Abstract
The outbreak of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has
rapidly reached pandemic levels. Sufficient testing for SARS-CoV-2 remains essential
for tracking and containing the rapid spread of the virus. However, due to increased
global demand, kits and proprietary reagents for RNA extraction are limited, which
markedly reduce SARS-CoV-2 testing capabilities in many countries. Here, we explore
the use of conventional acid guanidinium thiocyanate-phenol-chloroform (AGPC)-based
RNA purification as an alternative to commercial automated systems for detection of
SARS-CoV-2 by RT-qPCR. 87 clinical oropharyngeal or nasopharyngeal swab
specimens were extracted by AGPC and compared to the commercial platforms, the
Promega Maxwell
® RSC 48 instrument for automated RNA extraction and the fully
integrated diagnostic system, the Cobas ®6800 apparatus. Our results show that RNA
extracted using the AGPC method is fully comparable to modern automated systems
regarding analytical sensitivity, specificity and accuracy with respect to detection of
SARS-CoV-2 as evaluated by RT-qPCR. Moreover, we find that the AGPC method is
easily scalable and implemented in conventional laboratories. Taken together, these
data identify conventional AGPC-based RNA extraction as a low cost and suitable
alternative to automated systems for the detection of SARS-CoV-2, when automated
systems, kits and reagents are not readily available.
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
NOTE: This preprint reports new research that has not been certified by peer review and should not be used to guide clinical practice.
Introduction
The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) causing the
coronavirus disease 2019 (COVID-19) has rapidly reached pandemic levels, with
COVID-19 related morbidities and mortalities rising in many countries (1-3). All efforts
are needed to regain control and one important aspect is tracking the spread of
infections, both in the healthcare system and in the general public. However, the
increased spread of the virus has markedly increased global demand for materials and
reagents needed for adequate testing. The limited supply of commercial RNA extraction
kits, consumables and reagents has posed serious limitations on testing capacities in
many countries, especially those relying on automated systems where commercial kits
are indispensable. Furthermore, commercial systems and kits are expensive and not
readily available in all countries. Without the ability to test more widely it is difficult to
adequately evaluate the spread of the disease. In the wake of this, several new
Methods
have been proposed to overcome the bottleneck posed by RNA purification in
an effort to detect the novel coronavirus (SARS-CoV-2) and clinically diagnose the
COVID-19 (4, 5). As such the acid guanidinium thiocyanate-phenol-chloroform (AGPC)
Method
of RNA extraction has recently been found suitable to allow for SARS-CoV-2
PCR detection (5). However, it is unclear how it compares to automated systems
currently running at clinical laboratories and if levels of analytical sensitivity, specificity
and accuracy are comparable. A good accuracy is especially important since false
negative samples could lead to inadvertent spread of the COVID-19 disease within the
healthcare systems and general communities.
The AGPC is a simple method used in many laboratories worldwide. The method itself
has a long-known track record (6) and relies on an acid guanidinium thiocyanate-phenol
mixture that can be used to extract RNA with the addition of chloroform. Commercially
mixed reagents needed for this step are readily available from several vendors as e.g.
TRIzol™ (Invitrogen) and TRI Reagent
/i1 (Sigma-Aldrich, now Merck) or the reagents
can be produced locally from base chemicals at a low cost (6). The method is simple to
implement in conventional laboratory settings and is already routinely used at many
research institutions. One caveat in comparison with automated systems is the
additional hands-on sample preparation time needed. For COVID-19 testing it is
currently unknown whether the AGPC-based RNA extraction method can perform to a
level similar to automated testing systems, especially when reagents for the latter
systems are sparse.
We therefore aimed to compare the AGPC method to automated systems commonly
used in clinical detection of SARS-CoV-2 and to evaluate its suitability as a replacement
for conventional automated systems when shortages of proprietary materials are
experienced or not readily available. Furthermore, we aimed to explore whether the
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
AGPC extraction method could be used for detection of SARS-CoV-2 on a larger scale.
Our current data shows that accuracy of this method is fully comparable to automated
systems. This is important since the AGPC is a simple method used in laboratories
worldwide to extract RNA and commercially formulated reagents needed for this
technique are readily available or can be easily prepared from basic laboratory
chemicals at a low cost.
