The sensitivity of ECG contamination to surgical implantation site in adaptive neurostimulation

preprint OA: closed
📄 Open PDF View at publisher
AI-generated summary by gemini-2.5-flash-lite, 2026-07-17

Left subclavicular implants cause over three times more ECG contamination than right-sided implants, while cranial implants are unaffected, impacting closed-loop neurostimulation reliability.

One-sentence paraphrase of the abstract; not a substitute for reading it. No clinical advice. How this works

Abstract

Background Brain sensing devices are approved today for Parkinson’s, essential tremor, and epilepsy therapies. Clinical decisions for implants are often influenced by the premise that patients will benefit from using sensing technology. However, artifacts, such as ECG contamination, can render such treatments unreliable. Therefore, clinicians need to understand how surgical decisions may affect artifact probability. Objectives Investigate neural signal contamination with ECG activity in sensing enabled neurostimulation systems, and in particular clinical choices such as implant location that impact signal fidelity. Methods Electric field modelling and empirical signals from 85 patients were used to investigate the relationship between implant location and ECG contamination.a Results The impact on neural recordings depends on the difference between ECG signal and noise floor of the electrophysiological recording. Empirically, we demonstrate that severe ECG contamination was more than 3.2x higher in left-sided subclavicular implants (48.3%), when compared to right-sided implants (15.3%). Cranial implants did not show ECG contamination. Conclusions Given the relative frequency of corrupted neural signals, we conclude that implant location will impact the ability of brain sensing devices to be used for “closed-loop” algorithms. Clinical adjustments such as implant location can significantly affect signal integrity and need consideration. Highlights Chronic embedded brain sensing promises algorithm-based neurostimulation Algorithms for closed-loop stimulation can be impaired by artifacts The relationship of implant location to cardiac dipole has relevant impact on neural signal fidelity; simple models can provide guidance on the sensitivity ECG artifacts are present in up to 50% of neural signals from left subclavicular DBS systems Implanting DBS in a right subclavicular location significantly reduces frequency of ECG artifacts Cranial-mounted implants are relatively immune to artifacts

My notes (saved in your browser only)

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-05-19T01:45:01.086888+00:00
unpaywall
last seen: 2026-09-10T06:36:06.991349+00:00