineclips), totaling 150 images for the image review. The images reviewed and scored were midsagittal images of the uterus ( n = 50), the right ovary ( n = 50) and the left ovary ( n = 50) (Figure 2 ). The images were obtained from a large, retrospectively collected data library obtained as part of a wider study, the IMAGENDO® study (trial registration number: ACTRN12623000646640). Images were from a range of ultrasound clinics in Australia and were acquired on ultrasound machines from various equipment manufacturers. All images reviewed were two‐dimensional B‐mode single‐screen images, free of measurement calipers and color Doppler boxes. Images did not contain any data related to patient identifiers, such as the clinic in which the image was obtained or the date of the examination. Details of the ultrasound machine used to obtain the images were removed from the images as much as was practicable. The reviewed images were prepared by an independent investigator (A.D.). The reviewers were blinded to any other images from the scan, clinical history, sonographer's report (i.e. technical limitations of the real‐time scan) or formal examination report. Reviewed images were uploaded to an online survey platform (Survey Monkey, 2023; Momentive.ai, San Mateo, CA, USA) for data capture. Within Survey Monkey, each of the six raters reviewed every image and allocated a score for each one, according to the scoring system in Table 2 , from a multiple‐choice selection. All images were viewed by each rater in unique locations and on personal devices. A free‐text box was included to capture any comments the raters had concerning any technical limitations they believed may have contributed to IQ. If a score of 1, 2 or 3 was assigned, raters were instructed to add free‐text comments detailing factors they believe may have contributed to reduced IQ. The image review process was repeated by each rater after an interval of at least 1 week, for a total of two scores per image for each rater. This study was conducted as part of the IMAGENDO® study for which ethics approval was obtained from the University of Adelaide Human Research Ethics Committee (protocol number: H‐2020‐051).
Example of one case (three ultrasound images) reviewed by raters. For each case, ultrasound images in the midsagittal plane of the uterus (a), right ovary (b) and left ovary (c) were reviewed. All images reviewed were single‐screen images, free of measurement calipers and color Doppler boxes. Anatomical annotations remained on images; however, information that could identify the patient or clinic from which the image was acquired was removed before starting the study. Lt, left; Rt, right.
Interobserver agreement was assessed using the intraclass correlation coefficient (ICC), which was calculated to measure levels of agreement between all six raters, where 0 = no agreement and 1 = perfect agreement (< 0.5 = poor agreement, 0.5–0.75 = moderate agreement, 0.75–0.9 = good agreement and 0.9 = excellent agreement)
25
. ICC estimates were based on a single rater value of k = 6, indicating absolute agreement, using a two‐way mixed‐effects model. The interobserver agreement of the paired professionals was also assessed using ICC and weighted Cohen's kappa coefficient, where kappa 0.90 = almost perfect agreement
26
.
Median, mode and mean scores from each rater at each review session were calculated. Weighted Cohen's kappa and ICC values were calculated to assess the intraobserver agreement for each rater, over the two review sessions. A biostatistician was consulted to verify that this analysis method was satisfactory. SPSS for Mac version 29.0.2.0 (IBM Corp., Armonk, NY, USA) was used for ICC calculations. A P ‐value < 0.05 was considered statistically significant. A sample size calculation was performed using an online sample size calculator ( http://wnarifin.github.io ), which confirmed that a minimum sample of 45 images was required for a minimum acceptable ICC of 0.5 and an expected ICC of 0.75 with a significance level of 0.05 to achieve a power of 80%
27
.
Results
Table 3 presents the interobserver agreement achieved among the six raters and that achieved when raters were paired by professional specialization. Poor levels of agreement were obtained between the six raters for all 150 images (ICC, 0.480 (95% CI, 0.363–0.586)), as well as when images were grouped into uterine images (ICC, 0.359 (95% CI, 0.204–0.523)). Moderate levels of agreement between all raters were achieved for images of the ovaries (ICC, 0.531 (95% CI, 0.417–0.636)). Agreement between the paired sonographers and sonologists was poor for all images (ICC, 0.336 (95% CI, −0.078 to 0.619) and 0.425 (95% CI, 0.014–0.665), respectively), as well as when images were grouped into uterine images (ICC, 0.253 (95% CI, −0.097 to 0.577) and 0.299 (95% CI, −0.094 to 0.606), respectively) and ovarian images (ICC, 0.400 (95% CI, −0.043 to 0.669) and 0.469 (95% CI, 0.088–0.689), respectively). Moderate levels of agreement were achieved between the paired radiologists for all images overall (ICC, 0.600 (95% CI, 0.487–0.693)) and for images of the uterus (ICC, 0.538 (95% CI, 0.311–0.707)) and ovaries (ICC, 0.621 (95% CI, 0.483–0.728)). All measures were statistically significant with a P ‐value < 0.001.
