The Impact of Actuator Nozzle and Surroundings Condition on Drug Delivery using Pressurized-Metered Dose Inhalers

preprint OA: closed CC-BY-4.0
📄 Open PDF View at publisher

Abstract

The most commonly used method to deliver aerosolized drugs to the lung is with pressurized metered-dose inhalers (pMDIs). The spray actuator is a critical component of a pMDI, since it controls the atomization process by forming aerosol plumes and determining droplet size distribution. Through computational fluid dynamics (CFD) simulations, this study investigated the effect of two different nozzle types (single conventional and twin nozzles) on drug deposition in the mouth-throat (MT) region. We also studied the behavior of aerosol plumes in both an open-air environment and the MT geometry. Our study revealed that spray aerosol generated in an unconfined, open-air environment with no airflow behaves distinctly from spray introduced into the MT geometry in the presence of airflow. In addition, the actuator structure significantly impacts the device's efficacy. In the real MT airway, we found that the twin nozzle increases drug deposition in the MT region and its higher aerosol velocity negatively affects its efficiency.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00
unpaywall
last seen: 2026-05-26T02:00:01.498150+00:00
License: CC-BY-4.0