Materials and methods
Sample material
Sample material was obtained from oropharyngeal or nasopharyngeal swabs collected
in ESwab™ tubes containing 1 ml of liquid Amies medium (Copan). Specimens were
obtained from patient material undergoing routine diagnostic analyses at the
Department of Clinical Microbiology, Odense University Hospital. The samples used for
this study were chosen based on SARS-CoV-2 status from the routine diagnostic
analyses and to reflect a broad range of viral titers in the sample material as determined
by the SARS-CoV-2 E gene (Ct value range 16-38) using the Cobas
®6800 system
(Roche). To distinguish between true negative results and reactions affected by
inadequate RNA isolation, presence of PCR inhibitors or instrument failure, Nobilis ND
C2 vaccine (NDV) against Newcastle Disease (Nobilis) was added to the ESwab™
media as an internal control prior to RNA purification using the AGPC method and the
Maxwell
® RSC 48 instrument (Promega). The amount of added NDV had previously
been titrated to yield a Ct value ~27.5 as measured by RT-qPCR and routinely used for
evaluating sample quality after RNA isolation at the Department of Clinical Microbiology.
To determine a Ct cutoff value for NDV after purification with the Maxwell ® RSC 48 and
the AGPC methods, 24 clinical swab samples with known SARS CoV-2 status based on
the Cobas
®6800 system were processed using both methods and the data for the in-
house NDV and SARS-CoV-2 RT-qPCR assays were evaluated. Based on these data
and applying a precautionary principle the NDV Ct cutoff value was determined to be
<29.5.
Cobas
®6800 system
The fully automated IVD-CE-labelled Cobas ®6800 system (Roche Diagnostics, Basel,
Switzerland) was used in this study (set as gold standard) for the evaluation of Maxwell®
and AGPC methods of RNA purification. Testing was performed using the Cobas ®
SARS-CoV-2 test assay with proprietary primers directed against the SARS-CoV-2
ORF1 and E -gene. 400 µl ESwab™ sample media was added to the Cobas ® machine
and eluted in a final volume of 50 µl. From this, 27 µl of eluted sample was added to 25
µl of Cobas
® SARS-CoV-2 PCR mix.
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
Maxwell® RSC 48 automated RNA extraction
RNA extraction was performed with the Maxwell ® RSC Viral Total Nucleic Acid
Purification Kit (Promega) using the Maxwell ® RSC 48 Instrument according to the
manufacturer’s recommendation, without the initial heat incubation step. In brief, lysis
buffer, Proteinase K and internal controls were added to the cartridge and 200 µl of
ESwab™ sample media was added. The extracted RNA was eluted in a total volume of
50 µl.
The acid guanidinium thiocyanate-phenol-chloroform extraction method
RNA extraction was carried out according to the instructions of the Tri Reagent
®
manufacturer with minor modifications. In brief, 200 µl of the sample specimens were
aliquoted into sterile 1.5 ml test tubes containing 800 µl of TRI Reagent
® (Catalog No.
T9424, Sigma-Aldrich). This step was performed under Class II conditions, while the
remainder of the RNA extraction, after inactivation of the virus, was performed in a
conventional laboratory. Once mixed and transported to the conventional laboratory, the
test tubes with sample material and Tri Reagent
® were added 200 µl chloroform and
mixed by vortexing (5 sec. at max speed). Samples were then incubated for 2 minutes
at room temperature and subsequently centrifuged at 14,000g for 15 minutes (4°C).
Only the aqueous phase (500 µl) of the resulting mixture located at the top of the tube
was processed further, while the remainder of the Tri Reagent
®/chlorofom mixture was
discarded. The aqueous phase was pipetted into a new tube containing 2 µl of
GlycoBlue
TM (Catalog No. AM9515 ThermoFisher). 600 µl of isopropanol was then
added to the tubes containing the aqueous phase, after which the samples were mixed
by vortexing (3 sec at max speed) and then incubated at room temperature (20-25°C)
for 20 minutes. The samples were then centrifuged at 14,000g for 15 minutes (4°C), the
supernatant removed while making sure the blue RNA pellet remained at the bottom of
the tube. The RNA pellets were then washed in 1 ml of 75% Ethanol, vortexed (3 sec. at
max speed) and centrifuged at 8,000g for 5 minutes (4°C). The supernatants were then
removed while making sure that the RNA pellets remained at the bottom of the tubes.
