How to use power Doppler ultrasound in transvaginal assessment of uterine fibroids.

OA: gold publisher-OA-unknown
AI-generated summary by claude@2026-07, 2026-07-31

This paper provides background on ultrasound physics and fibroid characteristics, outlines relevant machine settings, and offers a tutorial on optimal power Doppler settings for reproducible vascular evaluation of uterine fibroids.

One-sentence paraphrase of the abstract; not a substitute for reading it. No clinical advice. How this works

Abstract

Measuring vascularization in uterine fibroids is important for their diagnosis, treatment and prognosis. Vascularization can be measured by power Doppler ultrasound. The power Doppler signal depends on fibroid characteristics and on a variety of ultrasound-machine settings. Literature describing which machine settings influence the power Doppler signal is limited. Each manufacturer names settings and presets at their own discretion, with little information available publicly. Consistency of machine settings is important for correct interpretation of images in daily practice and is essential in yielding reproducible data for research. The aims of this paper, drawing from both a literature search and semistructured interviews with ultrasound-machine engineers and clinical experts in gynecological ultrasound, were: (1) to provide comprehensive background information on ultrasound physics and fibroid characteristics; (2) to present an overview of machine settings relevant to both two- and three-dimensional power Doppler, including power Doppler frequency, pulse repetition frequency, gain, wall-motion filter, acoustic power, persistence and signal rise; and (3) to provide a step-by-step tutorial on the optimal settings for vascular evaluation of uterine fibroids using power Doppler. The step-by-step tutorial comprises six steps to optimize the power Doppler signal, create a preset and acquire a reliable three-dimensional volume. This step-by-step tutorial should help research groups and clinicians to use power Doppler correctly and reproducibly in the evaluation of uterine fibroids. © 2022 The Authors. Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of International Society of Ultrasound in Obstetrics and Gynecology.
Full text 16,267 characters · extracted from pmc-nxml · 4 sections · click to expand