Interobserver agreement among all six raters and according to their professional group for assessment of image quality of 150 ultrasound images, overall and grouped by uterine and ovarian images
Values in parentheses are 95% CI.
ICC, intraclass correlation coefficient.
Intraobserver agreement is shown in Table 4 . Weighted Cohen's kappa revealed weak‐to‐moderate agreement for each of the raters for all images (Cohen's kappa ranged from 0.533 to 0.718) and for images of the ovaries (Cohen's kappa ranged from 0.467 to 0.751). Slightly better agreement was seen for uterine images, with weighted Cohen's kappa ranging from 0.568 to 0.808, indicating weak‐to‐strong agreement. The ICC revealed similar results, with moderate‐to‐good intraobserver agreement for all raters for all images overall (ICC ranged from 0.636 to 0.825), as well as for images of the uterus (ICC ranged from 0.546 to 0.893) and ovaries (ICC ranged from 0.596 to 0.862). All measures were statistically significant with a P ‐value < 0.001.
Intraobserver agreement of six raters for assessment of image quality of 150 ultrasound images, overall and grouped by uterine and ovarian images
Values in parentheses are 95% CI.
P < 0.001 was achieved for all measures.
ICC, intraclass correlation coefficient.
Table S1 shows the median, mode and mean scores assigned by each rater at the first and second review sessions. Overall, the level of criticism exercised by the raters was consistent across both sessions. However, three raters (G.C., S.M. and C.P.) assigned slightly higher mean scores across both the uterine and ovarian images on their second review. The spread of mean scores between raters across the two sessions (first review: mean scores ranged from 2.6 to 3.5; second review: mean scores ranged from 2.8 to 3.5) suggests that certain raters were more critical of IQ while other raters were more lenient.
Discussion
This study revealed that the interobserver agreement for assigning quantitative scores to TVS images using our scoring system ranged from poor to moderate. Meanwhile, intraobserver reliability was slightly better, demonstrating moderate‐to‐good ICC and weak‐to‐strong weighted Cohen's kappa values. All statistical measures were significant ( P < 0.001), indicating that these variations were unlikely to be due to random chance. This is the first intra‐ and interobserver study within the literature assessing the performance of sonographers, sonologists and radiologists in the scoring of the IQ of gynecological TVS images.
The scoring system we created, which was designed to function like a task‐related checklist
28
,
29
, was intended to assess IQ objectively
10
,
13
,
28
. However, our results suggest this system may not have been well‐enough defined, contributing to subjective interpretation among raters. The subjective nature of assessing ultrasound IQ by human operators is a well‐documented challenge
10
,
19
,
30
,
31
. For instance, Papageorghiou et al .
31
reported that experts agreed on the accuracy of fetal ultrasound images (as they adhered to established protocols) only 79.4% of the time. Although both Molloholli et al .
10
and Wanyonyi et al .
30
reported that utilizing an objective IQ scoring system resulted in better agreement between professionals than subjective assessment, other studies reported an element of subjectivity concerning IQ assessment, even with the utilization of a structured scoring system
13
. The inherent subjectivity in IQ assessment may explain our intraoperator results, which, although better than our interoperator agreement, were weaker than expected.
In the present study, we created a new IQ scoring system, as no published system existed within the literature that could be broadly applied suitably to TVS images. While three scoring systems were found within the literature relating to gynecological ultrasound
6
,
11
,
18
, these were either too specific to a single application
6
,
11
or assessed operator competence rather than IQ
18
. Our aim to make the scoring system applicable across various TVS scenarios may have affected its objectivity. While some similar systems in obstetric ultrasound have shown good results
10
,
12
,
18
,
30
,
32
,
33
,
34
, these are unlikely to translate into gynecological ultrasound, where normal anatomical appearances are more variable. Further refinement of the scoring instructions may improve levels of agreement in the future.
In addition to assessing IQ for quality assurance, IQ scoring systems can aid in teaching, training and providing feedback to professionals
11
,
12
,
32
,
34
, with more experienced operators shown to produce higher scores using several published systems
4
,
5
,
11
,
17
,
18
. The utility of a scoring system as a teaching tool was highlighted in the study of Salomon et al .
11
, which found that scores improved significantly when trainee doctors performing gynecological ultrasound scans in the emergency department were given additional training. It remains unknown whether our system would be suitable for use as a training tool, which is an area for further research.
AI systems offer significant advantages over human assessment of IQ because they are not affected by human factors, such as fatigue
3
. The development of such systems, however, is not without difficulty. First, the subjectivity of IQ assessment often results in noisy image labeling, a problem AI tools typically struggle to manage
35
. Furthermore, AI tools need high volumes of labeled data, requiring precise image review by experts, such as the method undertaken in this study, which is both labor‐intensive and costly
12
,
36
. To address these issues, innovative data‐labeling strategies need to be developed.