After removal of the Ethanol, samples were incubated at room temperature (20-25°C)
for 10 minutes with the lid open to allow excess ethanol to evaporate. Subsequently the
RNA pellets were resuspended in 30 µl of RNAse-free water and heated to 60°C for 5
minutes on a heating block. RNA samples were then vortexed 3 sec. and then
centrifuged briefly (10 sec. on table centrifuge) to collect all the liquid at the bottom of
the tube. Samples were stored at 4°C before transported on ice to the Department of
Clinical Microbiology for further analysis.
SARS-CoV-2 RT-qPCR analysis
The SARS-CoV-2 was detected according to the real-time PCR protocol established by
Corman et al. (7) . Detection of the internal control Newcastle disease virus NDV was
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
performed using primers and probe sequences kindly provided by dr. Kurt J. Handberg,
Department of Clinical Microbiology, Skejby, Denmark. In brief, TaqManTM Fast Virus 1-
Step Master Mix (ThermoFisher) was used for the amplification reaction. A final
concentration 1000 nmol of primers and 200 nmol of probes were added to a total
reaction volume of 20 ul, containing 6 µl of RNA (purified RNA obtained from either
Maxwell
® automated RNA extraction or from AGPC extraction). Samples were analyzed
on a LightCycler®480 II (Roche) using the following program 50°C for 5min, 95°C for
20sec followed by 45 cycles of 95°C for 15sec and 60°C for 1min.
E_Sarbeco_F1: ACAGGTACGTTAATAGTTAATAGCGT
E_Sarbeco_R2: ATATTGCAGCAGTACGCACACA
E_Sarbeco_P1: FAM_ACACTAGCCATCCTTACTGCGCTTCG_BHQ1
NDV-F: 5'-CAC TGT CGG CAT TAT CGA TGA-3’
NDV-R: 5'-GAG CAT CGC AGC GGA AA-3’
NDV-Probe: 5'-FAM-CCC AAG CGC GAG TTA-MGB-3’
Comparison tests and Statistics
A total of 87 clinical sample specimens were chosen based on SARS-CoV-2 status from
the Cobas
®6800 system and used to evaluate the analytical sensitivity, specificity and
accuracy of our in-house SARS-CoV-2 RT-qPCR assay after RNA purification using the
Maxwell® RSC 48 and AGPC methods. Only samples yielding Ct values below 29.5 for
the internal control NDV assessed by RT-qPCR on the Maxwell ® and TRI Reagent ®
platforms were included in the analyses. Results obtained on the Cobas ®6800 system
for the SARS-CoV-2 E gene were compared to the results obtained using our in-house
RT-qPCR assay for the SARS-CoV-2 E gene. 95% confidence intervals and Pearson
correlation coefficients (r) were calculated using Prism 8.3 (Graphpad Software).
Results
The AGPC method delivers high analytical sensitivity, specificity and accuracy for
SARS-CoV-2 testing
To evaluate whether conventional AGPC based extraction of RNA could serve as a
viable alternative to automated systems with respect to reliability and accuracy, we
isolated RNA using the AGPC method from 87 clinical specimens (oropharyngeal or
nasopharyngeal swabs) with known SARS-CoV-2 status (57 positive and 30 negative),
and performed a side-by-side comparison with the identical samples extracted on a
Maxwell
® RSC 48 instrument. The samples used had previously been analyzed using
the state-of-the-art fully integrated Cobas®6800 diagnostic system capable of analyzing
patient specimens from sample to test result without manual interference, using
proprietary primers directed against the SARS-CoV-2 ORF1 and E-gene. Analysis of
the sample specimens with RNA extracted by the Maxwell
® RSC 48 instrument showed
a 98.2% sensitivity, 96.4% specificity and 97.6% accuracy, but also reported a single
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
false positive and a false negative sample compared to the Cobas®6800 system (Figure
1A). Importantly, analysis of the sample specimens using the AGPC method for RNA
extraction displayed a 98.0% sensitivity, 100% specificity and 98.8% accuracy, with no
false positive and only 1 false negative compared to Cobas
®6800 system.
