A Comparative Study of Predictive model (ECG Buddy) and ChatGPT-4o for Myocardial Infarction Diagnosis via ECG image Analysis: Performance, Accuracy, and Clinical Feasibility

preprint OA: closed
📄 Open PDF Full text JSON View at publisher
AI-generated deep summary by claude@2026-06, 2026-06-23 · read from full text

This retrospective study compared ChatGPT-4o with a dedicated ECG AI tool (ECG Buddy) for classifying myocardial infarction (MI) using a publicly available 12-lead ECG dataset from Pakistan, with MI-positive (239) and MI-negative (689) recordings. ChatGPT was queried with five MI confidence options, while ECG Buddy analyzed images based on features including ST-elevation MI and related clinical/biomarker criteria. ECG Buddy substantially outperformed ChatGPT, achieving 96.98% accuracy and AUC 98.8% versus ChatGPT’s 65.95% accuracy and AUC 57.34%, with DeLong’s test confirming significant differences (all P < .001); the authors also report that ChatGPT provided clinically plausible explanations in only 7.5% of analyzed error cases, and many explanations were absent or incorrect. This paper does not explicitly discuss endometriosis or adenomyosis; it was included in the corpus via a keyword match in the upstream search index.

Read from the paper's body, not the abstract. Not a substitute for reading the paper. No clinical advice. How this works

Abstract

Background Accurate and timely electrocardiogram (ECG) interpretation is critical for diagnosing myocardial infarction (MI) in emergency settings. Recent advances in multimodal Large Language Models (LLMs), such as Chat Generative Pre-trained Transformer (ChatGPT), have shown promise in clinical interpretation for medical imaging. However, whether these models analyze waveform patterns or simply rely on text cues remains unclear, underscoring the need for direct comparisons with dedicated ECG artificial intelligence (AI) tools. Methods This retrospective study evaluated and compared AI models for classifying MI using a publicly available 12-lead ECG dataset from Pakistan, categorizing cases into MI-positive (239 images) and MI-negative (689 images). ChatGPT (GPT-4o, version 2024-11-20) was queried with five MI confidence options, whereas ECG Buddy for Windows analyzed the images based on ST- elevation MI, acute coronary syndrome, and myocardial injury biomarkers. Results Among 928 ECG recordings (25.8% MI-positive), ChatGPT achieved an accuracy of 65.95% (95% confidence interval [CI]: 62.80–69.00), area under the curve (AUC) of 57.34% (95% CI: 53.44–61.24), sensitivity of 36.40% (95% CI: 30.30–42.85), and specificity of 76.20% (95% CI: 72.84–79.33). However, ECG Buddy reached an accuracy of 96.98% (95% CI: 95.67–97.99), AUC of 98.8% (95% CI: 98.3–99.43), sensitivity of 96.65% (95% CI: 93.51–98.54), and specificity of 97.10% (95% CI: 95.55–98.22). DeLong’s test confirmed that ECG Buddy significantly outperformed ChatGPT (all P < .001). In an error analysis of 40 cases, ChatGPT provided clinically plausible explanations in only 7.5% of cases, whereas 35% were partially correct, 40% were completely incorrect, and 17.5% received no meaningful explanation. Conclusion LLMs such as ChatGPT underperform relative to specialized tools such as ECG Buddy in ECG image-based MI diagnosis. Further training may improve ChatGPT; however, domain- specific AI remains essential for clinical accuracy. The high performance of ECG Buddy underscores the importance of specialized models for achieving reliable and robust diagnostic outcomes.
Full text 4,425 characters · extracted from oa-doi-fallback · 4 sections · click to expand

Abstract

Background Accurate and timely electrocardiogram (ECG) interpretation is critical for diagnosing myocardial infarction (MI) in emergency settings. Recent advances in multimodal Large Language Models (LLMs), such as Chat Generative Pre-trained Transformer (ChatGPT), have shown promise in clinical interpretation for medical imaging. However, whether these models analyze waveform patterns or simply rely on text cues remains unclear, underscoring the need for direct comparisons with dedicated ECG artificial intelligence (AI) tools.