Practical

The ultrasound machine settings of importance for 2D and 3D power Doppler are discussed herein taking into account ultrasound features (Appendix  S1 ), fibroid characteristics (Appendix  S2 ) and the literature (Table  1 ). First, we discuss the essential, adjustable machine settings, as well as relevant, but mostly standardized, settings, for performing 2D power Doppler, including color score. It should be borne in mind that there may be other ultrasound machine settings that might influence the power Doppler signal. Second, we provide tips for acquiring a 3D power Doppler volume to calculate the vascularization index. Definitions of the machine settings and details of how they affect the Doppler signal are presented in Table  2 . Step‐by‐step instructions for optimal settings are summarized in Figure  1 . Illustrative examples of ultrasound views are presented in Figure  2 . Ultrasound (US) machine settings described in the literature and their reported effect on power Doppler (PD) outcome Only first author of each study is given. Reported as an index. NR, not reported; PRF, pulse repetition frequency; WMF, wall‐motion filter. Definitions of machine settings that influence power Doppler (PD) signal 3D, three‐dimensional; PRF, pulse repetition frequency; WMF, wall‐motion filter. Clutter filter removes low‐frequency Doppler signal. Flowchart summarizing step‐by‐step recommendations for adjusting settings to optimize power Doppler signal, creating a customized ‘FIBROID’ preset and acquiring a reliable three‐dimensional (3D) volume in transvaginal assessment of uterine fibroids. The two‐dimensional (2D) field‐of‐view is defined by the scanning area or box and the scanning angle of the probe. While the 3D region of interest (ROI) is also determined by the scanning area or box and the scanning angle, in the 3D ROI, these are perpendicular to each other, with the scanning area or box orientated left to right and the scanning angle from back to front. Examples of ultrasound views illustrating effect of adjusting different machine settings on power Doppler transvaginal imaging of uterine fibroids. In left column are sagittal power Doppler images, obtained using WS80 Samsung ultrasound machine; in middle column are three‐dimensional (3D) reconstructions performed onsite with same ultrasound machine; in right column are directional Doppler sagittal images or 3D reconstructions obtained using GE Voluson ultrasound machine. (a) Optimal imaging. (b–f) Effect of adjusting: (b) pulse repetition frequency (PRF); (c) gain; (d) wall‐motion filter (WMF); (e) power; (f) speed of acquisition or quality. MED, medium; Mid, mid range; PEN, penetration. Set the focus just behind the fibroid for optimal lateral resolution; features adjacent to it will be better distinguished if they lie within the area of convergence of the ultrasound waves 15 , 16 . Minimize the depth and the field‐of‐view, defined by the scanning area and angle, to maintain a frame rate that will achieve optimal temporal resolution. A wide field‐of‐view delays the display of the Doppler signal 17 , 18 . The field‐of‐view should contain the entire vascular pseudocapsule and the area close to the probe to prevent reverberation artifacts 18 . B‐mode and Doppler frequency should be set as low as possible, resulting in high penetration. The pulse repetition frequency (PRF) is the most important machine setting to adjust. The PRF is the rate of pulses transmitted consecutively, not to be confused with the ultrasound frequency of those pulses. The measureable blood flow velocity depends on the PRF according to the formula: velocity = distance × PRF, where distance is the difference between the depths of the target at two consecutive pulses. When the PRF is set too high, the power Doppler signal of low blood flow velocity, such as in fibroids, will not be detected. The PRF should not be set lower than half of the maximum velocity, to avoid aliasing (Appendix  S1 ) 15 . In general, the PRF should be set as low as possible 19 , 20 , 21 , 22 , 23 , 24 . The PRF is associated negatively with the power Doppler signal (Figure  2b ) 19–22,25 . The gain works as an amplifier of incoming Doppler signals 26 . Gain set too low might result in failure to visualize the blood flow, while gain set too high can result in electronic noise artifacts being displayed, even in the absence of flow. Also, mirror‐image artifacts can occur when the gain is set too high and the ultrasound machine can no longer distinguish the correct direction of flow, displaying the Doppler signal in the opposite or an incorrect direction 15 . Increasing the gain increases the 3D power Doppler signal (Figure  2c ) 21 , 23 , 26 , 27 , 28 . The signal can be fine‐tuned by increasing the gain until noise artifacts become visible and then lowering the gain until the artifacts have just disappeared. Standardization of gain values can be considered if, within a hospital, only one type of machine with one specific setting is used 28 . The wall‐motion filter removes low‐frequency Doppler signals 21 . It is used, for example, to filter out the reverberation of an arterial wall, which may interfere with the Doppler signal of arterial blood flow. However, it may remove true low‐frequency slow flow in small fibroid vessels 15 . The wall‐motion filter is related inversely to the power Doppler signal (Figure  2d ) 20 , 21 , 22 , 23 . When measuring low‐velocity blood flow, it is recommended to use the lowest wall‐motion filter possible 20 . Additional settings may be adjusted, although this depends on the type of machine that is used. In some machines, these settings are fixed and cannot be adjusted. The acoustic