Maassen LW
No ORCID on file
· 2 papers in corpus
We present a case about Essure® removal surgery in which the third markers of the device have torn off. The woman needed a second surgery for complete removal of the devices. Fluoroscopy during surgery is a good method to visualize the lost…
ObjectiveTo study the composition of intracavitary deposits on Essure® hysteroscopic sterilization devices.DesignCase report.SettingReproductive Medicine and Gynecology department of a University Hospital.PatientsA 39 years old patient pres…