Most of the scoring systems we reviewed (Table 1 ) assessed the skill and competency of the operator as a means of ultrasound quality assurance, rather than the IQ of documented images in isolation
4
,
7
,
9
,
13
,
15
,
16
,
17
. In clinical practice, the skill and competence of sonographers are essential for ensuring examination accuracy and optimal patient outcomes. However, AI models assess ‘data quality’ (i.e. stored images and videos) independent of operator skill. As literature suggests that AI is already playing a significant role in the IQ assessment of other (non‐gynecological) ultrasound applications
19
,
23
, it may be that we are perhaps entering a new paradigm in the assessment of ultrasound IQ, one that balances traditional human‐led methods alongside the emerging role of AI and digital systems.
This study has several strengths. A large dataset of TVS images acquired from various sonographers, clinics and ultrasound machines was used, resulting in statistically significant results. This also provided a variation in ‘data’, which must be considered when developing AI tools for either IQ assessment or image interpretation. Additionally, although all six raters are experienced professionals, by including sonographers, radiologists and sonologists, we were able to evaluate whether different professional perspectives affect how IQ is interpreted.
This work represents the first IQ assessment scoring system related broadly to gynecological imaging, applicable to both clinical practice and the design of AI tools. Despite its novelty, this study has several limitations, such as our decision to construct a scoring system from a review of published literature only, rather than a multicenter expert consensus or Delphi approach. Furthermore, our literature search was confined to studies on ultrasound only. Expanding our search to other imaging modalities reviewed more frequently offline may have identified concepts not explored previously in ultrasound. As all raters viewed images in unique locations on personal devices, environmental factors (e.g. screen settings and lighting) may have impacted image appearance. Although the system design underwent extensive review, we did not perform a pilot test of the system within the development phase, which could have allowed for further refinement and potentially improved the results of this study. Finally, although all raters were provided with uniform, written instructions, no calibration meeting was held prior to the study, which could have resulted in higher levels of agreement.
In conclusion, wide variation in inter‐ and intraobserver agreement was noted for IQ grading of gynecological TVS images utilizing our proposed scoring system. Although AI systems will likely be able to assess IQ more objectively than humans, the development of these systems will depend on human labeling, which will likely have labeling inconsistencies (creating noisy data) due to the inherently subjective nature of ultrasound IQ. As such, reliable methods of IQ scoring are essential for further progress in the development of AI systems in gynecological ultrasound. Additionally, consideration should be given to AI methods capable of accommodating noisy labeling.
Introduction
Accurate interpretation of ultrasound images is reliant upon the acquisition of high‐quality images. Artificial intelligence (AI) algorithms may interpret ultrasound images, yielding a proposed diagnosis, but they are similarly dependent upon high‐quality image acquisition, with poor image quality (IQ) reducing the accuracy of diagnosis
1
,
2
,
3
. For instance, Blaivas et al .
2
reported that the performance of deep learning systems reduced significantly when applied to images from low‐cost ultrasound machines with lower IQ. Therefore, it is prudent to consider AI tools to assess IQ, concurrent with the development of AI algorithms for diagnosis, ensuring AI systems receive data input of satisfactory quality to allow for accurate diagnosis.
Many scoring systems within the published literature propose ways of critically evaluating ultrasound IQ and allocating examinations a score for natural (human‐led) imaging
4
,
5
,
6
,
7
,
8
,
9
,
10
,
11
,
12
,
13
,
14
,
15
,
16
,
17
,
18
. However, many of these systems rate factors that are less relevant when developing AI tools, such as operator skill
4
,
8
,
11
and accuracy of annotations
7
,
9
,
13
. Two systems specific to gynecological ultrasound have been published previously that focus on antral follicle counting
6
and emergency gynecology presentations
11
, but there is currently no scoring system within the published literature for the assessment of transvaginal ultrasound (TVS) IQ.
AI algorithms have been developed to automate IQ assessment in medical imaging
19
,
20
,
21
,
22
. Although such tools can be utilized solely for thresholding, by identifying and then rejecting poor‐quality images
23
, a broader IQ assessment system has the potential to provide wider clinical assistance. Chen et al .
24
developed an AI system to classify intravascular ultrasound images as low, medium or high quality, which achieved a classification accuracy of 96.34%. This system was trained using a library of images whose quality was labeled subjectively by doctors, highlighting the need for suitable IQ labeling strategies in AI algorithm development. Additionally, such systems should be ideally trained on a full spectrum of data (ranging from high to low quality) to learn how to identify good vs poor IQ. This study aimed to propose a scoring system able to quantify the IQ of gynecological TVS images suitable for AI algorithm development and to assess the intra‐ and interobserver agreement of this tool among multiple ultrasound professionals.
Supplementary Material
Appendix S1 Various iterations of our objective scoring system throughout the development process
Table S1 Median, mode and mean scores given by each rater at the first and second review for assessment of ultrasound image quality of all images ( n = 150) and when grouped into uterine (Ut) ( n = 50) and ovarian (Ov) images ( n = 100)