AGPC based RNA extraction allows for SARS-CoV-2 E-gene detection comparable to
automated systems
To further validate our findings, we performed a direct comparison of the Ct values
measured for the SARS-CoV-2 E gene in the SARS-CoV-2 positive sample specimens
after AGPC extraction of the RNA, to the Ct values from identical samples reported
using the Cobas
®6800 system and the Maxwell® RSC 48 instrument (Figure 1B and C),
respectively. The Ct values attained for the SARS-CoV-2 E gene using the AGPC
Method
showed a highly significant correlation to those reported for the Cobas ®6800
system (r=0.97, p<0.0001) and the Maxwell® RSC 48 instrument (r=0.98, p<0.0001).
To distinguish between true negative results, and reactions affected by inadequate RNA
isolation, presence of PCR inhibitors or instrument failure, all sample specimens used
for RNA extraction via the AGPC method and the Maxwell ® RSC 48 instrument were
added Nobilis ND C2 vaccine (NDV) prior to RNA isolation. Thus, to assess the purity,
yield and efficiency of the RNA extraction using the AGPC method we compared the Ct
values reported for the NDV to those attained for the same identical samples extracted
using the Maxwell
® RSC instrument. Direct comparison of the Ct values revealed no
significant differences in Ct values for NDV between the AGPC method and the
Maxwell® RSC 48 instrument (Figure 1D).
The AGPC method is easily scalable and allows for reliable RNA extraction in a
conventional laboratory setting
The Department of Molecular Medicine at University of Southern Denmark is currently
aiding the Department of Clinical Microbiology in detection of SARS-CoV-2 using the
AGPC method. To examine the reliability of the AGPC-based isolation method in large
scale, we set up a pipeline for RNA isolation. Using standard laboratory equipment set
up in an empty teaching laboratory, we have adapted and modified the standard AGPC
protocol for use in a large-scale RNA extraction pipeline (Supplemental Figure 1 –
Pipeline setup ). Taking an iterative approach, we have scaled up the number of
samples processed daily, which has reduced the number of samples specimens that fail
the NDV control to 4.5% (Figure 1E). The full protocol for this setup has been described
in detail in supplemental materials (Supplemental data – Detailed protocol) and large-
scale extraction can easiest be run with 5 technicians with the potential output of 400-
600 samples specimens processed per day (7.5 hrs./day).
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
Discussion
We find that the AGPC method is a reliable method for RNA extraction fully comparable
to automated systems with respect to detection of SARS-COV-2 in oropharyngeal
samples isolated from patients with suspected cases of COVID-19. The AGPC method
of RNA extraction relies on simple chemical mixtures that can be easily obtained from
commercial vendors or mixed by combining the needed chemicals. Overall the method
is low-tech and routinely used in the many laboratories worldwide. The use for the
AGPC method as a substitute for automated systems is especially important when
automated systems are readily available but strained due to supply shortages.
However, in places where these automated systems are not available, the AGPC
Method
may allow for detection of viruses to levels comparable to those of the modern
automated systems. This also means that a large number of scientists and technical
personnel already are well-trained in using this method to address research questions
for which RNA extraction is required. If necessary, these experienced personnel can be
mobilized to aid in the RNA extraction, when either global demand reduces supply
chains for critical components needed in the automated systems to extract RNA or if
such automated systems are not readily available or overwhelmed in specific regions.
The AGPC method is robust and near the level of advanced commercial methods.
When compared to the Maxwell
®-based automation of RNA extraction using the same
individual samples and the same SARS-COV-2 realtime PCR assay run in parallel, the
Results
are virtually identical. For the Cobas
®6800 system, the Ct values reported for the
SARS-CoV-2 E gene were similarly comparable to those reported for the AGPC
method. However, the Cobas ®6800 system requires twice the volume of sample
specimen for isolation of the RNA, utilizes approximately four times the amount of RNA
product for the RT-qPCR analysis and is based upon proprietary primers and probes for
detection of the SARS-CoV-2 E gene. Thus, a direct comparison cannot be made
between the Cobas
®6800 system and the results obtained with the used in-house
SARS-COV-2 PCR assay. The detection sensitivity of the various methods is of utmost
importance as viral titers may vary and false negative samples would allow individuals
to spread COVID-19 infection inadvertently. A clear understanding of viral titers during
the course of the infection is still lacking but may fluctuate during the course of the
disease. While PCR based detection is not able to detect all, but nearly all COVID-19
cases (8, 9), it is important that the method used returns the highest amount of
analytical accuracy in for diagnostic purposes. Our data suggest that the AGPC method
is comparable in this respect.