or signal power has an influence on the power Doppler signal comparable to that of the gain 21 . The acoustic power is equal to the amplitude of the pulse pressure of the ultrasound beam multiplied by the displacement velocity 29 . Increasing the acoustic power may boost the Doppler signal, but most machines have an upper limit for reasons of safety. Some machines do not allow any adjustment. The acoustic power and 3D power Doppler signal are positively related (Figure  2e ) 21 . Persistence and signal rise are averaging techniques that influence detection of the power Doppler signal 21 . Persistence or ‘frame averaging’ is the averaging of frames acquired successively in order to reduce noise 15 . A higher index involves more averaging and therefore a greater number of previous frames is included in the averaged image 30 . When the persistence is increased, there is a delay in visualizing the Doppler flow. If the persistence is too low, blood flow cannot be detected and movements are missed. Details on which index corresponds to a particular number of frames on the weighted factor applied are not always shared by the manufacturer and this depends on the type of machine. Increasing the persistence moderately will result in a somewhat higher power Doppler signal 21 . Optimal persistence is mid‐range 30 . Signal rise is another type of averaging technique. Increasing the signal rise delays the time at which the Doppler signal first appears. Afterimages can occur when a high signal rise is used, while too low a signal rise results in an image with unnatural‐looking flow and noise artifacts. A very high signal rise results in a decrease in the power Doppler signal 21 . Yet, low‐velocity blood flow is better visualized with a higher signal rise, which corresponds to greater frame averaging. The speed of acquisition defines the resolution or sensitivity of the 3D scan to detect flow. Also known as ‘quality’, the acquisition speed, along with the 3D region of interest, determines the scanning time. A low speed of acquisition (high quality) of a large 3D region of interest results in a long scanning time. The longer the scanning time, the higher the probability of movement artifacts, for example resulting from the patient breathing. The 3D region of interest is limited and its size varies between selected ultrasound probes 17 . The scanning area visible on the displayed plane, in combination with the scanning angle perpendicular to this plane, defines the 3D region of interest. The speed of acquisition is negatively related to the 3D power Doppler signal (Figure  2f ) 21 . On most ultrasound machines, the speed of acquisition (quality) is expressed as an index. Details regarding which settings correspond to which resolutions and speeds are not made public by the manufacturers. It is important to check the acquired 3D power Doppler volume before saving it. There are multiple buttons on the ultrasound machine dashboard which allow adjustment of the volume in the sagittal, transverse and coronal planes. The 3D volume should include the entire fibroid; when this is not the case, or the 3D volume shows artifacts, the region of interest should be adjusted. This may affect the speed of acquisition. If artifacts remain or insufficient Doppler signal is visible, further adjusting one of the ultrasound settings discussed herein should be considered. After all settings have been optimized and the 3D image acquired, the data can be analyzed by VOCAL software. This software is applied mostly in the research setting to semiquantify vascularity and it is applicable to images acquired by machines from multiple manufacturers. The vascularization index represents the amount of vasculature per unit surface area, by calculating the proportion of color voxels (as a proxy for the blood vessels) relative to the total number of voxels, including both gray and colored ones (as a proxy for tissue and blood vessels) 21 . The software offers several options during the steps of the calculation and the choice of options will influence the value of the vascularization index. First, a region of interest can be defined automatically or manually, the latter being the preferred method 12 , using a certain number of virtual slices around a central axis. The most reliable is to use rotation steps of 9°, which is the smallest step size possible. For large uterine fibroids we advise using steps of 30° in order to be time‐effective while remaining sufficiently accurate. While delineating the region of interest, care should be taken to include the vascular pseudocapsule correctly. The vascular arcade of the uterus or uterine artery is easily mistaken for the fibroid's pseudocapsule on a Doppler image. Vascular indices are calculated automatically by a ‘histogram facility’. The histogram can include voxels of the total volume, or of a certain predefined ‘shell’, both based on the region of interest 31 . Whether to use the outer shell (Figure  3a ), symmetric shell (Figure  3b ), inner shell (Figure  3d ) or no shell (Figure  3c ) depends on the target 12 . For example, to calculate the fibroid pseudocapsule's flow, an outer shell of 0.5 cm is optimal 28 . Power Doppler image of uterine fibroid, illustrating calculation of vascularization index. Sagittal view (top) and three‐dimensional reconstructions of regions of interest (bottom). Black line (c) corresponds to region of interest; this contour is delineated manually by the operator. Remaining lines (a,b,d) indicate shells applied automatically by VOCAL™ software. (a) Outer shell of pseudocapsule; (b) symmetric shell of pseudocapsule; (c) total fibroid (shell off); (d) inner shell of pseudocapsule.

