A new cryo-EM system for electron 3D crystallography by eEFD
preprint
OA: closed
Abstract
A new cryo-EM system has been developed and investigated for use in protein electron 3D crystallography. The system provides parallel illumination of a coherent 300 kV electron beam to a sample, filters out energy-loss electrons through the sample with an in-column energy filter, and allows rotational data collection on a fast camera. It also possesses motorized cryo-sample loading and automated liquid-nitrogen filling for cooling of multiple samples. To facilitate its use, we developed GUI programs for efficient operation and accurate structure analysis. Here we report on the performance of the system and first results for thin 3D crystals of the protein complexes, catalase and membrane protein complex ExbBD. Data quality is remarkably improved with this approach, which we name eEFD (electron energy-filtered diffraction of 3D crystals), compared with those collected at 200 kV without energy filtration. Key advances include precise control of the microscope and recordings of lens fluctuations, which the programs process and respond to. We also discuss the merits of higher-energy electrons and filtration of energy-loss electrons in electron 3D crystallography.
My notes (saved in your browser only)
Citation neighborhood (no data yet)
We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.
Source provenance