While there are numerous advantages to the AGPC method, there are also inherent
limitations. In comparison to automated RNA extraction systems there is extensive
hands-on time and inadvertently risks of human errors. Furthermore, there is a
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
possibility for loss of the sample, if the pellet is lost during the isolation step or if the
sample is inadvertently mixed when transferring the aqueous phase between tubes.
However, well established workflows can minimize these risks to very low levels. By
spiking the sample specimens with the Nobilis ND C2 vaccine against Newcastle
Disease containing inactivated virus, we obtained a good measure of the extraction
efficiency and presence of potential inhibitors of the PCR analysis. This allowed us to
monitor the entire extraction process and in the rare case of sample loss or other
failures, repeat the RNA extraction and analysis, as only 200 µl out of the 1ml Eswab
TM
media was used for the initial analysis.
RNA isolation using the AGPC method is favored among scientists for small scale RNA
purification setups, due to its low cost, versatility and ease of use. Here we show that
the AGPC method is easily scalable to volumes usable for clinical diagnostics as a
supplement to conventional automated systems. The state-of-the-art Cobas
®6800
system has a capacity of 384 samples per 8 hours, which is roughly equivalent to the
throughput of our RNA isolation pipeline presented here. While the Cobas
®6800 system
is unsurpassed in ease, accuracy and sensitivity, the current COVID-19 pandemic and
worldwide shortage of kits and reagents for automated systems, underlines the
importance of redundant methods, which can be applied at a low cost, independent of
proprietary reagents and commercial interest.
As with any work with viruses there is a chance of viral contamination. However, the
collected sample specimens are mixed directly in organic solvent, which efficiently
inactivates coronavirus. Hence, only this initial step needs to be carried out in a
specialized class II facility (5, 10, 11). Nevertheless, AGPC solutions are hazardous and
the isolation procedure in general needs to be carried out in fume hoods. This requires
specialized workplaces commonly found across universities, hospitals, private research
companies and institutions. Furthermore, if needed, many laboratories worldwide have
RT-qPCR equipment that can be assembled at testing sites if necessary, for
maintaining adequate PCR detection capacity.
Further experiments should aim to see whether the AGPC method could be optimized
with respect to workflow, preparation time and to reduce manual handling of samples.
Of relevance, would be to test whether swabs could be taken and directly placed in TRI
Reagent
®, if viral transport media is in short supply or if just to inactivate the virus and
reduce the time spent aliquoting sample specimens into the organic solvent.
Furthermore, similar comparisons could be made for other viruses to allow detection for
various diagnostics in places not relying on automated systems.
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
Acknowledgements
We thank the laboratory and biomedical technicians at the Institute for Molecular
Medicine, University of Southern Denmark and Department of Clinical Microbiology,
Odense University Hospital for their expert technical assistance during testing,
validation and implementation of the AGPC method to expand the SARS-CoV-2 testing
capacity. We also thank dr. Kurt J. Handberg, Department of Clinical Microbiology,
Skejby, Denmark, for kindly sharing primers and probe sequences for the detection of
NDV.
Ethical statement
The study described herein was conducted at the Department of Clinical Microbiology,
Odense University Hospital, under the auspices of The Danish Health Authority.
Exception from review by the ethical committee system and informed consent was given
by the Regional Committees on Health Research Ethics for Southern Denmark in
accordance with Danish law on assay development projects.
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
References
1. Huang C, Wang Y, Li X, Ren L, Zhao J, Hu Y, et al. Clinical features of patients
infected with 2019 novel coronavirus in Wuhan, China. Lancet.
2020;395(10223):497-506.
2. Zhou F, Yu T, Du R, Fan G, Liu Y, Liu Z, et al. Clinical course and risk factors for
mortality of adult inpatients with COVID-19 in Wuhan, China: a retrospective
cohort study. Lancet. 2020;395(10229):1054-62.
3. Zhu N, Zhang D, Wang W, Li X, Yang B, Song J, et al. A Novel Coronavirus from
Patients with Pneumonia in China, 2019. N Engl J Med. 2020;382(8):727-33.