Background

Measurement of vascularization in uterine fibroids is important for their diagnosis and management. Evaluation of the vascularization pattern enables differentiation between fibroids and adenomyosis and between different histological types of fibroid 1 , 2 . The extent of a fibroid's vascularization is associated with clinical symptoms and growth potential 3 , 4 , 5 . Fibroid vascularization also has an influence on treatment response 6 , 7 , 8 . The importance of reporting fibroid vascularization has been emphasized by the ‘Morphological Uterus Sonographic Assessment’ (MUSA) working group 9 . Doppler ultrasound is a sonographic imaging technique used for blood flow measurements. Details on Doppler physics can be found in Appendix  S1 . Fibroids have a distinctive vascularization pattern, including a highly vascularized peripheral rim and a core containing many slow‐flow capillaries. Other characteristics of fibroids important for power Doppler imaging are summarized in Appendix  S2 . The degree of vascularization may be reported using a subjective color score on two‐dimensional (2D) power Doppler ultrasound imaging 9 . This color score is easy to apply in daily practice, but depends on the observer's subjective appreciation and experience. Compared with color Doppler, power Doppler is three times more sensitive in detecting low flow velocities 10 , 11 . In addition, three‐dimensional (3D) power Doppler represents in a single image the amount and distribution of blood vessels in a fibroid. Furthermore, in contrast to a subjective color score, 3D power Doppler can provide an objective measurement of vascularity using Virtual Organ Computer‐aided AnaLysis (VOCAL™) software 9 . This ‘vascularization index’ is a ratio that is calculated automatically, and represents the amount of vasculature per unit surface area. The index is reproducible, is able to discriminate vascularity and has good correlation with histology 12 , 13 , 14 . However, both subjective and objective quantification of vascularization depend on the settings and type of machine used. Transvaginal ultrasound is the primary technique for imaging uterine fibroids, and reveals features that differ from those on transabdominal ultrasound, the latter being required for large fibroids. Each manufacturer names settings and presets at their own discretion, with little public information available, adding to the complexity of the various different settings. Consistency of machine settings is important for correct interpretation of images in daily practice and is essential in yielding reproducible research data. Herein, we aim to present an overview of machine settings relevant to both 2D and 3D power Doppler in the transvaginal ultrasound assessment of uterine fibroids, and provide a step‐by‐step tutorial. This tutorial is based on a literature search combined with semistructured interviews with ultrasound‐machine engineers and clinical experts in gynecological ultrasound (Appendix  S3 ).

Conclusion

Using power Doppler ultrasound in the transvaginal assessment of uterine fibroids is challenging. We have composed step‐by‐step guidance to help research groups and clinicians to use power Doppler signals in the evaluation of uterine fibroids. We have identified and discussed relevant machine settings which influence the power Doppler signal. Some challenges will remain even after following this step‐by‐step guide. For example, some settings are interdependent: when keeping one setting constant and adjusting a second to increase the Doppler signal, a third setting may need to be adjusted to prevent the Doppler signal from becoming too high 23 . Moreover, the low blood flow velocity in vessels of small diameter 3 , 20 , 32 might be at the lower limit of what is technically possible to measure by power Doppler. Measuring multiple or large fibroids might be complex due to attenuation in fibroids more than 8–10 cm in diameter 4 and to the increased distance from the transducer to the proximal side of the fibroid 27 . Despite these challenges, this step‐by‐step tutorial should contribute to the correct use and improved reproducibility of power Doppler ultrasound in the transvaginal assessment of uterine fibroids.

Supplementary Material

Appendix S1 Background information on ultrasound and Doppler physics Click here for additional data file. Appendix S2 Transvaginal power Doppler assessment of uterine fibroids Click here for additional data file. Appendix S3 Methods for selection of machine settings that influence power Doppler signal Click here for additional data file.

Text is read by the "Ask this paper" AI Q&A widget below. Extraction quality varies by source — PMC NXML preserves structure cleanly, OA-HTML may include some navigation residue, and OA-PDF can have broken hyphenation. The publisher copy (via DOI) is the canonical version.

My notes (saved in your browser only)

Ask this paper AI returns verbatim quotes from the full text · source: pmc-nxml

Answers must be backed by verbatim quotes from this paper's full text. Hallucinated quotes are dropped automatically; if no verbatim passage answers the question, we say so. How this works

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. The paper's references may be in our DB but unresolved to ``paper_id`` (resolution happens at ingest when the cited DOI matches a row we already have). Run the cross-source citation reconcile pass to retry.

Source provenance

europepmc
last seen: 2026-08-02T06:10:09.037253+00:00
unpaywall
last seen: 2026-05-21T05:10:58.409756+00:00
License: publisher-OA-unknown · commercial use NOT OK · attribution required