A High-Level Control Algorithm Based on sEMG Signalling for an Elbow Joint SMA Exoskeleton

preprint OA: closed CC-BY-4.0
🔓 Open OA copy View at publisher

Abstract

A high-level control algorithm capable of generating position and torque references from surface electromyography signals (sEMG) has been designed. It is applied to a shape memory alloy (SMA) actuated exoskeleton used in active rehabilitation therapies for elbow joints. The sEMG signals are filtered and normalized according data collected online during the first seconds of~therapy sessions. The control algorithm uses the sEMG signals to promote active participation of patients during the therapy session. In order to generate the position reference pattern with good precision, the sEMG normalized signal is compared with a pressure sensor signal to detect the intention of each movement. The algorithm has been tested in simulations and with healthy people for control of an elbow exoskeleton in flexion–extension movements. The results indicate that sEMG signals from elbow muscles in combination with pressure sensors that measure arm–exoskeleton interaction can be used as inputs for the control algorithm, which adapts the reference for exoskeleton movements according a patient's intention.

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00
unpaywall
last seen: 2026-05-24T02:00:01.246996+00:00
License: CC-BY-4.0