4. Fomsgaard ASR, M.W. An alternative workflow for molecular detection of SARS-
CoV-2 - escape from the NA extraction kit-shortage. Medrxiv. 2020.
5. Won J, Lee S, Park M, Kim TY, Park MG, Choi BY, et al. Development of a
Laboratory-safe and Low-cost Detection Protocol for SARS-CoV-2 of the
Coronavirus Disease 2019 (COVID-19). Exp Neurobiol. 2020.
6. Chomczynski P, and Sacchi N. Single-step method of RNA isolation by acid
guanidinium thiocyanate-phenol-chloroform extraction. Anal Biochem.
1987;162(1):156-9.
7. Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DKW, et al.
Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro
Surveill. 2020;25(3).
8. Xie X, Zhong Z, Zhao W, Zheng C, Wang F, and Liu J. Chest CT for Typical
2019-nCoV Pneumonia: Relationship to Negative RT-PCR Testing. Radiology.
2020:200343.
9. To KK, Tsang OT, Chik-Yan Yip C, Chan KH, Wu TC, Chan JMC, et al.
Consistent detection of 2019 novel coronavirus in saliva. Clin Infect Dis. 2020.
10. Kumar M, Mazur S, Ork BL, Postnikova E, Hensley LE, Jahrling PB, et al.
Inactivation and safety testing of Middle East Respiratory Syndrome Coronavirus.
J Virol Methods. 2015;223:13-8.
11. Blow JA, Dohm DJ, Negley DL, and Mores CN. Virus inactivation by nucleic acid
extraction reagents. J Virol Methods. 2004;119(2):195-8.
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
Figure 1: Phenol-chloroform extraction of RNA is a viable alternative to automated
systems, showing similar levels of sensitivity, specificity and accuracy.
(A) 87 patient specimens with known SARS-CoV-2 status based upon routine testing for
SARS-CoV-2 using the Cobas®6800 platform were compared to results reported for the in-
house SARS-CoV-2 RT -qPCR assay after RNA isolation using the Maxwell ® RSC 48
instrument or TRI Reagent ®. Only RNA specimens passing the internal control for the
Newcastle Disease Virus vaccine strain (NDV, Ct value s <29. 5) were used for the
analyses (Maxwell®; 82 samples , TRI Reagent ®; 80 samples). True positive (TP), true
negative (TN), false positive (FP), false negative (FN), confidence interval (CI). Sensitivity
is defined as the probability a test result is positive for a SARS -CoV-2 positive sample.
Specificity is defined as the probability a test result is negative for a SARS -CoV-2 negative
sample. Accuracy is defined as the probability a patient sample is correctly evaluated for
SARS-CoV-2.
(B) Diagram showing a highly significant correlation (r = 0.970, p<0.001) between obtained
Ct values for the SARS-CoV-2 E gene in SARS-CoV-2 positive specimens when assessed
by RT -qPCR using the Cobas®6800 platform and the in -house SARS -CoV-2 RT -qPCR
assay after RNA isolation using TRI Reagent®.
(C) Diagram showing a highly significant correlation (r = 0.978, p<0.001) between obtained
Ct values for the SARS-CoV-2 E gene in SARS-CoV-2 positive specimens when assessed
by RT-qPCR using the in -house RT-qPCR assay after RNA isolation using the Maxwell ®
RSC 48 instrument and TRI Reagent®.
(D) Side-by-side comparison of Ct values obtained for the internal control NDV when
assessed by RT -qPCR using identical patient specimens from which RNA was isolated
using the Maxwell® RSC 48 instrument or the AGPC method. Not significant (NS).
(E) Pie chart showing the average rate of sample specimens that pass the NDV internal
control for RNA extraction and sample quality after isolation of the RNA using the AGPC
pipeline (n = 736).
. CC-BY-NC 4.0 International licenseIt is made available under a
is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)
The copyright holder for this preprint this version posted May 27, 2020. ; https://doi.org/10.1101/2020.05.26.20099440doi: medRxiv preprint
Text is read by the "Ask this paper" AI Q&A widget below.
Extraction quality varies by source — PMC NXML preserves structure
cleanly, OA-HTML may include some navigation residue, and OA-PDF can
have broken hyphenation. The publisher copy
(via DOI)
is the canonical version.