Estimation of Inferior Vena Cava Size from Ultrasound Imaging in X-Plane

preprint OA: closed CC-BY-4.0
🔓 Open OA copy View at publisher

Abstract

Ultrasound (US) scans of the inferior vena cava (IVC) provide useful information on the volume status of a patient. However, their investigation is user-dependent and prone to measurement errors. An important technical problem is the objective difficulty in studying a very compliant blood vessel like IVC, which makes large respirophasic movements and shows a complicated three dimensional geometry. Using bi-dimensional (2D) B-mode views either in long or short axis has improved the characterization of IVC dynamics compared to a measurement along a single direction (M-mode). However, specific movements of the IVC can challenge also the information provided by these 2D sections. Thus, these two orthogonal views, provided by an US system in X-plane, are here integrated by an innovative method. It is tested on simulated videos of the IVC performing complicated movements, that are compensated by the new method, overcoming the biased measurements provided by 2D scans. The method is then applied on example experimental data.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2024) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.

Source provenance

europepmc
last seen: 2026-05-20T01:45:00.602351+00:00
unpaywall
last seen: 2026-05-26T02:00:01.498150+00:00
License: CC-BY-4.0