Simple Calibration Method for Confocal Microscopy Used in Micro-Additive Manufacturing Processes

preprint OA: closed CC-BY-4.0
🔓 Open OA copy View at publisher

Abstract

Additive manufacturing (AM) is a promising new technology that is having a very fast growth from home workshops to high-tech cutting-edge factories. As any manufacturing technique, adequate metrology services are needed to assure the quality of items manufactured by AM. One of the most widely used instruments to measure the characteristics of surfaces manufactured with AM is the confocal microscope. In this paper, authors present a whole calibration procedure for confocal microscopes designed to be implemented preferably in workshops or industrial environments rather than in research and development departments. Because of that, it is as simple as possible. The procedure is designed without forgetting any of the key aspects that need to be taken into account and based on classical reference material standards. These standards can be easily found in industrial dimensional laboratories and easily calibrated in accredited calibration laboratories.