Methods

This retrospective study evaluated and compared AI models for classifying MI using a publicly available 12-lead ECG dataset from Pakistan, categorizing cases into MI-positive (239 images) and MI-negative (689 images). ChatGPT (GPT-4o, version 2024-11-20) was queried with five MI confidence options, whereas ECG Buddy for Windows analyzed the images based on ST- elevation MI, acute coronary syndrome, and myocardial injury biomarkers.

Results

Among 928 ECG recordings (25.8% MI-positive), ChatGPT achieved an accuracy of 65.95% (95% confidence interval [CI]: 62.80–69.00), area under the curve (AUC) of 57.34% (95% CI: 53.44–61.24), sensitivity of 36.40% (95% CI: 30.30–42.85), and specificity of 76.20% (95% CI: 72.84–79.33). However, ECG Buddy reached an accuracy of 96.98% (95% CI: 95.67–97.99), AUC of 98.8% (95% CI: 98.3–99.43), sensitivity of 96.65% (95% CI: 93.51–98.54), and specificity of 97.10% (95% CI: 95.55–98.22). DeLong’s test confirmed that ECG Buddy significantly outperformed ChatGPT (all P < .001). In an error analysis of 40 cases, ChatGPT provided clinically plausible explanations in only 7.5% of cases, whereas 35% were partially correct, 40% were completely incorrect, and 17.5% received no meaningful explanation.

Conclusion

LLMs such as ChatGPT underperform relative to specialized tools such as ECG Buddy in ECG image-based MI diagnosis. Further training may improve ChatGPT; however, domain- specific AI remains essential for clinical accuracy. The high performance of ECG Buddy underscores the importance of specialized models for achieving reliable and robust diagnostic outcomes. Competing Interest Statement Joonghee Kim, MD, PhD, developed the algorithm. He also founded a start-up company, ARPI Inc., where he serves as the CEO. Youngjin Cho, MD, PhD, and Dongbum Suh, MD, work for the company as research directors. Haemin Lee works for the company as a data scientist. The rest of the authors declare no conflict of interest. Funding Statement This research was partly supported by a grant from the Technological Innovation Research & Development Program (SCALEUP TIPS), funded by the Ministry of SMEs and Startups (grant number: RS-2024-00415492), and the Medical AI Clinic Program through the National IT Industry Promotion Agency, funded by the Ministry of Science Information Communication and Technology (grant number: H0904-24-1002). Author Declarations I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained. Yes The details of the IRB/oversight body that provided approval or exemption for the research described are given below: Khan AH, Hussain M. ECG images dataset of cardiac patients. Mendeley Data. 2021;V2. doi:10.17632/gwbz3fsgp8.2 I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals. Yes I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance). Yes I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable. Yes Data Availability All data produced in the present study are available upon reasonable request to the authors

Text is read by the "Ask this paper" AI Q&A widget below. Extraction quality varies by source — PMC NXML preserves structure cleanly, OA-HTML may include some navigation residue, and OA-PDF can have broken hyphenation. The publisher copy (via DOI) is the canonical version.

My notes (saved in your browser only)

Ask this paper AI returns verbatim quotes from the full text · source: oa-doi-fallback

Answers must be backed by verbatim quotes from this paper's full text. Hallucinated quotes are dropped automatically; if no verbatim passage answers the question, we say so. How this works

Citation neighborhood (no data yet)

We don't have any in-corpus citations linked to this paper yet. This is a recent paper (2025) — citers typically take a year or two to land, and the OpenAlex reference graph may still be filling in.

Source provenance

europepmc
last seen: 2026-05-20T01:45:00.602351+00:00
unpaywall
last seen: 2026-08-04T06:56:55.048231+00:00